For the Microscan MS-610 and Microscan MS-615

Page 91

Multidrop Communications select address, 1F hex, and an ACK (acknowledge). The concentrator then transmits an STX (start of text), the data (in this case a <T>), an ETX (end of text), and an LRC character. The scanner replies by transmitting its own address, followed by an ACK, acknowledging receipt of the command. Upon receipt of an ACK, the concentrator concludes the successful exchange with a RES. In the example above, the scanner only acknowledges a trigger counter request from the concentrator. It does not respond to the trigger counter request until a subsequent poll. For example, if the scanner’s trigger count was 12 at the time the trigger counter request was received, on a subsequent poll it would transmit 02T/00012. (The 02 at the beginning of the string is the scanner’s address.)

Select Reset • If the scanner receives bad data from the concentrator, it transmits a SEL (its select address) and a NAK to the concentrator. The concentrator retransmits the data up to three times. The concentrator will end the sequence with a RES (reset) if no ACK is received. 1

Appendices

1. For additional information on Multidrop, see the MS-5000 Multidrop Concentrator User's Manual. MS-610 Scanner User’s Manual

A-17

Page 92 read from a scan by OCR; it may hold reading errors

r) ® os c ® Q. a. <<

Appendices

A-18

Table A-7 Multidrop Address Characters

; Poll Select ; Poll Select Multidrop | Character Character Multidrop | Character Character Address Address

ASCII | HEX | ASCII | HEX ASCII | HEX | ASCII | HEX

01 A\ 1C | 1D 26 N 4E O 4F

02 ne 1E A. 1F 27 P 50 Q 51

03 SP 20 ! 21 28 R 52 S 53

04 " 22 # 23 29 T 54 U 55

05 $ 24 % 25 30 V 56 Ww 57

06 & 26 ' 27 31 Xx 58 Y 59

07 ( 28 29 32 Z 5A [ 5B

08 * 2A + 2B 33 \ 5C ] 5D

09 , 2C - 2D 34 us 5E 5F

10 2E / 2F 35 60 a 61

11 0 30 1 31 36 b 62 C 63

12 2 32 3 33 37 d 64 e 65

13 4 34 5 35 38 f 66 g 67

14 6 36 7 37 39 h 68 i 69

15 8 38 9 39 40 j 6A k 6B

16 3A ; 3B 41 | 6C m 6D

17 < 3C = 3D 42 n 6E fe) 6F

18 > 3E ? 3F 43 p 70 q 71

19 @ 40 A 41 44 r 72 s 73

20 B 42 Cc 43 45 t 74 u 75

21 D 44 E 45 46 Vv 76 Ww a

22 F 46 G 47 47 x 78 y 79

23 H 48 I 49 48 z TA { 7B

24 J 4A K 4B 49 | 7C } 7D

25 L 4C M 4D 50 o 7E A 7F

MS-610 Scanner User’s Manual

Page 93

Glossary of Terms

Appendix J, Glossary of Terms Autodiscriminate. The ability to decode several different bar code symbologies without changing configuration. AWG. Abbreviation for American Wire Gauge. This is a standard measuring system for wire sizes, listed from 0 (biggest) to 40 (smallest). A 26 AWG is equivalent to 0.0159 in. (0.405 mm) in diameter. Bar Code. Data that has been encoded into an array of parallel bars and spaces of varying widths. Bar Code Density. Number of characters per inch or other unit of measure. Baud Rate. The term used to describe the number of discrete signal events per second. In RS-232 and RS-422/485 systems, baud rate is the same as bits per second (bps). Clock Speed. Internal decoder timing relative to individual bars and spaces. Code 39. An alphanumeric bar code with a character set containing a start/ stop character, 10 numbers, 26 letters, 6 symbols, and a space. This code is discrete, variable length, and self-checking. Configuration. The method used to change factory default settings for operational features to match a specific application. Configuration can be done through menu selection or with serial commands. Connector. Physical device (plug or socket) on unit or cable to provide in/ out connectivity for various circuits and pins. Counter. Memory space provided to keep track of read cycle events. Depth of Field. The distance between the minimum and maximum range in which a scanner can read bar code labels. EPROM. Erasable, programmable, read only memory. End of Read Cycle. The time at which the scanner stops expecting label information to decode. This can be caused by a timeout, a trigger event, or a good read. Focal Length. The distance measured from the scanner to the center of the depth of field, or focal point.

Intercharacter Gap. The extra space between the last element of one character and the first element of the adjacent character of a specific bar code symbol. MS-610 Scanner User’s Manual

A-19

Appendices

Good Read. The event that occurs when a label's data is accurately scanned and decoded.

Page 94

Appendices Label Height. Regardless of orientation, the measurement taken along the length of a label’s individual bars. Label Length. Regardless of orientation, the measurement taken across the label's bars from one end to the other, including the quiet zone. Label Speed. The rate in inches or centimeters per second at which a label moves through the scan beam. Ladder Label Orientation. A bar code label in which the bars are parallel to the label’s direction of travel. Match Code. The ability to compare bar code labels being scanned against a master label that is stored in the memory of the scanner. Menu Configuration. The process of changing factory default settings via a sequence of menus displayed on a terminal monitor. Mil. One thousandths of an inch or 0.0254 mm. In bar-coding, a measurement that identifies a bar code label by the width of its narrowest element. Mismatch. An event that occurs when the scanned bar code label does not match the master label that is stored in the memory of the scanner. Multidrop. A communications protocol for networking two or more scanners or other devices with a concentrator (or controller) and characterized by the use of individual device addresses and the RS-485 standard. Narrow-bar-width. The width of the narrowest bar of a given label, expressed in thousands of an inch (or mils). Non-volatile RAM (NOVRAM). Random Access Memory that is available on power-on; that is, after power to the unit has been recycled. Noread. A non-read. A condition that occurs when the scanner is set up to decode labels, and no labels are scanned during the read cycle. Number of Scans Calculation. The number of times a bar code label is scanned by the scanner during one pass through the laser beam. Object Detector. A photo electric device used to sense to presence or absence of an object.

Appendices

Operational Commands. Serial commands from the host to the scanner that control current operating parameters, counters, master label operations, and program management that are not stored in non-volatile RAM. Oval Spot. An elongated laser beam involving custom optics that is designed to project lengthwise with the label's bars and spaces. Picket Fence Label Orientation. A bar code label in which the bars are perpendicular to the label’s direction of travel. A-20

MS-610 Scanner User’s Manual

Page 95

Glossary of Terms Pitch. Label (or scanner) rotation around the center a line perpendicular to the label’s bars. Point-to-Point. A protocol consisting of a single communications event, typically used to connect a bar code reader to a terminal or host computer . Port. Logical circuit for data entry and exit. (One or more ports may be included within a single connector.) Protocol. The rules for communication between devices, providing a means to control the orderly flow of information between linked devices. Quiet Zones. Specified “clear” (nonprinted) areas immediately before and after the bar code symbol. The area is usually white (for black and white bar code) and at least 10 times the width of the narrowest bar, as measured in thousands of an inch. The zones can be other than white as long as their densities remains consistent and they have the required contrast relative to the bars. RAM. Random Access Memory. Memory that is lost after power is recycled to the unit. Read Cycle. A programmed period of time or condition during which the scanner will accept bar code label input. Read Range. The distances in which a label can be reliably read, as measured from the front of the scanner. See “Depth of Field.” Relay Driver. A TTL signal sent by the scanner. The output is determined by the relay driver selection during configuration. ROM. Read Only Memory. Memory that cannot be changed. Scanner. A scanning device that is comprised of a scan head and a decoder integrated in one package. Scan Rate. Number of scans per second that the scanner projects. Scan Width. That portion of the scan line in which a label can be read. Serial Commands. On-line data strings (including configuration and operations) from a host or other terminal to the scanner, that are always preceded by a < left angle bracket symbol and followed by a > right angle bracket symbol.

MS-610 Scanner User’s Manual

A-21

Appendices

Serial Configuration (Host Configuration). Serial command specifically for changing configuration and distinguished from operational command by the fact that they modify the non-volatile for power up configuration. Most of these commands begin with the upper case K character immediately following the left angle bracket symbol, as in <Kg3>.

Page 96

Appendices Skew. Label (or scanner) rotation around the center of the skew axis. Specular Reflection. The direct, mirror-like reflection of laser light back to the scanner, causing over-light saturation. For the MS-610 scanner, specular reflection occurs at pitch angles between +3 and +5 degrees. Symbology. A set of bar code symbols, such as Code 39 or Code 128, that have special rules to define the widths and positions of bars and spaces to represent specific numeric or alphanumeric information. Tilt. Label (or scanner) rotation around the centerline of the scan beam. Timeout. A user-selected period of time that ends a scanner’s read cycle. Tracking. Parameter of analog-to-digital converter that affects the accuracy to the digitized signal. Trigger. A signal, either external or serial, that initiates the read cycle and causes the scanner to expect label input. TTL. Abbreviation for transistor-transistor logic, the standard for signal output.

Appendices

Wide-to-narrow Ratio. The ratio of the width of the widest (or wider) bar to the narrowest bar of a given bar code symbology. For example, 2:1, 2:5:1, 3:1, and 4:1.

A-22

MS-610 Scanner User’s Manual

Page 97

Index Numerics

A ACK-NAK Protocol 2-7 ADP A-3 Angled Picket Fence number of scans calculation 1-5 orientation A-7 Approvals vi As Soon As Possible by menu selection 2-11, 2-16 by serial command 3-9 ASCII table (with control characters) A-5 terminal communications 1-8, 2-1 Autodiscriminate (definition) A-19 Autodiscriminate All Codes 4-4

B Bar Code Output by menu selection 2-11, 2-16 by serial command 3-9 Bar Code Symbology A-13 Baud Rate by menu selection 2-5, 2-9 by serial command 3-7 Baud Rates A-2 Beeper (status, volume, speed) by menu selection 2-26–2-27 by serial command 3-12

C Cabling ground loops 1-16 noise interference 1-16 pin assignments 1-6 shield considerations 1-16

MS-610 Scanner User’s Manual

I-1

Index

15-pin Connector A-3 20372 5HeadingSp FIS Options A-4

Carriage Returns (as characters) 2-4 Caution Summary vii Check Digit of code 39 (menu) 2-19, 2-20 of code 39 (serial) 3-10 of interleaved 2 of 5 (menu) 2-20, 2-23 of interleaved 2 of 5 (serial) 3-11 Check Digit Output of code 39 (menu) 2-20, 2-21 of code 39 (serial) 3-10 of interleaved 2 of 5 (menu) 2-20, 2-23 of interleaved 2 of 5 (serial) 3-11 Checksum of EPROM (display) 4-5 Clock Speed by menu selection 2-26, 2-28 by serial command 3-12 Codabar by menu selection 2-20, 2-22 by serial command 3-10 definition A-13 operational command 4-4 Code 128 by menu selection 2-20, 2-25 by serial command 3-11 definition A-13 Code 39 3-10 by menu selection 2-19–2-21 by serial command 3-10 definition A-13 operational command 4-4 Code Length of codabar (menu) 2-20, 2-22 of codabar (serial) 3-10 of code 128 (menu) 2-20, 2-25 of code 128 (serial) 3-11 of code 39 (menu) 2-20, 2-21 of code 39 (serial) 3-10 Code Length #1 (I 2 of 5) by menu selection 2-20, 2-23 by serial command 3-11 Code Length #2 (I 2 of 5) by menu selection 2-20, 2-23 by serial command 3-11

Page 98

Index

Index Code Types by menu selection 2-19–2-25 by serial command 3-10–3-11 listing A-2 menu structure 2-20 operational commands 4-4 See also Codabar, Code 128, Code 39, Interleaved 2 of 5, Narrow Margins, and UPC/EAN Command Start Character by menu selection 2-26, 2-29 by serial command 3-13 Communications by menu selection 2-5–2-10 by serial command 3-5–3-8 default settings A-2 menu structure 2-5 See also Host Protocol Parameters, Protocol, Preamble, Postamble, Longitudinal Redundancy Check, Response Timeout, Intercharacter Delay, Host Port Parameters, Baud Rate, Parity, Stop Bits, Data Bits RS422 Concentrator A-15 Connector. See I/O Connector Continuous Read by menu selection 2-11, 2-12 by serial command 3-8 Continuous Read 1 Output by menu selection 2-11, 2-12 by serial command 3-8 Controller A-15 Counter Operational Commands 4-4

D Data Bits A-2 by menu selection 2-5, 2-10 by serial command 3-7 Defaulting the Scanner 2-4, 3-4, A-8 Defining Special Characters 2-4 Device Control Operational Commands 4-3

I-2

E Electrical Characteristics A-4 End of Read Cycle by menu selection 2-11, 2-14 by serial command 3-8 when to output option (menu) 2-11, 2-16 when to output option (serial) 3-9 External Edge by menu selection 2-11, 2-12 by serial command 3-8 External Level (triggering mode) by menu selection 2-11, 2-12 by serial command 3-8 External Trigger Level by menu selection 2-11, 2-15 by serial command 3-9

F FIS (final instruction sheet) A-4 Fixed Code Length of codabar (menu) 2-20, 2-22 of codabar (serial) 3-10 of code 128 (menu) 2-20, 2-25 of code 128 (serial) 3-11 of code 39 (menu) 2-20, 2-21 of code 39 (serial) 3-10 Full Screens 2-26, 2-27

G Glossary A-19 Good Decode Reads by menu selection 2-11, 2-17 by serial command 3-9 Good Read 2-28 Good Read Counter Commands. See Match Counter Commands Ground and Shield Considerations 1-16

H Host Port Parameters by menu selection 2-5, 2-9

MS-610 Scanner User’s Manual

Page 99

I I 2 of 5. See Interleaved 2 of 5 I/O Connector A-2 Installation 1-13 Intercharacter Delay by menu selection 2-5, 2-9 by serial command 3-7 Intercharacter Gap A-19 Interface A-2 Interface Standards A-14 Interleaved 2 of 5 by menu selection 2-20, 2-23 by serial command 3-11 definition A-13 operational command 4-4 of code 39 (serial) 3-10 Laser (enable/disable commands) 4-3 Laser On/Off by menu selection 2-26, 2-27 by serial command 3-12 Light Source A-2 Longitudinal Redundancy Check by menu selection 2-5, 2-8 by serial command 3-7 LRC. See Longitudinal Redundancy Check

M

L Label angled picket fence orientation A-7 density 1-11 dimensions 1-11 height 1-5 how to position 1-9 length 1-5 ratio 1-11 speed 1-5, 1-11 Label contrast A-3 Label Dimensions 1-11 Label Ratio 1-11 Ladder number of scans calculation 1-4 Ladder Orientation A-6 Large Intercharacter Gap of codabar (menu) 2-20, 2-22 of codabar (serial) 3-10 of code 39 (menu) 2-20, 2-21

Master Label Operational Commands 4-5–46 Match 2-28 Match Code by menu selection 2-11, 2-17–2-18 by serial command 3-9 definition A-20 operational command (enter and exit) 4-5 Match Counter Commands 4-4 Match Counter Reset 4-4 Mechanical Specifications A-2 Menu Configuration Program code types menu 2-19–2-25 communications menu 2-5–2-10 entering 2-2, 4-3 exiting 2-4 loss of communications/when to default 2-4 main menu 2-2 operations menu 2-11–2-18 saving menu changes 2-4 user outputs menu 2-26–2-29 Mirror Type A-2 Mismatch 2-28 Mismatch Counter Commands 4-4 Mismatch or Noread 2-28 Modulus 43 2-19 Mounting Holes 1-13 Mounting Plate Kit A-11 MS-610 Adapter plate A-12

MS-610 Scanner User’s Manual

I-3

Index by serial command 3-7 Host Protocol Parameters by menu selection 2-5, 2-6–2-9 by serial command 3-5–3-7 Host Relay Driver Pulse 4-3

Page 100

Index

Index Multidrop by menu selection 2-5, 2-7 by serial command 3-5 Multidrop Address A-18 Multidrop Communications A-15–A-17 multidrop addresses A-15 polling sequences A-15 select sequences A-16

N Narrow Margins by menu selection 2-19, 2-20 by serial command 3-10 Narrow–bar–width A-20 New Master Pin by menu selection 2-26, 2-27 by serial command 3-12 New Trigger by menu selection 2-11, 2-14 by serial command 3-8 Noise Interference 1-16, A-14 Noread 2-28 Noread Counter Commands 4-4 Noread Message by menu selection 2-11, 2-16 by serial command 3-9 Nulls (as characters) 2-4 Number of Scans 1-4

O Object Detector 1-14 Operating Environment A-2 Operating Life A-2 Operational Commands ??–4-6 Operations by menu selection 2-11–2-18 by serial command 3-8–3-9 menu structure 2-11 See also Triggering Mode, End of Read Cycle, Timeout in 10 ms incs, Serial Trigger Character, External Trigger Level, Noread Message, Bar Code

I-4

Output, When to Output, Good Decode Reads, Match Code Optical Characteristics A-3 Optional Equipment A-11 Output Wavelength A-2

P Parity A-2 by menu selection 2-5, 2-9 by serial command 3-7 Physical Specifications A-2 Picket Fence number of scans calculation 1-4 Picket Fence Orientation A-6 Pinouts 1-6, A-2 Pitch 1-9, A-3 Point-to-Point by menu selection 2-5, 2-6 by serial command 3-5 Point-to-Point with RTS/CTS by menu selection 2-5, 2-6 by serial command 3-5 Point-to-Point with RTS/CTS & XON/XOFF by menu selection 2-5, 2-6 by serial command 3-5 Point-to-Point with XON/XOFF by menu selection 2-5, 2-6 by serial command 3-5 Polarity. See External Trigger Level and Relay Polarity Poll Character A-18 Polling Mode D by menu selection 2-5, 2-6 by serial command 3-5 Polling Reset A-16 Postamble by menu selection 2-5, 2-8 by serial command 3-7 Power LED A-4 power supplies 1-6 Power Supply Requirements A-4 Preamble

MS-610 Scanner User’s Manual