Quick Check 890 Quick Check ImageData Barcode Verifier User's Guide, pages 11 to 20

Pages 11 to 20 of 40

The original document is the PDF (624 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Honeywell QC-890

Page 11

RS-232 Serial Communication Interface An RS-232 interface requires no additional drivers. Connect the QC890 Verifier to the host as shown in the illustration below.

RS-232 Selections Scan the following barcode to select RS-232 communications.

RS-232 Communication The RS-232 settings supported are: Baud Rate:

115,200

Parity:

None

Data Bits:

8

Stop Bits:

1

Flow Control:

None

Bluetooth Bluetooth implementation uses SPP (Serial Port Profile) to connect the QC890 Verifier to the host system. The QC890 SPP is a standard Bluetooth profile implementation. This allows the QC890 system to be compatible with most Bluetooth devices. The QC890 unit can be used with existing Bluetooth radios built into laptops, PCs, or PDAs.

Quick Check® 890 User’s Guide

2-3

Page 12

Using the QC890 Verifier with Bluetooth Devices Note: When using USB or RS-232 interfaces, the Bluetooth communication interface automatically turns off. Scan the Bluetooth barcode below to enable the Bluetooth option. Default = Off.

Bluetooth On Once the Bluetooth option is enabled, scan the appropriate barcode that follows to connect to a PC or laptop, or a PDA/Mobility Systems device. To establish a Bluetooth connection to a PC or a laptop, scan the barcode below, follow the instructions supplied with your Bluetooth device to locate the QC890 Verifier, and connect to it. If you go out of range with your QC890 Verifier, the QC890 Verifier automatically reconnects to the Bluetooth device once you come back in range. When Bluetooth communication is enabled, the default is PC/ Laptop.

* Bluetooth Connection - PC/Laptop To establish a Bluetooth connection to a PDA or a Hand Held Products Mobility Systems device, scan the barcode below, follow the instructions supplied with your Bluetooth device to locate the QC890 Verifier, and connect to it. If you go out of range with your QC890 Verifier, the QC890 Verifier will not reconnect. It will be up to the host device to reconnect to the QC890 Verifier.

Bluetooth Connection - PDA/Mobility Systems Device

2-4

Quick Check® 890 User’s Guide

Page 13

Enabling Your Device for Bluetooth After scanning the appropriate Bluetooth connection barcode, you need to run the Quick Check ImageData application to enable the Bluetooth radio in your device (PC, PDA, or Mobility Systems device) in order to look for the QC890 Verifier.

Once the QC890 Verifier is recognized, it appears in the list of available devices. Select the QC890 Verifier to which you want to bond and select Next.

Quick Check® 890 User’s Guide

2-5

Page 14

You will be asked if you need a passkey to add your device. The default passkey for all Hand Held Products QC890 verifiers is “1234.”

The Wizard loads the QC890 Verifier information into your device. You may click a checkbox to indicate the QC890 Verifier is the default device to which Quick Check ImageData connects. Once you select Finish, your device is ready to communicate with the QC890 Verifier via the assigned ports. Make a note of the assigned port number. You will need to know this port number in the following installation steps.

Enabling QCID for Bluetooth 1.

Once you open the QCID application, click on File -> Preferences.

2.

Choose the assigned incoming port number reported in the Add Bluetooth Device wizard and click OK.

3.

Your QC890 Verifier connects to the QCID application.

2-6

Quick Check® 890 User’s Guide

Page 15

Using Quick Check ImageData For the PC version of Quick Check ImageData, start Quick Check ImageData by clicking on Start->Programs->Hand Held Products->Quick Check ImageData or double click the Quick Check ImageData icon on your desktop (set up during installation). For the PDA version, start Quick Check ImageData by clicking on Start ->Programs. Select the Quick Check ImageData icon. The first time you use QCID, the application asks you to select the COM Port to which the QCID is attached. After you make the selection, QCID attempts to connect to the attached QC890 Verifier. When a connection is established, the main screen of the application, as shown below, displays. PC Main Screen

Quick Check® 890 User’s Guide

2-7

Page 16

PDA Main Screen

Help System The Quick Check ImageData program has an extensive help system to aid in the verification of your barcodes. Please refer to the help files for questions regarding the verification process and for specific questions regarding the QCID program.

2-8

Quick Check® 890 User’s Guide

Page 17

3 Using the Quick Check 890 Verifier System Scanning a Barcode Symbol Using the QC890 Verifier

!

Calibration is a requirement of the ISO/IEC standards. Without proper initial calibration, the values displayed may not be accurate. Please be sure to calibrate the QC890 Verifier prior to scanning barcodes and when you change apertures and/or external scanning devices.

1. Place the QC890 Verifier on top of the symbol. 2. Press the scan button on top of the unit to turn on illumination. Look through the unit’s window to align the unit with the barcode. Note: The best verification data is gathered by aligning the center of the symbol to the center arrows.

3. Press the scan button on top of the unit again while the illumination is on to scan and verify the barcode. 4. To scan PDF417 barcodes, hold the scan button down and move the verifier up and down so the aiming line sweeps from the top of the bar code to the bottom, and back up. This ensures that the entire code is scanned.

Scanning a Barcode Symbol Using Quick Check ImageData 1. Place the QC890 Verifier on top of the symbol. 2. If necessary, press the scan button on top of the unit to turn on illumination to aid in alignment of the symbol. 3. For the PC version, click the Scan button on the Main Display window.

Quick Check® 890 User’s Guide

3-1

Page 18

4. For the PDA version, select Device and then Scan from the Scan Data tab.

Calibrating the QC890 Verifier

!

Calibration is a requirement of the ISO/IEC standards and must be performed with each aperture change.

When calibrating the QC890 Verifier system, the QC890 Verifier must be connected to the Quick Check ImageData software. During calibration, it is important to align the center of the barcode symbol to the center arrows, and keep the QC890 Verifier flat and level to the surface of the reflectance page. The reflectance page is a precision standard that is made of photographic materials. To ensure proper calibration, you must replace the reflectance page if it becomes dirty or physically damaged. Make sure to use only the original calibration tools that were supplied with your QC890 Verifier kit or certified replacements obtained from Hand Held Products. 1. Place the verifier on top of the reflectance page. Use the symbol alignment window to center the reflectance page. 2. Press the scan button on top of the unit to turn on illumination to aid in alignment.

Note: The best calibration is performed by aligning the center of the symbol to the center arrows. 3. Press the scan button on top of the unit a second time while the illumination is on to scan the calibration barcode. You will hear a series of beeps and the

3-2

Quick Check® 890 User’s Guide

Page 19

illumination will turn off. If the calibration process is successful, the following screen displays:

4. Enter your name and the location where the calibration was performed and click on OK. The following screen displays:

5. If the unit does not calibrate, contact Hand Held Products Technical Support (see Technical Assistance on page 5-1). To access calibration logs, go to Settings->Calibration->Calibration Log. Enter the password visions07.

Scanning Properly Using a Pen/Mouse Wand Note: It is important to perform reflectance calibration prior to using any pen or mouse wand. See "Calibrating an External Pen/Mouse Wand" on page 3-4. 1. Turn on the verifier and confirm that the pen/mouse wand is securely connected. 2. Hold the pen wand as you would a pen or pencil, placing your fingers on each side of the foot near the tip:

or

Quick Check® 890 User’s Guide

3-3

Page 20

Hold the mouse wand horizontally so it moves across the barcode in the direction of the arrows:

3. Lightly touch the wand against either quiet zone (clear area) on either side of the barcode. For the pen wand, you will have the proper scanning angle when the tip and the reflectance guide (black plastic clamp) are flat on the scanning surface. 4. At about the same speed with which you would quickly underline a sentence in this manual, and without interruption, gently move the wand across the center of the barcode. You do not need to press hard on the wand while scanning the barcode. Gentle pressure on the wand will not scratch the contacted surface, but pressing down hard may damage the code and yield poor results. 5. Listen for audible tones. • •

If you hear a single beep tone followed by another single beep of the same tone, it signifies a good read. If you hear a single beep and then a beep of a different tone, the unit detected an error in the scanned symbol.

Calibrating an External Pen/Mouse Wand 1. Using the proper scan technique described in step 4 above, scan the reflectance page. You will hear a single beep after the first good scan and the following display appears:

2. Continue to scan the reflectance page for a total of 10 good scans. After each good scan, a single beep sounds and the display indicates the number

3-4

Quick Check® 890 User’s Guide