For the Microscan EZ Match
Page 251
Symbol Quality
ISO/IEC 16022 Symbol Quality Output Symbol Quality Parameters for Data Matrix symbols specified by ISO/IEC 16022.
Symbol Contrast Usage: Definition:
Serial Cmd: Default: Options:
Lets the user know if contrast settings are less than acceptable. All the pixels that fall within the area of the test symbol, including its required zone, will be sorted by their reflectance values to select the darkest 10% and the lightest 10% of the pixels. The arithmetic mean of the darkest and the lightest pixels is calculated and the difference of the two means is the Symbol Contrast. (ANSI) Symbol Contrast grading is defined this way: A (4.0) if SC > 70% B (3.0) if SC > 55% C (2.0) if SC > 40% D (1.0) if SC > 20% F (0.0) if SC < 20% If enabled, the symbol contrast is appended to the symbol data according to the ISO/IEC 16022 Symbol Quality Output Mode setting. <K709,symbol contrast,print growth,axial non-uniformity,unused ECC> Disabled 0 = Disabled 1 = Enabled
Print Growth Definition:
Serial Cmd: Default: Options:
The extent to which dark or light markings appropriately fill or exceed their module boundaries. These values are determined by counting pixels in the clock pattern of the binary digitized image, then comparing it to a nominal value and minimum and maximum values. The print growth grade is defined in this way: A (4.0) if -.050 < PG < 0.50 C (2.0) if -.085 < PG < 0.85 B (3.0) if -.070 < PG < 0.70 D (1.0) if -.10 < PG < 1.00 F (0.0) if PG < -1.00 or PG > 1.00 If enabled, the print growth is appended to the symbol data according to the ISO/IEC 16022 Symbol Quality Output Mode setting. <K709,symbol contrast,print growth,axial non-uniformity,unused ECC> Disabled 0 = Disabled 1 = Enabled
EZ Match User Manual
9-7
Page 252
ISO/IEC 16022 Symbol Quality Output
Axial Non-Uniformity (Data Matrix Only) Definition:
Serial Cmd: Default: Options:
Axial non-uniformity is a measure of how much the sampling point spacing differs from one axis to another, namely AN = abs (XAVG YAVG) / ((XAVG + YAVG)/2) where abs () yields the absolute value. If a symbology has more than two major axes, then AN is computed for those two average spacings which differ the most. (ANSI) axial non-uniformity grading is defined this way: A (4.0) if AN < .06 B (2.0) if AN < .08 C (2.0) if AN < .10 D (1.0) if AN < .12 F (0.0) if AN > .12 If enabled, the axial non-uniformity is appended to the symbol data according the ISO/IEC 16022 Symbol Quality Output Mode setting. <K709,symbol contrast,print growth,axial non-uniformity,unused ECC> Disabled 0 = Disabled 1 = Enabled
Unused Error Correction (Data Matrix Only) Definition:
Serial Cmd: Default: Options:
9-8
The correction capacity of Reed-Solomon decoding is expressed in the equation: e + 2d < d - p, where e is the number of erasures, d is the number of ECC code words, and p is the number of code words reserved for error detection. A (4.0) if UEC > .62 B (3.0) if UEC > .50 C (2.0) if UEC > .37 D (1.0) if UEC > .25 F (0.0) if UEC < .25 If enabled, the unused ECC is appended to the symbol data according to the ISO/IEC 16022 Symbol Quality Output Mode setting. <K709,symbol contrast,print growth,axial non-uniformity,unused ECC> Disabled 0 = Disabled 1 = Enabled
EZ Match User Manual
Page 253
Symbol Quality
ISO/IEC 16022 Symbol Quality Output by ESP ESP’s Symbol Quality interface allows you to evaluate Data Matrix symbols for compliance with ISO/IEC 16022 requirements. First, determine which ISO/IEC 16022 Parameters you need to evaluate, using the Symbol Quality tree control.
Once you’ve set these Data Matrix evaluation parameters, go to the Preferences tab and set your Data Matrix Grading Report output preferences.
After your output preferences are set, click the Data Matrix Grading button.
There will be a wait of a few seconds, and then the evaluation results will appear at the lower right of the Symbol Quality view, in a format similar to the one shown below.
If you want to view the results in a report format, click the Generate Report button. The report will be output in the format you chose in the Preferences dialog (Default Report Type).
EZ Match User Manual
9-9
Page 254
Microscan Symbol Quality Output
Microscan Symbol Quality Output Percent Cell Damage (Data Matrix Only) Definition: Serial Cmd: Default: Options:
When this feature is enabled, the cell damage percentage is appended to data output. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Total Read Time Definition: Serial Cmd: Default: Options:
The time that transpires between the image capture and the output of the decoded data, including locate time. When enabled, the total read time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Capture Time Definition: Serial Cmd: Default: Options:
Capture time (in milliseconds) is a fixed “overhead” that includes the time of capture and transfer of the image. When enabled, the capture time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Locate Time Definition: Serial Cmd: Default: Options: 9-10
The time in milliseconds from the start of image processing until the symbol has been located and is ready to be decoded. When enabled, the locate time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled EZ Match User Manual
Page 255
Symbol Quality
Decode Time Definition: Serial Cmd: Default: Options:
The time in milliseconds required to decode a symbol. When enabled, the decode time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Pixels Per Element (Data Matrix Only) Definition:
Serial Cmd: Default: Options:
The number of pixels for each element, either dark or light for both x and y directions. When enabled, the pixels per element value is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
ECC Level (Data Matrix Only) Definition: Serial Cmd: Default: Options:
Outputs the Data Matrix ECC level. When enabled, the ECC level is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Matrix Size (Data Matrix Only) Definition: Serial Cmd: Default: Options:
Defines the symbol matrix size, in number of pixels in both the x and y axis. When enabled, the matrix size value is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
EZ Match User Manual
9-11
Page 256
Microscan Symbol Quality Output
Quiet Zone (Data Matrix Only) Definition: Serial Cmd: Default: Options:
When this feature is enabled, the size of the quiet zone is evaluated and a PASS or FAIL message is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
Symbol Angle (Data Matrix Only) Definition: Serial Cmd: Default: Options:
9-12
When this feature is enabled, the symbol orientation is appended to data output as a degree value representing the angle of the Data Matrix symbol’s L-shaped finder pattern relative to the imager. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled
EZ Match User Manual
Page 257
Symbol Quality
Microscan Symbol Quality Output by ESP ESP’s Symbol Quality interface allows you to evaluate Data Matrix symbols for compliance with a rigorous set of standards, such as Locate Time, Capture Time, and Decode Time. First, determine which parameters you want to evaluate, using the Symbol Quality tree control.
Once you’ve set these evaluation parameters, go to the Preferences tab and set your Data Matrix Grading Report output preferences.
After your output preferences are set, click the Data Matrix Grading button.
There will be a wait of a few seconds, and then the evaluation results will appear at the lower right of the Symbol Quality view, in a format similar to the one shown below.
If you want to view the results in a report format, click the Generate Report button. The report will be output in the format you chose in the Preferences dialog (Default Report Type).
EZ Match User Manual
9-13
Page 258
Microscan Symbol Quality Output by ESP
9-14
EZ Match User Manual
Page 259
10 Matchcode Contents Matchcode by ESP ..................................................................................................................... 10-2 Matchcode Serial Commands .................................................................................................... 10-2 Overview of Matchcode .............................................................................................................. 10-3 Matchcode Type ......................................................................................................................... 10-4 Match Replace............................................................................................................................ 10-9 Mismatch Replace .................................................................................................................... 10-10 New Master Pin ........................................................................................................................ 10-11
This section explains the EZ Match Imager’s matchcode output functions and the master symbol database setup.
EZ Match User Manual
10-1
Page 260
Matchcode by ESP
Matchcode by ESP
Click the Parameters button and then the Matchcode tab.
To change a setting, double-click the setting and use your cursor to scroll through the options.
To open nested options, single-click the +.
Matchcode Serial Commands Matchcode Type Number of Master Symbols New Master Pin Enter Master Symbol Data Request Master Symbol Data Delete Master Symbol Data Match Replace Mismatch Replace
10-2
<K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on No Read,sequence on mismatch> <K224,number of master symbols> <K225,status> <K231,master symbol number,master symbol data> <K231?,>[for all] or <K231?,master symbol number> <K231,master symbol number,> <K735,status,match replacement string> <K736,status,mismatch replacement string>
EZ Match User Manual