Page 99
Chapter 6 Setting ■ When “Connect Unit” is set to “3) Printer” 3) Printer
Parameter Automatic printout * See p. 126. When this parameter is set to ON, data is automatically printed every time OK or NG is output.
LCD display
Set item
16 Auto Print 0) OFF
0) OFF
1) OK only 0) OFF
1) OK only
2) NG only 3) ON
2) NG only 3) ON
Automatic printout format
17 Auto Print= 0) No. + Data
* See p. 126.
0) No. + Data 1) No, only 2) Data only
Returns to 01 Port 1 Baud.
■ When “Connect Unit” is set to “4) Port 2 Direct” 4) Port 2 Direct
Parameter
LCD display
Set item
6 Infinite
Port 1 receive timer * When the specified amount of time has elapsed, received characters are automatically deleted. This parameter does not normally need to be changed.
Direct display
0.1 to 25.5 S
09 Rcv 1 Timer 5.0S
(Initial value) 5.0S
18 Direct Read
* Sets data from which the serial interface is displayed on the LCD display.
0) Port 2
ENTER key function
19 Enter Key=
* Determines whether a serial output is provided when the [ENTER] key is pressed
0) Out Port 2
0) Port 2 1) Port 1, 2
0) Out Port 1 1) Port OFF
Returns to
2) Out Port 2
01 Port 1 Baud.
* When “4) Port 2 Direct” is selected, direct communication is enabled. Direct communication is described on p.9. Before starting direct communication, the following setting procedure should first be performed. • •
Align the baud rate of serial port 1 to that of the connected personal computer. Align the data format (data length, parity and stop bit) of serial port 1 to that of the connected personal computer.
89
Page 100
Chapter 6
Setting
6.2.3 Communication set mode for serial port 2 ■ Menu list Communication settings for serial port 2 Port 2 Baud Port 2 Form (data length, parity and stop bit) Receive Form Rcv 2 Timer ■ Detailed settings
Parameter
Baud rate
Data format Data length: 7 or 8 Parity: n (none), e (even) or o (odd) Stop bit length: 1 or 2
LCD display
01
Port 2 Baud
3)
9600
02
Port 2 Form
0) 1) 2) 3) 4) 5) 6) 7)
Receive format
Port 2 receive timer * When the specified amount of time has elapsed, received characters are automatically deleted. This parameter does not normally need to be changed.
*
90
03
Receive Form
1)
[data]+CR
02
Rcv 2 Timer
0) stx+[data]+etx 1) [data]+CR 2) esc+[data]+CR
5.0S
Returns to 01 Port 2 Baud.
38400 31250 19200 9600 4800 2400 1200 600
00) 7n1 01) 7e1 02) 7o1 03) 7n2 04) 7e2 05) 7o2 06) 8n1 07) 8e1 08) 8o1 09) 8n2 10) 8e2 11) 8o2
01) 7e1
6
* Align this format to the send data format of the connected bar code reader.
Set item
Infinite 0.1 to 25.5 S (Initial value) 5.0S
Change the parameters according to the connected bar code reader. These parameters do not need to be changed when a BL series bar code reader is used. When a bar code reader manufactured by another company is connected, check its communication specifications and change only the required parameters.
Page 101
Chapter 6 Setting
6.2.4 Main setting mode ■ Menu list Main setting Normal (Normal operation) ............ See p. 32 Parallel output form Max. No. of Presets Auto increment Counter Counter protect Preset active function .................... See p. 46 Max. No. of Presets Auto increment Counter Counter protect Verify step function ........................ See p. 50 Max. No. of Presets Auto increment Counter Counter protect Serial No. check function ............... See p. 54 Serial No. direction Starting digit Max. No. of digits Amount of change Serial No. protect
6
Note: The parameters of the main setting mode cannot be changed when the connected device is set to "Serial 2 Direct" in the Serial 1 Communication setting mode.
91
Page 102
Chapter 6
Setting ■ Detailed settings for Normal (Multi-preset) operation, Preset active function, or Verify step function
Parameter Operation setting
LCD display 01 Mode Setup 0) Normal 0) Normal
Set item
0) Normal 1) Active 2) Varify Step 3) Serial No.
1) Active 2) Verify Step
Parallel output form * See p.45
Max. No. of Presets * Set the maximum number of data presets to be registered.
Auto increment * See p.33
02 Out 1. 0) Bit 03 Max Preset No. MAX 10 04 AutoIncrement 0) ON
0) Bit 1) Binary 2) BCD 0 to 399 (Initial value)10 0) ON 1) OFF
Counter * See p.42
05 Counter 1) OFF
0) ON 1) OFF
1) OFF
6 Counter protect 1 * See p.44
Counter protect 2 * See p.44
Counter protect 3 * See p.44
0) ON
06 Counter Mode1 1) Reset @ OFF 07 Counter Mode2 1) Hold @ Preset 08 Counter Mode3 2. 1) Hold @ Print
0) Reset @ OFF 1) Hold @ OFF 0) Reset @ Preset 1) Hold @ Preset 0) Reset @ Print 1) Hold @ Print
Returns to 01 Mode Setup
1. "02 Out" is displayed only when the "01 Mode Setup" is set to "0) Normal". 2. "08 Counter Mode 3" is displayed only when the device connected to serial port 1 is set to "Printer".
92
Page 103
Chapter 6 Setting ■ Detailed settings for the serial No. check function
Parameter Operation setting
LCD display 01 Mode Setup
Set item 0) Normal 1) Active 2) Verify Step 3) Serial No.
3) Serial No. 3) Serial No.
Serial No. direction
09 No. Direction
0) Forward
* See p. 54.
1) Reverse
1) Reverse
Serial No. start digit
10 Start Digit
1 to 32. 1
* See p. 54.
Serial No. effective digits
11 Max Digits
1 to 6. 6
* See p. 54.
Serial No. increment
(Initial value) 1
-128 to +127.
12 No. Change +1
* See p. 56.
Serial No. protect
13 No. Protect
* See p. 56.
0) Reset
(Initial value) 6
(Initial value) +1
0) Reset 1) Continue 2) Repeat
6
Returns to 01 Port 2 Baud.
93
Page 104
Chapter 6
Setting
6.2.5 Option set mode ■ Menu list Optional settings Beep Digit Specify Max. Digits Start Digit Key (Preset) Key (Cancel) Key (Enter) Input (Preset) Input (Cancel) Input (Enter) Input Type Input Time Input User Out ON Timer Out Delay OUT5-1 User, OUT10-6 User NG Output Preset Output ■ Detailed settings
6
Parameter Buzzer sound
LCD display 01 Beep
* Sets the buzzer output timing. When “ON” is selected, a buzzer sound is output when a key operation is performed or data is read. When “Key Only” is selected, a buzzer sound is output only when a key operation is performed.
0) ON
Digit designation function
02 Digit Specify
* See p. 68.
0) OFF
0) ON 1) Key 2) OFF
1) OFF
Digit designation effective digits
Set item
0) OFF 1) Forward 2) Reverse
Any other than 0) OFF
03 Max. Digits 16
* See p. 68. Continued on the next page.
94
Continued on the next page
Continued on the next page
1 to 32 (Initial value) 16
Page 105
Chapter 6 Setting
Parameter
LCD display
Digit designation start digit
04 Start Digit
Set item
1 to 32 (Initial value) 1
1
*See p. 68.
Key protect * Inadvertent pressing of each key is prevented. When the key protect time is set, each key press is accepted only when the key is pressed and held for the specified period (1 to 5 seconds).
05 Key(Preset) 0) 0 Sec Protect
06 Key(Cancel) 0) 0 Sec Protect
07 Key(Enter) 0) 0 Sec Protect
08 Input(Preset) 0) 0 Sec Protect
09 Input(Cancel) 0) 0 Sec Protect
10 Input(Enter) 0) 0 Sec Protect
Parallel input polarity
11 Input Type
* The input polarity can be changed to between I1 and I5 only.
0) N.O.
Continued on the next page
Continued on the next page
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
0) 1) 2) 3) 4)
0 Sec Protect 1 Sec Protect 3 Sec Protect 5 Sec Protect OFF
6
0) N.O.
1) N.C.
95
Page 106
Chapter 6
Setting
Parameter
Parallel input time constant
LCD display
Set item
12 Input Time
1 to 99 ms
10ms
(Initial value) 10 ms
Parallel input release to user
00000
0 : Released 1 : Not released * Order: IN5, IN4, IN3, IN2 and IN1
500ms
(Initial value) 500 ms
5ms
(Initial value) 5 ms
13 Input User
* See p.73.
Parallel output ON time
10 to 2550 ms
14 Out ON Timer
* See p.45.
OK/NG output delay time
0 to 255 ms
15 Out Delay
* See p. 45.
Parallel output release to user
16 Out5-1 User 00000
* See p.73.
NG output selection
18 NG Ouput
0 : Released 1 : Not released * Order: OUT10, OUT9, OUT8, OUT7 and OUT6
* Determines whether the NG output is turned on when the verification result is NG (At Verify NG) or when an error such as PLC link error has occurred (At Error). (See p. 144)
0) At Verify NG
0) At Verify NG
Preset output selection
19 Preset Output
* It can be output when the preset data is updated or the preset mode is selected. When “ON at Update” is selected, OUT10 and OK are turned on when the preset data is updated. When “ON in Preset” is selected, the OUT10 remains ON while the preset mode is selected.
0) OFF
17 Out10-6 User 00000
6
96
0 : Released 1 : Not released * Order: OUT5, OUT4, OUT3, OUT2 and OUT1
1) At Error
Returns to 01 Beep
0) OFF 1) ON at Update 2) ON in Preset
Page 107
Chapter 6 Setting
6.3
Initialization of Each Setting Each setting parameter in the BL-V35E can be initialized to the default value set at the time of shipment. ■ Initializing the setting Procedure 1. Select “7) Init” in “Setup”. 5 ) 7 )
Ou t No I n i t
6 )
Re f
2. Select an item to be initialized using the ▼ and ▲ key, and press the [ENTER] key. The following items can be initialized. 1) Init All: Initializes all settings to the default values at the time of shipment. 2) Init Preset: Clears all preset data. 1 ) 2 )
I n i t
A l l
I n i t
P r e s e t
3) Init Out: Initializes parallel settings to the default values at the time of shipment. 4) Init Counter: Initializes all counter values. 3 ) 4 )
I n i t I n i t
6 Ou t Coun t e r
5) nit Ref Data: Clears the reference data. 5 ) 1 )
I n i t I n i t
Re f A l l
Da t a
3. The message shown below is displayed to request confirmation for the initialization. Press the [ENTER] key to execute initialization. Press the [CANCEL] key to cancel initialization.
PRESET
TIM 1 TIM 2 4 3 2 1 OUT
I n i t ?
PRESET No.
DATA
SHIFT ABC..
OPTOIN
7
8
4
5 y e s = E n t e r
N o = C a n c e l
8 7 6 5 OK 10 9 NG
PRESET
KEYENCE
CANCEL
ENTER
1
2
0
CLEAR
PROGRAM
MONITOR
Executes initialization. Cancels initialization.
97
Page 108
Chapter 6
6
98
Setting