Page 191
I/O Parameters
Image Output Usage: Definition:
Serial Cmd: Default: Options:
This feature allows the user to select what read cycle condition triggers the image file output. Used to output an image file when a specified read cycle condition has been met, such as No Read or Good Read. When the read cycle is complete, the normal read cycle output will occur, including any configured postamble or preamble data. The read cycle will the check if image output is enabled, and if the specified read cycle condition has been met. If both these conditions are true, the image file will be output to the user-specified communication port in the user-specified image file format. <K739,image output mode,communication port,file format,JPEG quality> Disabled 0 = Disabled 1 = Good Read 2 = No Read 3 = No Read or Good Read 4 = Mismatch
Good Read If a Good Read condition is met during the read cycle, the image file corresponding with the first Good Read will be output. The image file will immediately follow the read cycle output and postambles. For a Good Read condition to occur, all symbols in the read cycle must qualify.
No Read If a No Read condition occurs during the read cycle, the image file corresponding with the first No Read will output. The image file will immediately follow the read cycle output and postambles. There are conditions in which the read cycle may have a No Read condition but the image will not be output. For example, if the read cycle is looking for two No Read symbols in the same image capture but one is a Good Read, the image will not be output, because it is not considered a No Read image.
No Read or Good Read If a No Read condition or a Good Read condition occurs, the first image of either kind will be output.
Mismatch If a Good Read occurs during the read cycle but the image is a Mismatch condition, the first Mismatch image will be output.
MINI Hawk Xi High Performance Imager User Manual
7-37
Page 192
Image Output
Communication Port Definition: Serial Cmd: Default: Options:
This setting selects the communication port through which the image file will be transferred to the host. <K739,image output mode,communication port,file format,JPEG quality> Host 0 = Host 1 = TCP Port 1 2 = TCP Port 2
File Format Definition:
Serial Cmd: Default: Options:
This setting selects the format of the image file to be output. Note: If JPEG is the chosen image file format, the resolution of the image will be determined by the setting in the JPEG Quality field. <K739,image output mode,communication port,file format,JPEG quality> JPEG 0 = Bitmap 1 = JPEG 2 = Binary
JPEG Quality Definition: Serial Cmd: Default: Options:
7-38
This setting determines the quality of JPEG image compression. The user may select a value from 1 (lowest) to 100 (highest). <K739,image output mode,communication port,file format,JPEG quality> 90 1 to 100 (percent)
MINI Hawk Xi High Performance Imager User Manual
Page 193
I/O Parameters
Image Output Example The following example assumes that the imager is set to output on No Read, and that it is in a rapid capture mode of 3 captures. Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Output Frame:
1 No Read No Read No Read No Read Frame # 1
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Output Frame:
1 No Read No Read Good Read Good Read None
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Output Frame:
2 No Read No Read Good Read, Symbol # 1 No Read Frame # 1
MINI Hawk Xi High Performance Imager User Manual
7-39
Page 194
Database Identifier Output
Database Identifier Output Usage: Serial Cmd: Default: Options:
Useful in keeping track of which database entries read which symbols. <K759,status,separator> 0 0 = Disabled 1 = Enabled
Output Status Disabled When this command is disabled, no database identifier information will be output.
Enabled When this command is enabled, the imager will append a two-digit number and the characters “DB” to the data output following the separator for each symbol decoded using Configuration Database. For example, if the separator is an underscore character and the second database entry reads a symbol encoded with “data capture” during the read cycle, the symbol data output will be “data capture_DB02”. If the database is not active, no identifiers will be attached to output.
Separator Usage: Serial Cmd: Default: Options:
The separator character separates the symbol data from the database identifier. <K759,status,separator> <space> Any ASCII character.
Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K759h,,3C> For >: <K759h,,3E> For , : <K759h,,2C>
7-40
MINI Hawk Xi High Performance Imager User Manual
Page 195
I/O Parameters
Quality Output Separator Definition: Serial Cmd: Default: Options:
The separator character separates quality output data from symbol data. <K704,separator,decodes per trigger status> , Any ASCII character.
Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K704h,3C> For >: <K704h,3E> For , : <K704h,2C>
Decodes per Trigger Status Definition:
Serial Cmd: Default: Options:
When this feature is enabled, the imager enters a state where it processes frames as configured until the end of a read cycle, with or without a successful symbol decode. When the read cycle ends, the imager outputs any decoded symbol data along with the decodes per trigger count. <K704,separator,decodes per trigger status> 0 0 = Disabled 1 = Enabled
MINI Hawk Xi High Performance Imager User Manual
7-41
Page 196
Configuring EZ Trax Output
Configuring EZ Trax Output After you have set up your imager for EZ Trax, you can configure the software in three different ways: using the EZ Trax interface, using the tree controls in ESP, and using serial commands. For detailed information about how to use the EZ Trax user interface for configuration, refer to the Help menu in EZ Trax software, or the EZ Trax Quick Start Guide, available on the Microscan Tools Drive.
Configuring EZ Trax Output by ESP To configure EZ Trax using ESP, first click the I/O tab in App Mode to display the I/O Parameters tree control.
7-42
MINI Hawk Xi High Performance Imager User Manual
Page 197
I/O Parameters
EZ Trax Output When EZ Trax output is enabled, the imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled.
Comm Port Selects the communication port that will be used to transfer image files.
Image Mode
Determines which read cycle condition triggers image file output.
Image Format
Selects the format of the image output file.
MINI Hawk Xi High Performance Imager User Manual
7-43
Page 198
Configuring EZ Trax Output
JPEG Quality Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality.
Object Info Output
If a symbol is decoded during an image capture, the symbol type and other supplementary symbol information is also included.
Grade Output
Symbol quality information will be output in the form of an abbreviated identifier for each read cycle.
7-44
MINI Hawk Xi High Performance Imager User Manual
Page 199
I/O Parameters
Configuring EZ Trax Output by Serial Command Usage: Definition: Serial Cmd: Default: Options:
Useful for when you want to configure EZ Trax output using serial commands instead of the EZ Trax graphic interface. When EZ Trax output is enabled, the imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Enabled
Comm Port Definition: Serial Cmd: Default: Options:
Selects the communication port that will be used to transfer image files. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Host 0 = Host 1 = TCP Port 1 2 = TCP Port 2
Host Port Sends output using the current Host Port.
Image Mode Definition: Serial Cmd: Default: Options:
Determines which read cycle condition triggers image file output. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Good Read 2 = No Read 3 = No Read and Good Read 4 = Mismatch
Good Read Outputs the first Good Read image in the read cycle. The image file will immediately follow the symbol data.
MINI Hawk Xi High Performance Imager User Manual
7-45
Page 200
Configuring EZ Trax Output
No Read Outputs the first No Read image in the read cycle. The image file will immediately follow the symbol data. Note: If two symbols are present in the same image capture and one of them is successfully decoded, no image will be output.
No Read and Good Read Outputs the first No Read or Good Read image in the read cycle.
Mismatch Outputs the image of the first mismatch condition.
Image Format Definition: Serial Cmd: Default: Options:
Selects the format of the image output file. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Compressed 0 = Lossless 1 = Compressed
Lossless A “lossless” image is one that is in a high resolution format, and that will not lose resolution even when manipulated and saved multiple times. When Lossless is enabled, the JPEG Quality setting has no effect.
Compressed Outputs the image in JPEG format. Image quality is determined by the JPEG Quality setting.
JPEG Quality Definition: Serial Cmd: Default: Options:
7-46
Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> 90 1 to 100 (percent)
MINI Hawk Xi High Performance Imager User Manual