The text of a manual · part 3 of 12
Quick Check 600/800 Series Bar Code Verifiers User's Guide, pages 21 to 30
Pages 21 to 30 of 120
The original document is the PDF (1.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the PSC QC-800, PSC QC-600, Datalogic QC-800 and HHP QC-800
Page 21
Documentation *QC600/800 User’s Guide & Programming Manual . . . . QC600-800-UG *QC600/800 Reflectance Page . . . . . . . . . . . . . . . . . . . . . . . . . QCRFPG Printers QC Standard Speed Dot Matrix Printer . . . . QCP68-USA/QCP68-EURO QC High Speed Dot Matrix Printer . . QCHSP68-USA/QCHSP68-EURO Printer supplies Power supply for QCP & QCHSP Printers. . . . . . . . . . . . . . . . . . QCPAC Printer ribbon for QCP & QCHSP . . . . . . . . . . . . . . . . . . . . . . . . . QCPR Printer paper for QCP & QCHSP (4 roll/pack). . . . . . . . . . . . . . . . QCPP Interface cable for QC600/800 to QCP or QCHSP . . . . . . . . . . . . QCPC Wand Scanner QCIW05V Pen Wand 5 mil visible red . . . . . . . . . . . . . . . . . . . . . .622301 QCIW06V Pen Wand 6 mil visible red . . . . . . . . . . . . . . . . . . . . . .622302 QCIW10V Pen Wand 10 mil visible red . . . . . . . . . . . . . . . . . . . . .622299 QCMWI03 Mouse Wand 3 mil infrared . . . . . . . . . . . . . . . . MOUSE3MIR QCMWV03 Mouse Wand 3 mil visible red . . . . . . . . . . . . MOUSE3MVR QCMWI05 Mouse Wand 5 mil infrared . . . . . . . . . . . . . . . . MOUSE5MIR QCMWV05 Mouse Wand 5 mil visible red . . . . . . . . . . . . MOUSE5MVR QCMWI06 Mouse Wand 6 mil infrared . . . . . . . . . . . . . . . . MOUSE6MIR QCMWV06 Mouse Wand 6 mil visible red . . . . . . . . . . . . MOUSE6MVR QCMWI10 Mouse Wand 10 mil infrared . . . . . . . . . . . . . . MOUSE10MIR QCMWV10 Mouse Wand 10 mil visible red . . . . . . . . . . MOUSE10MVR QCMWI20 Mouse Wand 20 mil infrared . . . . . . . . . . . . . . MOUSE20MIR QCMWV20 Mouse Wand 20 mil visible red . . . . . . . . . . MOUSE20MVR
Quick Check® 600/800 Series User’s Guide
1 - 15
Page 22
1 - 16
Quick Check® 600/800 Series User’s Guide
Page 23
2
Quick Start Guide
If you are familiar with bar code verifiers, this section will get you up and running in a few minutes. We assume the factory defaults are in effect. 1. Install the batteries • • •
Slide off battery cover on the bottom of unit. Insert four batteries into compartment, matching “+” on battery ends with “+” in case. Slide battery cover back in place.
2. Charge the batteries • • • •
• • •
• • •
•
•
Chapter 1
Scan reflectance page if you’re using either a pen or mouse wand. Three beeps (one long, two short) sound after a good scan. Continue to scan page for ten good scans.
5. Scan the bar code symbol •
Chapter 1
Press POWER. Unit emits four quick beeps. If unit displays: RECALIBRATE, calibrate unit. Go to step 4. LOW BATTERY!, charge batteries. Go to step 2. NO PRINTER ATTACHED... records may be stored in buffer. Use arrow keys to select an option. Either continue to scan using current AutoPrint/Store setting, clear (delete) records, or change AutoPrint/ Chapter 5 Store setting to Off (not store data). Go to step 5. PRINT BUFFER... records are stored in buffer. Use arrow keys to select an option. Either continue to scan and leave records in buffer, print records, or clear (delete) records. Go to step 5. SCAN, go to step 5. QUICK-CHECK Opening Menu, go to step 5.
4. Calibrate the unit •
Chapter 1
Plug AC charger into port on back of unit. Plug charger into a VAC wall outlet. Charge completely discharged batteries at least 12 hours; charge low charge batteries less time. After charging, remove AC charger from unit.
3. Turn on the verifier •
Chapter 1
Chapter 3
The scan grade display appears. (Extended Accuracy setting default is off). To obtain final symbol grade, scan symbol nine more times (# Scan/ Symbol setting default is 10).
6. Review the scan results
Quick Check® 600/800 Series User’s Guide
Chapter 4
2-1
Page 24
• •
The standard decode display appears. Use arrow keys to view data OR Scan SSTR Codes to view data.
7. Print your scan results • •
If the Quick Check 600/800 Verifier is connected to a printer and it is turned on, results print automatically (AutoPrint/Store setting default is On). If a printer is not connected or is turned off, results are stored in buffer.
8. Review the option settings: •
2-2
Chapter 3
Chapter 3
If data parameters are not customized the way you want, press SELECT to display Print Results menu and then press Ø three times to change the settings.
Quick Check® 600/800 Series User’s Guide
Page 25
3
Changing the Settings
There are four types of options on the Quick Check 600/800 Verifier you can customize: specifications (applications and symbologies), test criteria, scanning options, and output options. Each option has several items that you can turn on, turn off, or change a value. This section describes all the selectable option settings. You can easily review and change the settings using SELECT and the ↑, ↓, ← and → buttons. You save settings by displaying their name and then making the changes you want. A separate save command is not necessary. You can exit the settings menu at any point and scan a bar code symbol by placing the wand tip or the mouse wand on a quiet zone and scanning or by pressing POWER. It may be useful to create a customized Command Code for each type or set type of operations you perform. When you need to perform another procedure, just scan the Command Code that contains those settings. Refer to instructions for making and using a Command Code starting on "Using a Command Code" on page 3-16. Sometimes you may want to change just one or two of the settings, but not create a new Command Code that differs only slightly from the one you frequently use. You can easily change settings after you use a Command Code.
Chapter 10, Flow Charts contains a flow chart that displays all the settings changes you can make and how to access each of them using the menus.
Changing the Specifications The Quick Check 600/800 Verifier specification section consists of the industry application standards and symbologies. The Quick Check 600/800 Verifier can verify the quality of the following symbologies and industry application standards:
Symbologies
Industry Applications
• • • • • • • • • • • •
• • • • • • • • •
EAN/UPC Code 39 Interleaved 2 of 5 Code 128 Code 93 Codabar Code 11 Code 16K Code 49 MSI Code IATA 2 of 5 Regular 2 of 5
Quick Check® 600/800 Series User’s Guide
AAIG Comp Tia LOGMARS SCC HIBCC CCBBA SISAC BOOKLAND Coupon Code
3-1
Page 26
Symbologies You can enable (turn on) the symbologies available on your Quick Check 600/ 800 Verifier so they are recognized. When a symbology is enabled, you can modify some of its test parameters. "Symbology Specifications" on page 3-3 lists the parameters you can modify and their factory installed defaults. The parameters are described after the table.
3-2
Quick Check® 600/800 Series User’s Guide
Page 27
Symbology Specifications Symbology
Parameters (defaults are in bold)
EAN/UPC
Enable: On or off Magnification factor: (080/085/090/100/105/110/115/120/130/140/150/ 160/180/200) Expand U.P.C.-EO: Yes or No Enable: On or off Check Character: None or Checked Fixed W/N: None or a value between 1.4 and 3.9 + 0.2 Fixed Msg Len*: None or a value between 1 and 49 characters Enable: On or off Check Digit: None or Checked Fixed W/N: None or a value between 1.4 and 3.9 + 0.2 Fixed Msg Len*: None or a value between 1 and 78 digits Enable: On or off Modes Shown? Yes or No Fixed Msg Len*: None or a value between 1 and 70 characters Enable: On or off Fixed Msg Len*: None or a value between 1 and 80 characters (from Code 39 set) Enable: On or off Character Font & Tolerances: Trad’l,”USS” or ISO/IEC’83 Fixed W/N: None or a value between 1.4 and 3.9 + 0.2 Fixed Msg Len*: None or a value between 1 and 61 digits Enable: On or off 1 or 2 check digits Fixed Msg Len*: None or a value between 1 and 80 digits
Code 39
Interleaved 2 of 5
Code 128 Code 93
Codabar
Code 11
Code 16K
Code 49
MSI Code
IATA 2 of 5 Regular 2 of 5
Enable: On or Off Modes Shown? Yes or No Fixed Msg Len*: None or up to 154 characters, full symbol Enable: On or Off Fixed Msg Len*: None or up to 49 alpha (81 numeric characters, full symbol Enable: On or Off User Check: None, Mod 10 or Mod 11 Fixed Msg Len*: None or a value between 1 and 50 digits Enable: Yes or No Enable: Yes or No Enable: Checked or None
* The message length specification for each symbology is listed. Although you can enter any value between 0 and 255, the unit does not decode more than the values indicated by the specification.
Quick Check® 600/800 Series User’s Guide
3-3
Page 28
Magnification Factor The Magnification (Mag) Factor applies to the EAN/UPC symbologies. By setting the size of the bar code symbols that you test, you define appropriate tolerances. You can select a mag factor from the following values: 80%, 90%, 95%, 100%, 105%, 110%, 115%, 120%, 125%,130%, 140%, 150%, 160%, 180% or 200%. The default is 100%.
Expand U.P.C./EO The Yes setting for this parameter allows you to expand the compressed U.P.C. code. When you expand the scan data, the digits are shifted to the left and + appears on the display to indicate that additional data is available for your review. The No setting for this parameter presents the data in zero suppression format.
Check Options The check character, check digit and UserCheck are optional characters or digits in some symbologies that are placed in a predetermined position in a symbol. You can use them to validate that the correct data has been decoded. Check characters in a symbol may be mandated by a bar code symbology industry specification, by you, by both, or not at all. Refer to the relevant industry’s publication for specific information on check characters. • When you select Checked, the verifier looks for and checks the check character or digit. • When you select None, the verifier does not look for the check character or digit. None is the default setting. Selected symbologies offer additional options: • Code 39 contains an optional Mod 43 check character. • Interleaved 2 of 5 contains an optional 3-1-3-1 weighted check digit. • MSI Code can contain a second check digit (UserCheck), either the Modulo (Mod) 10 or Modulo (Mod) 11.
Modes Shown This parameter allows you to have additional information appear on the display and on the printout. This information includes special extra characters and control characters within the encoded message. Having this extra information displayed may help you analyze reading or printing problems. In Code 128, for example, you can display starting mode, mode changes and shifts. Code 16K allows you to display similar additional characters.
3-4
Quick Check® 600/800 Series User’s Guide
Page 29
The Yes setting for this parameter is the default. Some of the extra characters are represented by special character “strings” as shown in the following Character Translation Table. When you select the No setting, this information does not appear on the display or the printout.
Character Translation Table (Code 128, Code 16K, and Code 49) Character
On LCD
On Printout
Character
On LCD
On Printout
00(NUL) 01(SOH) 02(STX) 03(ETX) 04(EOT) 05(ENQ) 06(ACK) 07(BEL) 08(BS) 09(HT) 0A(LF) 0B(VT) 0C(FF) 0D(CR) 0E(SO) 0F(SI) 10(DLE) 11(DCI) 12(DC2) 13(DC3) 14(DC4)
0 1 2 3 4 5 6 7 8 9 A B C D E F 10 11 12 13 14
<NL> <SH> <SX> <EX> <ET> <EQ> <AK> <8L> <BS> <HT> <LF> <VT> <FF> <CR> <SO> <SI> <DE> <D1> <D2> <D3> <D4>
15(NAK) 16(SYN) 17(ETB) 18(CAN) 19(EM) 1A(SUB) 1B(ESC) 1C(FS) 1D(GS) 1E(RS) 1F(US) 7F(DEL) FCN1 FCN2 FCN3 FCN4 Mode A Mode B Mode C Shift Pad**
15 16 17 18 19 1A 1B 1C 1D 1E 1F * F1 F2 F3 F4 A B C Sh Pd
<NK> <SY> <EB> <CN> <EM> <SB> <ES> <FS> <GS> <RS> <US> <DL> <F1> <F2> <F3> <F4> <A:> <B:> <C:> <Sh> Pd
* No “special” character is displayed for <DEL>. ** For Code 16K only.
Character Font & Tolerances Codabar supports two options for Character font and tolerances. • Trad’l (Traditional Codabar) uses traditional print tolerances based on Monarch’s original print tolerances. This is the default setting. • USS Codabar uses the ISO/IEC 1983 MH10.8M standards. Refer to Codabar and ISO/IEC publications for specific information on print tolerances.
Quick Check® 600/800 Series User’s Guide
3-5
Page 30
Wide-to-Narrow Ratio The wide-to-narrow ratio (W/N) applies to symbologies with two element widths, including Codabar, Interleaved 2 of 5, and Code 39. “W/N” is the ratio of wide to narrow elements. Only one symbology can have a W/N ratio at a time. The range of values is 1.4 to 3.9 ± 0.2.
Fixed Message Length You can set a Fixed Message Length for any symbology except EAN/UPC, but only one symbology can have a fixed message length at a time. You can enter the message length for any value between 0 and 255. "Symbology Specifications" on page 3-3 lists the message length specification for each symbology.
♦ To modify the symbology options: 1. Display the menu that begins with Print Results (press POWER and then SELECT). 2. Set the cursor on the asterisk next to Change Settings and then press SELECT. A menu that begins with Specifications appears. 3. With the cursor on the asterisk next to Specifications, press SELECT. A menu that begins with Symbologies appears. 4. With the asterisk next to Symbologies, press SELECT. A list containing the names of the first four available symbologies appears. Next to each symbology name is either the term ON or OFF indicating its enable status. To review the entire list, press ↑ or ↓ and hold it down, or press ↑ or ↓ and let up each time. The names of the additional available symbologies appear. 5. To change the enable status of a symbology, position the cursor next to the symbology name and press SELECT. For all the symbologies except Code 93 and Code 49, a new menu appears with the name of the symbology on top and a list of parameters that you can change. With the cursor on the asterisk next to ENABLE, press → or SELECT. The symbology enable status changes between Yes and No. • • • •
To change a parameter, use ↓ to move the cursor to the asterisk next to the name of the first parameter. To change a numeric value such as the magnification factor, press → to move the cursor from the asterisk to the parameter value, then press ↑ or ↓ until the value you want appears. To change a selection such as Yes/No or Checked!/None, press → or SELECT. To return to the previous menu, press POWER.
When you exit any menu, all the displayed parameter choices are saved.
3-6
Quick Check® 600/800 Series User’s Guide