The text of a manual · part 4 of 6

MS9500 Voyager Series Single-Line Hand-Held Laser Scanner Installation and User's Guide, pages 28 to 36

Pages 28 to 36 of 52

The original document is the PDF (2.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Metrologic MK9520 / MS9520

Page 28

DEPTH OF FIELD

Minimum Bar Code Element Width A

B mm

.13

.15

-

- mils

5.2

5.7

-

-

24

C

D

E

F

G

H

J

K

.19

-

.25

.33

.53

-

7.5

-

10

13

21

-

Page 29 read from a scan by OCR; it may hold reading errors

IR ACTIVATION

2.5" [64 mm] Scan Width at Face Distance From Face __0.0" | [0 mm] Short Range IR | 6.2" [158 mm] 4, 4.0" +1.0" * | [102 mm +25.4 mm] Long Range IR 12.6" [320 mm] 11.0" +2.0", >.

Width of Scan Field

[229 mm +51 mm]

25

Page 30

APPLICATIONS AND PROTOCOLS The model number on each scanner includes the scanner number and factory default communications protocol. Scanner

Version Identifier

MS9520 MS9540

41

RS-232/Light Pen Emulation

MS9520 MS9540

47

Keyboard Wedge, Stand-Alone Keyboard and RS-232 Transmit/Receive

MS9520 MS9540

9

OCIA and RS-232 Transmit/Receive

MS9520 MS9540

11

IBM 468X/469X, RS232-TXD, RXD, RTS, CTS

MS9520 MS9540

00

Laser Emulation and RS-232 Transmit/Receive

MS9520 MS9540

14

RS232-TXD, RXD, RTS, CTS, DTR, DSR

Communication Protocol(s)

The MS9500 Series Hand-Held Laser Scanner with built-in PC Keyboard Wedge Interface is designed to be used for Keyboard emulation only. Many RS-232 programmable functions available in other Metrologic scanners are also available as keyboard wedge functions. The following are the most important selectable options specific to the keyboard wedge. Keyboard Type • • •

**AT (includes IBM® PS2 models 50, 55, 60, 80) XT IBM PS2 (includes models 30, 70, 8556) Keyboard Country Type

• • •

**USA Belgium French

• • •

German Italian Japanese

• • •

Spanish Swiss United Kingdom

**Default setting. Refer to pages 30-34 for default settings. Refer to the MetroSelect® Single-Line Programming Guide (MLPN 00-02544) or MetroSet®2’s help files for information on how to change the default settings.

26

Page 31

TROUBLESHOOTING GUIDE The following guide is for reference purposes only. Contact a Metrologic representative at 1-800-ID-Metro or 1-800-436-3876 to preserve the limited warranty terms. All Interfaces

MS9500 Series Troubleshooting Guide Symptoms

Possible Causes

Solution

No LEDs, beep or laser

No power is being supplied to the scanner

Check transformer, outlet and power strip. Make sure the cable is plugged into the scanner.

No LEDs, beep, or laser

No power is being supplied to the scanner from host

Some host systems cannot supply enough current to power Voyager. A power supply may be needed.

2 Beeps with alternately flashing LEDs on Power up

Possible ROM failure

Flash ROM Upgrade Required

3 Beeps on power up

Non-volatile RAM failure

Contact a Metrologic Representative, if the unit will not hold the programmed configuration.

Continuous razz tone on power up

RAM or ROM failure

Contact a Metrologic Representative, if the unit will not function.

Razz tone and green LED flash at power up

VLD failure

Contact a Metrologic Representative

Razz tone, red and green LEDs flash at power up

Scanning mechanism failure

Contact a Metrologic Representative

Unit scans, Communicates and beeps twice

Same symbol timeout set too short

Adjust same symbol timeout for a longer time.

The unit powers up, but does not scan/or beep

Beeper disabled. No tone selected

Enable beeper. Select tone. 27

Page 32

TROUBLESHOOTING GUIDE (CONTINUED) Symptoms

Possible Causes

Solution

The unit powers up, but does not scan and/or beep

Scanning a particular symbology that is not enabled

UPC/EAN, Code 39, interleaved 2 of 5, Code 93, Code 128 and Codabar are enabled by default. Verify that the type of bar code being read has been selected.

The unit powers up, but does not scan and/or beep

The scanner has been programmed for a character length lock, or a minimum length and bar code being scanned does not satisfy the programmed criteria

Verify that the bar code that is being scanned falls into the criteria (Typical of Non-UPC/EAN codes) The scanner defaults to a minimum of 3 character bar code.

The unit scans a bar code, but locks up after the first scan red LED stays on

The scanner is configured to support some form of host handshaking but is not receiving the signal

If the scanner is setup to support ACK/NAK, RTS/CTS, XON/XOFF or D/E, verify that the host cable and host are supporting the handshaking properly.

The unit scans, but the data transmitted to the host is incorrect

The scanner’s data format does not match the host system requirements

Verify that the scanner’s data format matches that required by the host. Most sure that the scanner is connected to the proper host port.

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The print quality of the bar code is suspect

Check print mode. The type of printer could be the problem. Change print settings. i.e. change to econo mode or high speed.

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The aspect ratio of the bar code is out of tolerance

Check print mode. The type of printer could be the problem. Change print settings. i.e. change to econo mode or high speed.

28

Page 33

TROUBLESHOOTING GUIDE (CONTINUED) Symptoms

Possible Causes

Solution

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The bar code may have been printed incorrectly

Check if it is a check digit/character/or border problem.

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The scanner is not configured correctly for this type of bar code

Check if check digits are set properly.

Scanner beeps at some bar codes and NOT for others of the same bar code symbology

The minimum symbol length setting does not work with the bar code

Check if the correct minimum symbol length is set.

The unit scans the bar code but there is no data

Configuration is not correct

Make sure the scanner is configured for the appropriate mode.

Configuration is correct

Make sure that the proper PC type AT, PS2 or XT is selected. Verify correct country code and data formatting are selected. Adjust inter-character delay symptom.

Configuration is not correct

Increase interscan code delay setting. Adjust whether the F0 break is transmitted. It may be necessary to try this in both settings.

The unit scans but the data is not correct

The unit is transmitting each character twice

29

Page 34

TROUBLESHOOTING GUIDE (CONTINUED) Symptoms

Possible Causes

Solution

Alpha characters show as lower case

Computer is in Caps Lock mode

Enable Caps Lock detect setting of the scanner to detect whether the PC is operating in Caps Lock.

Everything works except for a couple of characters

These characters may not be supported by that country’s key look up table

Try operating the scanner in Alt mode.

Power-up OK and scans OK but does not communicate properly to the host

Com port at the host is not working or configured properly

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232” data.

Power-up OK and scans OK but does not communicate properly to the host

Cable not connected to the proper com port

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232” data.

Power-up OK and scans OK but does not communicate properly to the host

Cable not connected to the proper com port

Check to make sure that the baud rate and parity of the scanner and the communication port match and the program is looking for “RS-232” data.

The host is receiving data but the data does not look correct

The scanner and host may not be configured for the same interface parameters

Check that the scanner and the host are configured for the same interface parameters

Characters are being dropped

Inter-character delay needs to be added to the transmitted output

Add some inter-character delay to the transmitted output by using the MetroSelect Single-Line Programming Guide MLPN 00-02544.

30

Page 35

RS-232 DEMONSTRATION PROGRAM If an RS-232 scanner is not communicating with your IBM compatible PC, key in the following BASIC program to test that the communication port and scanner are working. This program is for demonstration purposes only. It is only intended to prove that cabling is correct, the com port is working, and the scanner is working. If the bar code data displays on the screen while using this program, it only demonstrates that the hardware interface and scanner are working. At this point, investigate whether the application software and the scanner configuration match. If the application does not support RS-232 scanners, a software wedge program that will take RS-232 data and place it into a keyboard buffer may be needed. This program tells the PC to ignore RTS-CTS, Data Set Ready (DSR) and Data Carrier Detect (DCD) signals. If the demonstration program works and yours still does not, jumper RTS to CTS and Data Terminal Reading (DTR) to DCD and DSR on the back of your PC. 10 20 30 35 40 50 60 70 100 110 32766 32767

CLS ON ERROR GOTO 100 OPEN “COM1:9600,S,7,1,CSO,DSO,CD0,LF” AS#1 PRINT “SCAN A FEW BAR CODES” LINE INPUT #1, BARCODE$ PRINT BARCODE$ K$ = INKEY$: IF K$ = CHR$(27) THEN GOTO 32766 GOTO 40 PRINT “ERROR NO.”; ERR ;“PRESS ANY KEY TO TERMINATE.” KK$ = INKEY$: IF K$ = “”THEN GOTO 110 CLOSE: SYSTEM END

31

Page 36

DESIGN SPECIFICATIONS OPERATIONAL Light Source:

Visible Laser Diode 650 nm ± 10 nm

Laser Power:

Less than 1 mW (peak)

Depth of Scan Field:

0 mm – 203 mm (0” – 8”) for 0.330 mm (13 mil) bar code at default setting

Scan Speed:

72 scan lines per second

Scan Pattern:

Single scan line

Minimum Bar Width:

0.127 mm (5.0 mil)

InfraRed Activation:

Long Range: 0 mm – 279 mm ± 51 mm (0” – 11” ± 2”) Short Range: 0 mm – 102 mm ± 25 mm (0” – 4” ± 1”)

Decode Capability:

Autodiscriminates all standard bar codes for others call a Metrologic service representative

System Interfaces:

RS232, PC Keyboard Wedge, Stand-Alone Keyboard, OCIA, IBM 468X/469X, Light Pen Emulation, Laser Emulation, RS232 with DSR

Print Contrast:

35% minimum reflectance difference

Number Characters Read: Roll, Pitch, Yaw:

42°, 68°, 52°

Beeper Operation:

7 tones or no beep

Indicators (LED) Default Settings:

Up to 80 data characters (Maximum number will vary based on symbology & density)

Green = laser on, ready to scan; Red = good read; Yellow (MS9540 Only) = CodeGate button is inactive (on). CodeGate button is active (off)

MECHANICAL Length:

198 mm (7.8”)

Width:

Handle - 45 mm (1.8”), Head - 78 mm (3.1”)

Depth:

40 mm (1.6”)

Weight:

149 g (5.25 oz)

Input Voltage:

5VDC ± 0.25V

ELECTRICAL Power:

Operating = 0.825 W, Standby = 0.600 W

Current:

Operating = 165 mA @ 5VDC, Standby = 120 mA @ 5VDC

DC Transformers: Laser Class: EMC:

Class 2; 5.2V @ 650 mA CDRH: Class II; EN60825-1:1994/A11:1996 Class 1 FCC Class B

ENVIRONMENTAL Temperature: Humidity: Light Levels: Shock: Contaminants: Ventilation:

32

Operating = 0°C to 40° (32° to 104°F) Storage = -40°C to 60°C (-40°F to 140°F) 5% to 95% relative humidity, non-condensing Up to 4842 Lux (450 footcandles) Designed to withstand 1.5 m (5’) drops Sealed to resist airborne particulate contaminants None required