Page 141
Read Cycle
Number of Captures Usage: Definition:
Serial Cmd: Default: Options:
Number of Captures is used to specify the number of captures to be processed in Rapid Capture Mode. Sets the total number of captures that are processed during a read cycle in Rapid Capture Mode. This feature is used in conjunction with Capture Timing parameters to specify the capture sequence of a rapid capture read cycle. <K241,capture mode,number of captures,rapid capture mode,pre-processing> 1 1 to 64 Note: The range of maximum number of captures is dynamic. This range is dependent on the maximum image size in the system. A full-size image (1280 x 1024) reduces the maximum number of images to 6. The smaller the image size, the greater the maximum number of captures. Once the image size is reduced to small enough dimensions, the maximum number of captures will be capped at 64. If a user enters a maximum capture value greater than that allowed, the value will be limited to the number of system images. This command also affects the number of stored images allowed in the system. If the maximum number of captures is selected, the number of stored images allowed will be 0. Note: Number of Captures specifies the maximum number of captures as it pertains to a Multi-Operation read cycle. When configured to switch operations on a frame count or timeout, the total number of captures cannot exceed this maximum value. When configured to switch operations on a specified frame count and the operation completes before this maximum value is reached, the image frame count will increase automatically to match the frame switch count. This maintains image frame and timeout consistency for each operation index.
Rapid Capture Mode Definition: Serial Cmd: Default: Options:
In Rapid Capture Mode, one or multiple captures (as many as 32) can be taken at an interval specified by the time-between-captures parameter. In this mode, the only limiting time factor is integration and transfer timing. <K241,capture mode,number of captures,rapid capture mode,pre-processing> Timed Capture 0 = Timed Capture 1 = Triggered Capture
EZ Match User Manual
6-19
Page 142
Capture Mode
Timed Capture Usage: Definition:
Timed Rapid Capture is useful in fast-moving applications in which symbols are only in the field of view for a short time and precise timing is required. In Timed Rapid Capture, decoding occurs independent of and simultaneous with capturing, thus allowing precise timing or no delay at all between captures. Also, consecutive captures are regarded as the same symbol if the output data is the same. No Time Delay Between Captures
Start of Read Cycle
Diagram A n
End of read cycle
Time Before First Capture Time Delay Between Captures End of read cycle
Diagram B Processing
6-20
EZ Match User Manual
Page 143
Read Cycle
Triggered Capture Usage: Definition:
Useful in applications where each decode must be treated as a discrete event, regardless of symbol data. The first trigger event starts the read cycle, and subsequent triggers will continue until the predetermined Number of Captures is met, or until the predetermined End of Read Cycle condition is met-- whichever occurs first. Note: If End of Read Cycle is set for New Trigger and the read cycle qualifications have not been met, the read cycle will only end once it receives the first trigger after reaching the predetermined Number of Captures setting. First trigger starts the read cycle
Captures on every trigger
Capture Read cycle ends on New Trigger or Timeout, as configured.
Processing
EZ Match User Manual
6-21
Page 144
Pre-Processing Definition: Serial Cmd: Default: Options:
When enabled, this feature reduces the time between image capture and decode by beginning the processing task while an image is being captured. <K241,capture mode,number of captures,rapid capture mode,pre-processing> Enabled 0 = Disabled 1 = Enabled
Disabled When Pre-Processing is disabled, the imager can process previously captured images while image capture events are taking place. The benefit of this operation is that the image capture events have minimal impact on image processing. Therefore, if the imager is configured to capture 4 consecutive frames, but it is only looking for one decode, it could process and decode a symbol on the first image frame before all 4 image frames were captured. If Pre-Processing were enabled, the imager could not start processing an image frame until all 4 image frames were captured (since they are consecutive captures).
Enabled When Pre-Processing is enabled, the imager performs pre-processing tasks while an image is being captured. Since some of the processing is complete when the image has been captured, the time required to complete the processing of the image frame will be less (approx. 10 ms/frame). However, while an image is being captured, the imager is dedicated to operating on the current image being captured. This means that an image capture event will impact the processing time of previously captured image frames. The benefit of this mode is the decrease in capture-to-decode time.
6-22
EZ Match User Manual
Page 145
Read Cycle
Capture Timing Note: Capture Timing applies only to Rapid Capture Mode.
Time Before First Capture Usage: Definition: Serial Cmd: Default: Options:
In almost any moving line application, a time delay is needed to ensure that a symbol will be in the imager’s field of view at the beginning of the capture sequence. Time Before First Capture in a moving line application is the time between an external trigger event and the occurrence of the first capture. <K242,time before 1st capture,time between capture 1 and capture 2,,,,,,,,time between capture 7 and capture 8> 0 0 to 65535 (2.097 seconds, in 32 µS increments)
Start of Read Cycle Time Before First Capture
EZ Match User Manual
Processing
6-23
Page 146
Capture Timing
Time Between Captures Usage: Definition: Serial Cmd:
Default: Options:
This is useful in applications where more than one symbol can appear during a single read cycle (multisymbol), or where line speeds are slow enough that captured frames might overlap or miss a symbol. A time delay can be inserted between individual frame captures in Rapid Capture Mode. <K242,time before 1st capture,time between captures [time 1,time 2,...time7] Entering 0s will result in no time between captures. Entering a different value in each field will vary the time delays accordingly. Note: You must enter time values along with comma separators for each field you want to change. If you omit fields, or enter only commas, the fields will remain as previously set. 0 0 to 65535 (2.097 seconds, in 32 µS increments)
Note: Number of Captures and number of frame delays (Time Between Captures) must be the same. Start of Read Cycle
Diagram A
Time Delay Between Captures = 0 1 2 3 4 5
End of Read Cycle
Time Delay Between Captures varies.
Diagram B
6-24
EZ Match User Manual
Page 147
Read Cycle
Image Storage Image Storage Type Definition:
Serial Cmd: Default: Options:
Allows the user to store images from separate read cycles and to retrieve them later. The number of available slots for storage depends on the mode of operation. If the imager is in Rapid Capture Mode, the number of images that can be stored is equivalent to the maximum number of the rapid count (the current rapid count setting). If the imager is inContinuous Capture Mode, a number of images equivalent to the maximum number of the rapid count minus 3. <K244,image storage type,image storage mode> Disabled/Clear 0 = Disabled/Clear 1 = Store on No Read
Disabled / Clear Upon selection of this option, all saved images will be cleared and the imager will not store images for later viewing.
Store on No Read This option will cause the imager to store an image upon exiting the read cycle for retrieval at a later time. If multiple captures are present during the duration of a read cycle, the stored image will be the last image processed for that read cycle. This image is stored in RAM and can be retrieved as long as power is cycled to the imager, and as long as the imager has not been reset via a reset/save sequence. Other commands that can initialize storage in RAM are ones that change capture modes or put the imager in a test capture mode. Note: A Cap Absence or Tube Absence will qualify for image storage when Store on No Read is selected. In a Multi-Operation read cycle with output set to As Soon As Possible, only the results from the last operation are retained. Therefore, if two operations are active, the first resulting in a No Read condition and the second not resulting in a No Read condition, no image will be stored.
EZ Match User Manual
6-25
Page 148
Image Storage
Image Storage Mode Serial Cmd: Default: Options:
<K244,image storage type,image storage mode> First Mode 0 = First Mode 1 = Last Mode
First Mode This mode allows the imager to store images until the available image memory has been filled. At that point the imager will stop saving additional images. In this mode, you will always have the first image captured, because the saving process stops once memory has been filled.
Last Mode In this mode, image storage continues after available memory limits are reached. The oldest image in memory is overwritten, so you will always have the most recent stored image.
6-26
EZ Match User Manual
Page 149
Read Cycle
Image Storage Example
The following example assumes that the imager is in a rapid capture mode of 3 captures. Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
1 No Read No Read Good Read, Symbol # 1 Good Read None
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
1 No Read No Read No Read No Read Frame # 3
Number of Symbols: Frame # 1: Frame # 2: Frame # 3: Read Cycle Result: Stored Frame:
2 No Read No Read Good Read, Symbol # 1 No Read Frame # 2
EZ Match User Manual
6-27
Page 150
Minimum Good Reads
Minimum Good Reads Definition: Serial Cmd: Default: Options:
6-28
This value specifies the number of times a symbol needs to be read to qualify as a Good Read. <K221,minimum good reads> 1 1 to 255
EZ Match User Manual