The text of a manual · part 35 of 44
EZ Match Imager User Manual, pages 341 to 350
Pages 341 to 350 of 431
The original document is the PDF (3.5 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan EZ Match
Page 341
Utilities
Master Symbol Database Size Definition: Serial Cmd: Default: Options:
Number of Master Symbols allows you to select 1 to 10 master symbols for the master symbol database. <K231,master symbol database size> Note: You must follow this command with a save command <A> or <Z>. 1 1 to 10
Set Master Symbol Database Size here.
Caution: Since the total number of characters available for the master symbol database is 3000, changes to the Master Symbol Database Size will re-allocate the number of characters available for each master symbol and could cause existing master symbols to be deleted (except master symbol #1, unless it also exceeds the size limitation). The table below specifies the maximum number of characters available to each symbol according the number of master symbols defined, from 1 to 10. Master Symbol Number #1 #2 #3 #4 #5
EZ Match User Manual
Maximum Characters 3000 1500 1000 750 600
Master Symbol Number #6 #7 #8 #9 # 10
Maximum Characters 500 428 375 333 300
14-11
Page 342
Master Database
Enter Master Symbol Data Allows you to enter master symbol data for any enabled master symbol index number (1 to 10), provided the total number of characters does not exceed the maximum allowed. Serial Cmd: <K231,master symbol number,master symbol data> Options: Enter data for 1 to 10 symbols (any combination of ASCII text up to the maximum allowed. For example, to enter data for master symbol 9, after making certain that master symbol database size is enabled for 9 or more symbols, send <K231,9,data>. Usage:
Important: The ASCII characters <, >, and , can only be entered as hex pairs. See ASCII Character Entry Modifier on page 4-33 for a detailed explanation of how to enter ASCII characters as hex values. ESP:
Caution: If no data is entered, the existing data will be deleted. 1. Open the Utilities menu. 2. Set the number of master symbols you want to create in Master Symbol Database Size. 3. Double-click on each symbol number you want to set up and copy, or type your data in the popup dialog and click OK.
4. When all your data has been entered, click the Send Database to the Reader button. 14-12
EZ Match User Manual
Page 343
Utilities
Request Master Symbol Data Definition:
Serial Cmd:
Returns master symbol data for any enabled master symbols from 1 to 10. For example, to request master symbol # 5, enter <K231?,5>. The imager transmits master symbol # 5 data in brackets in the following format: <5/>. If no master symbol data is available, the output will be: <5/>. <K231?,master symbol number> Caution: Be sure to add the ? or you will delete the master symbol. Note: This command returns the number of master symbols if no number is included.
ESP:
1. Click the Utilities button and the Master Database tab. 2. Click the Receive Reader’s Database button.
Request All Master Symbol Data Definition: Serial Cmd:
This command will return master symbol data for all symbols enabled (up to 10). <K231?>
EZ Match User Manual
14-13
Page 344
Read Next Symbol as Master Symbol Definition: Serial Cmd:
ESP:
After you’ve set the size in the database, you can order the imager to read the next symbol as the master symbol for any given master symbol number. <Gmaster symbol number> To store the next symbol decoded as master symbol # 1, send: <G> or <G1>. To store the next symbol decoded as the master symbol for any other master symbol database number, send: <Gmaster symbol number [1-10]>. For example, <G5> will cause the next symbol read to be entered as master symbol # 5. In the Master Database tab under the Output Format menu, 1. Select the master symbol index number in which you want to store the symbol data. 2. Click the Read Symbol into Selected Index button. Caution: If you’ve selected an index which already has existing data, that data will be copied over by new decoded data when you use this command.
Request New Master Status Usage: Definition: Serial Cmd:
Informs the user when a new master symbol is pending and which position it is in. Returns the position in the master symbol database that will be loaded on the next read. <NEWM> The imager returns: <NEWM/next master to load> Once a symbol has been read and loaded, the status will be cleared and the response will be <NEWM/0>.
(See also New Master Pin in Chapter 10, Matchcode.)
14-14
EZ Match User Manual
Page 345
Utilities
Delete Master Symbol Data Definition: ESP:
Serial Cmd:
You can directly delete the master symbol data by serial command or by ESP. 1. Click the Utilities button to access the master symbol. 2. Click the Master Database tab and double-click the symbol number you want to delete. 3. Delete text and click OK. <K231,master symbol number,> To delete a master symbol, enter the database number and a comma, but leave the data field empty. For example, to delete master symbol # 5, send the following: <K231,5,>. The command is entered with a blank master symbol data field, which tells the imager to delete the selected master symbol from the database.
EZ Match User Manual
14-15
Page 346
Firmware
Firmware Firmware by ESP
Firmware Update Application code versions are specific to your imager. Consult with your sales representative before downloading application code. If needed, an application code will be sent to you in the form of a *.mot file. To download application code: 1. First, be sure that your imager is connected to the host. 2. Apply power to the imager. 3. Before updating, you should verify the current firmware. 4. Click in the Firmware Update text box and select the file type you want to download. This will open a file locator box. 5. Navigate to the appropriate file (it will be a *.mot file), open the file, and click Start. Caution: Do not interrupt power or disconnect the host cable while download is in progress. Be sure that each download is complete before moving on to the next. Important: When updating firmware, be sure that the application code and boot code versions are mutually compatible.
14-16
EZ Match User Manual
Page 347
Utilities
Firmware Verification Request Part Number by ESP
You can send a request to the imager for part numbers, checksums, boot code, and application code. 1. Click the Firmware tab. 2. From the dropdown selection box to the left of the Request Part No. button, make your selection. 3. Select the code type to see its part number displayed in the text field to the right of the Request Part No. button.
Request Part Number by Serial Command
• When you send <#> (a request for all product part numbers), the imager returns: <#b/BOOT_P/N><#a/APP_P/N><#p/PROFILE_P/N>. • When you send <#a> (a request for the application code part number), the imager returns: <#a/APP_P/N>. • When you send <#b> (a request for the boot code part number), the imager returns: <#b/BOOT_P/N>. • When you send <#p> (a request for profile module part numbers), the imager returns: <#p/PROFILE_P/N>.
EZ Match User Manual
14-17
Page 348
Firmware
Request Checksum by ESP
You can send a request to the imager for part numbers, checksums, boot code, and application code. 1. Click the Firmware tab. 2. From the dropdown selection box to the left of the Request Checksum button, make your selection. 3. Select the code type to see its checksum displayed in the text field to the right of the Request Checksum button.
Request Checksum by Serial Command
• When you send <!> (a request for all available firmware checksums), the imager returns: <!b/BOOT_CHECKSUM><!a/APP_CHECKSUM><!p/PROFILE_CHECKSUM> • When you send <!a> (a request for the application code checksum), the imager returns: <!a/APP_CHECKSUM> • When you send <!b> (a request for the boot code checksum), the imager returns: <!b/BOOT_CHECKSUM> • When you send <!p> (a request for profile module checksum), the imager returns: <!p/PROFILE_CHECKSUM>
14-18
EZ Match User Manual
Page 349
Utilities
Default / Reset / Save Understanding and controlling your imager’s active, saved, and default settings is critical to its successful operation. Function
Saved for Power-On
Resets (not saved for power-on)
Reset
Serial Cmd <A>
Reset and recall Microscan defaults <Ard> Reset and Recall Microscan White <ArdWB> Balance Defaults Reset and recall power-on parameters <Arp> Reset and recall customer default <Arc> parameters Reset and recall Microscan Tube/Cap <Arl> Library defaults
ESP (first, right click in any menu tree)a
EZ Button
Save to Reader, Send No Save No
No
No
No
No
No
No
No
No
No
Save to Reader, Send and Save Save to Reader, Send and Save Customer Defaults for Power-Onb
No
No
Save current settings for power-on
<Z>
Save current settings as customer default parameters for power-on
<Zc>
Recall Microscan default parameters and save for power-on Reset Microscan Default White Balance Parameters and Save for Power-On
<Zrd>
No
No
<ZrdWB>
No
No
Recall customer default parameters and save for power-on
<Zrc>
No
Press and hold while powering on imager
Save current Tube/Cap Library parameters for power-on Recall Microscan Tube/Cap Library defaults and Save for Power-On
<Zl>
Send and Save
No
<Zrl>
Send and Save
No
No a. When you right click in a menu tree and select Default Current Menu Settings or Default All ESP Settings, it is important to note that only ESP settings are defaulted. To save these defaults to the imager itself, you need to follow up with a Save to Reader, Send and Save command. b. Only available in ESP if enabled under the Options dropdown menu.
Resets
Resets (“A” commands) affect only the current settings (active memory) and are not saved for power-on.
Saved for Power-on
Power-on parameters (“Z” commands) are saved to NOVRAM and recalled and loaded into current parameters when power is cycled to the imager or the <Arp> command is issued.
Defaults
Defaults are Microscan firmware settings or saved customer settings that can be recalled, either by software or hardware reset. EZ Match User Manual
14-19
Page 350
Default / Reset / Save
Customer Default Parameters Customer default parameters (saved by <Zc>) are the same set of parameters as power-on parameters but are saved in a different, isolated section of NOVRAM. This allows a user essentially to create a backup set of parameters that can be recalled in the event that the current parameters or power-on parameters have been accidentally changed or are no longer desired. It is important to note that a hardware default does not affect customer default parameters. For example, a user that has inadvertently changed communication settings and saved them with a <Z> command, may not know the correct settings or have the capability to communicate within those settings. By first doing an EZ Button default or hardware default to restore the known Microscan defaults, the user can then recall the previous customer-saved settings with an <Arc> or <Zrc> command.
Microscan Default Parameters Microscan default parameters are contained in the firmware and cannot be changed.
Software Defaults Microscan default parameters can be recalled (loaded into current settings) with <Ard> command or recalled and saved for power-on with the <Zrd> command.
Hardware Default If a software default and reset is not possible, it may be necessary to reset the imager by shorting (connecting) specific pins. This procedure has the same effect as the <Zrd> software command. Important: For this reset to occur, this command must be executed within 60 seconds of a power-on or a reset. 1. Apply power to the imager. 2. If using an IB-131, locate pins 7 and 11 on the host connector. Caution: Be certain that the correct pins are located. Connecting the wrong pins could cause serious damage to the unit. 3. Momentarily connect these wires (or pins) and listen for a series of short beeps. 4. Within 3 seconds, connect them again. A longer beep should be heard. If not, repeat the process.
Default on Power-On You can also use the EZ Button to default the imager by holding it down while applying power to the imager, provided that this feature is enabled.
14-20
EZ Match User Manual