Laser Bar Code Reader BL-700 Series User's Manual: Chapter 1 Laser Safety Precautions, pages 84 to 93

Pages 84 to 93 of 161, from Laser Bar Code Reader BL-700 Series User's Manual

The original document is the PDF (1.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Keyence BL-700, Keyence BL-701 and Keyence BL-781

Page 84

Chapter 4

Installing the Setup Software

Comparing the settings currently edited with a saved file 1. To compare the settings currently edited with the setting file previously saved, click on [Compare].

4 2. Select a file to be compared, and click on [OK]. Then, the system compares the selected file with the settings currently edited, and displays the result. Matched

Not matched

Printing contents of a setting file 1. To print settings, click on [Print]. The following screen will appear.

2. Adjust “Printer Setup” as required, and click on [OK]. Printing will then start.

71

Page 85

Chapter 4

Installing the Setup Software

Restoring the settings currently edited to the default (initial) settings 1. To restore the settings currently being edited to the default settings, click on [Defaults (Initialize settings)]. 2. If you try to initialize the settings currently being edited without saving them, the following message appears. If you wish to execute initialization, click on [OK].

Note: The initialization procedure described in this section is used to initialize the settings being edited with the setup software. The settings on the BL-700 cannot be initialized.

4

72

Page 86

Chapter 4

4.4

Installing the Setup Software

Sending/Receiving Settings To send the updated settings to the BL-700 and to read the settings from the BL700, perform the following procedure.

Sending updated settings to the BL-700 To send updated settings to the BL-700, perform the following procedure. 1. Press the BL-700 TEST switch for 8 seconds. When the 1st, 3rd and 5th STABILITY LEDs from the top flash simultaneously, the communication protocol is temporarily set as indicated below (this status is called “setting data send/ receive waiting status”). • • • • • • Tips

Baud rate: 9600 bits/s Data length: 7 bits Parity: Even Stop bit length: 1 bit PLC link: Disabled Multi-drop link: Disabled

4

If you know the current communication parameter settings of the BL-700, this step is not necessary. You can send data to the BL-700 by setting the communication parameters of the host computer using [COM PORT] so that they conform to the current settings of the BL-700. However, if PLC link is enabled, you cannot send data to the BL-700 without performing this step because the handshaking protocol is set for PLC link only. 2. Click on [COM PORT] to set the communication parameters of the host computer according to the values listed in the step 1. (above procedure). Select an RS-232C port, and click on [OK].

73

Page 87

Chapter 4

Installing the Setup Software 3. Click on [Send settings (Save in BL EEP-ROM)]. The updated settings are then sent to the BL-700.

4. If the settings were successfully sent to the BL-700, the following message appears.

4

If the host computer failed to send the data to the BL-700, the following message appears.

If this message is displayed, check the following points: •

• •

74

Check that [COM PORT] is set as shown in (1) when the 1st, 3rd and 5th STABILITY LEDs from the top flash simultaneously (when in “setting data send/receive waiting status”). Check that the power to the BL-700 is ON. Check that the RS-232C cable pin assignment of the BL-700 is the same as that of the host computer.

Page 88

Chapter 4

Installing the Setup Software

Reading the current settings of the BL-700 To read the current settings of the BL-700, perform the following procedure. (This procedure is almost the same as the procedure for sending settings.) 1. Press the BL-700 TEST switch for 8 seconds. When the 1st, 3rd and 5th STABILITY LEDs from the top flash simultaneously, the communication protocol is temporarily set as indicated below (this status is called “setting data send/ receive waiting status”). • • • • • • Tips

Baud rate: 9600 bits/s Data length: 7 bits Parity: Even Stop bit length: 1 bit PLC link: Disabled Multi-drop link: Disabled

If you know the current communication parameter settings of the BL-700, step 1. is not necessary. You can send data to the BL-700 by setting the communication parameters of the host computer using [COM PORT] so that they conform to the current settings of the BL-700. However, if PLC link is enabled, you cannot send data to the BL-700 without performing step 1., because the handshaking protocol is set for PLC link only. 2. Click on [COM PORT] to set the communication parameters of the host computer according to the values listed in step 1. Select an RS-232C port and click on [OK].

3. Click on [Send settings (Save in BL EEP-ROM)]. When you click on [Read BL settings], the settings of the BL-700 can be read on the host computer.

75

4

Page 89

Chapter 4

Installing the Setup Software 4. If the settings of the BL-700 were successfully read, the following message appears.

If the host computer failed to read the data, the following message appears.

4 If this message is displayed, check the following points: •

• •

76

Check that [COM PORT] is set as shown in (1) when the 1st, 3rd and 5th STABILITY LEDs from the top flash simultaneously (when in “setting data send/receive waiting status”). Check that the power to the BL-700 is ON. Check that the RS-232C cable pin assignment of the BL-700 is the same as that of the host computer.

Page 90

Chapter 4

4.5

Installing the Setup Software

Using Monitor This setup software provides the “Monitor” program to check if the BL-700 can send data properly. The “Monitor” program allows you to display the data read by the BL series on the host computer’s monitor screen, and also to send a command from the host computer to the BL-700 or BL series. 1. Click on [COM PORT] to set the communication parameters of the host computer according to the current settings of the BL-700. 2. Click on [Monitor]. The MONITOR screen will appear.

4

3. When the BL series reads bar codes, the data read by the BL series is listed on the screen (in the [Received Data] field). A record of up to 1000 lines of data can be obtained. Using the scroll bar at the right of the [Received Data] field, you can see the previous data.

Note: All commands should be entered in all uppercase characters. The BL cannot accept lowercase characters.

77

Page 91

Chapter 4

Installing the Setup Software 4. If you click [Stop], received data is not listed. (The button name is changed to [Start].) If you try to display received data again, click on [Start]. 5. When you enter a command in the “Send command” field, you can send the command to the BL-700 (or BL series). The format of the command to be sent is “[Command] + [CR]”. Typing command and pressing [ENTER] key sends the command.

4 6. When you click on the [ARROW] button in the “Send command” field, the commands previously sent are listed. (Up to 100 commands can be listed.)

7. Click on [Quit] to quit this mode.

78

Page 92

Chapter 4

Installing the Setup Software

Starting the test mode from the Monitor Follow the procedure below to start the test mode from the Monitor. 1. Click [COM PORT] to match the computer’s communication protocol with the current BL-700 communication protocol. 2. Click [Monitor] to display the monitor screen.

3. Enter “TEST1” (uppercase) in the send command column and press the [RETURN] key. Readout rate measurement mode starts and the readout data is listed sequentially.

4. To quit the test mode, enter “QUIT” in the send command column and press the [RETURN] key. Note: When you try to enter the test mode by turning on trigger input, you cannot use the serial command to enter the test mode. Commands should be entered in all uppercase characters.

79

4

Page 93

Chapter 4

4.6

Installing the Setup Software

List of Error Messages The following table lists the error messages which may occur during operation of the setup software.

Errors during setup

Error message

Contents

“Entered data is incorrect. [OK]”

The entered data is incorrect. Re-enter the correct data.

Errors during “Communication with communication BL-700 failed. [OK]”

Error during communication with the BL-700 (for sending settings).

“Readout from BL-700 Error during communication with the BL-700 failed. [OK]” (for reading settings). “The specified model is The model set in the setup software is not the incorrect. [OK]” same as the model that is connected. Errors file editing

4

“Accessing file during rejected. [OK]”

No floppy disk is inserted. The floppy disk is write-protected. The floppy disk is full.

“File not found. [OK]”

The file name is incorrect. Enter a correct file name.

“File name incorrect. [OK]”

80