Page 111

Chapter 5 Codes

Pharma Code Definition:

Usage: Serial Cmd:

The bar code encodes up to five different numbers, each with its own color which may be entered in decimal or "binary" format with a 1 represented by a thick bar and a 0 represented by a thin bar. Bar width is independent of height. In decimal format, each part can be up to 999999. In binary format, each input have up to 19 ones and zeros. Pharma Code is used mostly with packaging for the pharmaceuticals industry. <K477,status,fixed code length status,fixed code length,min. no. of bars,bar widths,fwd/rev>

Note: It is recommended that you disable AGC before enabling Pharma Code.

Pharma Code Status Serial Cmd: Default: Options:

<K477,status,fixed bar length status,fixed bar length,min. no. of bars,bar widths,fwd/rev> Disabled 0 = Disabled 1 = Enabled

Fixed Bar Length Status (Pharma Code) Serial Cmd: Default: Options:

<K477, status,fixed bar length status,fixed bar length,min. no. of bars,bar widths,fwd/rev> Disabled 0 = Disabled 1 = Enabled

Fixed Bar Length (Pharma Code) Serial Cmd: Default: Options:

<K477, status,fixed bar length status,fixed bar length,min. no. of bars,bar widths,fwd/rev> 10 4 to 16

5–Codes

MS-880 Industrial Long Range Scanner User’s Manual

5-27

Page 112

Pharma Code

Minimum Number of Bars (Pharma Code) Serial Cmd: Default: Options:

<K477,status,fixed bar length status,fixed bar length,min. no. of bars, bar widths,fwd/rev> 4 4 to 16

Bar Widths (Pharma Code) Serial Cmd: Default: Options:

<K477,status,fixed bar length status,fixed bar length,min. no. of bars, bar widths,fwd/rev> Mixed 0 = Mixed 1 = Narrow 2 = Wide

Forward/Reverse Decoding Direction (Pharma Code) Definition: Serial Cmd:

5–Codes

Default: Options:

Specifies the direction that a bar can be read. <K477,status,fixed bar length status,fixed bar length,min. no. of bars, bar widths,fwd/rev> Forward 0 = Forward 1 = Reverse

5-28

MS-880 Industrial Long Range Scanner User’s Manual

Page 113

Chapter 5 Codes

Autodiscriminate Although Autodiscriminate is not a configuration command, but it is included here as a convenient tool for enabling most code types. Definition: Usage:

ESP: ESP:

Serial Cmd: Default: Options:

Enables all available symbology types except PDF417, UCC/EAN 128, and Pharma Code. The user may also individually disable/enable each symbology type. Commonly used for quick setup mode to detect bar code type. This is particularly useful for users who might be unfamiliar with their application’s symbology. Note: It does not alter individual fields such as Start/Stop, Fixed Length, etc. These need to be setup individually. To enable all codes, click the Autodiscrimination button in the Advanced Calibration menu, or type in <P> in the Terminal screen. In the Utilities menu, right-click on Autodiscriminate and select Enable. Note: There will be a few seconds delay while the scanner values are retrieved.

<P> Code 39 (only) <P> Enables all codes except noted above. <Q> Enable Code 36 only

Note: If using an I 2/5 label, verify that the number of characters in the label being scanned matches the code length enabled for the I 2/5 code type (default is 10 and 6). See “Interleaved 2 of 5” on page 5-14.

MS-880 Industrial Long Range Scanner User’s Manual

5-29

5–Codes

<R> Enable Codabar only <S> Enable I 2/5 only

Page 114 read from a scan by OCR; it may hold reading errors

Autodiscriminate wn o xe} je} nt ie)

5-30 MS-880 Industrial Long Range Scanner User’s Manual

Page 115

6 Chapter Contents Overview of Autocalibration .....................................................................6-2 Autocalibration by ESP ............................................................................6-4 Autocalibration by Serial Command.........................................................6-7 Autocalibration by Embedded Menu ........................................................6-8

The purpose of autocalibration is to optimize the scanner’s optical settings (focus, gain, and tracking) to obtain the highest read rates in one or several label positions as required by the user’s application. This section includes the overall steps in autocalibration and specific steps for autocalibration by ESP, serial command, and embedded menus. For most applications autocalibration is the only optical setup required. Changes to motor speed, focus, speed, gain, tracking, and AGC can also be made individually by manual command.

Autocalibration Methods There are three ways to access autocalibration: 1. ESP Autocalibration menu. 2. Serial command <@CAL>. 3. Embedded menu <@>.

MS-880 Industrial Long Range Scanner User’s Manual

6-1

6–Autocalibration

Autocalibration

Chapter

Page 116

6–Autocalibration

Overview of Autocalibration

Overview of Autocalibration When an autocalibration command is sent, the scanner first steps through a coarse grid of focus settings, looking for readable bar code symbols, and recording the highest read rates it encounters. This is “Pass 1,” as shown by the blue line in figure 6-1, a screen capture of an actual autocalibration routine in ESP. On “Pass 2” (red line) it returns to the focus position where it found the best read rate and steps through a finer grid of focus, gain and tracking settings. The software evaluates the results and selects the best data. If it passes, it displays an “Autocalibration Passed” message at the top of the window.

Figure 6-1 Typical Autocalibration Graph

6-2

MS-880 Industrial Long Range Scanner User’s Manual

Page 117

Chapter 6 Autocalibration

Follow this procedure for all three methods of autocalibration: 1. Set your scan rate at the appropriate speed for your application. 2. Position your bar code symbol at the distance used in your application and confirm that it is at a distance within the specifications for that symbol size (see “Position Scanner and Symbol” on page 1-12). 3. If you are not certain of the code type, enable all code types by sending a <P> command from the terminal window, or clicking on the Autodiscrimination button in Calibration/Advanced in ESP.1 4. Start the autocalibration procedure. Note: Once a successful autocalibration is completed, you may want to save those setting to the calibration database (see Chapter 7, “Calibration Database”). 5. Repeat autocalibration at other ranges and/or label types you might be scanning in your application. 6. Check your readings at each end of the scan width required for your application, keeping in mind that the scan width testing may not produce results that are entirely symmetrical. Note: If you are setting up for Laser Framing (see page 9-6), be sure that you only test for autocalibration in the active areas of the scan beam. 7. To save the new settings (including focus, gain, and tracking) for power-up, send a <Zp> command.

Figure 6-2 Autocalibration Setup

Note: To silence the beeper, send <K702,0> or remove or block the test label. After autocalibration has passed, the scanner will beep once, automatically exit the autocalibration routine and begin scanning for a label in Continuous Read mode, beeping for each good read.

1. Note: If using an I 2 of 5 label, verify that the number of characters in the label being scanned matches the code length enabled for the I 2 of 5 code type (default is 10 and 6). See “Interleaved 2 of 5” on page 5-14. MS-880 Industrial Long Range Scanner User’s Manual

6-3

6–Autocalibration

Autocalibration Steps

Page 118

6–Autocalibration

Autocalibration by ESP

Autocalibration by ESP 1. Click on the Autocalibration button on the overhead toolbar. 2. Click on the Auto Cal button to start autocalibration. An Autocalibration in Progress message will be displayed on your monitor along with the pass number and information on the focus, gain, and tracking. Allow some time for the scanner to cycle through the focus, gain, and tracking settings. Do not move or disturb scanner or label while Autocalibration is in progress. (It might take 2 to 3 minutes to complete.) 3. When completed, the settings that provided the highest read rate will be selected. You will see one of the following messages: Autocalibration Passed, Autocalibration Failed, or Original Settings Were Optimal.

Figure 6-3 Autocalibration In Process

If it passes, you will continue to see the read rate percentage and label data in the window below the scanner image. Note that in the figure 6-4 the focus position remains the same as it was in figure 6-3, but the Gain and Tracking have been finetuned during the second pass. The graph in figure 6-4 shows the results of this procedure. Figure 6-4 Autocalibration Passed

6-4

MS-880 Industrial Long Range Scanner User’s Manual

Page 119

Chapter 6 Autocalibration

In the advanced tab in the Autocalibration menu you can constrain the scan width of the scan beam with Laser Framing and set up the rules the scanner will follow when making the first pass of the autocalibration procedure.

Laser Framing Sets the length of time the laser is ON during each scan so that only a selected portion of the scan width is effectively scanned. To adjust the scan width: 1. First click on Framing Status. 2. Click on the pointers in Laser Off and Laser On to adjust the shape of your scan beam width to match only the area in which your bar code symbol will appear.

Focus Stepping After you’ve used autocalibration a time or two and observed the progress of the first and second passes in the graph and table in the Results tab, you will sense the relationship between focus position and the distance to the bar code symbol. See figure 6-7 for a table on this. To improve the speed and accuracy of your autocalibration setup: 1. Click on the Advanced tab in the Autocalibration menu. 2. If Auto Range is enabled, click on it to disable it. 3. From previous results, estimate the value of Starting Focus Position and drop down one or two positions. For example, if you estimate the distance to your label to correspond to a Starting Focus Position of 6, enter 5 or 4 to be certain of not missing the peak performance distance. 4. Enter a value in Focus Step Size. This is the number of focus increments that will be stepped through during the first autocalibration pass. The range is 2 to 5. The higher values will result in fewer readings during the first pass but will increase the risk of missing the peak performance position.

MS-880 Industrial Long Range Scanner User’s Manual

6-5

6–Autocalibration

Advanced ESP Autocalibration Features

Page 120

6–Autocalibration

Autocalibration by ESP Auto Range Enabling Auto Range will improve speed, but with a caveat. With Auto Range, the scanner will focus outward until it senses an object by signal strength and begin looking for a decodable symbol from there. This will result in quicker acquisition, but only if other objects closer to the scanner are not mistaken for the bar-coded object. Again, the user should experiment with this routine until satisfactory results are achieved. In any case, you can observe the Results tab and determine whether Auto Range will work smoothly with your application. To use Auto Range: 1. Make certain that no extraneous objects that might fall within the scan beam are between the bar code symbol and the scanner. 2. Click on the Advanced tab in the Autocalibration menu. 3. If Auto Range is not enabled, click on it to enable it. 4. Click on the Auto Cal button to start autocalibration.

6-6

MS-880 Industrial Long Range Scanner User’s Manual