Page 11

GENERAL FEATURES

2

2 GENERAL FEATURES 2.1

DISPLAY

The MX4000 is equipped with a 20 character x 4 line backlighted LCD. The LCD is used by the Standard Application Program to display programming and error messages and data sent to the host, to view current configuration parameters and user menus. If data is longer than the available number of characters in the LCD, it is not completely displayed, however it is completely transmitted to the host computer.

2.2

KEYPAD

There are four keys, marked with the symbol Up-arrow, Down-arrow, CLR and ENT; this keypad is mainly used to configure the MX4000 locally when remote configuration via a supervisor computer is not required. To enter into keyboard programming mode, press the UP-arrow and ENT keys simultaneously. In the MX4000 Standard Application Program, these particular keys have the following functions: This key scrolls functions or programmable data UP.

EN T

CL

R

This key scrolls functions or programmable data DOWN. This key allows you to return to the function of the previous menu. This key enables the function or selects the value of the data shown on the display.

3

Page 12

MX4000

2

For further information on the use of these keys refer to the "kmx4000" file provided with the MX4000 diskette.

2.3

LEDS

The 7 colored LEDs on the MX4000 front panel are shown in the following figure: EXT EXT TX RX TX TRIG 1 TRIG 2 MDROP MDROP MAIN

RX POWER MAIN ON

Figure 1 - MX4000 LEDs

The LEDs indicate: ETX TRIG1

(yellow) - input IN1 is active (available for Custom application programs).

ETX TRIG2

(yellow) - the input IN2 is active (available for Custom application programs).

TX MDROP

(green) - transmission of data or programming running on the multidrop interface. (for RS485 Network models only).

RX MDROP

(green) - reception of data or programming running on the multidrop interface. (for RS485 Network models only).

TX MAIN

(green) - transmission of data or programming running on the main serial interface (for Standard serial interface models only).

RX MAIN

(green) - reception of data or programming running on the main serial interface (for Standard serial interface models only).

POWER ON

(red) -

4 the MX4000 is powered.

Page 13

GENERAL FEATURES

2.4

2

REMOVABLE GLANDS PANEL

To make installation and replacement easier, the MX4000 is made up of two parts: 1) the body of the device containing all electronic components and optional boards 2) the removable panel housing the glands for all external communication cables This panel houses glands for serial and peripheral interface connections and for Profibus DP models, a 9-pin D-sub bus connector.

2.5

ON/OFF SWITCH

The on/off switch can be used to turn the device off if an emergency on the system occurs. The device can also be turned off for working on the internal cabling or other particular maintenance.

2.6

HOST COMMUNICATION INTERFACES

MX4000 models are available with the following communication interface types: •

Standard serial interface:

•

Profibus DP

•

InterBus-S

Details regarding use of the interfaces are given in the following paragraphs while connections are given in chapter 3 "Installation". Alternative commercially available Fieldbus interface boards can be mounted in place of the InterBus-S board. For more details refer to your local Datalogic distributor.

5

Page 14

MX4000

2

2.6.1

Standard Serial Interface

The MX4000 Standard serial interface models are equipped with the following serial communication types for the Main serial interface: RS232 RS485 HALF-DUPLEX RS485 FULL-DUPLEX 20 mA CURRENT LOOP

2.6.2

Profibus DP

The MX4000 Profibus DP interface models allow connection to a PROFIBUS DP network as a Slave Station, communicating with a baud rate up to 3 Mbit/s (see DIN 19245 Part 1, Part 2, Part 3). For details on the implementation of this interface for MX4000, refer to the document “mx4profi” on the MX4000 diskette. For additional details concerning PROFIBUS Master station configuration or general PROFIBUS specifications and implementation issues, please refer to the official Siemens PROFIBUS documentation (DIN 19245).

2.6.3

InterBus-S

The MX4000 InterBus-S interface models are equipped with an ANYBUS InterBus module. The InterBus interface is normally used for industrial automation, normally for the control of valves, sensors and I/O units. The ANYBUS module for InterBus is a remote bus slave. This slave node can be read and written to from an InterBus master. The ANYBUS module for InterBus will not initiate communication to other nodes, it will only respond to incoming commands. The MX4000 only implements the Process Data Service for the InterBus network. The Process Data is updated every InterBus cycle. Fieldbus data is accessed through parallel DPRAM.

6

Page 15

GENERAL FEATURES

2.7

2

AUXILIARY INTERFACE

All MX4000s have an RS232 auxiliary interface which can be connected to another host computer or an external system. This interface is mainly used as a control station or for system configuration using the WinHost utility program. Diagnostics and program downloading can be performed from this interface.

2.8 2.8.1

PERIPHERAL NETWORK COMMUNICATIONS Multidrop Line

These MX4000 models are supplied with an RS485 Half-Duplex interface for peripheral device connections.

2.9

I/O

Two input signals (NPN, PNP), and three output signals (open emitter, open collector), which are not used in the Standard Application Program, are available for Custom Application Programs. For more details refer to your local Datalogic distributor.

7

Page 16

MX4000

3

3 INSTALLATION 3.1

PACKAGE CONTENTS

Verify that the MX4000 and all the parts supplied with the equipment are present and intact when opening the packaging; the list of parts includes: 1)

MX4000 multiplexer

2)

Installation manual

3)

MX4000 configuration program disk

4)

Terminal blocks, power supply connector, and multidrop line termination resistors

4

1

3

2

Figure 2 - MX4000 package contents

8

Page 17

INSTALLATION

3.2

3

GUIDE TO INSTALLATION

The Installation of the MX4000 depends on the model and type of system implemented. The following paragraphs summarize complete installations for different systems and can be used as checklists to verify all of the steps necessary for complete installation. For all systems, alternatives to software configuration through the WinHost interface utility program is achieved by using the Keyboard Mode through the keypad and the display or by using the Host Mode programming procedure, by ESC sequences via the serial interface. For more details about these programming modes refer to the documents on the configuration program disk.

3.2.1

Standard Serial Interface

1) Open and mount the MX4000 according to the information in paragraphs 3.3 and 3.4. 2) Set the main serial interface type as required using jumpers JP1-JP3 (see paragraph 3.6.1 under “Main Serial Interface Selection”, figure 3.9). 3) Provide correct system cabling (see paragraph 3.3.1, 3.5, all applicable subparagraphs under 3.6.1, paragraphs 3.7.1 and 3.8). 4) Configure the software parameters from a host computer using the Winhost interface utility program as follows: Host Int. • • •

Verify that the Interface Configuration is Serial. In the Serial Configuration select the correct Protocol type and line parameters for your application. In Options, select the data formatting desired (described in Appendix A). Device Int.

• •

Define the Baud Rate for the Multidrop line. Enable the Devices, according to address number, which must correspond to actual devices with those addresses on the network.

The installation is now complete.

9

Page 18

MX4000

3

3.2.2

Profibus DP Interface

1) Open and mount the MX4000 according to the information in paragraphs 3.3 and 3.4. 2) Set a valid Profibus DP address (see paragraph 3.6.2). 3) Provide correct system cabling (see paragraphs 3.3.1, 3.5, 3.6.2, 3.7.1 and 3.8). 4) Configure the software parameters from a host computer using the Winhost interface utility programs follows: Host Int. • • •

Verify that the Interface Configuration is Profibus-DP. Select whether to enable Profibus Flow Control or not. In Options, select the data formatting desired (described in Appendix A). Device Int.

• •

Define the Baud Rate for the Multidrop line. Enable the Devices, according to address number, which must correspond to actual devices with those addresses on the network.

The installation is now complete.

3.2.3

Interbus-S Interface

1) Open and mount the MX4000 according to the information in paragraphs 3.3 and 3.4. 2) Provide correct system cabling (see paragraphs 3.3.1, 3.5, 3.6.3, 3.7.1 and 3.8). 3) Configure the software parameters from a host computer using the Winhost interface utility programs follows: Host Int. • • • •

Verify that the Interface Configuration is Anybus-DT. Select whether to enable Anybus Flow control or not. Verify that the Anybus Areas = 6. In Options, select the data formatting desired (described in Appendix A). Device Int.

• •

Define the Baud Rate for the Multidrop line. Enable the Devices, according to address number, which must correspond to actual devices with those addresses on the network.

The installation is now complete.

10

Page 19

INSTALLATION

3.3

3

OPENING THE DEVICE

Before installing the MX4000 it is necessary to open it to perform the following operations:

•

Mount the device on a panel or a wall. (par. 3.4)

•

Provide system wiring through the glands panel. (par. 3.3.1 and relative connection paragraphs: 3.5, 3.6, 3.7)

You have to remove the top panel to perform these operations; you also have to remove the glands panel to change cable connections. Refer to the following instructions and diagram below when opening the device: 1) Turn the MX4000 off. 2) Disconnect the power supply cable. 3) Unscrew the four screws as shown below to open the device. 4) Carefully remove the cover of the decoder while paying attention to the cable that is connected to the electrical ground; this cable is long enough to allow you to lay the cover to the side while operating inside the device.

Figure 3 - Opening the MX4000

11

Page 20

MX4000

3

3.3.1

Removing and Wiring the Glands Panels

After opening the MX4000 you must remove the panel that houses the glands to connect or change the internal cabling. To connect an external cable to the MX4000, proceed as follows: 1)

Turn the MX4000 off.

2)

Disconnect the power supply cable.

3)

Open the MX4000 as described in paragraph 3.3.

4)

Using a screwdriver, unlock the screws that fix the glands panel and remove it from the MX4000 body.

5)

Unscrew the glands and pass the cables through them.

6)

Cut the insulation of the cable on the MX4000 side by a few centimeters.

7)

Strip the individual wires by a few millimeters.

8)

Loosen the terminal block connector screw of the selected position, insert the wire and tighten the screw (see Figure 4).

9)

After inserting all the wires into the terminal block connectors, pass them through the glands panel hole and insert them into the base again (see Figure 5 and Figure 6).

10) Affix the glands panel to the MX4000 body with the proper screws adjusting the cables to the proper length through each gland and then tightening the glands.

12