Page 28
SCAN PATTERN SPECIFICATIONS (CONTINUED) Cross-Sectional Scan Pattern 2.00” Dia. [50.8mm]
3.50” Dia. [88.9mm]
MS6720 Scan Pattern at Face of Scanner
MS6720 Scan Pattern at 4.00” (101.6mm) 4.66” [118.4mm]
MS6720 Scan Pattern at 8.00” (203.6mm)
Drawings have been scaled
22
Page 29
TROUBLESHOOTING GUIDE The following guide is for reference purposes only. Contact a Metrologic representative at 1-800-ID-METRO or 1-800-436-3876 to preserve the limited warranty terms. All Interfaces
MS6720 Troubleshooting Guide SYMPTOMS
No LEDS, beep or motor spin
POSSIBLE CAUSE(S)
No power at power up
SOLUTION
Check transformer, outlet and power strip
NOTE: If it’s a direct connect scanner, the problem could be inadequate power being supplied to the scanner. It may be necessary to use a Metrologic transformer for power. Check the fuse in the MCA. 3 beeps on power up
Non-volatile RAM failure
Contact a Metrologic Rep, if the unit will not hold the programmed configuration
Continuous razz tone on power up
RAM or ROM failure
Contact a Metrologic Rep, if the unit will not function
Razz tone and red LED flash at power up
VLD failure
Contact a Metrologic Rep
Razz tone and both LEDs flash at power up
Scanner motor failure
Contact a Metrologic Rep
Unit scans, Communicates and beeps twice
Same symbol timeout set too short
Adjust same symbol timeout for a longer time
23
Page 30
TROUBLESHOOTING GUIDE (CONTINUED) All Interfaces continued SYMPTOMS
Unit goes to sleep and does not wake up
POSSIBLE CAUSE(S)
SOLUTION
IR sensor used to wake up the scanner is always sensing an object
Adjust scanner positioning so that the IR senses an object when a bar code is presented. For example; point to the ceiling and then back to the object. Disable IR sleep mode
Unit scans bar code too slowly upon waking up
Wake up requires full motor spin up for operation
Disable IR sleep mode
NOTE: Disabling the IR sleep mode allows the scanner to respond to bar codes quicker
24
Page 31
TROUBLESHOOTING GUIDE (CONTINUED) SYMPTOMS The unit powers up properly, lasers come on, but the unit does not scan and does not beep when a bar code is presented POSSIBLE CAUSE(S)/SOLUTION(S) Improper settings can be the cause for a scanner not to scan. The following would be typical examples: 1. Scanning a particular symbology that is not enabled. (UPC/EAN, Code 39, Interleaved 2 of 5, Code 93, Code 128 and Codabar are enabled by default.) Verify that the type of bar code being read has been selected. 2. The scanner has been programmed for a character length lock, or a minimum length and the bar code being scanned does not satisfy the programmed criteria. Verify that the bar code that is being scanned falls into the criteria. (Typical of Non-UPC/EAN codes.) 3. The scanner scans a bar code but the scanner locks up (green LED comes on and stays on) after the first scan. The scanner is configured to support some form of host handshaking but is not receiving the signal. If the scanner is setup to support ACK/NAK, RTS/CTS, XON/XOFF or D/E, verify that the host cable and host are supporting the handshaking properly. 4. The scanner scans and transmits but the data is not correct at the host. Verify that the scanner’s data format matches that required by the host. Make sure that the scanner is connected to the proper host port. SYMPTOMS Scanner beeps at some bar codes and NOT for others of the same bar code symbology POSSIBLE CAUSE(S)/SOLUTION(S) 1. The print quality of the bar code is suspect. 2. The aspect ratio of the bar code is out of tolerance. 3. The bar code may have been printed incorrectly. (check digit/character/or border problem.)
25
Page 32
TROUBLESHOOTING GUIDE (CONTINUED) Keyboard Wedge Only SYMPTOMS
Unit scans the bar code but there is no data
POSSIBLE CAUSE(S)
SOLUTION
Configuration is not correct
Make sure the scanner is configured for keyboard wedge mode - Section F of the ScanSelect Guide [MLPN 00-0 2186] or MetroSet’s help files Make sure that the proper PC type AT, PS2, XT is selected
Unit scans but data is not correct
Configuration is not correct verify correct country code and data formatting are selected Adjust inter character delay Increase the inter scan code delay setting
Unit is transmitting each character twice
Configuration is not correct
Alpha characters show as lower case
Computer is in Caps Lock mode
Enable Caps Lock detect setting of the scanner to detect whether PC is operating in Caps Lock
Everything works except for a couple of characters
These characters may not be supported by that country’s key look up table
Try operating the scanner in Alt mode
26
Adjust whether the F0 break is transmitted. It may be necessary to try this in both settings.
Page 33
TROUBLESHOOTING GUIDE (CONTINUED) RS-232 only SYMPTOMS Power-up OK Scans OK Does not communicate properly to the host
POSSIBLE CAUSE(S)
SOLUTION
Com port at the host not working or not configured Check to make sure the baud properly rate and parity of the scanner Cable not connected to and the communication port the proper com port match and the program is looking for “RS-232" data. Com port not operating properly
NOTE: Metrologic’s default of 7 data bits, space parity is usually what is expected when an IBM compatible type PC is configured for 8 data bits, no parity. SYMPTOMS
SOLUTION
Host receiving data but data does not look correct
Check that the scanner and host are configured for the same interface format
Characters are being dropped
Add some intercharacter delay to the transmitted output. This can be programmed into the scanner through ScanSelect
RS-232 Demonstration Program If an RS-232 scanner is not communicating with your IBM compatible PC, key in the following BASIC program to test that the communication port and scanner are working. This program is for demonstration purposes only. It is only intended to prove that cabling is correct, the com port is working, and the scanner is working. If the bar code data displays on the screen while using this program, it only demonstrates that the hardware interface and scanner are working. At this point, investigate whether the application software and the scanner configuration match. If the application does not support RS-232 scanners, a software wedge program that will take RS-232 data and place it into a keyboard buffer may be needed. This program tells the PC to ignore RTS-CTS, Data Set Ready (DSR) and Data Carrier Detect (DCD) signals. If the demonstration program works and yours still does not, jumper RTS to CTS and Data Terminal Reading (DTR) to DCD and DSR on the back of your PC. 10 20 30 35 40 50 60 70 100 110 32766 32767
CLS ON ERROR GOTO 100 OPEN “COM1:9600,S,7,1,CS0,DS0,CD0,LF” AS #1 PRINT “SCAN A FEW BAR CODES” LINE INPUT #1, BARCODE$ PRINT BARCODE$ K$ = INKEY$: IF K$ = CHR$(27) THEN GOTO 32766 GOTO 40 PRINT “ERROR NO.”; ERR; “ PRESS ANY KEY TO TERMINATE.” K$ = INKEY$: IF K$ = “” THAN GOTO 110 CLOSE: SYSTEM END
27
Page 34
APPLICATIONS AND PROTOCOLS The model number on each scanner includes the scanner number and communications protocol. Scanner 6720 6720 6720 6720 6720
Version Identifier 9
11 14 15 47
Communication Protocol(s) OCIA (OCIA) IBM (46XX) RS-232 (232) Light Pen Emulation (LTPN) Keyboard Wedge (KBW)
The MS6720 Hand-Held Laser Scanner with Built-in PC Keyboard Wedge Interface is designed to be used for keyboard emulation only. However, many RS-232 programmable functions that are available in other Metrologic scanners are also available as keyboard wedge functions. The most important selectable options specific to the keyboard wedge are the following:
• • •
Keyboard Type **AT (includes IBM® PS2 models 50, 55, 60, 80) XT IBM PS2 (includes models 30, 70, 8556)
• • • •
Keyboard Country Type ** USA • Italian United Kingdom • Spanish French • Belgium German • Swiss
** Default setting. Refer to Appendix B for default settings. Refer to the ScanSelect® Scanner Programming Guide [MLPN 00-02186] or the MetroSet2 on line help for information on how to change the default settings.
28
Page 35
APPENDIX A MS6720 Design Specifications Operational Application:
Fixed Projection/Hand-Held Laser Bar Code Scanner
Light Source:
VLD 640-660nm
Laser Power:
0.681 milliwatt (peak)
Scan Speed:
0" - 4"; 0"- 8" (programmable) 0 - 101.6 mm; 0 - 203.2 mm 1000 scan lines per second
Scan Pattern:
5 fields of 4 parallel lines
Scan Lines:
20
Depth of Field:
Min Bar Width:
5.2 mil
Indicators (LED):
red = laser on, ready to scan green = good read, decoding
Beeper Operation:
3 tones or no beep)
Maintenance:
Clean output window periodically
Decode Capability:
Autodiscriminates all standard bar codes; for other symbologies call Metrologic
System Interfaces:
Keyboard Wedge, RS-232, OCIA, Light Pen, IBM 46xx
Print Contrast: Roll, Pitch, Yaw:
35% minimum reflectance difference 360°, 60°, 60°
Mechanical Dimensions: Weight: Termination:
110 mm (L) x 75 mm (W) x 160 mm (H) 450 grams ~ 1 meter retracted coil cable
Continued on next page
29
Page 36
APPENDIX A (CONTINUED) MS6720 Design Specifications Electrical Laser Class: EMC: Input Voltage: Power : Operating Current :
CDRH: Class IIa; Nach EN60825-1:1994/A11:1996 Class 1 FCC, ICES-003 and EN 55022 Class B 5VDC ± .25V 1.1 watt 225 mA typical @ 5VDC (468x – 300 mA)
Standby Current:
74 mA (468x – 165 mA)
DC Transformers:
Class 2; 5VDC @ 300 mA
Environmental Storage Temperature: Operating Temperature: Humidity:
-40°C to 60°C (-40°F to 140°F) 0°C to 35°C (32°F to 95°F) 5% to 95% relative humidity, non-condensing
Light Levels:
Up to 300 LUX
Ventilation:
None required
Shock:
1m. (40") drop
Contaminants:
Sealed to resist airborne particulate contaminants
Specifications subject to change without notice.
30