Page 51

User Outputs Menu

User Outputs Menu The User Outputs menu allows you to configure the decoder’s output. – Bold text represents default settings.

User Outputs Beeper Status

Beeper Volume – Level 4

– Disabled

– Level 1

– On Noread

– Level 5

Full Screens

Laser On/Off

– Enabled

– Fast

– Disabled

– Disabled

Reverse Video

– Enabled

– Disabled

– Slow

– Enabled

– Level 2 – Level 3

Relay Driver Usage – Good Read – Noread

Data Line Usage – Good Read – Noread

Data Line Polarity – Positive

– Negative

Data Line Pulse Width* – 50 ms

– User Definable

(0 to 255)

Command Start Character – <

– User Definable

(ASCII char.)

* Divide the number entered on the command line by 100 for time in s econds.

Figure 2-7 User Outputs Menu Structure

Beeper Default: On Good Options: On Good, On Noread, Disabled A beep sounds after each good read of a bar code label, after each noread, or not at all, according to the setting. Note: The beep period will be short for triggered modes where a new trigger occurs immediately or the output is delayed to the end of the read cycle on edge and serial triggers.

Beeper Volume Default: Level 4 Options: Level 4, Level 5, Level 1, Level 2, Level 3

Beeper Speed Default: Fast Options: Fast, Slow

MS-3000 Dual Head Decoder User’s Manual

2-27

2–Menu Config.

– On Good

Beeper Speed

Page 52

Chapter 2

Menu Configuration Allows you to set the relative time for the beeper to respond. When Slow is enabled the beeper takes approximately 80 ms to sound. If your application requires a quicker response than 80 ms, enable Fast.

Full Screens

2–Menu Config.

Default: Enabled Options: Enabled, Disabled When enabled, the full menu screen is displayed. When disabled, only the command line is displayed.

Laser On/Off Default: Disabled Options: Disabled, Enabled Note: A serial or external trigger (see “System Trigger Type,” on page 2-15) must be enabled for Laser On/Off to take effect. When enabled, the laser is ON only during the read cycle. When disabled, the laser operates continuously. Note: Laser On/Off does not relate to the <H> (Enable Laser Scanning) or <I> (Disable Laser Scanning) operational commands in “Device Control Commands,” on page 5-3.

Reverse Video Default: Disabled Options: Disabled, Enabled When enabled, the decoder will read bar code labels with bars that are lighter in color than their corresponding backgrounds.

Relay Driver Usage Default: Good Read Options: Good Read, Noread Allows you to determine whether the occurrence of a good read or a noread will output a TTL pulse of 5V. If and only if the decoder is configured for Sequenced and One Label, the pulse will output to pin 2 on trigger connector #1. All other configurations will direct the output to pins 6 and 8 on the 25-pin host port. (See the following item, “ Data Line Usage.”)

2-28

MS-3000 Dual Head Decoder User’s Manual

Page 53

User Outputs Menu

Data Line Usage

Data Line Polarity Default: Positive Options: Positive, Negative Allows you to choose between positive and negative output signals to pins 6 and 8 on the host connector and pin 2 on trigger connector #1.

Data Line Pulse Width Default: 5 (50 ms) Options: 1 to 255 (10 ms to 2.55 seconds) Allows you to change the duration of the pulses applied to pins 6 and 8 on the host connector and pin 2 on trigger connector #1. Caution : Too wide of a pulse can cause missed labels, especially if it exceeds End of Read Cycle Timeout (see page 2-15).

Command Start Character Default: < Options: Any ASCII character. Allows you to define a new ASCII start character in a serial command.

MS-3000 Dual Head Decoder User’s Manual

2-29

2–Menu Config.

Default: Good Read Options: Good Read, Noread Allows you to determine whether the occurrence of a good read or a noread will output a TTL pulse of 5V. If the decoder is configured for Independent or if configured for Sequenced and Two Labels, the pulse will be sent to the host via pin 6 (for scan head #1) or pin 8 (for scan head #2) on the 25-pin host port.

Page 54

Chapter 2

Menu Configuration

Raster Setup Menu Note: Raster Setup is applicable to the MS-1280 raster scan head only.

The Raster Setup menu allows you to set top and bottom offset values as well as the raster motor speed. – Bold text represents default settings.

2–Menu Config.

Raster Setup (page 2-30)

Status

– Enabled

– Disabled

(page 2-31)

Top Offset in Degrees – 0

– User

Definable (0 to 45)

(page 2-31)

(page 2-31)

Bottom Offset in Degrees

Sweeps per Second

– 45

– 14

– User Definable

– User

(0 to 45)

Definable (0 to 135)

Figure 2-8 Raster Setup Menu Structure

MS-1280 Raster Scan Head

Top Offset = 0°

Bottom Offset = 45° Figure 2-9 Raster Sweep Arc

Table 2-1 Raster Settings

Sweeps per Second

Raster Arc

Minimum

Maximu m

1° 5° 10° 15° 20° 25° 30° 35° 40° 45°

1 1 1 1 1 1 1 1 1 1

135 75 45 35 29 23 21 19 17 14

Status Default: Enabled Options: Enabled, Disabled For additional information on raster setup see your raster scan head user manual.

2-30

MS-3000 Dual Head Decoder User’s Manual

Page 55

Raster Setup Menu

Top Offset in Degrees Default: 0 (degrees) Options: 0 to 45 (in one-degree increments) Note: The top offset must always be less than the bottom offset or the resulting arc will be 0 degrees. Top Offset in Degrees, together with Bottom Offset in Degrees, allows you to set the raster sweep arc. (See figure 2-9 on page 2-30.) Default: 45 (degrees) Options: 0 to 45 (in one-degree increments) Note: The bottom offset must always be greater than the top offset or the resulting arc will be 0 degrees. Bottom Offset in Degrees, together with Top Offset in Degrees, allows you to set the raster sweep arc. See figure 2-9 on page 2-30.

Sweeps per Second Default: 14 Options: 0 to 135 Allows you to control the raster motor speed. A “sweep” is defined as a single pass, up or down, describing the raster image. Table A-1, “Raster Settings,” on page A-30 shows minimum and maximum possible sweep speeds at selected raster arcs. The minimum possible speed at any arc is always one sweep per second. Maximum sweep speeds are the maximums to which the scan head defaults - even when you enter higher speeds. To maximize the number of scans a label will receive, select as few as possible sweeps per second. However, to ensure that the minimum required scan lines cross the bar code label, it is essential that at least two full sweeps occur during the time it takes for the label to pass through the readable scan width area. For example, if the label is readable in the read range for only 1/10th of a second, then the number of sweeps per second should be at least 20. If readable for only 1 second, then sweeps per second (SPS) should be at least 2. To see the formula for calculating sweeps per second, see your raster scan head user manual.

MS-3000 Dual Head Decoder User’s Manual

2-31

2–Menu Config.

Bottom Offset in Degrees

Page 56 read from a scan by OCR; it may hold reading errors

Chapter 2 Menu Configuration o7) = ¢ ° oO S c o + N

2-32 MS-3000 Dual Head Decoder User’s Manual

Page 57

Chapter

3

Serial Configuration Chapter Contents

This chapter describes how to configure the MS-3000 dual decoder by serial commands from a host. 1 All of the configuration parameters that can be changed in the menus with the exception of Full Screens, as described in Chapter 2, “ Menu Configuration,” can also be changed by serial configuration commands. All of the serial configuration parameters, with the exception of Trigger Filter Timing Value, also appear in the menus. In addition, most of these configuration parameters can also be changed with a Microscan profile card (P/N 99-500011-01), obtainable from your Microscan representative. (See Chapter 4, “ Profile Card Configuration.”)

1. The decoder communicates in full duplex, terminal mode with no handshake. It also recognizes carriage returns and line feeds. The host or ASCII terminal with must match the following default settings before any communication can take place: 9600 Baud Rate, Seven Data Bits, Even Parity, and One Stop Bit. If communicating from an auxiliary terminal via the auxiliary port (pins18 and 19 of the 28 pin connector, the serial command <D> (Enter Menu Configuration Program) is the only command that the decoder will recognize.

MS-3000 Dual Head Decoder User’s Manual

3-1

3–Serial Config.

Summary of Serial Configuration Commands .....................................3-2 Serial Configuration Command Format ..............................................3-3 Concatenating Serial Commands........................................................3-4 Loss of Communications .....................................................................3-4 Trigger Filter Timing Value ..................................................................3-5 Communications Commands...............................................................3-5 Operations Commands........................................................................3-9 Code Types Commands ....................................................................3-11 User Outputs Commands ..................................................................3-13 Raster Setup Commands...................................................................3-15

Page 58

The text of this page cannot be shown here; see it in the PDF.

Page 59

Serial Configuration Command Format The format for a serial configuration command is, <Kparameterdata,data,...etc.><initializing command>

Where: • Less than < and greater than > symbols are included as part of the commands.1 • “parameter,” as used here, are those character(s) that precede the data. • The “initializing command” <A> or <Z> is sent after configuration is complete. <Z> resets and saves for power up. <A> initializes the change to RAM.2 (For more information, see “Program Management Commands,” on page 5-3.) For example, the following command enables UPC and saves the change for power on: <Ks1><Z>.

1. Command start character by default is a left angle bracket, <. It may be redefined by menu or serial command. However, the end character, a right angle bracket, >, cannot be changed. 2. See “Initializing serial configuration commands” on page 3-4 for definitions and examples.

MS-3000 Dual Head Decoder User’s Manual

3-3

3–Serial Config.

When using serial configuration commands, note also the following conventions: • Parameters and data are “case sensitive.” That is, characters must be entered as upper or lower case, as specified. • All data fields (except the last) must be followed by a comma (without a space). • If there is no change in a given field, then commas can be entered alone, or with the existing data (for example, <Ka,,,0> or <Ka4,1,0,0>). • All fields preceding the modified field must be included. For example, in the RS-232 port, to change Data Bits to Eight without changing any other field, enter either: <Ka,,,1> or <Ka4,1,0,1>. • All fields following the modified field can be left out. For example, in the RS-232 port, to change Baud Rate to 4800, enter <Ka3>. (See examples on following pages.)

Page 60

Chapter 3

Serial Configuration Initializing serial configuration commands To ensure that a serial configuration command will take effect, you need to follow it with one of the operational commands below: <A> To reset but not save changes for power on <Z> To reset and save changes for power on For example, to change Baud Rate and reset without saving changes for power-up, enter <Ka3><A>. To change Baud Rate and reset, saving the changes to NOVRAM, enter <Ka3><Z>.

Concatenating Serial Commands

3–Serial Config.

Commands can also be concatenated (added together) to a maximum of 64 characters in a single string or data block. Additional data blocks of 64 or less characters can be sent provided there is at least a 10 ms pause between blocks. 1 For example, the following command, <Kc1><KA24><Ko1><A> enables LRC, sets the Host Protocol Response Timeout to 24 ms, enables Narrow Margins and resets the data buffers (without saving the changes for power-on).

Loss of Communications Assigning a multidrop address to a decoder or making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, try gaining access to the decoder by use of a profile card 2 or by entering a <D> command from an auxiliary terminal via the auxiliary RS-232 port. If neither of these methods is available, try defaulting the decoder. See Appendix C, “ Defaulting the Decoder,” on page A-4.

Trigger Filter Timing Value Format: <KF?> Returns the current trigger filter timing value in milliseconds. Format: <KFone tenth trigger filter timing value> Allows you to set trigger filter timing. Divide the desired number of milliseconds by 10 and enter the quotient. 1. Data in excess of 64 characters will reset the buffer, causing the first 64 characters of the string to be lost and indicator #1 to illuminate red 2 seconds after the end of each read cycle, until the decoder’s memory is reset. 2. See mode 0 in Chapter 4, “Profile Card Configuration.”

3-4

MS-3000 Dual Head Decoder User’s Manual