Page 261
Diagnostic Test Menu r, Reset Unit The “r” command resets the unit via software reset.
Syntax >> r j, Jump to Application The “j” command jumps to the application code.
Syntax >> j e, Display Last Logged Error The “e” command displays the last error logged by the bootloader/diagnostics.
Syntax >> e h, Display Command Help The “h” command displays help for each of the commands in the Bootloader Menu.
6
>> h
Diagnostic Test Menu The Diagnostic Test Menu displays a list of tests that are currently supported for HawkEye™ 1500 diagnostic debug and verification development. Each test can be executed individually or as part of a group that is determined by attributes associated with the test.
v2.4.1, Nov 2008
HawkEye™ 1500 Series User Manual
6-13
The Bootloader
Syntax
Page 262
Chapter
6
The Bootloader
Note: The attributes are set internal to the software and cannot be modified at runtime. Test suites are grouped in increments of 10 starting with test 100. For example, bus tests are part of the suite numbered 100109, flash related tests are 110-119, and so forth. 10 All Tests 40 All Power-on Tests 100 All Bus Tests 103 Memory Cell 110 All Flash Tests 120 All RS232 Tests 130 All Ethernet Tests 133 Ethernet PHY Lpbk
20 All Mfg Tests
30 All Internal Tests
101 Data Line 104 Inverted Cell 112 Flash Bootloader 121 RS232 Internal 131 Ethernet Reset 134 Ethernet EXT Lpbk
102 Address Line 113 Flash Blocks 122 RS232 Interrupt 132 Ethernet MAC Lpbk 135 Ethernet Interrupt
Table 6–2 enumerates which tests are executed for groups 10 (“All Tests”), 20 (“All Mfg Tests”), 30 ( “All Internal Tests”), and 40 (“All Power-on Tests”). TABLE 6–2. The Tests That Are Executed
Test Name
All Tests
All Mfg Tests
All Internal Tests
All Power-on Tests
Data Line
X
X
X
X
Address Line
X
X
X
X
Memory Cell
X
X
X
Inverted Cell
X
X
X
Flash Bootloader
X
Flash Blocks
X
X
X
X
RS232 Internal Loopback
X
X
X
X
Ethernet Reset
X
X
X
X
Ethernet MAC Loopback
X
X
X
X
Ethernet PHY Loopback
X
X
X
X
Ethernet EXT Loopback
X
X
6-14
X
HawkEye™ 1500 Series User Manual v2.4.1, Nov 2008
Page 263
LEDS
LEDS The system LEDs visually convey power-on status and error codes.
Power-on Sequence Each stage of the power-on sequence drives the LEDs in a binary up-count fashion according to the table below. The LEDs are lit before the test is executed and remain in that pattern until the next test is run or an error condition is detected and displayed. TABLE 6–3. Power-on Sequence
Fail
Pass
Trig
Test
X
Data Line Test
X
Flash Bootloader Test
X X
Address Line Test
X
Flash Block Test
X
X
X
X
X
X
X
RS232 Internal Loopback Test Ethernet Reset Test
X
Ethernet MAC Loopback Test Ethernet PHY Loopback Test
Error Codes In the event of an error being detected, the system beeper will sound five times and an error code representing the test that failed will flash on the LEDs. The LEDs continue to flash until the QuicSet button is pressed, at which point an error message is logged to the serial port and the Diagnostic Monitor is launched.
v2.4.1, Nov 2008
HawkEye™ 1500 Series User Manual
6-15
6 The Bootloader
Mode
Page 264
Chapter
6-16
6
The Bootloader
HawkEye™ 1500 Series User Manual v2.4.1, Nov 2008
Page 265
A
APPENDIX A
HawkEye™ 1510
A HawkEye™ 1510
This appendix contains information specific to the HawkEye™ 1510 Smart Camera-Based Reader. FIGURE A–1. HawkEye™ 1510 v2.4.1, Nov 2008
HawkEye™ 1500 Series User Manual
A-1
Page 266
Appendix
A
HawkEye™ 1510
Optics The HawkEye™ 1510 uses the following lenses. •
CS Mount Lenses
•
C Mount Lenses (requires 5 mm extension (Microscan part number 928-0047-1)): CS mount + 5mm = C mount
TABLE A–1. Lens Sizes and Microscan Part Numbers
Part Number
Size
C-Mount Lenses (Requires 5mm extension: Microscan P/N: 928-0047-1) 928-0057-1
Lens: 8mm, F/1.4-16, Filter Thread: 25.5mm P 0.5mm
928-0057-2
Lens: 12mm, F/1.8-16, Filter Thread: 25.5mm P 0.5mm
928-0057-3
Lens: 16mm, F/1.4-16, Filter Thread: 25.5mm P 0.5mm
928-0057-4
Lens: 25mm, F/1.6-16, Filter Thread: 25.5mm P 0.5mm
928-0057-5
Lens: 35mm, F/2.1-22, Filter Thread: 25.5mm P 0.5mm
928-0057-6
Lens: 50mm, F/2.8-22, Filter Thread: 25.5mm P 0.5mm
928-0057-7
Lens: 75mm, F/3.9-32, Filter Thread: 25.5mm P 0.5mm
CS-Mount Lenses
A-2
928-0049-1
Lens: 2.8mm, F/1.3-C
928-0049-2
Lens: 4mm, F/1.2-C
928-0049-3
Lens: 8mm, F/1.2-C
928-0049-4
Lens: 16mm, F/1.4-C
HawkEye™ 1500 Series User Manual v2.4.1, Nov 2008
Page 267
Table A–2 contains working distances and fields of view with various extensions and lenses. TABLE A–2. Working Distance and Fields of View
1
Extension (mm)
8 mm
12 mm
16 mm
25 mm
35 mm
50 mm
75 mm
Horiz. FOV (mm) Camera Face Distance (mm)
5.0
108 203
91 248
81 305
38 234
36 328
36 495
28 582
Horiz. FOV (mm) Camera Face Distance (mm)
5.5
41 94
44 140
51 197
28 201
43 297
33 460
26 561
Horiz. FOV (mm) Camera Face Distance (mm)
6.0
28 74
34 107
38 157
29 193
29 282
32 439
25 546
Horiz. FOV (mm) Camera Face Distance (mm)
10.0
9 562
13 75
14 118
17 197
21 333
20 461
Horiz. FOV (mm) Camera Face Distance (mm)
15.0
7 61
9 94
11 159
14 269
15 395
Horiz. FOV (mm) Camera Face Distance (mm)
20.0
6 87
9 144
11 232
13 357
Lens Focal Length
Notes:
1 2
Lens focus ring adjusted to close position. Lens focus ring at mid point position.
External Lighting Mounting Options •
• v2.4.1, Nov 2008
HawkEye™ Light Mount Assembly (HELTMA) Brackets –
HELTMA-1S thru HELTMA-4S (Short - 6.5 inches (165.1mm))
–
HELTMA-1L thru HELTMA-4L (Long - 8.5 inches (215.9mm))
–
HELTMA-5 (30 Degree outrigger bracket)
–
HELTMA-6 (30 Degree outrigger assembly)
Adjustable brackets simplify mounting for Microscan NER lights –
DOAL 25
–
DOAL 50
–
DF-150-3…
HawkEye™ 1500 Series User Manual
A-3
A HawkEye™ 1510
External Lighting Mounting Options
Page 268
Appendix
A
HawkEye™ 1510
Figure A–2 shows light mount bracket HELTMA-1L with NER Doal 50 attached. FIGURE A–2. Light Mount Bracket HELTMA-1L with NER Doal 50
Figure A–3 shows light mount bracket HELTMA-1L with NER DF-150-3 dark field illuminator and mounting bracket attached.
A-4
HawkEye™ 1500 Series User Manual v2.4.1, Nov 2008
Page 269
FIGURE A–3. Light Mount Bracket HELTMA-1L with NER DF-150-3
TABLE A–3. HawkEye™ Light Mount Assembly Brackets
Part Numbers
Description
External Lighting Brackets for On-Axis Lights HELTMA-1L HELTMA-1S
Adjustment range: 156.7mm [6.17”] Adjustment range: 100.1mm [3.94”] Compatible with NER DF-100 and DF-150 series Dark-Field Illuminators. Note: DF-150-3 requires NER mounting bracket part number 090-000100
HELTMA-2L HELTMA-2S
Adjustment range: 156.7mm [6.17”] Adjustment range: 100.1mm [3.94”] Compatible with NER R-100 “V2” series Ring Illuminators.
HELTMA-3L HELTMA-3S
Adjustment range: 156.7mm [6.17”] Adjustment range: 100.1mm [3.94”] Compatible with NER R-60 “V2” series Ring Illuminators.
v2.4.1, Nov 2008
HawkEye™ 1500 Series User Manual
A HawkEye™ 1510
External Lighting Mounting Options
A-5
Page 270
Appendix
A
HawkEye™ 1510
TABLE A–3. HawkEye™ Light Mount Assembly Brackets (Continued)
Part Numbers
Description
HELTMA-4L HELTMA-4S
Adjustment range: 156.7mm [6.17”] Adjustment range: 100.1mm [3.94”] Compatible with NER Doal-50 “V2” series Diffused On-Axis Lights.
External Lighting Brackets for Off-Axis Lights HELTMA-5
30 Degree outrigger bracket Compatible with NER AR-50 9 Pin-D series Area Array Illuminators and BL-50 9 Pin-D series Backlights.
HELTMA-6
30 Degree outrigger assembly Includes HE1515 Bright Field Light. Requires external 12 volt power supply, compatible with NER 010-502400 (US cord), 010-502500 (EU cord) power supplies.
Power for Lights Options •
External Independent Power
•
Internal Power Connector –
Continuous – 12v, 0.5A •
–
Strobe – Use external power and strobe modules •
A-6
DOAL 25, DOAL 50, DF100-1 ONLY
When LIGHTING:EXTERNAL is selected TTL O/P 1 (Pin 5) becomes Strobe Trigger
HawkEye™ 1500 Series User Manual v2.4.1, Nov 2008