For the Metrologic IS6520

Page 10

INSTALLING THE IS6520-41/MS6520-41 TO THE HOST SYSTEM 1.

Turn off the host system.

2.

Connect the 10-pin RJ45 male connector into the jack on the Cubit® IS6520/MS6520. You will hear a ‘click’ when the connection is made. Note: If the Cubit IS6520 is receiving power from the host system, skip to step #5. (See caution statement below)

3.

Connect the L-shaped plug of the power supply into the power jack on the PowerLink cable.

4.

Connect the power supply into an AC outlet. Make sure the AC input requirements of the power supply match the AC outlet.

5.

Connect the PowerLink cable to the proper port on the host system.

6.

Turn on the host system.

q o p n s r

Note: Plugging the scanner into a port on the host system does not guarantee that scanned information will be communicated properly to the host system. The scanner is shipped from the factory programmed with default settings. Please refer to the MetroSelect® Programming Manual (MLPN 2407) for instructions on changing the scanner’s configuration. In addition, please check that the scanner and host system are using the same communication protocol.

Caution: To maintain compliance with applicable standards, all circuits connected to the scanner must meet the requirements for SELV (Safety Extra Low Voltage) according to EN 60950. To maintain compliance with standard CSA C22.2 No. 950/UL 1950 and norm EN 60950, the power source should meet applicable performance requirements for a limited power source.

4

Page 11

INSTALLING THE IS6520-47/MS6520-47 TO THE PC 1.

Turn off the PC.

2.

Connect the 10-pin RJ45 male connector into the jack on the Cubit® IS6520/MS6520. You will hear a ‘click’ when the connection is made.

3.

Connect the L-shaped plug of the power supply into the power jack on the PowerLink cable. Refer to Manufacturer’s Recommendation below.

4.

Connect the power supply into an AC outlet. Make sure the AC input requirements of the power supply match the AC outlet.

5.

Disconnect the keyboard from the PC.

6.

The PowerLink cable is terminated with a 5-pin DIN female connector on one end, and a 6-pin mini DIN male on the other. Metrologic will supply an adapter cable with a 5-pin DIN male connector on one end and a 6-pin mini DIN female connector on the other. According to the termination required, connect the appropriate end of the adapter cable to the PowerLink cable, leaving the necessary termination exposed for connecting to the keyboard and the keyboard port on the PC. Refer to Appendix C page 31 for pin assignments.

7.

Connect to the PowerLink cable to the keyboard and the keyboard port on the PC.

8.

Power up the PC.

q o p s&t n u r Manufacturer’s Recommendation: Metrologic recommends the use of an external power supply with IS652047/MS6520 Keyboard Wedge applications. Powering the IS6520-47/MS6520 directly from the computer keyboard connector could interfere with the operation of the scanner or the computer. Not all computers supply the same current through the keyboard port, explaining why a scanner would work on one computer and not another. Caution: To maintain compliance with applicable standards, all circuits connected to the scanner must meet the requirements for SELV (Safety Extra Low Voltage) according to EN 60950. To maintain compliance with standard CSA C22.2 No. 950/UL 1950 and norm EN 60950, the power source should meet applicable performance requirements for a limited power source.

5

Page 12

DISCONNECTING THE POWERLINK CABLES FROM THE SCANNER Before removing the cable from the scanner, Metrologic recommends that the power on the host system is off and the power supply has been disconnected from the PowerLink cable.

o& p q n 1.

Locate the small ‘pin-hole’ on the back of the scanner.

2.

Bend an ordinary paperclip into the shape shown above.

3.

Insert the paperclip (or other small metallic pin) into the small ‘pin-hole’.

4.

You will here a faint ‘click’. Pull gently on the strain-relief of the PowerLink cable and it will slide out of the scanner.

6

Page 13

SCANNER PARTS A. Speaker Serves as an audible indicator.

B. Green & Red LEDs C

B

A

During normal operation, the green LED is on. This indicates that the laser is on and the unit is ready to scan. On a successful read of a bar code, the red LED will turn on. After communication to the host is complete, the red LED will turn off. The LEDs are also used as diagnostic indicators and mode indicators.

C. Output Window Laser light emits form this aperture.

D

D. PowerLink Cable The IS6520/MS6520 scanner has a 10-pin modular jack. The 10-pin modular plug on the PowerLink cable connects into the IS6520/MS6520.

MOUNTING HOLES Each IS6520/MS6520 Cubit® is equipped with three threaded mounting holes located on the back of the scanner. The following diagram shows the exact location of each mounting hole. The mounting holes are 6 millimeters deep and are threaded for use with 3 millimeter screws. Caution: Using a screw that is longer than 6 millimeters can potentially damage the scanner.

7

Page 14

STANDS Fixed Vertical Stand (MLPN 46-46145) Optional purchase.

a. Fixed Vertical Stand ........ Qty. 1 e. d.

b. #7 x 1.00” Wood Screw ... Qty. 3 c. M3 x 0.5 x 6 mm Screw ... Qty. 3 Tilt Stand (MLPN 46-46486) a. c.

Shipped with the MS6520 only.

b.

c. Tilt Stand .......................... Qty. 1 d. #8 x 1.00” Wood Screw ... Qty. 3

Fixed Vertical Stand Installation (MLPN 46-46145) Step 1: Drill three #43 pilot holes in the countertop. Step 2: Attach the stand to the countertop with the three #7 wood screws provided. Step 3: Attach the scanner to the stand with the three M3 screws provided. Tilt Stand Installation (MLPN 46-46486) Step 1: Drill three #39 pilot holes in the countertop. Step 2: Insert the scanner into the rubber boot (the MS6520 is shipped assembled). Step 3: Tilt the scanner forward in the stand. Step 4: Attach the stand to the countertop using the three #8 wood screws provided. Step 5: Tilt the scanner back into place. 8

Page 15

AUDIBLE INDICATORS When the IS6520/MS6520 scanner is in operation, it provides audible feedback. These sounds indicate the status of the scanner. Eight settings are available for the tone of the beep (normal, 6 alternate tones and no tone). To change the tone, refer to the MetroSelect® Programming Guide MLPN 2407. One Beep When the scanner first receives power, the green LED will turn on, then the red LED will flash and the scanner will beep once. (The red LED will remain on for the duration of the beep.) The scanner is ready to scan. When the scanner successfully reads a bar code, the red LED will flash and the scanner beeps once (if programmed to do so). If the scanner does not beep once and the green light does not flash, then the bar code has not been successfully read. Razzberry Tone This tone is a failure indicator. Refer to “Failure Modes" page 10. Three Beeps - During Operation During operation of the scanner, the red LED will flash while the scanner simultaneously beeps three times (while going into programming mode). The red LED will continue to flash until the unit exits program mode. Upon exiting program mode, the scanner will beep three times and the red LED will stop flashing. When configured, 3 beeps can also indicate a communications timeout during normal scanning mode. When using one-code-programming, the scanner will beep three times (the current selected tone), followed by a short pause then by a high tone and a low tone. This tells the user that the single configuration bar code has successfully configured the scanner. Three Beeps - On Power Up This is a failure indicator. Refer to “Failure Modes” on page 10.

9

Page 16

FAILURE MODES Flashing Green and One Razzberry Tone This indicates the scanner has experienced a laser subsystem failure. Return the unit for repair at an authorized service center.

Flashing Green and Red and Two Razzberry Tones This indicates the scanner has experienced a motor failure. Return the unit for repair at an authorized service center.

Continuous Razzberry Tone with Both LEDs Off If, upon power up, the scanner emits a continuous razzberry tone, then the scanner has an electronic failure. Return the unit for repair at an authorized service center.

Three Beeps - On Power Up If the scanner beeps 3 times on power up then, the nonvolatile memory that holds the scanner configuration has failed. Return the unit for repair at an authorized service center.

10

Page 17

VISUAL INDICATORS There is a red LED and a green LED on IS6520/MS6520. When the scanner is on, the flashing or stationary, activity of the LEDs indicates the status of the current scan and the scanner. No Red or Green LED The LEDs will not be illuminated if the scanner is not receiving power from the host or transformer. Steady Green When the laser is active, the green LED is illuminated. The green LED will remain illuminated until the laser is deactivated. During the power save mode, the laser will turn on and turn off. During this period, the green LED remains illuminated. Steady Green and Single Red Flash When the scanner successfully reads a bar code, the red LED will flash and the scanner will beep once. If the red LED does not flash or the scanner does not beep once, then the bar code has not been successfully read. Steady Green and Steady Red After a successful scan, the scanner transmits the data to the host device. Some communication modes require that the host inform the scanner when data is ready to be received. If the host is not ready to accept the information, the scanner's red LED will remain on until the data can be transmitted. Steady Green and Flashing Red This indicates the scanner is in program mode. A razzberry tone indicates that an invalid bar code has been scanned in this mode. Steady Red, Green Off This indicates the scanner may be waiting for communication from the host.

11

Page 18

PROGRAMMING MODES The IS6520/MS6520 Cubit® has 3 modes of programming. Bar Codes: Cubit can be configured by scanning the bar codes located in the MetroSelect® Configuration Guide (MLPN 2407). Please refer to this guide for instructions. This manual can be downloaded for FREE from Metrologic’s website (www.metrologic.com). MetroSet®: This user-friendly Windows-based configuration program allows you to simply ‘point-and-click’ at the desired scanner options. This program can be downloaded for FREE from Metrologic’ website (www.metrologic.com), or set-up disks can be ordered by calling 1-800-ID-METRO. Serial Programming: This mode of programming is ideal for OEM applications. This mode gives the end-user the ability to send a series of commands using the serial port of the host system. The commands are equivalent to the numerical values of the bar codes located in the MetroSelect Configuration Guide (MLPN 2407). How does Serial Programming work? 1.

Each command sent to the scanner is the ASCII representation of each numeral in the configuration bar code. The entire numeric string is framed with an ASCII [stx] and an ASCII [etx]. EXAMPLE #1: Command for Disabling Codabar Command = [stx]100104[etx] String Sent to Scanner = 02h 31h 30h 30h 31h 30h 34h 03h (All values are hexadecimal).

2.

If the command sent to the scanner is valid, the scanner will respond with an [ack].

3.

If the command sent to the scanner in invalid, the scanner will respond with a [nak]. NOTE: If this occurs, the end-user must start over at the very beginning of the configuration sequence. Simply re-tranmsitting the invalid command will not work, you must start over.

4.

During programming, the motor and laser remain active. YOU CANNOT SCAN ANY BAR CODES WHILE IN PROGRAM MODE.

5.

There is a 20 second window between commands. If a 20 second timeout occurs, the scanner will send a [nak] and you must start over.

12