For the Keyence LC-2400A

Page 91

9-3 Specifications

9-3 Specifications Model

Controller Sensor head

LC-2420 ±0.2 mm

Measuring range

LC-2400A/LC-2400W/LC-2401 LC-2430 LC-2440 ±0.5 mm ±3 mm

LC-2450 ±8 mm

10 mm 30 mm 30 mm 50 mm Visible red semiconductor laser, Maximum output 1.9 mW, Pulse duration: 10 µs 670 nm Class ΙΙ Class 2 Klasse 2 20 x 12 µm 30 x 20 µm 35 x 20 µm 45 x 20 µm 0.01 µm 0.02 µm 0.2 µm 0.5 µm 0.32 µm 0.4 µm 0.5 µm ±0.05% of F.S. ±0.1% of F.S. ±0.05% of F. S. 50 kHz

Response frequency Response time Number of averaging measurements OFFSET range Gain adjustment Analog output Displacement data output Intensity data output 96-pin Displacement LC-2400A/ connector 3. data output LC-2400W LC-2401

20 kHz (- 3 dB, Number of averaging measurements: 1) 100 µs 1 to131072 (18 selectable settings) ±199.99 µm ±499.98 µm ±2.9998 mm ±7.9995 mm AUTO/MANUAL (4 settings) ± 10 V, Output impedance: 0 Ω, 6 settings, Resolution: 5 mV 0 to 5 V, Output impedance: 0 Ω TTL level, positive logic 1LSB=0.01 µm 1LSB=0.02 µm 1LSB=0.2 µm 1LSB=0.5 µm TTL level, positive logic 1LSB=0.32 µm 1LSB=0.4 µm 1LSB=0.5 µm TTL level, positive logic, 1LSB = 2 TTL level, negative logic [Output timing]

Digital I/O

Operating distance Light source Wavelength Class FDA IEC 825-1 11.1993 DIN EN 60825-1 07.1994 Minimum spot diameter Resolution 1. LC-2400A/LC-2400W LC-2401 Linearity 2. LC-2400A/LC-2400W LC-2401 Sampling frequency

Intensity data output Control input Control output 50-pin connector 4.

Displacement data output

LC-2400A/ LC-2400W LC-2401

Intensity data output Control input Control I/O

Control input

Control output Interface

RS-232C

GB-IB5. Measuring stability (±5°C (±9°F))6. Power supply voltage Fuse Power consumption Ambient temperature Relative humidity Noise Weight Sensor head Controller 1, 2: 3: 4, 5: 6:

TTL level, negative logic [Upper/lower limit, INTENSITY alarm, AREA OVER alarm] 16-bit parallel NPN open-collector, negative logic 1LSB=0.01 µm 1LSB=0.02 µm 1LSB=0.2 µm 1LSB=0.5 µm Paralle NPN open-collector, negative logic 11 bits 12 bits 15 bits 16 bits 1LSB=0.32 µm 1LSB=0.4 µm 1LSB=0.5 µm 16-bit parallel NPN open-collector, negative logic 1LSB=2 TTL level, negative logic [Output timing] Non-voltage input (contact, solid-state) [HOLD timing input, AUTO-ZERO ON/OFF input, Program selection input, 1CH/2CH selection input, LASER REMOTE input] NPN open-collector, 100 mA (30 V) max. [Upper/lower limit output, Intensity alarm output,AREA OVER alarm output] Displacement/intensity data output and control input (baud rate: 75 to 19200 bps selectable) Displacement/intensity data output and control input ±0.2% of F.S. ±0.03% of F.S. 100 to 240 VAC ±10%, 50/60 Hz (when using the control unit separately; ±15 VDC ±3%, 0.5 A, +5 V±5%, 3 A) 250 V ~ 3.15 A FAST High Breaking 70 VA max. 0 to 40°C (32 to 104°F), No freezing 35 to 85%, No condensation 70 dB max. Approx. 500 g Approx. 250 g Approx. 6.8 kg (Approx. 2 kg, not including the display unit)

When using a standard specular-reflective object (LC-2420/LC-2430) or white diffuse-reflective object (LC-2440/LC-2450) as a target and the number of averaging measurements is set to 512: The 96-pin connector can be used only when the control unit is used separately. Cannot be used with the LC-2400W. Optionally available for LC-2400A only. Contact us for details. The data was obtained when the standard target and the sensor head were fixed to a steel plate (LC-2420/LC-2430) or an aluminum plate (LC-2440/LC2450). (20°C (68°F) as reference temperature)

84 CHAPTER 9 Appendix

Page 92

9-4 Dimensions

9-4 Dimensions

[Unit: mm]

Display Unit (LC-D1A/LC-D1W)

123

110

(47)

234 320

232 23

297

22

Control Unit (LC-C1A/LC-C1W/LC-C2)

Control Unit LC-C1A 215

• Front view

207.4 1.5

29.1 33

166.7

• Bottom view

• Top view

5.4 4xM5

89.8 4 x M5 159 255.4

194 199

120

67.2 15.2

29.2

29.2

22 5.2 3.4

12.2

194.5

180.2

3.4 230.2

2 x ø4.5

60

30

• Rear view 237

CHAPTER 9 Appendix

85

Page 93

9-4 Dimensions

Sensor head

[Unit: mm]

LC-2420/LC-2430 Regular-reflective Model

LC-2440/LC-2450 Diffuse-reflective

LC-2420

LC-2430

Reference distance (X)

10.5

30.5

Angle formed by transmitted and received beams (θ)

63°

45°

Model

LC-2440

LC-2450

Reference distance (X)

30.5

50.5

Angle formed by transmitted and received beams (θ)

30°

22°

18

18

58.5

58.5

68

60 25 ø10

45.5

1.5

22.5

70

ø7 Cable length: 20 cm

25 ø10

31 1.5

19

ø7 Cable length: 20 cm

61

70

55.5

0.5

12.5

0.5 q

5.5 X

11

(X – 0.5)

3 x ø 4.5 (Mounting hole)

X

4.5

4

(X – 0.5)

60

52

θ

35.5

15

36 22.2

12.5 15.5

17.8

0.5

0.5

2

2

Connecting Cable (Model: OP-21412)

40 58.8

19

86 CHAPTER 9 Appendix

ø7, Cable length: 3 m

51

3 x ø4.5 (Mounting hole)

Page 94

9-5 Characteristics

9-5 Characteristics the target inclination) (Target: Block gauge)

Measurement value (µm) 200

200

200

0

Measurement value (µm)

±0.5°

±2 µm

+0.5° –0.5°

–200

・LC-2420 (Traversal direction with respect to the target inclination) (Target: Block gauge)

0.8 µm

Measurement Error due to Target Inclination ・LC-2420 (Longitudinal direction with respect to (Typical)

–200

0

200

Distance (µm)

Distance (µm)

–200

–200 +0.5° –0.5°

+0.5° –0.5°

・LC-2430 (Longitudinal direction with respect to the target inclination) (Target: Block gauge) Measurement value (µm)

±0.5°

600

Measurement value (µm) 2 µm

600

±5 µm

+0.5° –0.5°

・LC-2430 (Traversal direction with respect to the target inclination) (Target: Block gauge)

–600

600

0

0

–600

600

Distance (µm)

–600

–600

+0.5° –0.5°

・LC-2440 (Longitudinal direction with respect to the target inclination) (Target: Standard white diffuse-reflective object)

3

+0.5° –0.5°

・LC-2440 (Traversal direction with respect to the target inclination) (Target: Standard white diffuse-reflective object)

Measurement value (mm)

±10°

3

Measurement value (mm)

–3

0

3

10 µm

±10 µm

+10° –10°

Distance (µm)

–3

0

3

Distance (mm)

–3

+10° –10°

Distance (mm)

–3

+10° –10°

CHAPTER 9 Appendix

87

Page 95

9-5 Characteristics

・LC-2450 (Longitudinal direction with respect to the target inclination) (Target: Standard white diffuse-reflective object) Measurement value (mm)

Measurement value (mm)

12

30 µm

µm

12

±10°

±25

+10° –10°

・LC-2450 (Traversal direction with respect to the target inclination) (Target: Standard white diffuse-reflective object)

–12

0

–12

12

0

12

Distance (mm)

–12

Distance (mm)

–12

+10° –10°

+10° –10°

Distance vs. Spot Diameter (Typical) ・LC-2430

・LC-2420

Receiver

Spot diameter [µm]

Spot diameter [µm]

30

30 x

Transmitter x y

20

20 y

Target y

10

10 x (Regular-reflective type)

0

–200

0

+200 [µm] Distance from the center of the measuring range ・LC-2440 Spot diameter [µm] Receiver Transmitter

0

–500

0

+500 [µm] Distance from the center of the measuring range

・LC-2450 Spot diameter [µm] y 400 y

300 300 200

Target

200 x y x

100 x

100

(Diffuse-reflective type)

0

+3 [mm] Distance from the center of the measuring range

88 CHAPTER 9 Appendix

–3

0

0

–8

0

+8 [mm] Distance from the center of the measuring range

Page 96

9-6 Initial Setting List

9-6 Initial Settings List Sensor head Offset Low-intensity limit High-intesity limit Low-intensity limit time High-intensity limit time Gain Number of averaging measurements Measurement mode Auto-zero Filter Cutoff frequency Baud rate Data format Upper/lower limits in each mode Calibration Analog ratio Digit suppress Function

Initializing the LC-2400 Series

LC-2420

LC-2430

LC-2440 LC-2450 0 50 49750 0 ms 0 ms AUTO 2048 NORMAL OFF OFF 100 Hz 1200 bps 8 bits, Parity check: None, Stop bit length: 1 bit (Display panel: ) +200.00 µm +500.00 µm +3.0000 mm +8.0000 mm -200.00 µm -500.00 µm -3.0000 mm -8.0000 mm 200.00 µm 500.00 µm 3.0000 µm 8.0000 µm 50 µm/V 100 µm/V 1000 µm/V 2000 µm/V Full 1CH Follow the procedure below to reset the LC to initial settings.

Note

When you initialize your LC, all settings stored in programs will also be deleted. [LC-2400A series] • Turn the power OFF. • Turn the power ON while pressing and holding the key. • The LC will be initialized and reset to the above initial settings. [LC-2400W series] • Turn the power switch at the rear of the controller off. • Turn the key-operated switch to the ON position while pressing and holding the key. • The LC will reinitialize with the delfault settings shown above.

CHAPTER 9 Appendix

89

Page 97

9-7 Glossary

9-7 Glossary Resolution

The LC series outputs a target displacement value by detecting the displacement of the laser beam reflected on the target surface. The S/N ratio of the electric signal generated by the beam displacement varies depending on the intensity of the received laser beam. The displacement output value fluctuates due to internal noise. The amount of the fluctuation is called resolution.

Linearity

Ideally, the displacement output from the LC would be identical to the target displacement, and the relationship between the output value and the target displacement would be represented by a straight line. Actual measurements, however, deviate slightly from the ideal line. The tolerance range in relation to the ideal line is called linearity.

Taeget displacement

Linearity

Ideal line

Actual measurement

Displacement measurement value

Sampling Frequency

Sampling frequency is defined as the number of samples taken per second.

Response frequency

If a target rotates or oscillates above a certain frequency, the LC cannot detect the target displacement. Response frequency is defined as the frequency when the output level is 3 dB lower than (approx. 70% of) the stable level.

Response Time

Response time is defined as the time required for output voltage to reach ± 10% of the stable level when a target instantaneously moves between two points.

Operating Distance

Operating distance is defined as the distance between the front face of the sensor head and the middle of the measuring range.

90 CHAPTER 9 Appendix

Page 98

9-7 Glossary

Calibration

Calibration is performed to compensate for the measurement deviation from the ideal line that is caused by the difference in target color, material and surface condition. For example, the LC may output 1.990 mm, although actual displacement is 2.000 mm. This deviation results from differences in the target surface, since the laser beam reflects differently depending on the target surface. This is a characteristic of sensors using the principle of laser beam reflection. By calibrating, the measurement deviation between two measurement points can be corrected. The amount of deviation between two measurement points varies depending on the linearity for each target.

Measurement value Deviation Linearity Point A when measuring an actual target Ideal line 0

Deviation

Number of Averaging Measurements

Measurement value Point A

Target displacement

0

Target displacement

Point B

Point B Before calibration

After calibration

The LC samples 50000 pieces of data per second to output measurement values. By averaging the sample data, fluctuation of measurement values can be reduced. The averaging methods are as follows: Number of averaging measurements 1 to 2048 4096 to 131072

Averaging method Moving average of one sampling Moving average of N samplings

N = Number of averaging measurements 2048 Example: Number of averaging measurements: 16 (Shifting one measurement before each new average) Sampling 1, 2, 3, 4, ……… 15, 16, 17, 18, 19, ……

1

2

3

4

20 µs x Number of averaging measurements 1

2

3

Measurement output timing

20 µs 20 µs 20 µs

CHAPTER 9 Appendix

91

Page 99

9-7 Glossary

Example: Number of averaging measurements: 4096 (Shifting "N" measurements before each new average: N =

Sampling 1,

2,

3,

4096 = 2) 2048

5, 6, ……… 4096, 4097, 4098, 4099, 4100, 4101, ……

4,

1

20 µs x Number of averaging measurements

2

3

1

2

20 µs x N

20 µs x N

Measurement output timing

HIGH-INTENSITY LIMIT/LOWINTENSITY LIMIT

The LC is set to the DARK mode 100 µs max. after the intensity monitor output falls below the LOW-INTENSITY LIMIT. Then, the LC returns to the measurement mode 100 µs max. + (20 µs x Number of averaging measurements) after the intensity monitor output exceeds the LOW-INTENSITY LIMIT.

HIGH-INTENSITY LIMIT Intensity monitor output LOW-INTENSITY LIMIT Response time ( 100 µs max.)

Response time+data time sampling time (100 µs max. ) + (20 µs x Number of averaging measurements)

+10V DARK mode

(Example) Analog displacement output 0V

Note

INTENSITY LIMIT TIME

When the intensity monitor output exceeds the HIGH-INTENSITY LIMIT, measurement data is output at the same timing as above.

The BRIGHT and DARK modes can be canceled for a specified time duration. This duration is called the INTENSITY LIMIT TIME. The INTENSITY LIMIT TIME can be set individually for the BRIGHT and Dark modes. Using the INTENSITY LIMIT TIME, a measurement value just before the BRIGHT or DARK mode is retained for a specified time duration. LOW-INTENSITY LIMIT TIME=t

DARK mode=t'

HIGH-INTENSITY LIMIT t''

Intensity monitor output LOW-INTENSITY LIMIT

+10V

Response time ( 100 µs max.)

Response time+data time sampling time (100 µs max. ) + (20 µs x Number of averaging measurements)

t < t' (Example) Analog displacement output 0V t t > t'' (Example) Analog displacement output 0V

Response time

( 100 µs max.)

t

92 CHAPTER 9 Appendix

Page 100

9-7 Glossary

When the LOW-INTENSITY LIMIT TIME (t) is longer than the DARK mode time (t’’), the LC returns to the measurement mode 100 µs max. after the intensity monitor output exceeds the LOW-INTENSITY LIMIT. Note

As for the HIGH-INTENSITY LIMIT TIME, measurement data is output at the same timing as above.

CHAPTER 9 Appendix