Chapter 10 Diagnostics
Laser High Warning Transmits a message whenever the laser current exceeds a factory-calibrated reference value which cannot be altered. The message repeats once every 30 minutes until the condition is corrected.
Laser High Status Definition: Usage: Serial Cmd: Default: Options:
Enables the Laser High message. Alerts the user to impending laser failure. (Contact Microscan Service.) <K”warning message status,laser high status,laser low status, novram/reset warning status> Enabled 0 = Disabled 1 = Enabled
Laser High Message Definition: Serial Cmd: Default: Options:
Defines the Laser High message. <K;laser high message> HI-LASER Any 1 to 10 character ASCII string except NUL, <, or >.
10–Diagnostics
MS-820 Industrial Bar Code Scanner User’s Manual
10-5
Page 152
Warning Messages
Laser Low Warning When enabled a message up to 10-characters is transmitted whenever the laser current falls below a factory-calibrated reference value which cannot be altered. The message repeats once every 30 minutes until the condition is corrected.
Laser Low Status Definition: Usage: Serial Cmd: Default: Options:
Enables the Laser Low message. Alerts the user to impending laser failure. (Contact Microscan Service.) <K” warning message status,laser high status,laser low status, novram/reset warning status> Enabled 0 = Disabled 1 = Enabled
Laser Low Message Definition:
10–Diagnostics
Serial Cmd: Default: Options:
When enabled a message up to 10-characters is transmitted whenever the laser current falls below a factory-calibrated reference value which cannot be altered. The message repeats once every 30 minutes until the condition is corrected. Defines the Laser Low message. <K:laser low message> LO-LASER Any 1 to 10 character ASCII string except NUL, <, or >.
10-6
MS-820 Industrial Bar Code Scanner User’s Manual
Page 153
Chapter 10 Diagnostics
NOVRAM Corrupt Warning Status When enabled, a 7 character message is sent whenever the system detects a warning condition. Important Note: Warning Corrupt Message Status must be enabled before the NOVRAM message in this menu can be transmitted. Function:
Usage: Serial Cmd: Default: Options:
When enabled, a 7 character message is sent whenever the system detects a warning condition. However, if you suspect that erroneous defaults are occurring, enable NOVRAM Messages by sending <K1,,,1>. This will allow the NOVRAM messages, if present, to be viewed. Alerts the user to NOVRAM failure. (Contact Microscan Service.) <Kwarning message status,laser high status,laser low status, novram reset warning status> Disabled 0 = Disabled 1 = Enabled
Warning messages for NOVRAM can appear in one of five formats: <NVRAM-C> This message indicates an error in user settings. The message repeats once every 30 minutes. If this message is received, restore customer settings and send a <Z> command. If the condition persists, call customer service.
<NVBAD-D> Indicates that diagnostic lifetime hours, power-on counts, and reset counts have been lost. The message repeats once every 30 minutes. (Contact Microscan Service.)
<NVBAD-L> Indicates a problem with laser current factory reference. The message repeats once every 30 minutes. (Contact Microscan Service.)
<DFALT-C> Indicates that the customer NOVRAM settings have been defaulted either by a customer initiated default or by a problem with the scanner. This message is sent ONLY ONCE. Indicates that the factory NOVRAM settings have been defaulted. This message is sent ONLY ONCE upon reset. (Contact Microscan Service.)
MS-820 Industrial Bar Code Scanner User’s Manual
10-7
10–Diagnostics
<DFALT-F>
Page 154
High Temperature Threshold
High Temperature Threshold A message can be defined that will display when a user-set High Temperature Threshold has been reached. The message repeats once every 30 minutes until the condition is corrected. This value is a reference to the external ambient temperature in the environment around the scanner.
High Temperature Degrees Definition: Usage:
Serial Cmd: Default: Options:
A user can set a temperature value that when exceeded will cause a high temperature message to be displayed. If High Temperature Threshold is set to zero, then the warning message will be disabled Helps ensure that the scanner is being used within its temperature specification. Also, since hotter environments tend to shorten the life of electronics components, a user may want to set the temperature threshold closer to the nominal temperature of 25°. <K+degrees,message> 50° Celsius 0° to 50°, zero disables.
High Temperature Message Definition:
10–Diagnostics
Serial Cmd: Default: Options:
The user can enter a 10-character message that will be displayed whenever High Temperature Threshold is exceeded. <K+degrees,message> HI_TEMP Any 1 to 10 character ASCII string except NUL, <, or >.
10-8
MS-820 Industrial Bar Code Scanner User’s Manual
Page 155
Chapter 10 Diagnostics
Low Temperature Threshold A message can be defined that will display when a user-set Low Temperature Threshold has been reached. The message repeats once every 30 minutes until the condition is corrected. This value is a reference to the external ambient temperature in the environment around the scanner.
Low Temperature Degrees Definition:
Usage: Serial Cmd: Default: Options:
The user can set a low temperature value that whenever the ambient temperature falls below it causes a low temperature message to be displayed. If Low Temperature Threshold is set to zero, the warning message will be disabled. Helps ensure that the scanner is being used within its temperature specification. <K – degrees,message> 1° Celsius 0° to 50°, zero disables.
Low Temperature Message Definition: Serial Cmd: Default: Options:
The user can enter a 10-character message that will be displayed whenever the ambient temperature falls below the Low Temperature Threshold. <K– degrees,message> LO_TEMP Any 1 to 10 character ASCII string except NUL, <, or >.
10–Diagnostics
MS-820 Industrial Bar Code Scanner User’s Manual
10-9
Page 156
Lifetime Hours
Lifetime Hours A message can be defined that will be transmitted whenever the scanner’s elapsed time clock exceeds the set hours threshold. The message repeats one time every 30 minutes.
Hours Threshold Definition:
Usage: Serial Cmd: Default: Options:
A user can set a time in hours that will cause a message to be sent whenever the scanner’s run time exceeds this value. The elapsed time is the time the scanner is powered up. Useful for setting up a preventive maintenance program. <K$ hours10,message> 2500 (actually 25,000 hours) 0 to 65534
Note: Hours entered are 1/10 of the actual hours.
Message Definition:
10–Diagnostics
Serial Cmd: Default: Options:
A user can define message up to 10-characters that will be displayed whenever the hours threshold is reached or exceeded. <K$hours10,message> SERVICE Any 1 to 10 character ASCII string except NUL, <, or >.
10-10
MS-820 Industrial Bar Code Scanner User’s Manual
Page 157
Chapter 10 Diagnostics
Present Ambient Temperature (deg. C) (This command is READ ONLY.) Definition:
Usage: Serial Cmd:
Close estimation of external temperature. A thermal sensor on the main PCB measures the internal temperature of the scanner. From this a case-differential constant is subtracted to yield an estimated external temperature. Indicates temperature of the environment immediately outside of the scanner. Useful for setting Over Temperature Threshold and Under Temperature Threshold. Send: <K%?> Returns <K% degrees Celsius>.
Note: When viewing the Present Operating Temperature, allow at least 20 minutes after startup for the scanner’s inside and outside temperatures to stabilize.
10–Diagnostics
MS-820 Industrial Bar Code Scanner User’s Manual
10-11
Page 158
Counts
Counts (These commands are READ ONLY.)
Powerups Definition Usage: Serial Cmd: Read Only Ranges:
Returns the number of times the scanner has been re-powered or a watchdog reset occurs. Useful for detecting unwanted resets caused by power supply problems or ESD transients. Send <K_> Returns <K_ powerups,resets> 0 to 65,535 powerups, 0 to 65,535 resets.
Resets Definition Usage: Serial Cmd:
Resets include watchdog reset, <A>, <Z>, <Zd>, and hardware defaults. A watchdog reset is a reset that is forced whenever the software locks-up. Useful for detecting unwanted resets caused by power supply problems or ESD transients. Send <K_> Returns <K_ powerups,resets> 0 to 65,535 powerups, 0 to 65,535 resets.
10–Diagnostics
Read Only Ranges:
10-12
MS-820 Industrial Bar Code Scanner User’s Manual
Page 159
Chapter 10 Diagnostics
Time Since Last Reset (This command is READ ONLY.) Definition: Usage: Serial Cmd:
Records the number of hours and minutes of operation since the last system reset. Useful as a troubleshooting tool that can help pinpoint the cause of a reset. Send <K@?> Returns <K@ hours,minutes>
Read Only Ranges:
0 to 23 hours, 0 to 59 minutes.