The text of a manual · part 9 of 13
Ultra High Accuracy Laser Displacement Meter LC-2400 Series Instruction Manual, pages 81 to 90
Pages 81 to 90 of 121
The original document is the PDF (805 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Keyence LC-2400A
Page 81
8-3 Monitor Output (Standard Equipment) 8-4 DIGITAL I/O (Standard Equipment on Control Unit)
Note
To interrupt entry of the analog ratio, press the PRM key, and then press the key several times until the LC returns to the measurement mode. Any new parameter settings are ignored, and the LC returns to the measurement mode. 3. Press the key. The specified analog ratio is entered, and the LC returns to the measurement mode.
Intensity Monitor Output
The LC outputs an analog voltage proportional to the intensity. Signal output
Output range: 0 to + 5 V Output impedance: 0 Ω Current capacity: 10 mA
0V BNC connector
Note
• The intensity monitor output range is 10000/V for all sensor head models • When an error message is displayed, the output voltage is retained to 0 V.
8-4 Digital I/O (Standard Equipment on Control Unit) Cannot be used with the LC-2400W series. The digital I/O function can be used when the control unit is used separately. Displacement and intensity measurement values and alarm signals for indicating abnormal measurement conditions are output as digital signals.
Specifications
Output TTL level (74LS652 or equivalent) +5 V
3 kΩ D
Q
OUT 6.2 kΩ
Input TTL level (74LS19 or equivalent) +5 V 3 kΩ IN
6.2 kΩ
74 CHAPTER 8 External Input/Output
Page 82
8-4 DIGITAL I/O (Standard Equipment on Control Unit)
Output Signal
Displacement data output The displacement data output signal is a positive logic signal in the form of a 16-bit twos complement binary numbers representing displacement measurement values. When the number of averaging measurements is set to between 1 and 2048, the displacement data is output every 20 µs. Sensor head LC-2420 LC-2430 LC-2440 LC-2450
Displacement data output 1LSB = 0.01 µm 1LSB = 0.02 µm 1LSB = 0.2 µm 1LSB = 0.5 µm
Intensity data output The intensity data output signal is a positive logic signal in the form of 16-bit twos complement binary numbers representing intensity. For all sensor head models: 1LSB= 2 Strobe (STB) The strobe signal is output to specify data output timing. It is a negative logic signal with a pulse duration of approx. 6 µs. The strobe signal goes LOW approx. 1.6 µs after the measurement data output signal is prepared. Alarm output The LC alarm signal output is LOW. Alarm (OR logic) All alarm outputs except for [GO] are ORed. The alarm signal goes LOW when at least one alarm signal is output.
Input Signal
External trigger enable input (ETE IN) When the external trigger enable input (ETE IN) is LOW and the external trigger input (EXT TRIG IN) signal is received, the LC outputs measurement data at a specified frequency. Input Level Output HI (The input is HIGH even if When the number of the input pin is open.) averaging measurements is set to between 1 and 2048, measurement data is output ETE IN every 20 µs. LOW Measurement data is output only when the external trigger input (EXT TRIG IN) signal is received.
Note
When the number of averaging measurements is set to between 1 and 2048, measurement data is output every 20 µs. When the number of averaging measurements is set to 4096 or more, measurement data is output at the following frequencies.
CHAPTER 8 External Input/Output
75
Page 83
8-4 DIGITAL I/O (Standard Equipment on Control Unit)
Number of averaging measurements Frequency 1 to 2048 20 µs (40 µs) 4096 40 µs (80 µs) 8192 80 µs (160 µs) 16384 160 µs (320 µs) 32768 320 µs (640 µs) 65536 640 µs (1280 µs) 131072 1280 µs (2560 µs) Figures in ( ) apply when the ADD or SUB function is selected. Timing chart 1. When the ETE IN signal is HIGH: 20 µs Displacement/ intensity data Alarm
STB
1.6 µs
Approx. 6.4 µs
The strobe (STB) signal is output at a frequency of 20 µs, regardless of the number of averaging measurements.
Note
2. When the ETE IN signal is LOW: 1 µs min. EXT TRIG IN
Displacement/ intensity data ❋ Alarm output
30 µs max.
STB
1.6 µs
Note
Approx. 6.4 µs
* When the ETE IN signal is LOW and the number of averaging measurements is set to between 1 and 2048, the alarm signal is output every 20 µs (40 µs), regardless of the EXT TRG IN signal level. • Data in ( ) apply when the ADD or SUB function is selected.
76 CHAPTER 8 External Input/Output
Page 84
8-4 DIGITAL I/O (Standard Equipment on Control Unit)
External trigger input (EXT TRIG IN) When the ETE IN signal is LOW, measurement data is output a maximum of every 30 µs. After the EXT TRG IN signal goes LOW. The output data is retained until the EXT TRIG IN signal goes LOW again.
Pin Names and Functions
DIGITAL I/0
1
V
W x y
fl
Pin No.
Signal name
Pin No.
Signal name
Pin No.
1
Displacement data output DBIN1
32
Earth ground terminal
65
Signal name Alarm output (HIGH)
(LSB)
33
Intensity data output IBIN1 (LSB)
66
Alarm output (LOW)
2
Displacement data output DBIN2
34
Intensity data output IBIN2
67
Alarm output (+ AREA OVER)
3
Displacement data output DBIN3
35
Intensity data output IBIN3
68
Alarm output (- AREA OVER)
4
Displacement data output DBIN4
36
Intensity data output IBIN4
69
Alarm output (BRIGHT)
5
Displacement data output DBIN5
37
Intensity data output IBIN5
70
Alarm output (DARK)
6
Displacement data output DBIN6
38
Intensity data output IBIN6
71 alarm output (ERROR)
7
Displacement data output DBIN7
39
Intensity data output IBIN7
72
Alarm output (GO)
8
Displacement data output DBIN8
40
Intensity data output IBIN8
73
Alarm (OR logic)
9
Displacement data output DBIN9
41
Intensity data output IBIN9
74
*
10
Displacement data output DBIN10
42
Intensity data output IBIN10
75
ETE IN (input)
11
Displacement data output DBIN11
43
Intensity data output IBIN11
76
*
12
Displacement data output DBIN12
44
Intensity data output IBIN12
77
EXT TRG IN (input)
13
Displacement data output DBIN13
45
Intensity data output IBIN13
78
STB (output)
14
Displacement data output DBIN14
46
Intensity data output IBIN14
79
*
15
Displacement data output DBIN15
47
Intensity data output IBIN15
80
*
16
Displacement data output DBIN16
48
Intensity data output IBIN16 (MSB)
81
*
(MSB)
49
*
82
*
17
OPEN
50
*
83
*
18
0 V (for ±5 V)
51
*
84
*
19
OPEN
52
*
85
*
20
*
53
*
86
*
21
*
54
*
87
*
22
*
55
*
88
*
23
OPEN
56
*
89
+5 V
24
OPEN
57
+5 V
90
0 V (for +5 V)
25
+5 V
58
0 V ( for +5V)
91
0 V (for +5 V)
26
+5 V
59
0 V ( for +5 V)
92
+15 V
27
+15 V
60
+15 V
93
0 V (for ±15 V)
28
+15 V
61
0 V (for ±15 V)
94
0 V (for ±5 V)
29
0 V (for ±15 V)
62
-15 V
95
Earth ground terminal
30
-15 V
63
-15 V
96
Earth ground terminal
31
-15 V
64
Earth ground terminal
* Through the reserve pins (pin Nos. 49 to 56, and 79 to 88), input signals from the display unit are received. When the control unit is used separately, the reserve pins must remain open (without connection to any device).
CHAPTER 8 External Input/Output
77
Page 85
78 CHAPTER 8 External Input/Output
Page 86
CHAPTER 9
APPENDIX 9-1
Trouble-Shooting Guide ................................................80
9-2
Hints on Highly Accurate Measurement ........................ 83
9-3
Specifications ................................................................84
9-4
Dimensions ....................................................................85
9-5
Characteristics ...............................................................87
9-6
Initial Setting List ...........................................................89
9-7
Glossary ........................................................................90
9-8
Expansion I/O ................................................................94
9-9
Index ............................................................................111
Page 87
9-1 Trouble-Shooting Guide
9-1 Trouble-Shooting Guide If the LC series does not display a measurement value or displays an inaccurate measurement value, or if an error occurs during communication with external equipment, refer to the following list to identify the problem and reinstall or readjust the LC. If the LC does not work properly after taking the appropriate actions, please contact KEYENCE.
Troubleshooting Problem
Cause
Action
Nothing is displayed on the The key-operated power switch is Turn ON the key-operated power display panel. turned OFF. switch. The power switch on the rear panel is turned OFF. (LC-2400W only)
"- - - -" is displayed on the display panel.
Measurement value fluctuates.
Inaccurate measurement value is displayed.
p.12 p.14
The power supply cable is disconnected.
Connect the power supply cable.
p.22
The fuse is blown.
Replace the fuse.
–
The power has just been switched on or a parameter setting has been just changed.
No problem –
Measurement error has occurred. Refer to the "Error Message List" for appropriate actions.
p.82
The transmitter/receiver of the sensor head is not clean.
Wipe the dust off the transmitter/ receiver.
p.18
The preset gain is not correct.
Set the gain to a correct value.
p.30
The preset number of averaging measurements is too small.
Set the number of averaging measurements to an appropriate value.
p.36
The measurement point is not clean.
Wipe the dust off the measurement point.
–
The sensor head orientation is not correct.
Make sure that the sensor head orientation is correct.
p.25
The sensor head or target is tilted.
Check the sensor head and target setup positions.
p.25
The control unit and the sensor head have different serial numbers.
Use the control unit and sensor head having the same serial number.
p.22
Calibration was improperly performed.
Re-calibrate.
The sensor head or target is tilted. The target is out of the measuring range. The preset measurement mode is not correct.
80 CHAPTER 9 Appendix
Reference page p.32
Check the sensor head and target setup positions.
p.25
Set the measurement mode properly according to the setting.
p.30
Page 88
9-1 Trouble-Shooting Guide
Problem
Cause
Reference page
Action
The measurement mode is set to Press the INTENSITY key to change INTENSITY. the measurement mode to DISPLACEMENT.
Analog output voltage exceeds 10 V.
p.29
OFFSET value is entered.
Make sure that the OFFSET value is correct.
p.35
The preset mode is not correct.
Be sure to choose the correct display/output mode.
p.31
The transmitter/receiver of the sensor head are not clean.
Wipe the dust off the transmitter/ receiver.
p.18
The measurement point is not clean.
Wipe the dust off the measurement point.
–
An error message is displayed.
Referring to the "Error Message List", take the appropriate actions.
p.82
The analog output voltage exMake sure that the preset analog ceeds the specified output range. ratio is correct.
p.73
Analog output value is not correct.
The preset analog ratio is not correct.
Make sure the preset analog ratio is correct.
p.73
The LC cannot communicate with the external device connected to the RS-232C connector.
A null modem cable was used.
Use a standard serial cable.
p.50
Set the baud rate and data format to that of the external device.
p.51
Transmit the correct character or ASCII code.
p.67
The preset baud rate is not correct. The preset data format is not correct. The transmitted character or ASCII code is not correct.
CHAPTER 9 Appendix
81
Page 89
9-1 Trouble-Shooting Guide
Error Message List Display
Cause
Action
Reference page
Measurement error
The measurement value exceeds the specified output range.
Change the OFFSET value and the sensor head and target setup positions so that the measurement value does not exceed the specified output range.
p.25 p.35
Insufficient intensity
Intensity is insufficient or the laser beam shield is closed.
Make sure that the LOW-INTENSITY LIMIT and gain are set to correct values in relation to the intensity. Make sure that the laser beam shield is opened.
p.4 p.30 p.37
Make sure that the HIGH-INTENSITY LIMIT and gain are set to correct values in relation to the intensity.
p.30 p.37
The sensor-to-target distance is too long. Check the sensor head and target setup The sensor-to-target position.
p.25
Error message
Intensity too high Intensity is too high.
+ AREA OVER - AREA OVER distance is too short. Calibration error
The output value exceeds the specified calibration range.
Re-calibrate. p.32
Filter setting error The preset CUTOFF Set the CUTOFF frequency to 0100 or frequency is not more. correct.
p.36
Connection error Pins 1 and 2 of the LASER REMOTE connector are not shorted. No sensor head is connected. The connecting cable is disconnected.
Reconnect the sensor head. Short the LASER REMOTE connector pins 1 and 2 by connecting the LASER REMOTE input connector to the LASER REMOTE connector. Check that the connecting cable is not disconnected.
p.15 p.22
GP-IB error
Install the GP-IB board properly to the GP-IB connector.
Battery error BAUD RATE LED is flashing
Note
The GP-IB board is not properly connected to the GP-IB connector.
Memory back-up Replace the battery. battery is exhausted.
p.24 p.19
"Far" and " Near" are displayed when the sensor-to-target distance exceeds the measuring range by at least 5%. Example: When using the LC-2440: "Far" is displayed when the sensor-to-target distance exceeds the measuring range by at least + 3.3 mm. "Near" is displayed when the sensor-to-target distance exceeds the measuring range by at least - 3.3 mm.
82 CHAPTER 9 Appendix
Page 90
9-2 Hints on Highly Accurate Measurement
9-2 Hints on Highly Accurate Measurement This section describes the measurement conditions for improving the LC2400 series measurement accuracy.
Performance of the LC-2400 Series
The resolution given in the LC specifications was obtained under the following conditions: • A regular-refective object (for LC-2420/LC-2430) or white diffuse-reflective object (for LC-2440/LC-2450) was used as the target. Sufficient intensity was received. • The number of averaging measurements was set to 512. • Each target was measured in static conditions. As the number of averaging measurements and the intensity level were reduced, the resolution was reduced. The resolution was also reduced when measuring a moving target with an uneven surface such as a metal-worked or hair-line finished target. For example, when measuring the surface runout of a hair-line finished rotating disk, the resolution is reduced. Linearity is also reduced under such conditions.
How to Improve the Measurement Accuracy
Under the following conditions, the LC enables highly accurate measurements using a target with an uneven surface such as a hair-line finished target. • Increase the intensity. Apply a white acrylic resin board or mirror-surfaced object to the measurement point to increase the intensity. • Increase the laser beam diameter. Place a target at the end of the measuring range to increase the laser beam diameter. When using the LC, for example, set the sensor-to-target distance to 28 mm or 32 mm. In this case, be sure to place the target within the measuring range. • Increase the number of averaging measurements. As the number of averaging measurements is increased, the response frequency is reduced. Be sure to set the number of averaging measurements to an appropriate value in relation to the target travel speed. Number of averaging measurements vs. response frequency Response frequency (Hz) =
20000 Number of averaging measurements
* The response frequency is defined as the value when the analog output level is 3 dB lower than the stable level. For other methods for improving the measurement accuracy, consult KEYENCE.
CHAPTER 9 Appendix
83