Page 111
China Postal Code
Check Digit Mode Serial Cmd: Default: Options:
<K498,status,check digit mode,min length,max length> Disabled 0 = Disabled 1 = Enable, No Output 2 = Enable, Output
Bar Code Configuration Mode
Disabled (Default)
Enable, No Output
Enable, Output
Save Settings
5-54
HS-1 Handheld Scanner User’s Manual
Page 112
Symbologies
Minimum Symbol Length Serial Cmd: Default: Options:
<K498,status,check digit mode,min length,max length> 11 1 – 48
Bar Code Configuration Mode
11 (Default)
Save Settings
Maximum Symbol Length Serial Cmd: Default: Options:
<K498,status,check digit mode,min length,max length> 48 1 – 48
Bar Code Configuration Mode
HS-1 Handheld Scanner User’s Manual
48 (Default)
Save Settings
5-55
Page 113
PDF417
PDF417 Note: The HS-1 may make a “grinding” sound while attempting to decode PDF417 symbols. This is normal, and does not cause damage to the scanner. Usage:
Definition:
Serial Cmd: Default: Options:
Used in applications where a large amount of information (over 32 characters) needs to be encoded within a symbol, typically where the symbol is transported from one facility to another. For example, an automobile assembly line might use a single symbol with multiple fields of information that will be read at several stations along the way, without reference to a database. A two-dimensional, multi-row (3 to 90), continuous, variable length symbology that has high data capacity for storing up to 2,700 numeric characters, 1,800 printable ASCII characters, or 1,100 binary characters per symbol. Each symbol character consists of 4 bars and 4 spaces in a 17-module structure. <K476,status> Enabled 0 = Disabled 1 = Enabled
Bar Code Configuration Mode
Disabled
Enabled (Default)
Save Settings
5-56
HS-1 Handheld Scanner User’s Manual
Page 114
Symbologies
Symbology Identifier Definition:
Serial Cmd: Default: Options:
The Symbology Identifier is a standard prefix set of characters that identifies the symbol type. When enabled, the HS-1 analyzes and identifies the symbology and adds a three-character identifying prefix to the data: • ] (closed bracket character) indicating the presence of a symbology identifier. • A, C, E, F, G, I, e, P, M (A = Code 39; C = Code 128/EAN-128; E = UPC/EAN; F = Codabar; G = Code 93; I = Interleaved 2 of 5; e = GS1 DataBar (RSS); P = Plessey Code; M = MSI Code) • Modifier <K753,0,symbology identifier status> Disabled 0 = Disabled 1 = Enabled
Bar Code Configuration Mode
Disabled (Default)
Enabled
Save Settings
HS-1 Handheld Scanner User’s Manual
5-57
Page 115
Symbology Identifier
Explanation of Modifiers for Code 39, Codabar, and I-2/5 • For Code 39, Codabar, and Interleaved 2 of 5, the modifier indicates Check Character and Check Character Output status. • For Code 39 only, Full ASCII must be enabled to see modifiers 4, 5, and 7. Modifier
Check Character
0 1 3 4 5 7
Disabled Enabled Enabled Disabled Enabled Enabled
Check Character Output N/A Enabled Disabled N/A Enabled Disabled
Full ASCII conversion performed (Code 39 only) No No No Yes Yes Yes
Example: ]A5 indicates a Code 39 symbol with Check Character and Check Character Output enabled and Full ASCII conversion performed.
Explanation of Modifiers for Other Symbologies • For Code 128/EAN-128, a 1 indicates EAN-128; otherwise the modifier is 0. • For all other symbologies, the modifier is 0.
5-58
HS-1 Handheld Scanner User’s Manual
Page 116
6 Terminal Contents Terminal Window.......................................................................................................................... 6-2 Find .............................................................................................................................................. 6-3 Send ............................................................................................................................................. 6-4 Macros.......................................................................................................................................... 6-5 Terminal Window Menus.............................................................................................................. 6-6
This section describes the Terminal window and Macro functions in ESP.
HS-1 Handheld Scanner User’s Manual
6-1
Page 117
Terminal Window
Terminal Window Click this button to display the Terminal view. The following view will appear: Copy, paste, save
Type text here to find matching text in the Terminal window.
Type serial commands here, or directly in the terminal screen.
Clear screen Terminal screen
Click on desired Macro to run.
Macros listed on this bar.
Click on arrows to Add Macro, Remove Macro, or Edit Macro.
The Terminal allows you to send serial commands to the HS-1 by typing commands in the Send text field or directly in the Terminal screen, by copying and pasting, or by using Macros. The Terminal screen also displays symbol data or information from the HS-1. You can right-click on the screen to bring up a menu of several options for managing data in the Terminal view.
6-2
HS-1 Handheld Scanner User’s Manual
Page 118
Terminal
Find The Find function allows you to enter text strings to be searched for in the Terminal window. For example, data from a series of symbols has been sent to the Terminal and you want to determine if a particular symbol whose data begins with “ABC” has been decoded. • Type “ABC” into the Find field.
• Press Enter. The first instance of “ABC” will be highlighted in the Terminal window. • Click the Find button to the left of the text field to locate additional instances of “ABC”.
HS-1 Handheld Scanner User’s Manual
6-3
Page 119
Send
Send The Send function allows you to enter serial commands and then send them to the HS-1. • Type the command into the Send box.
• Press Enter to send the command to the HS-1. • Click the Send button to the left of the text field to send the command again.
6-4
HS-1 Handheld Scanner User’s Manual
Page 120
Terminal
Macros Macros can be stored in a macro selection bar, edited in a separate window, and executed by clicking on the macro name.
Click on Next Row to see the next row of macros
Click on the first arrow here to see Add Macro or Default Macros. When you default macros, the entire macros set is restored to their original macro commands.
Click on subsequent arrows to edit macros.
When you click on the macro name, the macro is executed in the Terminal window. If the macro is a command, it is sent to the HS-1 at the same time that it is displayed.
Editing a Macro When you click the arrow next to any macro and select Edit Macro, the following dialog appears:
You can edit an existing Macro Name or Macro Value by typing a new name or value in the corresponding text field and clicking OK.
HS-1 Handheld Scanner User’s Manual
6-5