Page 111
Read Cycle
New Trigger Usage: Definition:
New Trigger is an effective way to end a read cycle when objects move past the reader at irregular intervals (not timing-dependent). New Trigger ends the current read cycle and initiates a new one when a new trigger occurs. New Trigger refers only to a rising edge trigger. With External Edge, Serial Data, or Serial Data or External Edge enabled, an edge or serial trigger ends a read cycle and initiates the next read cycle. In the case of External Level, a falling edge trigger ends the read cycle but the next read cycle does not begin until the occurrence of the next rising edge trigger.
Timeout or New Trigger Usage: Definition:
Useful in applications that require an alternative way to end the read cycle. For example, if an assembly line should stop completely or the intervals between objects are highly irregular. Timeout or New Trigger is identical to Timeout except that a timeout or a new trigger (whichever occurs first) ends the read cycle.
Last Frame Usage: Definition:
Useful in applications in which the number of captures needed can be defined but the timeout duration varies. Last Frame only applies to Rapid Capture Mode.
Last Frame or New Trigger Usage: Definition:
Useful in applications in which line speeds are irregular and a new labelled object could appear before the last frame in a Rapid Capture sequence. Last Frame or New Trigger is identical to New Trigger except that a new trigger or last frame (whichever occurs first) ends the read cycle.
Read Cycle Timeout Definition: Serial Cmd: Default: Options:
Read Cycle Timeout is the duration of the read cycle. <K220,end of read cycle,read cycle timeout> 200 (x10mS) 1 to 65,535
MS-4 Imager User Manual
5-15
Page 112
Capture Mode
Capture Mode Definition: Serial Cmd: Default: Options:
Capture Mode relates to the way that images are captured and processed. <K241,capture mode,number of captures,rapid capture mode> Continuous Capture 0 = Rapid Capture 1 = Continuous Capture
Rapid Capture Definition:
In a rapid capture mode, one or multiple captures (as many as 32) can be taken at an interval specified by the time-between-captures parameter. In this mode, the only limiting time factor is integration and transfer timing.
Continuous Capture Usage: Definition:
Continuous Capture is useful in applications with slower line speeds or where symbol spacing may be random or not time-dependent. In Continuous Capture Mode, image captures are taken throughout the read cycle in a multi-buffered format (see diagram below). The Imager begins processing the first captured image at the same time that it captures the second image. Captures will continue occurring throughout the read cycle until an end condition occurs, such as a timeout, a new trigger, the last frame in a capture sequence, or a combination of the above. Capture Start of Read Cycle
1 2 3
4
5
End of read cycle
Processing time
5-16
MS-4 Imager User Manual
Page 113
Read Cycle
Number of Captures Usage: Definition: Serial Cmd: Default: Options:
Used to increase the opportunities for good reads and to “extend” the field of view in dynamic applications. Sets the total number of captures that are processed during a read cycle in Rapid Capture Mode when Switching Mode is set to Number of Captures. <K241,capture mode,number of captures,rapid capture mode> 1 1 to 32
Note: The range of maximum number of captures is dynamic. This range is dependent on the maximum image size in the system. A full-size image (640 x 480) reduces the maximum number of images to 6. The smaller the image size, the greater the maximum number of captures. Once the image size is reduced to small enough dimensions, the maximum number of captures will be capped at 32. If a user enters a maximum capture value greater than that allowed, the value will be limited to the number of system images. This command also affects the number of stored images allowed in the system. If the maximum number of captures is selected, the number of stored images allowed will be 0.
Rapid Capture Mode Definition: Serial Cmd: Default: Options:
In Rapid Capture Mode, one or multiple captures (as many as 32) can be taken at an interval specified by the time-between-captures parameter. In this mode, the only limiting time factor is integration and transfer timing. <K241,capture mode,number of captures,rapid capture mode> Timed Capture 0 = Timed Capture 1 = Triggered Capture
MS-4 Imager User Manual
5-17
Page 114
Timed Capture Usage: Definition:
Timed Capture is useful in fast-moving applications in which symbols are only in the field of view for a short time and precise timing is required. In Timed Rapid Capture, decoding occurs independent of and simultaneous with capturing, thus allowing precise timing or no delay at all between captures. Also, consecutive captures are regarded as the same symbol if the output data is the same. No Time Delay Between Captures
Start of Read Cycle
Diagram A n
End of read cycle
Time Before First Capture Time Delay Between Captures End of read cycle
Diagram B Processing
5-18
MS-4 Imager User Manual
Page 115
Read Cycle
Triggered Capture Usage: Definition:
Useful in applications where each decode must be treated as a discrete event, regardless of symbol data. When Rapid Capture Mode is set to Triggered Capture, each decode is considered a separate event, even if the current decode data is identical to the previous decode data. The first trigger event starts the read cycle, and subsequent triggers will continue until the predetermined Number of Captures is met, or until the predetermined End of Read Cycle condition is met-- whichever occurs first. Note: If End of Read Cycle is set for New Trigger and the read cycle qualifications have not been met, the read cycle will only end once it receives the first trigger after reaching the predetermined Number of Captures setting. First trigger starts the read cycle
Captures on every trigger
Capture Read cycle ends on New Trigger or Timeout, as configured.
Processing
MS-4 Imager User Manual
5-19
Page 116
Capture Timing
Capture Timing Note: Capture Timing applies only to Rapid Capture Mode.
Time Before First Capture Usage: Definition: Serial Cmd: Default: Options:
In almost any moving line application, a time delay is needed to ensure that a symbol will be in the Imager’s field of view at the beginning of the capture sequence. Time Before First Capture in a moving line application is the time between an external trigger event and the occurrence of the first capture. <K242,time before 1st capture,time between capture 1 and capture 2,,,,,,,,time between capture 7 and capture 8> 0 0 to 65535 (2.097 seconds, in 32 µS increments)
Start of Read Cycle
Time Before First Capture Processing
5-20
MS-4 Imager User Manual
Page 117
Read Cycle
Time Between Captures This is useful in applications where more than one symbol can appear during a single read cycle (multisymbol), or where line speeds are slow enough that captured frames might overlap or miss a symbol. A time delay can be inserted between individual frame captures in Rapid Capture Mode. <K242,time before 1st capture,time between captures [time 1,time 2,...time7] Entering 0s will result in no time between captures. Entering a different value in each field will vary the time delays accordingly. Note: You must enter time values along with comma separators for each field you want to change. If you omit fields, or enter only commas, the fields will remain as previously set. 0 0 to 65535 (2.097 seconds, in 32 µS increments)
Usage: Definition: Serial Cmd:
Default: Options:
Note: Number of Captures and number of frame delays (Time Between Captures) must be the same.
Start of Read Cycle
Diagram A
Time Delay Between Captures = 0
1 2 3 4 5
End of Read Cycle
Time Delay Between Captures varies.
Diagram B
MS-4 Imager User Manual
5-21
Page 118
Image Processing Timeout
Image Processing Timeout Caution: Image Processing Timeout, if not properly set, can have a negative impact on good reads. Usage: Definition:
Serial Cmd: Default: Options:
Useful in higher speed applications where image processing time is long enough that not all captures have an opportunity to be processed. Specifies the maximum amount of time to process a captured image. When the timeout expires, the image processing is aborted. This timeout works in both Rapid Capture and Continuous Capture modes, as well as with the Configuration Database. <K245,image processing timeout> 5,000 mS (5 sec.) 1 to 65535 (in 1mS increments)
Notes: • The timeout period does not include capture time. • If a timeout occurs during processing and no symbols in the field of view have been decoded, the image will be recorded as a No Read. For this reason, a longer timeout should be tried to ensure that the symbol is decoded successfully.
5-22
MS-4 Imager User Manual
Page 119
Read Cycle
Image Storage Image Storage Type Definition:
Serial Cmd: Default: Options:
Allows the user to store images from separate read cycles and to retrieve them later. The number of available slots for storage depends on the mode of operation. If the Imager is in Rapid Capture Mode, the number of images that can be stored is equivalent to the maximum number of the rapid count (the current rapid count setting). If the Imager is in Continuous Capture Mode, a number of images equivalent to the maximum number of the rapid count minus 3. <K244,image storage type,image storage mode> Disabled/Clear 0 = Disabled/Clear 1 = Store on No Read
Disabled / Clear Upon selection of this option, all saved images will be cleared and the Imager will not store images for later viewing.
Store on No Read This option will cause the Imager to store an image upon exiting the read cycle for retrieval at a later time. If multiple captures are present during the duration of a read cycle, the stored image will be the last image processed for that read cycle. This image is stored in RAM and can be retrieved as long as power is cycled to the Imager, and as long as the Imager has not been reset via a reset/save sequence. Other commands that can initialize storage in RAM are ones that change capture modes or put the Imager in a test capture mode.
Image Storage Mode Serial Cmd: Default: Options:
<K244,image storage type,image storage mode> First Mode 0 = First Mode 1 = Last Mode
First Mode This mode allows the Imager to store images until the available image memory has been filled. At that point the Imager will stop saving additional images. In this mode, you will always have the first image captured, because the saving process stops once memory has been filled.
Last Mode In this mode, image storage continues after available memory limits are reached. The oldest image in memory is overwritten, so you will always have the most recent stored image.
MS-4 Imager User Manual
5-23
Page 120
Image Storage
Image Storage Example
The following example assumes that the Imager is in a rapid capture mode of 3 captures. Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
1 No Read No Read Good Read, Symbol # 1 Good Read None
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
1 No Read No Read No Read No Read Frame # 3
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
2 No Read No Read Good Read, Symbol # 1 No Read Frame # 2
5-24