Page 49
Leuze electronic 7
Commissioning
Commissioning Attention Laser Radiation! Follow the safety notices in chapter 2.3 on page 7!
7.1
Measures to be performed prior to the initial commissioning ! Before commissioning, familiarise yourself with the operation and configuration of the device(s)! ! Before switching on, recheck all connections and ensure that they have been properly made.
Configuration elements using the example of the MA 2 Jumper for setting the device address range: Jumper above: addr. 0 … 15 Jumper below: addr. 16 … 31
Coding switch 0 … F for setting the device address switch positions 0 … F correspond to the device address 0 … 15 or 16 … 31
Figure 7.1:
TNT 35/7-24V
Operating mode switch: Switch in upper pos.: operation Switch in lower pos.: service
Configuration elements in the MA 2
Setting the device address The device address is set in the MA 2/MA 4… using a rotary code switch. The setting is carried out as follows: ! Set the device address to • 0, if the combined BCL 21 / MA 2/MA 4… unit will not be operated in a network, • 1…30, if several BCL 21 / MA 2/MA 4… units will be operated in a network. Each multiNet plus network device must have a different device address assigned to it. If the MA 2/MA 4… is connected to the multiNet Master, the device combination BCL 21 with MA 2/MA 4… automatically turns into a multiNet plus slave device. Notice! Address 31 is used for resetting the parameters.
Leuze electronic
BCL 21/22
47
Page 50
Leuze electronic
Commissioning Notice!
From the hardware address, the BCL 21 detects that networking is required. With the BCL 21/22, it is possible to perform a reset using the software and the online commands. In addition, it is possible to perform a reset by switching off the supply voltage. The parameters are not lost as a result of the reset. For information on the reset commands, see chapter 9. The LED remains dark during a reset; the green LED illuminates when the device is ready for operation.
7.2
Function Test "Power On" test After connecting the operating voltage, the devices carry out an automatic "Power On" function test. The LED on the front side of the MA 2/MA 4… lights up. In the case of factory settings, the green LED flashes. In the additional service mode, the orange LED lights up.
Interface Proper function of the interface can be tested easiest in service operation using the service interface with the "BCLConfig" programming software and a notebook computer. For order numbers, see chapter 5.
"Online commands" Using the "Online" commands, important device functions can be checked, e.g. proper functioning of the laser.
Problems Should problems occur during device commissioning, refer first to chapter 8.2. Should a problem persist after checking all electrical connections and settings on the devices and host, please contact a Leuze service office near you (see the back page of this operating manual).
48
BCL 21/22
Leuze electronic
Page 51
Leuze electronic 7.3
Commissioning
Setting the Parameters In the device combination BCL 21/22 with MA 2/MA 4…, two different parameter sets are managed: • factory default parameter set • current parameter set Before a parameter set is loaded into the BCL 21/22 processor memory, the validity of the parameter set is checked using checksums.
Factory default parameter set This parameter set contains the default settings made ex works for all BCL 21/22 parameters. It is permanently stored in the ROM of the BCL 21/22. The parameter set with the default settings is loaded into the memory of the BCL 21/22, • the first time the device is commissioned after delivery • following the command "Factory Default" in the parameterisation program • if the checksums of the current parameter set are invalid.
Current parameter set In this parameter set, the current settings for all device parameters are stored. When the BCL 21/22 is in operation, the parameter set is stored in the EEPROM of the BCL 21/22. The current set can be stored: • by copying a valid parameter set from the host computer • by means of an off-line setup with the PC setup program BCLConfig The current parameter set is loaded into the memory of the BCL 21/22,
7.3.1
Service Operating Mode Setting the required parameters is carried out easiest in the 'Service' operating mode.
Service interface By switching the operating mode switch in the MA 2/MA 4… from operation (switch in upper pos.) to service (switch in lower pos.), the connection to the host computer is interrupted and the RS 232 interface is activated. At the BCL 22, the interface is activated via a bridge between the pins 7 and 15 on the 15-pin sub-D connector. The connector unit MA 2/MA 4… offers a service interface for commissioning the reading stations in a network. This interface can be accessed by removing the housing cover. It has a 9-pin Sub-D connector (see figure 7.1).
Leuze electronic
BCL 21/22
49
TNT 35/7-24V
• each time the supply voltage is connected • following a software reset
Page 52
Leuze electronic
Commissioning Connection
A PC or terminal can be connected to the MA 2/MA 4… via the RS 232/V.24 serial interface and used to set the parameters of the BCL 21/22. The connection is made using a crossed RS 232 connection cable that establishes the RxD, TxD and GND connections. A hardware handshake via RTS, CTS is not supported at the service interface.
Service Operating Mode
PC/terminal COM interface Figure 7.2:
MA 2/MA 4… service interface
Connecting the MA 2 service interface to a PC or terminal
Notice! The service interface contains a predefined transmission protocol with the following parameters: • • • • • •
50 transfer rate 9600 baud no parity 8 data bits 1 stop bit prefix: STX postfix: CR, LF