Page 161

I/O Parameters

Symbol Quality to Output 1 (ISO/IEC 15415) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.

Under Output 1 Parameters, expand the Symbol Quality (ISO/IEC 15415) tree.

Output on Symbol Contrast Usage:

Lets the user know if symbol quality is less than acceptable.

Definition:

If enabled, toggles Output 1 active when Symbol Contrast Threshold is met.

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Output on Print Growth Usage:

Lets the user know if symbol quality is less than acceptable.

Definition:

If enabled, toggles Output 1 active when Print Growth Threshold is met.

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Disabled

Options:

0 = Disabled

Quadrus EZ Reader User’s Manual

1 = Enabled

5-45

Page 162

Output 1 Parameters

Output on Axial Non-Uniformity Usage:

Lets the user know if symbol quality is less than acceptable.

Definition:

If enabled, toggles Output 1 active when Axial Non-Uniformity Threshold is met.

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Output on Unused Error Correction Usage:

Lets the user know if symbol quality is less than acceptable.

Definition:

If enabled, toggles Output 1 when Unused ECC Threshold is met.

Serial Cmd:

<K800,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Symbol Contrast Threshold Usage:

Lets the user set the acceptable level of symbol quality.

Definition:

Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Grade C

Options:

0 = Grade A 1 = Grade B

5-46

2 = Grade C 3 = Grade D

Quadrus EZ Reader User’s Manual

Page 163

I/O Parameters

Print Growth Threshold Usage:

Lets the user set the acceptable level of symbol quality.

Definition:

Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Grade C

Options:

0 = Grade A 1 = Grade B

2 = Grade C 3 = Grade D

Axial Non-Uniformity Threshold Usage:

Lets the user set the acceptable level of symbol quality.

Definition:

Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).

Serial Cmd:

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on unused ecc,unused ecc threshold>

Default:

Grade C

Options:

0 = Grade A 1 = Grade B

2 = Grade C 3 = Grade D

Unused ECC Threshold Usage:

Lets the user set the acceptable level of symbol quality.

Definition:

Conforms to ISO/IEC 15415 symbol quality grading (A,B,C,D).

Serial Cmd:

<K800,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>

Default:

Grade C

Options:

0 = Grade A 1 = Grade B

Quadrus EZ Reader User’s Manual

2 = Grade C 3 = Grade D

5-47

Page 164

Output 1 Parameters

Symbol Quality to Output 1 (Inkjet/Direct) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.

Under Output 1 Parameters, expand the Symbol Quality (Inkjet/Direct) tree.

Output on Dot Center Offset Definition:

When enabled, dot center offset of decoded symbols are compared to the dot center offset threshold. If the value does not meet the threshold standard, the output is asserted.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Output on Cell Fill Definition:

When enabled, cell fill of decoded symbols are compared to the cell fill threshold. If the value does not meet the threshold standard, the output is asserted.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Default:

Disabled

Options:

0 = Disabled

5-48

1 = Enabled

Quadrus EZ Reader User’s Manual

Page 165

I/O Parameters

Output on Dot Ovality Definition:

When enabled, dot ovality of decoded symbols are compared to the dot ovality threshold. If the value does not meet the threshold standard, the output is asserted. This has no effect when the dot is specified as a square.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Output on Angle of Distortion Definition:

When enabled, the angle of distortion of decoded symbols are compared to the angle of distortion threshold. If the value does not meet the threshold standard, the output is asserted.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Dot Center Offset Definition:

This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Default:

Grade B (>2-5%)

Options:

0 = Grade A (0-2%) 2 = Grade C (>5-10%)

Quadrus EZ Reader User’s Manual

1 = Grade B (>2-5%) 3 = Grade D (>10%)

5-49

Page 166

Output 1 Parameters

Cell Fill Definition:

This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>

Default:

Grade B (>2-5%)

Options:

0 = Grade A (0-2%) 2 = Grade C (>5-10%)

1 = Grade B (>2-5%) 3 = Grade D (>10%)

Dot Ovality Threshold Definition:

This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=2% (Grade A) is from Aerospace Direct Part Marking Standard Spec. This has no effect when the dot is specified as a square.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>

Default:

Grade B (>2-5%)

Options:

0 = Grade A (0-2%) 2 = Grade C (>5-10%)

1 = Grade B (>2-5%) 3 = Grade D (>10%)

Angle of Distortion Threshold Definition:

This adopts the ISO/IEC 15415 letter grades to be compared against to determine whether or not to assert the output. The <=3.5 degree (Grade A) and >3.5 to 7 degree (Grade B) are from Aerospace Direct Part Marking Standard Spec.

Serial Cmd:

<K820,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion,angle of distortion threshold>

Default:

Grade C (>7-10.5)

Options:

0 = Grade A (0-3.5 degree) 2 = Grade C (>7-10.5)

5-50

1 = Grade B (>3.5-7) 3 = Grade D (>10.5-14)

Quadrus EZ Reader User’s Manual

Page 167

I/O Parameters

Diagnostics

Note: When Diagnostic Warning is enabled under Output 1 Parameters, Output Mode configuration has no effect.

Usage:

Alerts user to critical conditions.

Definition:

Sets up specific warnings that will cause activation on output 1. The output will remain active as long as one of the diagnostic conditions is met. The output will go inactive as soon as it detects no active diagnostic warning.

Over Temperature Definition:

Setups the output to toggle active when an over-temperature condition is detected.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Service Unit Definition:

Setups the output to toggle active when the service timer has expired. This condition will only be held for one service timer tick.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

External Camera Disconnect Note: This feature cannot be used if in a Continuous Read mode. Definition:

Setups the output to toggle active if the external camera goes off-line.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

Quadrus EZ Reader User’s Manual

1 = Enabled 5-51

Page 168

Output 2 Parameters

Output 2 Parameters Note: Output 2 has the same parameters and default settings as Output 1.

Serial Cmd:

<K811,output on,output state,pulse width,output mode>

Trend Analysis to Output 2 Note: Output On under Output 2 Parameters must be set to Trend Analysis for this output to function.

Serial Cmd:

<K781,trend analysis mode,number of triggers,number to output on>

Symbol Quality to Output 2 (ISO/IEC 15415) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.

Serial Cmd:

5-52

<K801,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>

Quadrus EZ Reader User’s Manual

Page 169

I/O Parameters

Symbol Quality to Output 2 (Inkjet/Direct) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.

Serial Cmd:

<K821,output on dot center offset,dot center offset,output on cell fill,cell fill,output on dot ovality,dot ovality threshold,output on angle of distortion, angle of distortion threshold>

Diagnostic Warnings to Output 2 Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.

Serial Cmd:

<K791,over temp,service unit,external camera disconnect>

Quadrus EZ Reader User’s Manual

5-53

Page 170

Output 3 Parameters

Output 3 Parameters Output 3 has the same parameters and default settings as Output 1.

Serial Cmd:

<K812,output on,active state,pulse width,output mode>

Trend Analysis to Output 3 Note: Output On under Output 3 Parameters must be set to Trend Analysis for this output to function.

Serial Cmd:

<K782,trend analysis mode,number of triggers,number to output on>

Symbol Quality to Output 3 (ISO/IEC 15415) Note: Output On under Output 3 Parameters must be set to Symbol Quality for this output to function.

Serial Cmd:

5-54

<K802,Output on Symbol Contrast,Symbol Contrast Threshold,Output on Print Growth,Print Growth Threshold,Output on Axial Non-Uniformity,Axial Non-Uniformity Threshold,Output on Unused ECC,Unused ECC Threshold>

Quadrus EZ Reader User’s Manual