The text of a manual · part 17 of 37
Quadrus EZ Reader User's Manual, pages 161 to 170
Pages 161 to 170 of 366
The original document is the PDF (2.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Microscan EZ 6700 EX
Page 161
I/O Parameters
Symbol Quality to Output 1 (ISO/IEC 15415) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.
Under Output 1 Parameters, expand the Symbol Quality (ISO/IEC 15415) tree.
Output on Symbol Contrast Usage:
Lets the user know if symbol quality is less than acceptable.
Definition:
If enabled, toggles Output 1 active when Symbol Contrast Threshold is met.
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Output on Print Growth Usage:
Lets the user know if symbol quality is less than acceptable.
Definition:
If enabled, toggles Output 1 active when Print Growth Threshold is met.
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Disabled
Options:
0 = Disabled
Quadrus EZ Reader User’s Manual
1 = Enabled
5-45
Page 162
Output 1 Parameters
Output on Axial Non-Uniformity Usage:
Lets the user know if symbol quality is less than acceptable.
Definition:
If enabled, toggles Output 1 active when Axial Non-Uniformity Threshold is met.
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Output on Unused Error Correction Usage:
Lets the user know if symbol quality is less than acceptable.
Definition:
If enabled, toggles Output 1 when Unused ECC Threshold is met.
Serial Cmd:
<K800,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Symbol Contrast Threshold Usage:
Lets the user set the acceptable level of symbol quality.
Definition:
Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Grade C
Options:
0 = Grade A 1 = Grade B
5-46
2 = Grade C 3 = Grade D
Quadrus EZ Reader User’s Manual
Page 163
I/O Parameters
Print Growth Threshold Usage:
Lets the user set the acceptable level of symbol quality.
Definition:
Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Grade C
Options:
0 = Grade A 1 = Grade B
2 = Grade C 3 = Grade D
Axial Non-Uniformity Threshold Usage:
Lets the user set the acceptable level of symbol quality.
Definition:
Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).
Serial Cmd:
<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on unused ecc,unused ecc threshold>
Default:
Grade C
Options:
0 = Grade A 1 = Grade B
2 = Grade C 3 = Grade D
Unused ECC Threshold Usage:
Lets the user set the acceptable level of symbol quality.
Definition:
Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).
Serial Cmd:
<K800,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>
Default:
Grade C
Options:
0 = Grade A 1 = Grade B
Quadrus EZ Reader User’s Manual
2 = Grade C 3 = Grade D
5-47
Page 164
Output 1 Parameters
Symbol Quality to Output 1 (Inkjet/Direct) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.
Under Output 1 Parameters, expand the Symbol Quality (Inkjet/Direct) tree.
Output on Dot Center Offset Definition:
When enabled, dot center offset of decoded symbols are compared to the dot center offset threshold. If the value does not meet the threshold standard, the output is asserted.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Output on Cell Fill Definition:
When enabled, cell fill of decoded symbols are compared to the cell fill threshold. If the value does not meet the threshold standard, the output is asserted.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Default:
Disabled
Options:
0 = Disabled
5-48
1 = Enabled
Quadrus EZ Reader User’s Manual
Page 165
I/O Parameters
Output on Dot Ovality Definition:
When enabled, dot ovality of decoded symbols are compared to the dot ovality threshold. If the value does not meet the threshold standard, the output is asserted. This has no effect when the dot is specified as a square.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Output on Angle of Distortion Definition:
When enabled, the angle of distortion of decoded symbols are compared to the angle of distortion threshold. If the value does not meet the threshold standard, the output is asserted.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Dot Center Offset Definition:
This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Default:
Grade B (>2-5%)
Options:
0 = Grade A (0-2%) 2 = Grade C (>5-10%)
Quadrus EZ Reader User’s Manual
1 = Grade B (>2-5%) 3 = Grade D (>10%)
5-49
Page 166
Output 1 Parameters
Cell Fill Definition:
This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>
Default:
Grade B (>2-5%)
Options:
0 = Grade A (0-2%) 2 = Grade C (>5-10%)
1 = Grade B (>2-5%) 3 = Grade D (>10%)
Dot Ovality Threshold Definition:
This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec. This has no effect when the dot is specified as a square.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>
Default:
Grade B (>2-5%)
Options:
0 = Grade A (0-2%) 2 = Grade C (>5-10%)
1 = Grade B (>2-5%) 3 = Grade D (>10%)
Angle of Distortion Threshold Definition:
This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=3.5 degree (Grade A) and >3.5 to 7 degree (Grade B) are from Aerospace Direct Part Marking Standard Spec.
Serial Cmd:
<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>
Default:
Grade C (>7-10.5)
Options:
0 = Grade A (0-3.5 degree) 2 = Grade C (>7-10.5)
5-50
1 = Grade B (>3.5-7) 3 = Grade D (>10.5-14)
Quadrus EZ Reader User’s Manual
Page 167
I/O Parameters
Diagnostics
Note: When Diagnostic Warning is enabled under Output 1 Parameters, Output Mode configuration has no effect.
Usage:
Alerts user to critical conditions.
Definition:
Sets up specific warnings that will cause activation on output 1. The output will remain active as long as one of the diagnostic conditions is met. The output will go inactive as soon as it detects no active diagnostic warning.
Over Temperature Definition:
Setups the output to toggle active when an over-temperature condition is detected.
Serial Cmd:
<K790,over temp,service unit,external camera disconnect>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Service Unit Definition:
Setups the output to toggle active when the service timer has expired. This condition will only be held for one service timer tick.
Serial Cmd:
<K790,over temp,service unit,external camera disconnect>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
External Camera Disconnect Note: This feature cannot be used if in a Continuous Read mode. Definition:
Setups the output to toggle active if the external camera goes off-line.
Serial Cmd:
<K790,over temp,service unit,external camera disconnect>
Default:
Disabled
Options:
0 = Disabled
Quadrus EZ Reader User’s Manual
1 = Enabled 5-51
Page 168
Output 2 Parameters
Output 2 Parameters Note: Output 2 has the same parameters and default settings as Output 1.
Serial Cmd:
<K811,output on,output state,pulse width,output mode>
Trend Analysis to Output 2 Note: Output On under Output 2 Parameters must be set to Trend Analysis for this output to function.
Serial Cmd:
<K781,trend analysis mode,number of triggers,number to output on>
Symbol Quality to Output 2 (ISO/IEC 15415) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.
Serial Cmd:
5-52
<K801,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>
Quadrus EZ Reader User’s Manual
Page 169
I/O Parameters
Symbol Quality to Output 2 (Inkjet/Direct) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.
Serial Cmd:
<K821,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>
Diagnostic Warnings to Output 2 Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.
Serial Cmd:
<K791,over temp,service unit,external camera disconnect>
Quadrus EZ Reader User’s Manual
5-53
Page 170
Output 3 Parameters
Output 3 Parameters Output 3 has the same parameters and default settings as Output 1.
Serial Cmd:
<K812,output on,active state,pulse width,output mode>
Trend Analysis to Output 3 Note: Output On under Output 3 Parameters must be set to Trend Analysis for this output to function.
Serial Cmd:
<K782,trend analysis mode,number of triggers,number to output on>
Symbol Quality to Output 3 (ISO/IEC 15415) Note: Output On under Output 3 Parameters must be set to Symbol Quality for this output to function.
Serial Cmd:
5-54
<K802,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>
Quadrus EZ Reader User’s Manual