Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 13 of 23

MS-850 Raster Scanner User's Manual, pages 121 to 130

Pages 121 to 130 of 222

The original document is the PDF (4.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan MS-850

Page 121

Raster Control

Chapter

8

Chapter Contents Raster Setup ............................................................................................... 8-3

This section explains the setup and adjustments for the raster setup. Note: The characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands.

8–Raster Control

Raster Setup by Serial Command Command Title cmds

Format

Raster

KR

<KRstatus,raster top,raster bottom,raster sweep rate,read cycle on/off>

Scanner Setup Status Request

KU?

<KU?>

MS-850 Raster Scanner User’s Manual

8-1

Page 122

Raster Setup by ESP Menus

Click on this button to bring up the Raster menu. To change a setting, double-click on the setting and use your curser to scroll through the options.

Note: Setup commands in the ESP-MP are accessed from both the Read Cycle/Triggering menu and the Raster button on the toolbar.

Raster Setup by Embedded Menu

8–Raster Control

To access the Raster operations, select the Scanner Setup menu:

8-2

MS-850 Raster Scanner User’s Manual

Page 123

Chapter 8 Raster Control

Raster Setup Definition:

The user can enable raster scanning and adjust the sweep range (the upper and lower limits of the raster sweep).

Usage:

Raster scanning is useful when labels vary in placement and a single scan line cannot be counted on to cross the label. Raster scanning is also useful in reading PDF417.

Serial Cmd:

<KRstatus,top offset,bottom offset,raster sweep rate,read cycle on/off>

Raster Status Definition:

When disabled, the raster mirror will park at the top of the raster sweep.

Serial Cmd:

<KRstatus,top offset,bottom offset,raster sweep rate,read cycle on/off>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

Note: Raster sweeps are only counted during a read cycle (continuous, triggered, and read rate). Raster sweeps are NOT counted when the raster motor is parked, disabled, set to straight line operation (top and bottom are equal), or when performing a selfcalibration.

Top Offset Decreasing the Top Offset value causes the top of the raster pattern to move up.

Serial Cmd:

<KRstatus,top offset,raster bottom,raster sweep rate,read cycle on/off>

Default:

85

Options:

0 to 255

MS-850 Raster Scanner User’s Manual

8–Raster Control

Definition:

8-3

Page 124

Raster Setup

Bottom Offset Definition:

Increasing the Bottom Offset value causes the bottom of the raster pattern to move down.

Serial Cmd:

<KRstatus,top offset,bottom offset,raster sweep rate,read cycle on/off>

Default:

170

Options:

0 to 255

8–Raster Control

Note: For single scan line operation, set Top Offset equal to Bottom Offset or click on the Static Line option. Note: Assigning a larger value to Top Offset than Bottom Offset will park the raster mirror in its “home” (default) position.

8-4

MS-850 Raster Scanner User’s Manual

Page 125

Chapter 8 Raster Control

Raster Sweep Rate Definition:

The raster sweep rate is the number of sweeps per second the raster mirror completes. A sweep is one pass of the raster, up or down. The maximum sweep rate is limited by the size of the arc the raster must move through.

Usage:

If your application allows it, a slower sweep rate can produce more scans per second and better read rates. To determine the minimum read rate for your application, see “Raster Picket Fence Calculation” on page A-26.

Serial Cmd:

<KRstatus,top offset,bottom offset,raster sweep rate,read cycle on/off>

Default:

10

Options:

0 to 30 The table below shows the maximum speeds at selected raster sweeps. To maximize the number of scans per label, use the lowest effective sweep rate required for the application. Table 8-1 Maximum Sweep Rates at Selected Sweep Arcs Raster Sweep Arc 0 to 10° 11 to 20° 21 to 30°

Maximum Sweeps per Second 30 20 10

8–Raster Control

MS-850 Raster Scanner User’s Manual

8-5

Page 126 read from a scan by OCR; it may hold reading errors

Raster Setup

° Sem -_ < re) O Son ® . od 7) © a re)

8-6 MS-850 Raster Scanner User’s Manual

Page 127

Outputs

Chapter

9 Chapter Contents

This section explains the commands for controlling data and beeper outputs. Note: The characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands.

MS-850 Raster Scanner User’s Manual

9-1

9–Outputs

Noread Message .........................................................................................9-5 Bad Bar Code Message ..............................................................................9-6 No Bar Code Message ................................................................................9-7 Bar Code Output .........................................................................................9-8 Laser Control .............................................................................................9-11 Serial Verification ......................................................................................9-12 Beeper Status............................................................................................9-14 Partial Output ............................................................................................9-15 Operation Command Output .....................................................................9-16

Page 128

Outputs by Serial Command Command Title

Beeper Partial Output No Bar Code Bad Bar Code cmds Format Kk <Kkstatus,message> Kla <Klstatus,when to output> <KSserial command echo status,serial command beep KS status,control/hex output> Ku <Kustatus> KY <KYstatus,start postion,length> KN <KNstatus,message> K’ <K’status,message>

Laser On/Off

KC

<KCstatus,laser on/off status>

Operation Command Output

K/

<K/status>

Noread Message Bar Code Output Serial Verification a. Lower case L

9–Outputs

Outputs by Embedded Menu For information on accessing the embedded menus, see “Embedded Menus” on page A-18. From the Main menu, scroll down through the options and select the following screen:

9-2

MS-850 Raster Scanner User’s Manual

Page 129

Chapter 9 Outputs

Output Data by ESP Menu

Click this button to bring up the Outputs menu.

9–Outputs

To change a setting, double-click the setting and use your curser to scroll through the options.

MS-850 Raster Scanner User’s Manual

9-3

Page 130

Message Output The following flow chart explains the logic paths taken for Noread, Bad Bar Code and No Bar Code outputs.

Note: When Transition Counter Status is enabled, the scanner will test for a minimum number of bar and space transitions. If that minimum threshold is met, a Bad Bar Code message is sent; otherwise a No Bar Code message is sent.

Start of read cycle

Scans until end of read cycle

Does the scanner decode the symbol?

Yes

No

9–Outputs

Is Transition Sample Status enabled?

No

Scanner send Noread message (if enabled)

Yes

Scanner sends the Bad Bar Code message (if enabled)

Yes

Is Transition Sample Threshold met? No

Scanner Scanner outputs sendsthe the No NoBar Label Code message message (if(ifenabled) enabled) End of read cycle

Figure 9-1 Output Message Flow

9-4

MS-850 Raster Scanner User’s Manual