Page 61

Communications Commands

Communications Commands Note: Changes in Communications parameters or assigning an address to the decoder can cause loss of communications with the configuration terminal when you exit the menu program (whether or not changes are saved for power-on).

Protocol If selecting one of the options from 0 to 4 (Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, or Polling Mode D), use this format:

Example: To change the Protocol to Polling Mode D, enter <Kf4>. If selecting Multidrop (5), you must define an address and add it to the format (data B). Format: <Kfprotocol,address> protocol: 5 = Multidrop (requires address)

address: Any number from 1 to 50.

Example: To change the Protocol to Multidrop with an address of 33, enter <Kf5,33>. If selecting User Defined (6) or User Defined Multidrop (7), you must complete the format by either choosing new parameters or concatenating unchanged data fields (separate by commas). Tip: For User Defined, you first select Point-to-Point <Kf0> and then User Defined <Kf6...>. For user Defined Multidrop, you first select Multidrop <Kf5>, then User Defined Multidrop <Kf7...>.

MS-3000 Dual Head Decoder User’s Manual

3-5

3–Serial Config.

Format: <Kfprotocol> protocol: 0 = Point-to-Point 1 = Point-to-Point with RTS/CTS 2 = Point-to-Point with XON/XOFF 3 = Point-to-Point with RTS/CTS & XON/XOFF 4 = Polling Mode D 5 = Multidrop (requires address) 6 = User Defined 7 = User Defined Multidrop

Page 62

Chapter 3

Serial Configuration Format: <Kfprotocol,RES,address,REQ,EOT,STX,ETX,ACK,NAK > Protocol: 6 = User Defined 7 = User Defined Multidrop Note: Address can be assigned any ASCII character except a null. Control characters are used to define RES through NAK (except Address). Table 3-2 lists the control characters used for these data fields.

Example: To select an unpolled ACK/NAK User Defined protocol with LRC disabled, enter <Kf0><Kf6,,,,,,,^F,^U><Kc0>.1 ACK and NAK will be displayed in the menu. 2 Table 3-2 Protocol Commands

Protocol Command

Control Characters

(Mnemonic displayed on (Entered in menu or serial Microscan menu) command)

3–Serial Config.

RES

Effect of Command

^D

Reset

REQ

^E

Request

EOT

^D

Reset

STX

^B

Start of Text

ETX

^C

End of Text

ACK

^F

Acknowledge

NAK

^U

Negative Acknowledge

Preamble Format: <Kdstatus,preamble character(s)> preamble character(s): status: Enter one or two preamble characters from table A-2 on 0 = Disabled page A-3, except a null (00H). Default is ^M. 1 = Enabled Example: To enable Preamble with just one character, an FF (form feed), enter <Kd1,^L>.

1. <Kf0> nulls the address and <Kc0> disables LRC. 2. A control character, although conventionally represented here and in the ASCII table on page A-3 as two characters (^F or ^U, etc.), is actually a single ASCII character that is entered on the keyboard by holding down the control key while pressing the desired letter.

3-6

MS-3000 Dual Head Decoder User’s Manual

Page 63

The text of this page cannot be shown here; see it in the PDF.

Page 64

Chapter 3

Serial Configuration

Host Port Note: Changes made in the decoder’s communications parameters such as baud rate, parity, stop bits, LRC, etc., must be matched in the other device(s) or communications will be lost. If this occurs, default the decoder as described in Appendix C, “Defaulting the Decoder.”

Format: <Kabaud rate,parity,stop bits,data bits> baud rate: parity: 0 = 600 4 = 9600 3 = 4800 0 = None 1 = 1200 5 = 19.2K 1 = Even 2 = 2400 6 = 38.4K 2 = Odd stop bits: data bits: 1 = Two 0 = One 0 = Seven

1 = Eight

Example: To change the baud rate to 2400, enter <Ka2>. Example: To change Data Bits to Eight without changing any other fields, enter either: <Ka,,,1> or <Ka4,1,0,1>.

3–Serial Config.

RS-422 Format: <KbRS-422 status> RS-422 status: 1 = Enabled 0 = Disabled Example: To enable RS-422, enter <Kb1>. This command assumes the decoder is in RS-232 before RS-422 is enabled. If a multidrop address has been already assigned, the decoder will be in RS-485 communications, regardless of RS-422 status.

Auxiliary Port Format: <Kystatus,baud rate,parity,stop bits,data bits> status: baud rate: parity: 0 = 600 4 = 9600 0 = None 0 = Disabled 1 = 1200 5 = 19.2K 1 = Transparent 1 = Even 2 = 2400 6 = 38.4K 2 = Odd 2 = Half Duplex 3 = 4800 3 = Full Duplex stop bits: data bits: 1 = Two 1 = Eight 0 = One 0 = Seven Example: To enable Half Duplex and to change the baud rate to 2400, enter <Ky2,2>.

3-8

MS-3000 Dual Head Decoder User’s Manual

Page 65

Operations Commands

Operations Commands Independent Mode Format: <Kgstatus,scanner #1,scanner #2> status: scanner #1: 0 = Disabled 0 = Continuous Read 1 = Continuous Read 1 Output 1 = Enabled 2 = External 3 = Serial 4 = External and Serial scanner #2: 0= External 1= Serial 2= External and Cereal

Example: To enable Independent mode and change Scanner #1 and Scanner #2 to Serial, enter: <Kg1,3,1>

Sequenced Mode

Example: To enable Sequenced mode, change System Trigger to Serial, Number of Passes to Multiple, enable System Timeout, and set System Timeout to 12 seconds, enter: <Kh1,1,,1,1,1200>.

Level or Edge Format: <Kimode> mode: 0 = Edge 1 = Level Example: To change to Edge mode, enter: <Ki0>.

MS-3000 Dual Head Decoder User’s Manual

3-9

3–Serial Config.

Format: <Khstatus,system trigger,# labels, # passes,system timeout status, system timeout> status: system trigger: # labels 0= External 0 = Disabled 0 = One 1= Serial 1 = Enabled 1 = Two 2= External and Serial # passes system timeout status: system timeout: 0 = One 0 = Disabled Any number from 1 to 65,535 (10 ms inc.) Default is 600 1 = Multiple 1 = Enabled (6 seconds).

Page 66

Chapter 3

Serial Configuration

Trigger Polarity Format: <Kjpolarity> polarity: 0 = Negative 1 = Positive Example: To change External Trigger Polarity to Negative, enter <Kj0>.

Scanner #1 and #2 Setup

3–Serial Config.

Format: (#1) <Kkend of read cycle,timeout,trigger character,noread message status,noread message,scanner identifier status,scanner identifier> Format: (#2) <Knend of read cycle,timeout,trigger character,noread message status,noread message,scanner identifier status,scanner identifier> end of read cycle: timeout: serial trigger character: noread message 0 = Timeout Any number from Scanner #1: status: 1 to 65,535 default is ^] 1 =New Trigger 0 = Disabled (10 mS increments) Scanner #2: 2 = Timeout & 1 = Enabled default is ^^ New Trigger noread message: scanner identifier Any ASCII string status: 0 = Disabled up to 7 digits. Default is Noread. 1 = Enabled scanner identifier: Any ASCII character(s) up to 2 digits. Scanner #1: default is 1/ Scanner #2: default is 2/

Example: To enable Timeout and New Trigger, set Timeout to 5 seconds, and disable Scanner Identifier, enter: <Kk2,500,,,,0>.

Bar Code Output Format: <Klstatus,when to output> status: when to output: 0 = Disabled 0 = As Soon As Possible 1 = Enabled 1 = End of Read Cycle Example: To set Bar Code Output to End of Read Cycle, enter <Kl1,1>.

Number of Reads Before a Good Decode Format: <Kmnumber of reads> number of reads: Any number from 1 to 31. Default is 1. Example: To change Number of Reads to 3, enter <Km3>.

3-10

MS-3000 Dual Head Decoder User’s Manual

Page 67

Code Types Commands

Code Types Commands Narrow Margins Format: <Kostatus> status: 0 = Disabled 1 = Enabled Example: To enable Narrow Margins, enter <Ko1>.

Code 39

Example: To set Fixed Code Length to 30, enter <Kp,,,,1,30> or <Kp1,0,0,0,1,30>.

Codabar Format: <Kqstatus #1,start & stop match,start & stop output,large interchar. gap,fixed code length,code length,status #2> status #1: start & stop match start & stop status: output status: 0 = Disabled 0 = Disabled 0 = Disabled 1 = Enabled 1 = Enabled 1 = Enabled large interchar. fixed code code length: status #2: gap status: length status: Any number 0 = Disabled 0 = Disabled 0 = Disabled from 1 to 31. 1 = Enabled Default is 10. 1 = Enabled 1 = Enabled Example: To enable Codabar and set Fixed Code Length to 9, enter <Kq1,,,,1,9> or <Kq1,1,1,0,1,9>.

MS-3000 Dual Head Decoder User’s Manual

3-11

3–Serial Config.

Format: <Kpstatus #1,check digit status,check digit output,large interchar. gap status,fixed code length status,code length,status #2> status #1: check digit status: check digit output status: 0 = Disabled 0 = Disabled 0 = Disabled 1 = Enabled 1 = Enabled 1 = Enabled large interchar. fixed code length code length: status #2: gap status: status: Any number 0 = Disabled 0 = Disabled 0 = Disabled from 1 to 31. 1 = Enabled Default is 10. 1 = Enabled 1 = Enabled

Page 68

Chapter 3

Serial Configuration

I 2 of 5 Format: <Krstatus #1,check digit status,check digit output status,code length #1,code length #2,status #2> status #1: check digit status: check digit output status: 0 = Disabled 0 = Disabled 0 = Disabled 1 = Enabled 1 = Enabled 1 = Enabled code length #1: code length #2: status #2: Zero or any even Zero or any even number 0 = Disabled number from 2 to 30. from 2 to 30. Default is 6 1 = Enabled Default is 10. Example: To enable I 2 of 5 and to set Fixed Code Length #1 to 8 and Fixed Code Length #2 to 4, enter <Kr1,,,8,4> or <Kr1,0,0,8,4>.

UPC/EAN 3–Serial Config.

Format: <Ksstatus #1,EAN status,supplementals status,separator status,separator character,status #2> status #1: EAN status: supplementals status: 0 = Disabled 0 = Disabled 0 = Disabled 1 = Enabled 1 = Enabled 1 = Enabled (UPC must also be enabled) 2 = Required separator status: separator character: status #2: 0 = Disabled Any ASCII character (except 0 = Disabled NUL). Default is a comma (,). 1 = Enabled 1 = Enabled Example: To enable UPC and EAN, change supplementals to required, and change separator character to a dash (–), enter <Ks1,1,2,1,–> or <Ks1,1,2,,–>.

Code 128 Format: <Ktstatus #1,fixed code length status,code length,status #2> status #1: fixed code length code length: status #2: status: 0 = Disabled Any number 0 = Disabled 0 = Disabled from 1 to 31. 1 = Enabled 1 = Enabled Default is 10. 1 = Enabled Example: To enable Code 128, enable Fixed Code Length, and set Code Length to 9, enter <Kt1,1,9>.

3-12

MS-3000 Dual Head Decoder User’s Manual

Page 69

The text of this page cannot be shown here; see it in the PDF.

Page 70

Chapter 3

Serial Configuration

Relay Driver Usage Format: <Kvrelay driver usage> relay driver usage: 0 = Good Read 1 = Noread Outputs to pin 2 on trigger port #1. Example: To change Relay Driver Usage to Noread, enter <Kv1>.

Data Line Usage Format: <Kzrelay driver usage> data line usage: 0 = Good Read 1 = Noread Outputs to pins 6 and 8 on host port and corresponding to inputs from scan heads #1 and #2 respectively. Example: To change Data Line Usage to Noread, enter <Kz1>.

3–Serial Config.

Data Line Polarity Format: <Kwpolarity> polarity: 0 = Negative 1 = Positive Applies to pins 6 and 8 on host connector and pin 2 on trigger port #1. Example: To change Data Line Polarity to Negative, enter <Kw0>.

Data Line Pulse Width Format: <Kxpulse> duration of pulse: Any number from 0 to 255. Default is 5 (50 ms). Example: To set Good/Bad Pulse Width to 600 ms, enter <Kx60>. Applies to pins 6 and 8 on host connector and pin 2 on trigger port #1.

Command Start Character Format: <KEcommand start character> command start character: Any ASCII character. Default is <. Example: To change Command Start Character to a colon (:), enter <KE:>. Note: Subsequent commands must start with a colon (:). For example, enter :D> to access the menu, or :KE<> to change back to the default character. 3-14

MS-3000 Dual Head Decoder User’s Manual