For the Microscan MS-820
Page 171
Chapter 11 Utilities Enter Percent PDF Read Rate Test Sending <CPDF> instructs the scanner to transmit the number of decodes per second on the full PDF label. This test includes both capture and decode and translation time. Note that the output format of the data is done as a running average of 750 mS windows. This means that you can remove the label from the view of the scanner and it will continue to transmit some history of the label for a few outputs.
End Read Rate Test Sending <J> ends both the Percent test and the Decodes/Second test for both single and multi-label. Sending the <a1> check box will cause PDF417 symbology data to be prefaced with information consisting of error correction level, number of rows, number of columns, number of informative code words and the number of data characters. Explanation of above PDF Info sample of a PDF417: Level = 5 tells you that the error correction level is level 5. There are 13 rows and 6 columns. There are 14 info code words, and 16 characters in the data. This feature can be disabled by using the exit read rate command <J>.
MS-820 Industrial Bar Code Scanner User’s Manual
11-11
11–Utilities
Enable PDF Information
Page 172
Device Control
Device Control You can control laser scanning by ESP or serial command.
By ESP You can access Device Control from the Utilities menu.
11–Utilities
For any of the Device Control commands, right-click on the command and make your selection.
By Serial Command Enable Laser Scanning Sending <H> will turn the laser on continuously. Note: Enable Laser Scanning does not relate to Laser On/Off command.
Disable Laser Scanning Sending <I> will turn the laser off. This feature is useful during extended periods of time when no bar code labels are being scanned or the scanner is being configured. Disabling laser scanning will not affect any downloaded commands to the scanner. Note: Disable Laser Scanning does not relate to Laser On/Off, which when enabled also turns off the laser but only between read cycles.
Motor On Sending <KE> turns the spinning mirror on (if not already running). Note: the spinning mirror reaches full speed after a short time delay.
Motor Off Sending <KF> turns the spinning mirror motor off. This command is useful for long idle periods. Note: Laser turns off whenever motor is off.
11-12
MS-820 Industrial Bar Code Scanner User’s Manual
Page 173
Chapter 11 Utilities Output #1 Sending <L1> activates the link between Output 1(+) and Output 1 (–) of the 25-pin host connector (regardless of Master Label or Output 1 status).
Output #2 Sending <L2> activates the link between Output 2(+) and Output 2 (–) of the 25-pin host connector (regardless of Master Label or Output 2 status).
Output #3 Sending <L3> activates the link between Output 3(+) and Output 3 (–) of the 25-pin host connector (regardless of Master Label or Output 3 status).
11–Utilities
Code Type See also “Autodiscriminate” on page 5-28 for a discussion of Utility commands for Code Types and Autodiscrimination. Default: Options:
Code 39 (only) <P> Enables most code types. <Q> Enable Code 39 only <R> Enable Codabar only <S> Enable I 2/5 only
Master Label Database See “Master Label Database” on page 7-10.
Defaulting/Saving/Initializing See Appendix H, “Defaulting/Saving/Initializing” on page A-20.
MS-820 Industrial Bar Code Scanner User’s Manual
11-13
Page 174
11–Utilities
Defaulting/Saving/Initializing
11-14
MS-820 Industrial Bar Code Scanner User’s Manual
Page 175
Appendices
Chapter Contents
MS-820 Industrial Bar Code Scanner User’s Manual
A-1
Appendices
Appendix A General Specifications ............................................................ A-2 Appendix B Electrical Specifications .......................................................... A-5 Appendix C Serial Configuration Commands........................................... A-11 Appendix D Serial Command Format....................................................... A-14 Appendix E ASCII Table........................................................................... A-16 Appendix F Embedded Menus ................................................................. A-18 Appendix G Defaulting/Saving/Initializing................................................. A-19 Appendix H Position Scanner and Bar Code ........................................... A-23 Appendix I Test Decode Rate .................................................................. A-24 Appendix J Formulas for Number of Decodes ............................................ A-25 Appendix K Operational Tips.................................................................... A-29 Appendix L Interface Standards ............................................................... A-30 Appendix M Multidrop Communications................................................... A-31 Appendix N Glossary of Terms ................................................................ A-36
Page 176
Appendix A
Appendix A, General Specifications Key Features • Multiple label and symbology reading • Industrial IP65 rated enclosures
Back .93" (2.36 cm)
Mechanical Depth: 2.59” (65.8mm) Width: 3.47” (88.1mm) Height: 1.38” (35.0mm) Weight: 7.5 oz. (212g)
.54" (1.37 cm) 2 x M4 x .7 .18 deep
1,10" (2.79 cm)
Top
Appendices
Environmental Ambient Light Immunity: 450 foot candles (indoor: fluorescent, incandescent, mercury vapor, sodium vapor); 1800 foot candles (soft outdoors) Enclosure rating: IP65 Operating temperature: 0° to 50°C Storage Temperature: –29° to 70°COperating temperature: Humidity: Up to 90% (non-condensing)
Emissions and Immunity
.53" (1.35 cm)
3.47" (8.81 cm)
Side 1.10" (2.7 cm)
.62" (1.57 cm)
Laser Light
1.40" (3.56 cm)
A-2
Front
1.38" (3.50 cm)
General immunity for ITE equipment: EN 55024:98 General immunity for heavy industry: EN 61000-6-2:98 Radiated and conducted emissions of ITE equipment: EN 55022:98, Class A
Semiconductor visible laser diode: 650nm nominal; 780nM nominal (optional) Safety class: Visible laser: CDRH Class II, 650nm Infrared laser: CDRH Class 1, 780nm Operating Life: 50,000 hours @25°C
2.40" (6.10 cm)
2.59" (6.58 cm)
.54" (0.37cm)
2.59" (6.58 cm)
2 pl, M4.0 x 0.7 3.8 mm deep
2.59" (6.58 cm)
Bottom
Figure A-1 MS-820 Dimensions
MS-820 Industrial Bar Code Scanner User’s Manual
Page 177
General Specifications Scanning Parameters Scanning mirror type: Rotating, 10-faceted Options: Single line/raster Scan width angle: Typically 60°; Pitch: ±50° max.; Skew: ±40° max. Label contrast: 25% min. absolute dark to light differential at 650nm wavelength
Communication RS-232, RS-422/485, Daisy Chain , DeviceNet and Ethernet (Modicon) can be used with an accessory IB-131 Interface.
Code Types Standard offering: Code 128, Code 39, Code 93, Codabar, Interleaved 2 of 5, UPC/EAN, PDF417 Application standards: AIAG, UCC/EAN-128
FIS Options Light
Density
Line
001
Visible
Low Density
Single Line
002
Visible
High Density
Single Line
003
Visible
Low Density
Raster
004
Visible
High Density
Raster
Appendices
FIS #
Read Ranges Table A-1 MS-820 Read Rangesa Narrow-bar-width .0033" (.084mm) .005" (.127mm) .0075" (.190mm) .010" (.254mm) .015" (.381mm) .020" (.508mm) .040" (1.020mm)
High Density Call Microscan 4" to 5.5" (102 to 140mm) 3.5" to 6.75" (89 to 171mm) 3.25" to 8" (82 to 203mm) 3" to 9" (76 to 229mm) -
Low Density 10" to 12" (254 to 305mm) 7" to 16" (178 to 406mm) 6" to 19" (152 to 483mm) 5" to 22" (127 to 558mm) 4" to 30" (102 to 762mm)
a. Range data is based on a grade A Code 39 label.
MS-820 Industrial Bar Code Scanner User’s Manual
A-3
Page 178
Appendix A Beeper Beeps on output (good read, noread, on/off.
LED Indicators Table A-2 MS-820 Status Lights LED PWR RDY GD/RD
State Red-On Green-On Green-On
Status Scanner has power Scanner is ready to read Label is qualified as good
Safety Certifications
Appendices
FCC, CDRH, CE, UL/cUL, TüV, BSMI
Product specifications are given for typical performance at 25°C (77°F) using grade A labels. Some performance characteristics may vary at high temperatures or other environmental extremes. © Microscan Systems, Inc., Specifications subject to change Note: Specifications subject to change.
A-4
MS-820 Industrial Bar Code Scanner User’s Manual
Page 179
Electrical Specifications
Appendix B, Electrical Specifications Maximum Operating Power: 3.6 Watts Power Input: 10 to 28VDC, 200mV p-p max. ripple, 120 mA @ 24VDC (typical) Trigger, New Master: 4.5 to 28 VDC (optoisolated) Outputs (1,2,3): 1 to 28VDC (optoisolated) rated (ICE <100mA @24VDC, curren7t limited by user)
Pin Assignments Table A-3 MS-820 Host Connector, 15-pin Pin
In/ Out In
Out In
6
1
5
In Out
Appendices
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Host Host & Aux Host RS232 RS232 RS422/485 Power +10 to 28 VDC Host TxD Host TxD TxD (–) Host RxD Host RxD RxD (–) Power/Signal ground Trigger (–) RTS Aux TxD TxD (+) Output 1 (+) Default configurationa Trigger (+) CTS Aux RxD RxD (+) Output 3 (+) New master pin Chassis groundb Output 2 (+) Outputs 1, 2, 3 (–)
Out
In In In Out In
11
15
10
Out Out a. The default is activated by connecting pin 8 to ground pin 4. See “By Hardware Default” on page A-21. b. Chassis ground: Used to connect chassis body to earth ground only. Not to be used as power or signal return.
Caution: If using your own power supply, verify correct connection of power and ground lines. Incorrect connections or use of “Chassis ground,” “Power ground,” and “Signal ground” lines could cause equipment and/or software failure.
MS-820 Industrial Bar Code Scanner User’s Manual
A-5
Page 180
Appendix B
Optoisolator Outputs Optoisolator circuits can transfer pulses between the scanner and peripherals with no direct connection with the scanner’s internal circuitry. However, not every optoisolator configuration provides complete isolation. The following diagrams show both fully optoisolated and non-optoisolated circuits. They are only examples and do not represent all the possible wiring configurations. Generic Output Waveform Characteristics Iload =5mA Iload =50mA
Iload =100mA
V Out-On
0.5V
0.5V
1.0V t On-Typ
5mS
0.7mS
0.8mS t Off-Typ
5µS
5µS
5µS
Outputs (+)
Scanner Output
ILoad Outputs (–)
Output Circuit Examples
Appendices
Fully Optoisolated
Host
Isolated Scanner
This circuit is fully optoisolated and the recommended configuration. It allows the user to apply 1 to 28 VDC to the circuit. Caution: The maximum current that can pass through the optoisolator is 100 mA.
Power (+) 1 to 28 V Scanner
Outputs (+)
Output
Input Outputs (–) Power ground
Not Optoisolated, Scanner Grounded In this diagram, power is applied externally, but the scanner’s power ground is used to complete the circuit. This setup involves some risk to the optoisolator if excessive voltages are applied. Caution: The maximum current that can pass through the optoisolator is 100 mA.
A-6
Non-isolated Scanner
Host
Power (+) 1 to 28 V Scanner
Outputs (+)
Output
Input Outputs (–)
Power ground