Page 21

Depth of Field by Minimum Bar Code Element Width (continued) Optimal High Density Depth of Field (default)

17

Page 22

Depth of Field by Minimum Bar Code Element Width (continued) Close Depth of Field

18

Page 23

Depth of Field by Minimum Bar Code Element Width (continued) Normal Depth of Field

19

Page 24

Depth of Field by Minimum Bar Code Element Width (continued) Far Depth of Field

20

Page 25

Maintenance Smudges and dirt can interfere with the proper scanning of a bar code. Therefore, the output window will need occasional cleaning. 1. Spray glass cleaner onto lint free, non-abrasive cleaning cloth. 2. Gently wipe the scanner window.

21

Page 26

Troubleshooting Guide The following guide is for reference purposes only. Contact a Metrologic representative at 1-800-ID-METRO or 1-800-436-3876 to preserve the limited warranty terms on page 39. All Interfaces

MS7120 Series Troubleshooting Guide SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

No LEDs, beep or motor spin

No power is being supplied to the scanner

Check transformer, outlet and power strip. Make sure the cable is plugged into the scanner

No LEDs, beep

No power is being supplied to the scanner from host

Some host system’s cannot supply enough current to power Orbit. Use the power supply included with the scanner.

3 beeps on power up

Non-volatile RAM failure

Contact a Metrologic Representative, if the unit will not hold the programmed configuration

Continuous razz tone on power up

RAM or ROM failure

Contact a Metrologic Representative, if the unit will not function

Razz tone and green LED flash at power up

VLD failure

Contact a Metrologic Representative

Razz tone and both LEDs flash at power up

Scanner motor failure

Contact a Metrologic Representative

Unit scans, Communicates and beeps twice

Same symbol timeout set too short

Adjust same symbol timeout for a longer time

22

Page 27

Troubleshooting Guide (continued) SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

The unit powers up, but does not scan and/or beep

Beeper disabled. No tone selected

Enable beeper. Select tone

The unit powers up, but does not scan and/or beep

Scanning a particular symbology that is not enabled

UPC/EAN, Code 39, interleaved 2 of 5, Code 93, Code 128 and Codabar are enabled by default. Verify that the type of bar code being read has been selected

The unit powers up, but does not scan and/or beep

The scanner has been programmed for a character length lock, or a minimum length and bar code being scanned does not satisfy the programmed criteria

Verify that the bar code that is being scanned falls into the criteria. (Typical of Non-UPC/EAN codes.) (The scanner defaults to a minimum of 4 character bar code)

The unit scans a bar code, but locks up after the first scan (red LED stays on)

The scanner is configured to support some form of host handshaking but is not receiving the signal

If the scanner is setup to support ACK/NAK, RTS/CTS, XON/XOFF or D/E, verify that the host cable and host are supporting the handshaking properly

The unit scans, but the data transmitted to the host is incorrect

The scanner’s data format does not match the host system requirements

Verify that the scanner’s data format matches that required by the host. Make sure that the scanner is connected to the proper host port

23

Page 28

Troubleshooting Guide (continued) SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The print quality of the bar code is suspect

Check print mode. The type of printer could be the problem. Change print settings. For example change to econo mode or high speed

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The aspect ratio of the bar code is out of tolerance

Check print mode. The type of printer could be the problem. Change print settings. ie change to econo mode or high speed

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The bar code may have been printed incorrectly

Check if it is a check digit/character/or border problem

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The scanner is not configured correctly for this type of bar code

Check if check digits are set properly

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The minimum symbol length setting does not work with the bar code

Check if the correct minimum symbol length is set

24

Page 29

Troubleshooting Guide (continued) Keyboard Wedge Only SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

The unit scans the bar code but there is no data

Configuration is not correct

Make sure the scanner is configured for the appropriate mode. Check internal jumper

The unit scans but the data is not correct

Configuration is not correct

Make sure that the proper PC type AT, PS2 or XT is selected. Verify correct country code and data formatting are selected. Adjust intercharacter delay SYMPTOM

The unit is transmitting each character

Configuration is not correct

Increase the interscan code delay setting. Adjust whether the F0 break is transmitted. It may be necessary to try this in both settings.

Alpha characters show as lower case

Computer is in Caps Lock mode

Enable Caps Lock detect setting of the scanner to detect whether the PC is operating in Caps Lock

Everything works except for a couple of characters

These characters may not be supported by that country’s key look up table

Try operating the scanner in Alt mode

25

Page 30

Troubleshooting Guide (continued) RS-232 only SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

Power-up OK and scans OK but does not communicate properly to the host

Com port at the host is not working or configured properly

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232" data

Power-up OK and scans OK but does not communicate properly to the host

Cable not connected to the proper com port

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232" data

Power-up OK and scans OK but does not communicate properly to the host

Com port not operating properly

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232" data

The host is receiving data but the data does not look correct

The scanner and host may not be configured for the same interface font

Check that the scanner and the host are configured for the same interface font

Characters are being dropped

Intercharacter delay needs to be added to the transmitted output

Add some intercharacter delay to the transmitted output by using the MetroSelect Programming Guide MLPN 2407