14 Utilities Contents Serial Utility Commands ............................................................................................................. 14-2 Read Rate .................................................................................................................................. 14-4 Counters ..................................................................................................................................... 14-5 Device Control ............................................................................................................................ 14-8 Differences from Default............................................................................................................. 14-9 Master Database ...................................................................................................................... 14-10 Firmware................................................................................................................................... 14-16 Default / Reset / Save............................................................................................................... 14-19 Imager Status Requests ........................................................................................................... 14-21 Other Operational Serial Commands ....................................................................................... 14-23
Utility commands are generally commands that are performed during imager operations to check or determine Read Rates, or to perform miscellaneous operations on imager hardware. Serial utility commands are not prefaced with a “K” and a numeric code, nor do they require an initialization command (<A> and <Z>). They can be entered from within any terminal program or from ESP’s Terminal or Utilities windows.
EZ Match User Manual
14-1
Page 332
Serial Utility Commands
Serial Utility Commands Note: For a list of all K commands, see Serial Configuration Commands on page A-13. Type Read Rate
Counters
Device Control
Master Database
Part Number / Checksum
14-2
Command <C> <Cp> <J> <N> <O> <T> <U> <V> <W> <X> <Y> <CAP_A> <CAP_P> <CAP_U> <CAP_R> <TUBE_A> <TUBE_P> <TUBE_R> <L1> <L2> <L3> <11> <10> <I> <H> <a1> <G> <Gn> <NEWM> <#> <#a> <#b> <#p> <!> <!a>
Name Enter Decode Rate Test Enter Percent Rate Test Exit Decode Rate and Percent Rate Test No Read Counter No Read Counter Reset Trigger Counter Trigger Counter Reset Good Read/Match Counter Good Read/Match Counter Reset Mismatch Counter Mismatch Counter Reset Cap Absence Counter Cap Presence Counter Cap Unknown Counter Cap Counter Reset Tube Absence Counter Tube Presence Counter Tube Counter Reset Programmable Output 1 Programmable Output 2 Programmable Output 3 Targeting System On Targeting System Off Disable Imager Enable Imager Include PDF Information in Preamble (PDF417) Store Next Symbol Read to Database Index 1 Store Next Symbol Read to Database Index n New Master Load Status Display All Firmware Part Numbers Display Application Code Part Number Display Boot Code Part Number Display Profile Module Part Numbers Display All Available Firmware Checksums Display Application Code Code Checksum EZ Match User Manual
Page 333
Utilities Part Number / Checksum
<!b> <!p> <A> <Ard> <ArdWB> <Arp> <Arc> <Arl> <Z>
Default/Reset/Save
<Zc> <Zrd> <ZrdWB> <Zrc> <Zl> <Zrl>
Imager Status Request Commands
Other Operational Commands
EZ Match User Manual
<?> <K?> <K??> <K?#> <Knnn?> <Knnn??> <Knnn?#> <uy,path image> <dy,image name> <BCCFG> <@CAL> <@LCAP> <@WB>
Display Boot Code Checksum Display Profile Module Checksum Reset (does not save for power-on) Reset and Recall Microscan Defaults Reset and Recall Microscan White Balance Defaults Reset and Recall Power-On Parameters Reset and Recall Customer Default Parameters Reset and Recall Microscan Tube/Cap Library Defaults Save Current Settings for Power-On Save Current Settings as Customer Default Parameters for Power-On Recall Microscan Default Parameters and Save for Power-On Reset Microscan Default White Balance Parameters and Save for Power-On Recall Customer Default Parameters and Save for Power-On Save Current Tube/Cap Library Parameters for Power-On Recall Microscan Tube/Cap Library Defaults and Save for Power-On Imager Status Byte All Status Request All Descriptor Status Request All Range Status Request Single Status Request Single Descriptor Status Request Single Range Status Request Y-Modem Upload Y-Modem Download Enter Bar Code Configuration Mode Calibrate Learn Tube or Cap White Balance Calibration
14-3
Page 334
Read Rate
Read Rate Read Rate by ESP
You can access Read Rate from the Utilities view under the Read Rate tab. To see the number of decodes per second, click the Decodes/sec radio button and click the Start button. To see the percentage of decodes, click the Percent radio button and then the Start button. To end a Read Rate test, click the Stop button (the Start button becomes a “Stop” button during an active Read Rate test.) Important: When Read Rate is enabled, the IP Database is not active, regardless of how the imager is configured.
Read Rate by Serial Command Enter Decodes / Second Test Sending <C> instructs the imager to transmit the decodes per second and symbol data (if any). The decode rate can vary dramatically due to the angle and location of the symbol in relation to the field of view. This test is very useful in aligning and positioning the imager during setup.
Enter Percent Test Sending <Cp> instructs the imager to transmit the percentage of decodes and any decoded symbol data.
End Read Rate Test Sending <J> ends both the Percent test and the Decodes/Second test.
Tube/Cap Detection Notes
1. If Tube/Cap Detection operations are enabled in the <K260> Multi-Operation command, they will be part of Read Rate output. The formatted message will follow the Tube/Cap Detection message setup. 2. Active symbologies for Read Rate are determined in the same way, regardless of how Multi-Operation is configured. 3. Tube Angle is not part of Read Rate output. See the the <K850> Configurable Output 1 command for further details. 4. If enabled, Library Index is part of Read Rate output. Tubes or caps with different library indexes are treated as different objects. See the the <K850> Configurable Output 1 command for further details. 14-4 EZ Match User Manual
Page 335
Utilities
Counters Counter commands can be a numeric value from 00000 to 65535. After reaching the maximum numeric limit of 65535, an error message will be displayed and the counter will automatically roll over and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65536 per each rollover (the imager does not keep track of the number of rollovers) to the current count. Note: All counter values will be lost if power is recycled to the imager, or if the imager receives a Reset or Save command.
Counters by ESP You can access Counters from ESP’s Utilities menu. Click the Request button to display the appropriate count or Clear to set the counter to zero.
EZ Match User Manual
14-5
Page 336
Counters
Counters by Serial Command No Read Counter Sending <N> displays the total number of No Reads that have occurred since the last reset.
No Read Counter Reset Sending <O> sets the No Read Counter to 00000.
Trigger Counter Sending <T> displays the total number of triggers since the last reset.
Trigger Counter Reset Sending <U> sets the trigger counter to 00000.
Good Read / Match Counter (or Good Read Counter) Sending <V> displays the total number of Good Reads matching the master symbol, or, if Master Symbol is not enabled, the number of Good Reads since the last reset. This counter is always enabled, but will only work as a match count when Master Symbol is enabled. If Master Symbol is not enabled, this counter records the number of Good Reads. This count can be requested at any time.
Good Read / Match Counter Reset Sending <W> sets the Match Counter to 00000.
Mismatch Counter Sending <X> displays the number of decoded symbols since the last reset that do not match the master symbol.
Mismatch Counter Reset Sending <Y> sets the Mismatch Counter to zero.
Cap Absent Counter Sending <CAP_A> displays the total number of Cap Absences that have occurred since power-on, or since the last Cap Counter Reset occurred. Imager response format: <CAP_A/xxxxx>
Cap Present Counter Sending <CAP_P> displays the total number of Cap Presences that have occurred since power-on, or since the last Cap Counter Reset occurred. Imager response format: <CAP_P/xxxxx>
14-6
EZ Match User Manual
Page 337
Utilities
Cap Unknown Counter Sending <CAP_U> displays the total number of Cap Unknowns that have occurred since power-on, or since the last Cap Counter Reset occurred. Imager response format: <CAP_U/xxxxx>
Cap Counter Reset Sending <CAP_R> sets all Cap Counters to 00000.
Tube Absent Counter Sending <TUBE_A> displays the total number of Tube Absences that have occurred since power-on, or since the last Tube Counter Reset occurred. Imager response format: <TUBE_A/xxxxx>
Tube Present Counter Sending <TUBE_P> displays the total number of Tube Pesences that have occurred since power-on, or since the last Tube Counter Reset occurred. Imager response format: <TUBE_P/xxxxx>
Tube Counter Reset Sending <TUBE_R> sets all Tube Counters to 00000.
EZ Match User Manual
14-7
Page 338
Device Control
Device Control Device Control by ESP
Device Control by Serial Command Output 1 Pulse Sending <L1> activates the link between Output 1 (+) and Output 1 (-) of the host connector (regardless of Master Symbol or Output 1 status).
Output 2 Pulse Sending <L2> activates the link between Output 2 (+) and Output 2 (-) of the host connector (regardless of Master Symbol or Output 2 status).
Output 3 Pulse Sending <L3> activates the link between Output 3 (+) and Output 3 (-) of the host connector (regardless of Master Symbol or Output 3 status).
Disable Reader Sending <I> will turn the imager OFF, end the current read cycle, and will not allow the imager to enter another read cycle until turned ON. This feature is useful during extended periods of time when no symbols are being decoded, or the imager is being configured. Disabling the imager will not affect any commands that have already been downloaded.
Enable Reader Sending <H> will turn the imager ON and allow it to enter read cycles.
Enable PDF Information Sending <a1> will cause PDF417 data to be prefaced with information consisting of ECC level (ECC level n), number of rows (n rows), number of columns (n columns), number of informative code words (n info code words), and the number of data characters (n data bits). Send <a1> again to disable PDF Information. 14-8
EZ Match User Manual
Page 339
Utilities
Differences from Default Clicking the Differences from Default button will cause ESP to check all stored configuration settings and compare them to default settings. All settings that are different than default will appear in the left column (shown below), and descriptions of those settings will appear in the right column. Click this button for a list of ESP configuration settings that are different than default settings.
Click Generate Barcode to bring up the Bar Code Dialog. Then create configuration symbols containing the configuration commands of your choice.
Send configuration settings to the imager without saving by clicking Send to Reader.
Click Save As to save the report as plain text or a tab-delimited text file.
Send configuration settings to the imager and save in ESP by clicking Send and Save.
• To create a symbol containing any of the command settings in the table, click Generate Barcode. This will bring up the Bar Code Dialog. • To save the Differences from Default report, either as plain text or as a tab-delimited text file, click Save As. • Click Send and Save to send the settings to the imager and save them, or Send to Reader to send the settings without saving them. Important: The use the Differences from Default feature, you must connect to the imager and Receive Reader Settings via the Send/Recv button on the toolbar.
EZ Match User Manual
14-9
Page 340
Master Database
Master Database Important: The Master Database is used for all Matchcode modes except Sequential and Wild Card, both of which use Master Database Index # 1.
Master Database Overview Usage: Definition:
Used where more than one master symbol is required, as in a Multisymbol setup, for matching and other Matchcode operations. Allows the user to define up to 10 master symbols as the master symbol database, which can be entered by keyboard, scanned in, displayed, or deleted by serial or ESP commands.
1. Click the Master Database tab. 2. Enable Matchcode Type. 3. Set the Master Symbol Database Size. 4. Select the database index in which you want to enter the master symbol. 5. Do one of the following to enter master symbol data: a. Double-click the index row to type data directly into the index. b. Click the Read Symbol into Selected Index to enter the next decoded symbol. Set Master Symbol Database Size here.
Double-click a row to enter data into popup dialog box.
14-10
Loads the next symbol decoded into the selected index.
Loads the imager’s saved master symbols into ESP.
Saves the displayed database to the imager.