Page 201

8 Diagnostics Contents Diagnostics by ESP.......................................................................................................................8-2 Diagnostics Serial Commands ......................................................................................................8-2 Counts (Read Only).......................................................................................................................8-3 External Camera Message............................................................................................................8-4 Over Temperature Message .........................................................................................................8-6 Service Message...........................................................................................................................8-7

This section describes warning and operating messages and their settings. Note: The characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands.

Quadrus EZ Reader User’s Manual

8-1

Page 202

Diagnostics by ESP

Diagnostics by ESP

Click this button to bring up the Diagnostics menu. To open nested options, single-click the +.

To change a setting, double-click the setting and use your cursor to scroll through the options.

Diagnostics Serial Commands <K406,power-on,resets,power-on saves,customer default saves> <K410,disconnect msg status,disconnect message, External Camera Message connect msg status,control message> Over Temperature Message <K402,over temperature status,warning message> Service Message <K409,status,service message,threshold,resolution> Power-on/Reset Counts

8-2

Quadrus EZ Reader User’s Manual

Page 203

Diagnostics

Counts (Read Only) Counts for Power-on/Reset/Saves are stored in the reader and can be displayed at any time in response to serial commands listed here, in the embedded Diagnostic menu, or in ESP by requesting reader settings. If you did not choose to receive reader settings upon connection in ESP, you can right-click anywhere in the Diagnostic window and select Receive Reader Settings.

Power-on Definition:

Displays a count of the number of times power to the reader is recycled.

Serial Cmd:

<K406,power-on,resets,power-on saves,customer default saves>

Resets Definition:

Displays a count of the number of times the reader is reset.

Serial Cmd:

<K406,power-on,resets,power-on saves,customer default saves>

Power-on Saves Definition:

Displays a count of the number of power-on saves <Z> command.

Serial Cmd:

<K406,power-on,resets,power-on saves,customer default saves>

Custom Default Saves Definition:

Displays a count of the number of power-on saves (customer defaults) to flash memory (<Zc> command)

Serial Cmd:

<K406,power-on,resets,power-on saves,customer default saves>

Quadrus EZ Reader User’s Manual

8-3

Page 204

External Camera Message

External Camera Message Note: This feature cannot be used if in a Continuous Read mode.

Disconnect Status Definition:

When enabled, a message is sent when the system detects that the external camera is not connected.

Conditions:

On power-on, if enabled and the external camera is not connected, the “disconnect message” will be output. The message will not be output again unless power is cycled or a “disconnect” condition occurs after a “connect” condition with connect message enabled. Note: If the external camera is configured in “Interlaced” mode instead of “Progressive,” this will be detected as a Disconnect condition.

Serial Cmd:

<K410,disconnect msg status,disconnect message,connect msg status, connect message>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Disconnect Message Serial Cmd:

<K410,disconnect msg status,disconnect message,connect msg status, connect message>

Default:

EX_CAM_OFF

Options:

Any 1 to 10 ASCII characters except NULL < , or >.

Connect Status Definition:

When enabled, a message is sent when the system detects that the external camera is connected.

Conditions:

On power-on, if enabled and the external camera is connected, the “connect message” will be output. The message will not be output again unless power is cycled or a “connect” condition occurs after a “disconnect” condition with connect message enabled. Note: If the external camera is configured in “Interlaced” mode instead of “Progressive,” this will not be detected as a connect condition.

Serial Cmd:

<K410,disconnect msg status,disconnect message,connect msg status, connect message>

Default:

Disabled

Options:

0 = Disabled

8-4

1 = Enabled

Quadrus EZ Reader User’s Manual

Page 205

Diagnostics

Connect Message Serial Cmd:

<K410,disconnect msg status,disconnect message,connect msg status, connect message>

Default:

EX_CAM_ON

Options:

Any 1 to 10 ASCII characters except NULL < , or >.

Quadrus EZ Reader User’s Manual

8-5

Page 206

Over Temperature Message

Over Temperature Message Over Temperature Status Usage:

Helps ensure that the reader is being used within its temperature specification.

Definition:

When enabled, a message is sent whenever the system detects that the internal temperature has exceeded it's specified operating temperature.

Serial Cmd:

<K402,over temperaturestatus,warning message>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Warning Message Serial Cmd:

<K402,over temperature status,warning message>

Default:

OVER_TEMP

Options:

Any 1 to 10 ASCII characters except NULL <, or >.

8-6

Quadrus EZ Reader User’s Manual

Page 207

Diagnostics

Service Message Service Message Status Definition:

When enabled, a message is sent whenever the system detects that a user-set service time has expired. The service timer is reset on power-on, thus the timer only records the time that has elapsed since the last reset. The message is sent every time the timer expires.

Serial Cmd:

<K409,status,service message,threshold,resolution>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Service Message Serial Cmd:

<K409,status,service message,threshold,resolution>

Default:

SERVICE

Options:

Any 1 to 10 ASCII characters except NULL < , or >.

Threshold Definition:

Sets the number of hours or minutes that will transpire before the service message is output.

Serial Cmd:

<K409,status,service message,threshold,resolution>

Default:

300 (5 minutes)

Options:

1 to 65,535

Resolution Definition:

Records time in seconds or minutes.

Serial Cmd:

<K409,status,service message,threshold,resolution>

Default:

Seconds

Options:

0 = Seconds

Quadrus EZ Reader User’s Manual

1 = Minutes

8-7

Page 208

Service Message

8-8

Quadrus EZ Reader User’s Manual

Page 209

9 Camera Setup Contents Camera Setup by ESP ................................................................................................................. 9-2 Camera Setup Serial Commands................................................................................................. 9-2 Video ............................................................................................................................................ 9-3 Evaluation..................................................................................................................................... 9-5 Region of Interest (ROI) ............................................................................................................... 9-8 Dynamic Setup ........................................................................................................................... 9-12 CCD Image Sensor .................................................................................................................... 9-14 CMOS Image Sensor ................................................................................................................. 9-15 Illumination Source ..................................................................................................................... 9-17 Thresholding............................................................................................................................... 9-18 Image Processing Settings......................................................................................................... 9-19 Hollow Mode............................................................................................................................... 9-21 Mirrored Image ........................................................................................................................... 9-22 Other Camera Parameters ......................................................................................................... 9-23

This section provides adjustment to the physical controls of the internal camera, image acquisition, database settings, and image diagnostics. Note: Camera Setup only applies to the Quadrus EZ internal camera. If using an external camera, images are received on an as-ready basis. Allow a 33mS delay for the first image to appear from an external camera. Note also that the Quadrus EZ may have one of two types of image sensors, CCD or CMOS. Even though some of the parameters between the sensors are similar, they are configured independently with different configuration commands. Therefore, the user needs to know which sensor type is installed in the Quadrus EZ. This can be determined with the <op,10> command.

Quadrus EZ Reader User’s Manual

9-1

Page 210

Camera Setup by ESP

Camera Setup by ESP

Click this Button to bring up the Camera Setup menu

To open nested options, single-click the +.

To change a setting, double-click the setting and use your cursor to scroll through the options.

Camera Setup Serial Commands Region of Interest CCD Image Sensor CMOS Image Sensor Illumination Source Thresholding Image Processing Mode Multiple Symbols in Fast Linear Mode Image Processing Timeout Hollow Mode Mirrored Image

9-2

<K516,top,left,height,width> <K540,shutter speed,gain> <K541,shutter speed,gain,contrast,offset> <K535,illumination source> <K512,threshold mode,threshold value> <K513,processing mode> <K518,number of symbols> <K245,image processing timeout> <K517,hollow status> <K514,mirrored image>

Quadrus EZ Reader User’s Manual