MS-3000 Dual Head Decoder User's Manual, pages 21 to 30

Pages 21 to 30 of 126

The original document is the PDF (779 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan MS-3000

Page 21

The text of this page cannot be shown here; see it in the PDF.

Page 22

1–Setup and Inst.

Chapter 1

Setup and Installation If the unit does not have housing, use the measurements provided in figure 1-13 to locate the centers of the mounting holes. Front

1.0" (25.4)

Measurements in inches

4.5" (114.30)

5.8" (147.32)

3.8" (96.52)

0.160" (4.06) DIA (4 places)

5.2" (132.08)

0.030" (7.62) 5.8" (147.32) Figure 1-13 Mounting Diagram (without housing, not full size)

1-14

3000 Dual Head Decoder User’s Manual

Page 23

Ground and Shield Considerations

Proper grounding is necessary for operator safety, noise reduction, and the protection of equipment from voltage transients. Buildings–including any steelwork, all circuits, and all junction boxes–must be grounded directly to an earth ground in compliance with local and national electrical codes. RS-232 signals have a common signal ground (pin 7 of the 25-pin connector). Pin 7 is normally connected to pin 1 (chassis ground) in the decoder; however, under certain conditions (e.g., when potential differences exist between power outlet grounds) signal and chassis grounds can be isolated from each other inside the decoder by Microscan technicians. Any data line, as necessary, can be shielded. If shielding is used, isolate it from the decoder and ground only to the host earth ground.

Noise Interference Noise interference can be minimized if cabling subject to HOST noise interference is twisted and/or Shielding shielded or encased Communications Cable in grounded conduit, and the conduit or Earth Ground shielding (“drain” line) is grounded only to DECODER earth ground at the host, as shown in figure 1-14. You might Figure 1-14 Grounding Diagram, Decoder-Host need to examine and if necessary cut the shielding connection at or near the concentrator cable connector.

Ground Loops Ground loops, signal degradation due to different ground potentials in communicating devices, can be eliminated or minimized by ensuring that the host, concentrator, and their power supplies are connected to a common earth ground.

3000 Dual Head Decoder User’s Manual

1-15

1–Setup and Inst.

Ground and Shield Considerations

Page 24 read from a scan by OCR; it may hold reading errors

Chapter1 Setup and Installation me) o = so} c © 2. E} Sond ® 2 =

1-16 3000 Dual Head Decoder User’s Manual

Page 25

Chapter

2

Menu Configuration Communications Menu ........................................................................2-5 Host Protocol Parameters....................................................................2-6 Host Port Parameters ........................................................................2-10 Aux Port Parameters..........................................................................2-11 Operations Menu ...............................................................................2-12 Code Types Menu..............................................................................2-21 User Outputs Menu............................................................................2-27 Raster Setup Menu............................................................................2-30

This chapter describes how to configure the MS-3000 dual head decoder with on-screen menu commands from a host or auxiliary terminal. All keystrokes are in bold typeface. Default parameters in the menu structures are also in bold typeface. All of these parameters, with the exception of Full Screens, can also be changed by serial commands (see Chapter 3, “ Serial Configuration”). In addition, most of these configuration parameters can also be changed with a Microscan profile card (P/N 99-500011-01, see Chapter 4, “ Profile Card Configuration”).

Communicating with an ASCII Terminal The MS-3000 dual head decoder communicates in full duplex, terminal mode with no handshake. It also recognizes carriage returns and line feeds. The host or ASCII terminal with must match the following default settings before any communication can take place: 9600 Baud Rate, Seven Data Bits, Even Parity, and One Stop Bit. A PC or Macintosh computer can be used as an ASCII terminal if connected as shown under “Host Connector” and running a communications program set to the above defaults. See your computer manual for communication’s port pinouts.

MS-3000 Dual Head Decoder User’s Manual

2-1

2–Menu Config.

Chapter Contents

Page 26

Chapter 2

Menu Configuration

Entering the Menu Configuration Program

2–Menu Config.

To see the Main menu (figure 2-1), from an ASCII terminal that is connected to the decoder, send the serial command <D> (enter the < > brackets as well as the upper case D).1

Figure 2-1 Configuration Program - Main Menu2

If the menu does not appear, see Appendix D, “ Troubleshooting,” on page A-6.

Using the Menu Configuration Program The bottom line on the screen is called the command line. The command line identifies your place in the menu program, shows current status and allows you to review and change options. Use the designated keys 3 to scroll to and select the parameter you wish to change; press SP (space bar key) or N to scroll ahead, B to scroll back, CR (carriage return key) to select, and M to return to the previous higher level menu. To return to the Main menu at any time, press ESC (escape key) and M. 1. Command start character by default is a left angle bracket, <. It may be redefined by menu or serial command. However, the end character, a right angle bracket, >, cannot be changed. 2. Item 5, Raster Setup, applies only to the MS-1280. 3. The menu navigational keys are displayed in each menu.

2-2

MS-3000 Dual Head Decoder User’s Manual

Page 27

For example, to enable LRC (see figure 2-2, “ Communications Menu Structure,” on page 2-5 and “Longitudinal Redundancy Check ” on page 2-9), you would use the following command line path: From the Main menu, press CR at the Communications prompt (this is the first prompt displayed in the Menu Configuration Program) to access the Communications menu.

Communications, > Host Protocol Communications, > Host Port Communications, > Aux Port

Since LRC is a subtopic of Host Protocol, press CR to access the Host Protocol parameters.

Host Protocol, > Protocol Host Protocol, > Preamble = ^M Host Protocol, > Preamble = Disabled Host Protocol, > Postamble = ^M^J Host Protocol, > Postamble = Disabled Host Protocol, > LRC = Disabled Host Protocol, > Response Timeout in 1ms incs = 12 Host Protocol, > Intercharacter Delay in 1ms incs = 0 Host Protocol, > LRC Host Protocol, > LRC

= Disabled = Enabled

Protocol is the first parameter under Host Protocol. Press SP until you reach LRC, then press CR.

To enable LRC, press CR, SP, and CR.

To view LRC’s new status in the menu, press M to refresh the screen. To return to the Main menu, press M again. You can make additional changes within another menu before exiting the program. Simply follow the same method of scrolling to and selecting each main topic, then its subtopics, until you reach the parameter you want to change. Remember, to return to the Main menu at any time, press ESC (escape key) and M. Some parameters are user defined, in which case they prompt you with an arrow for data, such as:

At the prompt, redefine the parameter within the allowable range, and press CR to enable.

MS-3000 Dual Head Decoder User’s Manual

2-3

2–Menu Config.

Main, > Communications Main, > Operations Main, > Code Types Main, > User Outputs Main, > Raster Setup

Page 28

Chapter 2

Menu Configuration

Saving Menu Changes Press ESC (escape key) to see the following on the command line: Press M to return to the Main menu, or press E to exit the Menu Configuration program. If E is pressed, the following question will appear:

2–Menu Config.

Press N to exit without saving changes, or press Y to retain the current settings to non-volatile RAM for power up. If Y is selected, a beep will indicate the save has been carried out.

Loss of Communications Defaulting might be necessary if communications between the decoder and another device are interrupted or if using incompatible equipment (for example, a terminal is set to communicate at 9600 baud, but the decoder is configured at 38.4K baud). Communication can also be lost if an address has been assigned to the decoder. To reset parameters to default values, see Appendix C, “ Defaulting the Decoder,” on page A-4.

Defining Special Characters To define any control character from the ASCII table: Press SP once, then enter the control character by holding down the control key and simultaneously pressing the desired character. For example to define a line feed, press SP, then Control and J simultaneously. It is displayed as ^J on the command line and as <LF> in the menu when the screen is refreshed. To define CR as a character: Press SP, then CR. It is displayed as ^M on the command line and as <CR> in the menu when the screen is refreshed. To define a space as a character : Press SP twice. It is displayed as a blank space in the menu when the screen is refreshed. While it appears that nothing has been assigned, the hex value 20 will be sent during data transmission. To select NUL as the character: Press SP, then a 0 (zero). It is displayed as <NUL> in the menu when the screen is refreshed.

2-4

MS-3000 Dual Head Decoder User’s Manual

Page 29

Communications Menu

Communications Menu The Communications menu allows you to set the communication protocols between the decoder and the host. Communications

– Bold text represents default settings.

Host Protocol Protocol

Preamble

– Point-to-Point w/RTS/CTS

Preamble

– ^M

– User w/XON/XOFF

Postamble

– ^M ^J

– Enabled

Definable (ASCII char.)

– Point-to-Point

Postamble

– Disabled

– Disabled

– User

– Enabled

Definable (ASCII char.)

– Point-to-Point w/RTS/CTS & XON/XOFF – Polling Mode D – Multidrop – Address = 1 – User Definable (1 to 50) – User Defined – RES – Address = ^A

– REQ

– EOT

– STX – ETX

– ACK

– NAK

– REQ

– EOT

– STX – ETX

– ACK

– NAK

– User Defined Multidrop –

RES

– Address = ^A

Response Timeout

LRC

Interchar Delay

– 12

– Disabled

– 0

– User Definable (0 to 65,000)

– Enabled

– User Definable (0 to 255)

Host Port Baud Rate

Parity

– 1200

– Even

– 38.4K – 4800

– None

– 9600

– 19.2K – 2400 – 600

Stop Bits

– Odd

Data Bits

RS-422

– One

– Seven

– Disabled

Parity

Stop Bits

Data Bits

– One

– Seven

– Two

– Eight

– Enabled

Aux Port Aux Mode

Baud Rate

– Disabled

– 9600

– 1200

– Even

– Half Duplex

– 38.4K – 4800

– None

– Transparent – Full Duplex

– 19.2K – 2400 – 600

– Odd

– Two

– Eight

Figure 2-2 Communications Menu Structure

MS-3000 Dual Head Decoder User’s Manual

2-5

2–Menu Config.

– Point-to-Point

Page 30

Chapter 2

Menu Configuration There are three subtopics in this menu: Host Protocol, Host Port, and Aux Port. To help visualize the menu’s organization and to locate the page number where each topic is described, see figure 2-2, “ Communications Menu Structure,” on page 2-5. Note: Changes in Communications parameters or assigning an address to the decoder can cause loss of communications with the configuration terminal when you exit the menu program (whether or not changes are saved for power-on).

2–Menu Config.

Host Protocol Parameters Protocol Default: Point-to-Point Options: Point-to-Point, Point-to-Point with RTS and CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, Polling Mode D, Multidrop, User Defined, User Defined Multidrop Protocols define the sequence and format in which information is transferred between devices. Generally there are two basic protocol modes: unpolled and polled. In unpolled mode (all of the Point-to-Point protocols), a device sends information without being asked for by the host. In polled mode (Multidrop, Polling Mode D, and User Defined Multidrop), a device has an address and waits for a request from the host before sending data. Note: User Defined and User Defined Multidrop have more options available to them. See individual topics. Selecting:

Has this effect:

Point-to-Point

Has no address and sends data to the host (RS-232) whenever it is available and without any request or handshake from the host.

Point-to-Point with Used only with RS-232. This is a simple handshaking RTS/CTS (Request-to- protocol that allows a device to initiate data transfers to Send/Clear-to-Send) the host with an RTS (request-to-send) transmission. The host, when ready, responds with a CTS (clear-tosend) and the data is transmitted. RTS and CTS signals are transmitted over two dedicated wires (pins 4 and 5) as defined in the RS-232 standard.

2-6

MS-3000 Dual Head Decoder User’s Manual