The text of this page cannot be shown here; see it in the PDF.

Page 12

1–Setup and Inst.

Chapter 1

Setup and Installation Trigger Connector The trigger connector (TRIG) is a 6-pin DIN socket (figure 1-3) that mates with a 240 degree 6-pin DIN plug. Pin assignments are shown in table 1-2 . Table 1-2 Trigger Connector Pin Assignments 6

1

Pin

5 TRIG

4 3

2

Figure 1-3 Trigger Connector Socket

Function

1

Trigger Input +3 VDC to + 24 VDC

2

TTL Relay driver output signal 5 VDC

3

+5 VDC

4

+12 VDC

5

Ground

6

Not used

Pin 1 is the input from the object detector. When operating the decoder in external trigger mode, a toggle at this pin causes the decoder to begin a read cycle. Pin 2 of the trigger connector is a programmable relay driver. The MS-3000 software can be programmed to set this pin high or low upon a good read, a no read, a good match, or a mismatch. This pin can source or sink 4 mA (maximum) and can be used to drive a small relay to operate an alarm, diverter, etc. Microscan offers an optical package detector (P/N 99-440001-03) that plugs directly into this connector and a user-customized trigger port connector (P/N 20-600090-02). 1 Figure 1-4 shows examples of positive and negative external trigger inputs that could be applied to the trigger connector. (Shown in untriggered state.)

1. Trigger sources other than the Microscan object detector can be used. Mechanical switches, relays, etc. - which tend to be slow and bouncy and produce multiple trigger signals - are not recommended unless equipped with optical sensors or filtered transitions (optical, Hall effect, or DC solid state relays).

1-4

3000 Dual Head Decoder User’s Manual

Page 13

This page is a scan with no text; see it in the PDF.

Page 14

1–Setup and Inst.

Chapter 1

Setup and Installation

Back Panel Connectors On the rear panel (figure 1-6) there are the following four connectors: a. Power (5-pin DIN socket) b. LAN (modular RJ-11) c. Host (25-pin D-subminiature socket) d. Monitor (9-pin D-subminiature socket)

HOST

POWER

MONITOR

LAN

Figure 1-6 Rear Panel of MS-3000 Dual Head Decoder

Power Connector The power connector (figure 1-7) has a 5-pin DIN socket with pin assignments shown in table 1-3. You may also supply the required DC voltages yourself. A mating connector (Switchcraft #05BL5M plug) is required. Caution: Switching power supplies for Microscan equipment with a switching noise of 20 mV or greater with ±12 VDC are not recommended due to excessive ripple characteristics. Table 1-3 Power Connector Pin Assignments

1

Pin

3 5

POWER

2

4

Figure 1-7 Power Connector Socket

1-6

Function

1

DC Ground

2

Chassis Ground

3

+5 VDC

4

–12 VDC

5

+12 VDC

3000 Dual Head Decoder User’s Manual

Page 15

Attach Cabling The Local Area Network (LAN) connector allows the MS-3000 dual head decoder to communicate with a multidrop concentrator via a 4-wire cable (preferably shielded) with a 6-pin, 6-wire, RJ-11 type connector. Pin assignments are as shown in table 1-4. The LAN connector is used when a MS3000 dual head decoder is configured for RS-485 multidrop communications.

1

6

Table 1-4 LAN connector Pin Assignments

Pin

TRIG

SCAN

Figure 1-8 LAN Connector

Function

1

Ground

2

Receive Data (+)

3

Receive Data (–)

4

Transmit Data (+)

5

Transmit Data (–)

For proper operation of RS-485 multidrop communications, the main cable should not exceed 4000 feet, with each drop no more than 10 feet. The Microscan MS-5000 multidrop concentrator can support up to 50 decoders or other multidrop devices on one RS-485 line. Note that the last device must be terminated correctly. RS-485 pinouts are also available at the host connector.

Host Connector The host connector (figure 1-9 on page 1-9), a 25-pin D-subminiature socket, allows the decoder to be connected to a host a concentrator, or other communications device such as a PLC (programmable logic controller), a monitor, a PC, a relay, a diverter, an alarm, etc.

Note: All Microscan products are configured as DTE at the host connector when in RS-232 operation.

Caution: Do not use a host cable with more wires connected than are required for the application. The host connector of the decoder has many outputs that could cause damage or interfere with normal operation if connected and improperly used.

3000 Dual Head Decoder User’s Manual

1-7

1–Setup and Inst.

LAN Connector

Page 16

1–Setup and Inst.

Chapter 1

Setup and Installation Table 1-5 Host Connector Pin Assignmentsa

Pin

Function

1 2

Chassis ground Transmit data

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

Comment

Connected to DC ground +12V data output from the decoder ±12V signal indicating data from the host to the Receive data decoder ±12V signal asserted high by the decoder when it Request-to-send has data to send to the host ±12V signal asserted low by the host to stop the Clear-to-send decoder from sending data to it (data transmission will resume when the signal is asserted high.) Indicates a good read from scan head #1 TTL #1 ±5V signal (source -10 mA, sink +0.5 mA) Signal Ground Connected to chassis ground Indicates a good read from scan head #2 TTL #2 ±5V signal (source -10 mA, sink +0.5 mA) TTL +5 VDC Auxiliary supply Trigger input +0 VDC to +24 VDC Default (reset) Resets decoder to default configuration Aux Input (future use) RXDB+ Receive Data B+ TXDA– Transmit Data A– Reserved RXDA– Receive Data A– Profile Card Reserved TXDB+ Transmit Data B+ Data Terminal Ready (asserted high on power-up +12 VDC to indicate decoder is on) Profile Card Input Signal Ground –12 VDC +12 VDC Not used a. The default communications mode does not support pin 4 (RTS) and pin 5 (CTS). If RTS and CTS are not required by the host port, their respective pins should not be wired, as the results will be unpredictable.

A good read will cause a 5V signal (high or low) to be output to pin 6. A noread will cause a 5V signal (high or low) to be output to pin 8.

1-8

3000 Dual Head Decoder User’s Manual

Page 17

Attach Cabling 1–Setup and Inst.

DB-25 DTE to DB-25 DTE Connection

1

De co d er

HOST

Host

Transmit

2

2

Transmit

Rece ive

3

3

Rece ive

Sig n al Grnd

7

7

Sig n al Grnd

25

DB-25 DTE to DE-9 DCE Connection

Figure 1-9 Host Connector

De co d er

Host

Transmit

2

2

Rece ive

Rece ive

3

3

Transmit

Si g n al Gr nd

7

5

Sig n al Grnd

Figure 1-10 DTE and DCE Host Connections

Monitor Connector The monitor connector (auxiliary terminal) connects the decoder to an auxiliary monitor via a 9-pin D-subminiature socket and cable. Table 1-6 shows auxiliary terminal pin assignments. Communication at this connector is RS-232 only.

1 MONITOR

9

Table 1-6 Monitor Connector Pin Assignments

Pin

Function

2

Receive Data (in)

3

Transmit Data (out)

5

Signal Ground

Figure 1-11 Monitor Connector

3000 Dual Head Decoder User’s Manual

1-9

Page 18

Chapter 1

Setup and Installation

1–Setup and Inst.

3 Configure Decoder

Settings for Communications, Operations, Code Types User Outputs, and Raster Setup are stored in non-volatile memory and can be configured by menu, serial command, or profile card commands. For explanations of configuration settings, see Chapter 2, “ Menu Configuration.” To establish communications you will need to match the host’s or auxiliary terminal’s communication settings with your decoder’s settings (see Communications Menu” on page 2-5). Also make certain that the code type enabled in the decoder matches that of the label being used (see Code Types Menu” on page 2-21).

Hint: The <P> command allows your decoder to read all of the listed code types. Communicating with and ASCII Terminal The MS-3000 dual head decoder communicates in full duplex, terminal mode with no handshake. It also recognizes carriage returns and line feeds. The host or ASCII terminal must match the following default settings before any communication can take place: 9600 Baud Rate, Seven Data Bits, Even Parity, and One Stop Bit. A PC or Macintosh computer can be used as a dumb terminal if connected as shown in table 1-5 on page 1-8 and running a communications program set to the above defaults. See your computer user’s manual for communication’s port pinouts.

1-10

3000 Dual Head Decoder User’s Manual

Page 19

Position Scan Head and Label

Before testing, you will need to position the scan head and label in a manner that matches as nearly as possible the actual conditions of your application. Consult your scan head user’s manual for important setup specifications.

3000 Dual Head Decoder User’s Manual

1-11

1–Setup and Inst.

4 Position Scan Head and Label

Page 20

Chapter 1

Setup and Installation

1–Setup and Inst.

5 Do Read Rate Test

This test calculates the percentage of scans decoded. It is useful during setup procedures to help optimize scan head-to-label orientation. a. Place the label used in your application in front of the scan head and within the desired read range.

Note: Ensure that the label being scanned is of a code type enabled in the decoder’s software. Hint: Read rates are easier to read on the screen if Postamble is enabled.

100%FIFTY 100%FIFTY 100%FIFTY 100%FIFTY b. Enter < > to the decoder to start testing (see the “Enter Read Rate Test” command on page 5-5). A percentage number from 0 to 100 will be displayed on the monitor in the read rate test indicating the ratio of good reads per total number of scans.

Read Rate Percentage Label Data c. Find the correct read rate area by moving the label in and out and right and left while observing read rate on the monitor. Avoid the specular reflection area (see scan head user’s manual). d. Record the range area measurements and file the test results for future reference. If the results are not satisfactory, reposition or re-configure the decoder or choose a different narrow-bar-width label size. e. Enter < > to the decoder when testing is complete (see the “Exit Read Rate Test” command on page 5-5).

1-12

3000 Dual Head Decoder User’s Manual