Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 8 of 12

Quick Check 600/800 Series Bar Code Verifiers User's Guide, pages 71 to 80

Pages 71 to 80 of 120

The original document is the PDF (1.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the PSC QC-800, PSC QC-600, Datalogic QC-800 and HHP QC-800

Page 71

5. Slide the paper’s straight edge into the slot between the paper compartment and the printer compartment until about one-quarter of an inch is inserted.

6. Press the power switch to the center On position and wait a few seconds. 7. While holding the paper in place, press and hold the power switch in the Paper Feed (down right) position until an inch of paper emerges from the printer and then release the switch. 8. Pull the paper through the printer until several inches are exposed. 9. Put the paper spindle into the paper roll and place the roll and spindle onto the grooves near the back of the printer. 10. Roll the paper to take up any slack inside the printer. 11. Ensure the roll turns freely. If the paper roll does not turn freely, the paper may jam and damage the mechanism. 12. Slide the paper’s straight edge through the slot in the printer cover. 13. Replace the printer cover by pushing the back of it down into place and pressing the front down to lock it in place.

♦ To remove the ribbon from the QCP or QCHSP: 1. Press the printer power switch to the Off (down left) position. 2. Remove the printer cover by locating the four small grooves embossed on each side of the printer cover and tilting the cover open by pushing down gently on both sets of these grooves. The ribbon cartridge should be visible. 3. Push down on the right side of the ribbon cartridge (marked PUSH) and remove the cartridge from the printer compartment.

Quick Check® 600/800 Series User’s Guide

5-9

Page 72

4. Slide the paper through the gap between the inked ribbon and the frame of the new ribbon cartridge. 5. Set the new cartridge into the printer compartment and press both sides to secure it in place. The cartridge must be properly seated and aligned for the best print quality. If you get ribbon ink on the printer’s plastic case, wipe it off immediately. Once the ink dries, it is difficult to remove. 6. Replace the printer cover by pushing the back of it down into place and pressing the front down to lock it in place. 7. Turn the cartridge ribbon advance knob (marked by an arrow on the cartridge) clockwise to remove slack from the ribbon. 8. Replace the printer cover.

♦ To connect the QCP or QCHSP to the Quick Check 600/800 Verifier: 1. Press the printer power switch to the Off (down left) position. 2. Plug the transformer into a VAC outlet. 3. Plug the interface cable into the socket on the side of the verifier. 4. Connect the 25-pin connector of the interface cable into the rear of the printer. Secure the connector by tightening the screws on each side of the connector. 5. Press the power switch to the On (center) position. The red LED lights and a Ready message appears indicating that the printer’s built-in microprocessor has verified its internal operation.

♦ To perform a QCP or QCHSP self-test: The self-test tests the print head and ribbon. 1. Ensure the power switch is in the Off (down left) position and there is paper in the printer. 2. Plug the power source into the printer. 3. Press and hold down the power switch in the Paper Feed (down right) position until the LED lights and the printer starts to operate. 4. Release the power switch. The printer performs the self-test and prints its full set of characters.

5 - 10

Quick Check® 600/800 Series User’s Guide

Page 73

Troubleshooting the QCP or QCHSP Printing difficulties may be caused by selecting the incorrect printer interface configuration for the type of printer you are using. • For the QCP, select QCP. • For the QCHSP, select HSP. For instructions on selecting the printer interface setting, see Chapter 3, Changing the Settings. If you use your printer infrequently, the print impression can become weak if the ribbon dries out. If the printed material is difficult to read and you suspect this is the cause of the problem, press the power switch to the Paper Feed position and advance to a properly inked portion of the ribbon.

Quick Check® 600/800 Series User’s Guide

5 - 11

Page 74

5 - 12

Quick Check® 600/800 Series User’s Guide

Page 75

6

Background Information

General Bar Code Information Bar code technology allows you to get information into a computer much more accurately than keying. Quick Check 600/800 Verifiers measure the quality of the bar code symbols to ensure the proper information is encoded and can be read. This instrument is both a quick, easy-to-use pass/fail tester for bar code symbols and a powerful measuring tool for detailed testing of these same symbols. You can set your verifier to use either traditional or ISO/IEC parameters to analyze bar code symbols. The quality information from the verifier is presented through an easy-to-read liquid crystal display (LCD), light emitting diodes (LEDs), audible tones and printouts. Bar code symbols consist of a series of parallel, adjacent bars and spaces. Data is encoded in the width patterns of the symbol. These patterns, when interfaced with computer software, provide a large amount of information. Bar code systems offer a high degree of confidence that data entry is accurate. The substitution error rate, the rate at which a character may contain an error, is very low. The first read rate, the probability (percentage) that an attempt to input data will result in data being captured on the first attempt, is very high. Bar code technology dates back to the 1960s and early 1970s, but did not come into widespread use as we know it today until the later 1970s. Retail applications, primarily grocery stores, saw the need for automatic identification technology and pushed development ahead. As many other industries began to take advantage of bar code technology, they often needed to develop their own symbology, or type of bar code, with unique specifications and standards. Depending on the data that needs to be recorded, several different symbologies can be used. Communication cannot occur unless the reading and printing tools use a compatible symbology. The symbologies tested by the verifier include: • • • • •

EAN/UPC Code 39 Interleaved 2 of 5 Code 128 Code 93

• • • • •

Codabar Code 11 Code 16K Code 49 MSI Code

Quick Check® 600/800 Series User’s Guide

6-1

Page 76

The industry standards (applications) tested by the verifier include: • • • • •

AIAG BOOKLAND CCBBA COMPTIA/CTIA HIBCC

• LOGMARS • SCC (SCS/ITF, UCC/ EAN-128) • SISAC • UPC Coupon Code

Symbology Specifications The reading and printing specifications for each symbology are defined by various organizations. The verifier uses these specifications to give you quantitative and qualitative information. You can set the verifier to provide different standards to suit your company’s internal quality control guidelines.

Symbologies The following table lists the symbologies that your Quick Check 600/800 Verifier can verify, the major industries that use each symbology, and the organizations that hold the standards. Following the table is a list of the abbreviations for each organization, along with an address and phone number you can write or call for additional information. Symbology

Industries

Organization

UPC grocery, identifies product and manufacturer blood banking, Library of Congress, photo finishing, air parcel delivery automotive, health, industry, non-retail

GS1

Codabar & USS Codabar

CCBC AIM-USA

Code 11 Interleaved 2 of 5 industrial shipping containers for grocery warehousing and distribution

DOD, AIAG, HIBCC, AIM-USA AIM-USA GS1 AIM-USA

Code 16K health, microcomputer serials industry, UCC shipping containers, health, industry health, industrial industrial shelf marking

HIBCC,ABCD AIM-USA SISAC, GS1, HIBCC AIM-USA AIM-USA

Code 39

Code 128 Code 49 Code 93 MSI Code

6-2

Quick Check® 600/800 Series User’s Guide

Page 77

ABC (formerly CCBC) - America's Blood Centers, 725 15th Street NW, Suite 700, Washington, DC 20005, Tel: (202) 393 5725, Fax: (202) 393 1282, http://www.americasblood.org/ AIAG - Automotive Industry Action Group, 26200 Lahser Rd., Suite 200, Southfield, MI 48033, Tel: (248) 358-3570, http://www.aiag.org/Index.cfm AIM North America - A chapter organization of AIM, Inc., 125 WarrendaleBayne Road, Warrendale, PA 15086, Tel: (724) 934 5688, Fax: (724) 934 4495, http://www.aimusa.org/ BASIC - Book and the Serial Industry Systems Advisory Committee, SubCommittee of the Book Industry Study Group (BISG), 19 West 21st Street, Suite 905, New York, NY 10010, Tel: 646-336-7141, Fax: 646-336-6214, email: info@bisg.org Comp TIA - Computing Technology Industry Association, 1815 S. Meyers Road, Suite 300, Oakbrook Terrace, IL 60181-5228, Tel: (630) 678 8300, Fax: (630) 678 8384, http://www.comptia.org/association_home.aspx DoD Logistics AIT Office - Automatic Identification Technology, 8725 John J. Kingman Road, Fort Belvoir, VA 22060, Tel: (703) 767 6200, http://www.dodait.com/ GS1 (formerly UCC and EAN) Brussels - Blue Tower, Avenue Louise, 326, BE 1050 Brussels Belgium, Tel: +32 2 788 7800, Fax: +32 2 788 7899 GS1 US - Princeton Pike Corporate Center, 1009 Lenox Drive, Suite 202, Lawrenceville, New Jersey, 08648 USA, Tel: (609) 620 0200, Fax: (609) 620 1200, E-mail: info@gs1us.org, http://www.gs1us.org HIBCC - Health Industry Business Communications Council, 2525 E Arizona Biltmore Circle, Suite 127, Phoenix, AZ 85016, Tel: (602) 381 1091, Fax: (602) 381 1093, E-mail: info@hibcc.org, http://www.hibcc.org/

Quick Check® 600/800 Series User’s Guide

6-3

Page 78

Symbology Characteristics Your Quick Check 600/800 Verifier can handle bar code symbologies with a variety of characteristics. To help understand symbologies, you should be familiar with the following commonly used terms: •

Code refers to the actual data contained in the bar code symbol, such as a part number, serial number, transaction code or other type of data. • Symbol refers to the actual arrangement of parallel bars and spaces that encode the data. • The character set describes the range of data characters that can be encoded in a given symbol. For example, UPC uses only numbers and is called a numeric symbology, whereas Code 39 uses alphabetical characters, numbers and special characters. • There are two symbology types, discrete and continuous. • In a discrete code, each character stands alone and can be decoded. Between characters is a loosely toleranced intercharacter gap which contains no information. Every discrete character has a bar on each end. One example of a discrete code is Code 39. • In a continuous code, there are no intercharacter gaps. Every character begins with a bar and ends with a space. The end of one character is indicated by the start of the next character. An example of a continuous code is UPC. • The bar and space widths can vary within or between symbologies; those with only two-element widths (wide and narrow) versus those that use multiple widths. • Two-width symbologies, such as Code 39, have a ratio of wide to narrow typically between two and three. • Multiple-width symbologies, such as UPC, allow the bars and spaces to assume more than two different widths. Most multiple-width symbologies have characters whose length is subdivided into a predetermined number of modules. In addition, the width of each bar or space is always an integral number of modules. Bar code symbologies vary on the amount of information that can be encoded in a given length. Usually, only characters have a specified density, since the overall length of a symbol must include other characters. These other characters may include a start/stop code and a check character. • A start code is a pattern of bars and spaces that appears at the beginning of a symbol to inform the reading tool where the symbol begins. • A stop code is a pattern placed at the end of a symbol for marking the end of the data characters. Sometimes the start and/or the stop characters also indicate the scanning direction. Quiet zones are areas at the beginning and end of a bar code symbol that allow the optical equipment to differentiate a bar code from other printed material. Most of the symbologies in use today are bidirectional; this means that they can be read by a scanner either left-to-right or right-to-left without affecting the decoded data.

6-4

Quick Check® 600/800 Series User’s Guide

Page 79

A check character is a character placed in a predetermined position in a symbol and whose value is based on some mathematical relationship of the other characters in the symbol. It is used by the scanner to validate that the correct data has been decoded. If a numeric symbology contains a check character, it is often called a check digit. The following illustration shows many of these bar code parameters.

Bar code symbol parameters

Quick Check® 600/800 Series User’s Guide

6-5

Page 80

The following table lists some characteristics of the nine bar code symbologies decoded by the Quick Check 600/800 Verifier.

Symbology

Character Set

UPC EAN Codabar/USS Codabar numeric numeric numeric with 6 special characters alphanumeric, 26 uppercase letters, 10 digits, 7 special characters numeric and --numeric alphanumeric with all function codes and special characters (full ASCII set) alphanumeric with all function codes and special characters (full ASCII set) alphanumeric (all 128 ASCI characters)

Code 39

Code 11 Interleaved 2 of 5 Code 16K

Code 128

Code 49

Code 93 MSI Code alphanumeric (all 128 ASCI characters) numeric

Element widths 4 widths 4 widths 2 widths 2 widths

3 widths 2 widths 2 widths

4 widths

6 widths (stacked symbology) 4 widths 4 widths

How a Bar Code Verifier Works A bar code reader or scanner extracts the information that is optically encoded in a bar code symbol and converts it into computer-compatible digital data. Hand Held Products provides a variety of handheld wand or pen scanners with the verifier that are designed to minimize scratching the symbol. When you place the wand on a bar code symbol and move it across the symbol, electro-optical techniques read the symbol. All wands contain a light emitting diode (either visible or infrared light) which illuminates the symbol. The area of the symbol that is actually being scanned at any given time is referred to as the “spot” and is defined by an aperture. The spot size varies depending on which wand is used. The spot is formed by flooding the symbol with light. A tightly focused aperture collects the light. The light reflected from the spot is directed to a photodiode detector which generates a small current

6-6

Quick Check® 600/800 Series User’s Guide