Page 191

I/O Parameters

1D/Stacked Symbology Qualification Note: 1D/Stacked Symbology Qualification is not supported when Q-Mode is enabled.

Minimum Number of Bars Definition: Serial Cmd: Default: Options:

Sets the minimum number of bars to qualify linear symbols. <K717,minimum number of bars,minimum number of qualified scan lines,start/stop status> 0 0 to 255 (0 = disabled)

Minimum Number of Qualified Scan Lines Definition:

Serial Cmd: Default: Options:

Sets the minimum number of scans that have the required number of bars set in Minimum Number of Bars. Note: This is a percentage value, from 0 to 100. <K717,minimum number of bars,minimum number of qualified scan lines,start/stop status> Disabled 0 to 100 (0 = disabled).

Start/Stop Status Definition:

Serial Cmd: Default: Options:

When enabled, both Start and Stop characters need to be present to qualify as a decodable symbol. In the case of PDF417, only a Stop needs to be present. Note: This parameter does not apply to UPC, GS1 DataBar (RSS), MicroPDF417, or Pharmacode. <K717,minimum number of bars,minimum number of qualified scan lines,start/stop status> Disabled 0 = Disabled 1 = Enabled

Quadrus MINI Imager User’s Manual

7-11

Page 192

2D Symbology Qualification

2D Symbology Qualification Note: 2D Symbology Qualification is not supported when Q-Mode is enabled. Note: In the cases where Symbol Size Status, Dimension Status, or Orientation Status is enabled, the imager will always check the finder pattern regardless of Finder Patter Status setting.

Finder Pattern Status Definition: Serial Cmd: Default: Options:

Checks for the presence of a finder pattern. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> Disabled 0 = Disabled 1 = Enabled

Symbol Size Mode Note: If Symbol Size 1 is larger than Symbol Size 2, it will be automatically reversed in the algorithm. Definition: Serial Cmd: Default: Options:

Specifies the outputs resulting from the searches for Symbol Size 1 and Symbol Size 2. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> Disabled 0 = Disabled 1 = Bad/No Symbol Output 2 = Object Qualification 3 = Enable Both

Disabled Decoded messages are output, but no attempt to qualify symbols or output Bad or No Symbol messages is undertaken.

Bad/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol size. If both Symbol Size 1 and Symbol Size 2 requirements are met, a Bad Symbol message is output. If neither are met, a No Symbol message is output.

Object Qualification Checks first for matches to symbol size. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.

Enable Both Same as Object Qualification, except that if no objects qualify, then a No Symbol message is output; if at least one object qualifies, then a Bad Symbol message is output. 7-12

Quadrus MINI Imager User’s Manual

Page 193

I/O Parameters

Symbol Size 1 Definition: Serial Cmd:

Default: Options:

Checks for symbol sizes for a specified number of elements. In the case of rectangular symbols, checks the longer side. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 10 8 to 88 elements

Symbol Size 2 Definition: Serial Cmd:

Default: Options:

Checks for symbol sizes for a specified number of elements. In the case of rectangular symbols, checks the shorter side. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 10 8 to 88 elements

Symbol Size Tolerance Definition: Serial Cmd:

Default: Options:

Sets the allowable deviation, up or down, for symbol sizes specified in Symbol Size 1 and Symbol Size 2. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 2 0 to 10

Quadrus MINI Imager User’s Manual

7-13

Page 194

2D Symbology Qualification

Dimension Mode Definition: Serial Cmd:

Default: Options:

Enables searches for symbol Dimension 1 and Dimension 2. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> Disabled 0 = Disabled 1 = Bad/No Symbol Output 2 = Object Qualification 3 = Enable Both

Disabled Decoded messages are output, but no attempt to qualify symbols or output Bad Symbol or No Symbol messages is undertaken.

Bad/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol dimension. If both Dimension 1 and Dimension 2 requirements are met, a Bad Symbol message is output. If neither are met, a No Symbol message is output.

Object Qualification Checks first for matches to symbol dimension. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.

Enable Both Same as Object Qualification, except that if neither Dimension 1 nor Dimension 2 qualifies, then a No Symbol message is output; if either Dimension 1 or Dimension 2 qualifies, then a Bad Symbol message is output.

7-14

Quadrus MINI Imager User’s Manual

Page 195

I/O Parameters

Dimension 1 Definition: Serial Cmd:

Default: Options:

Checks symbol dimension for a specified number of pixels. In the case of rectangular symbols, checks the longer side. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 0 0 to 1280 pixels

Dimension 2 Definition: Serial Cmd:

Default: Options:

Checks symbol dimension for a specified number of pixels. In the case of rectangular symbols, checks the shorter side. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 0 0 to 1024 pixels

Dimension Tolerance Definition: Serial Cmd:

Default: Options:

Sets the allowable deviation, in percentage, for symbol sizes specified in Dimension 1 and Dimension 2. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 10 0 to 100

Quadrus MINI Imager User’s Manual

7-15

Page 196

2D Symbology Qualification

Orientation Mode Definition: Serial Cmd:

Default: Options:

Sets the orientation value. <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> Disabled 0 = Disabled 1 = Bad/No Symbol Output 2 = Object Qualification 3 = Enable Both

Disabled Decoded messages are output, but no attempt is made to qualify symbols or to output Bad Symbol or No Symbol messages.

Bad/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol orientation. If Orientation Value is satisfied, a Bad Symbol message is output. If not satisfied, a No Symbol message is output.

Object Qualification Checks first for matches to symbol Orientation Value. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.

Enable Both Same as Object Qualification, except that if Orientation Value is satisfied but the symbol is not decoded, then a Bad Symbol message is output; if not satisfied, a Bad Symbol message is still output.

7-16

Quadrus MINI Imager User’s Manual

Page 197

I/O Parameters

Orientation Value Usage: Definition:

Instructs the imager to ignore symbols not in the specified orientation. Sets the orientation of the symbol, in degrees clockwise from the default 0 orientation. “L” finder pattern For Data Matrix symbols, sets the (Data Matrix) finder “L” pattern, with the example to the right showing the default 0 orientation. For QR Code, sets the locator pattern (three squares), with the Three-square example to the right showing the finder pattern default 0 orientation. (QR Code) The other position settings will rotate clockwise from the defaults shown above.

Serial Cmd:

<K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> 0 0 to 359

Default: Options:

Note: A symbol’s orientation can miss the precise Orientation Value by several degrees and still be qualified. To be certain of the extent of the allowable tolerance, you should experiment with symbols in various orientations. In no case can they be more than 45° from the assigned orientation value and still be qualified.

Quadrus MINI Imager User’s Manual

7-17

Page 198

Read Duration Output

Read Duration Output Usage: Definition:

Useful in evaluating actual read cycle timing results, especially when initially setting up an application to determine maximum line speed (obtainable based on spacing between symbols.) When enabled the duration of the read cycle (in milliseconds) is appended to the symbol data. The read duration is the time from the beginning of the read cycle until data is output.

Read Duration Output Mode Serial Cmd: Default: Options:

<K706,status,separator> Disabled 0 = Disabled 1 = Enabled

Important: To measure the entire read cycle when in External Level trigger mode, set When to Output Symbol Data to End of Read Cycle. This output can measure over 49 days’ worth of duration; if exceeded, the “OVERFLOW” message will be output in place of the duration.

Read Duration Output Separator Definition: Serial Cmd: Default: Options:

User defined character that separates the symbol information from the Read Duration Output. <K706,status,separator> [space character] Any ASCII character.

Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K706h,,3C> For >: <K706h,,3E> For , : <K706h,,2C> See ASCII Character Entry Modifier on page 3-32 for a detailed explanation of how to enter ASCII characters as hex values.

7-18

Quadrus MINI Imager User’s Manual

Page 199

I/O Parameters

Output Indicators The Quadrus MINI Imager has a beeper and three LED arrays, as follows: 1. A target pattern of blue LEDs for centering the field of view, which is projected from the front of the imager . 2. An array of green LEDs projected from the front of the imager that can be programmed to flash in response to user-defined conditions. 3. A row of five status LEDs on the side of the imager.

Green Flash Mode Usage: Definition: Serial Cmd: Default: Options:

Used as a visual verification that a good read has occurred. An array of green LEDs in the front of the imager can be programmed to flash in response to user-defined conditions. <K750,green flash mode,target pattern status,green flash duration> Good Read 0 = Disabled 1 = Good Read 2 = Static Presentation 3 = Match 4 = Mismatch 5 = Strobe

Disabled Green flash LEDs are disabled.

Good Read Green flash LEDs will flash when a good read condition is met or when Matchcode is enabled and a match occurs.

Quadrus MINI Imager User’s Manual

7-19

Page 200

Output Indicators

Static Presentation Mode Static Presentation Mode is used in conjunction with Continuous Read Mode: <K200,0>. When operating in Static Presentation Mode, the red LEDs will illuminate while the imager is searching for a symbol in Continuous Read Mode. When a symbol is placed in the field of view and a good read occurs, the green LEDs will illuminate and stay on for the duration of time set in Green Flash Duration. Only one read will occur during that time unless more than one symbol is enabled in Number of Symbols. Note: If Static Presentation Mode is selected but the imager is not in Continuous Read, the Green Flash will not occur. To use Static Presentation: 1. Enable Continuous Read. 2. Select the number of symbols. 3. Enable Static Presentation in Green Flash Mode. 4. Select the read time in Green Flash Duration.

Match The green LEDs will flash when a match condition is met. If multisymbol is enabled, then green flash LEDs will illuminate only if all symbols qualify as a match. If matchcode is disabled, then this mode will activate the LEDs on a good read.

Mismatch Same as Match, except that LEDs will illuminate on a mismatch.

Strobe Green flash LEDs will act as an illumination strobe for image capture. If it is required that the green flash LEDs be the only illumination for image capture, then the internal illumination LEDs can be disabled.

7-20

Quadrus MINI Imager User’s Manual