Page 9
MATRIX 400™ QUICK GUIDE
STEP 6 – X-PRESS™ CONFIGURATION Once Matrix 400™ has calibrated image density, you can configure it for optimal code reading relative to your application. This configuration can be performed either through the X-PRESS™ Interface or the VisiSet™ configuration program.
LOCATE READY
1. Enter the Focus function by pressing and holding the X-PRESS™ push button until the Focus LED is on.
green
2. Release the button to enter the Focus function. The Blue Diamonds™ turn on.
green
3. Select a code from your application. Position the code at the center of the FOV (equidistant from the Blue Diamonds™).
yellow
4. Exit the Focus function by pressing the X-PRESS™ push button once. The Blue Diamonds™ turn off.
yellow
LEARN
GOOD SETUP
TRIGGER FOCUS
COM TEST
STATUS red
Figure 12 – X-PRESS™ Interface: Locate Function
SETUP 5. Enter the Setup function by pressing and holding the X-PRESS™ push button until the Setup LED is on. 6. Release the button to enter the Setup function. The Setup LED will blink until the procedure is completed. The Setup procedure ends when the Image Acquisition parameters are successfully saved in the reader memory, the Setup LED will remain on continuously and Matrix 400™ emits 3 high pitched beeps. If the calibration cannot be reached after a timeout of about 5 (five) seconds Matrix 400™ will exit without saving the parameters to memory, the Setup LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep. 7. Exit the Setup function by pressing the X-PRESS™ push button once.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red
Figure 13 – X-PRESS™ Interface: Setup Function
LEARN 8. Enter the Learn function by pressing and holding the X-PRESS™ push button until the Learn LED is on. 9. Release the button to enter the Learn function. The Learn LED will blink until the procedure is completed. The Learn procedure ends when the Image Processing and Decoding parameters are successfully saved in the reader memory, the Learn LED will remain on continuously, the Green Spot is activated and Matrix 400™ emits 3 high pitched beeps. If the calibration cannot be reached after a timeout of about 3 (three) minutes Matrix 400™ will exit without saving the parameters to memory, the Learn LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep. 10. Exit the Setup function by pressing the X-PRESS™ push button once.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red
Figure 14 – X-PRESS™ Interface: Learn Function
If you have used this procedure to configure Matrix 400™ go to step 9.
9
Page 10
MATRIX 400™ QUICK GUIDE
STEP 7 – INSTALLING VISISET™ CONFIGURATION PROGRAM VisiSet™ is a Datalogic reader configuration tool providing several important advantages: •
Setup Wizard for rapid configuration and new users;
•
Defined configuration directly stored in the reader;
•
Communication protocol independent from the physical interface allowing to consider the reader as a remote object to be configured and monitored.
To install VisiSet™, turn on the PC that will be used for the configuration, running Windows 98, 2000/NT, XP or Vista, then insert the VisiSet™ CD-ROM, wait for the CD to autorun and follow the installation procedure. This configuration procedure assumes a laptop computer, running VisiSet™, is connected to the reader's auxiliary port. After installing and running the VisiSet™ software program the following window appears:
Figure 15 - VisiSet™ Opening Window
Set the communication parameters from the "Options" menu. Then select "Connect", the following window appears:
Figure 16 - VisiSet™ Main Window After Connection
10
Page 11
MATRIX 400™ QUICK GUIDE
STEP 8 – CONFIGURATION USING SETUP WIZARD The Setup Wizard option is advised for rapid configuration or for new users. It allows reader configuration in a few easy steps. 1.
Select the Setup Wizard button from the Main menu.
2.
Remove the lens cover in order to focus the reader and loosen the two Locking Knobs on the lens. 1
Adjust the Focus ring to the "Far position" and the Diaphragm ring to the "F4" number setting which is the preferred setting for installation. Place the Grade A Barcode Test Chart in front of the reader at the correct reading distance (see step 3 and the Optical Accessory Selection table in the Appendix of this Quick Reference Guide).
1
For far reading distances, the Diaphragm ring can be set to values between F2 and F4 to increase image lighting.
11
Page 12
MATRIX 400™ QUICK GUIDE
3.
Press the "Positioning" button. The reader continuously acquires images and gives visual feedback in the view image window. Select the largest code from the chart that completely fits into the view image window. Move the reader (or code) to center it. The code must be aligned across the X-axis reference line at the center of the FOV. See figure below. Press the Positioning button again to stop positioning.
3
4.
Select a Calibration Mode choice and press the "Calibrate" button. The reader flashes once acquiring the image and auto determines the best exposure and gain settings. If the code symbology is enabled by default, the code will also be decoded.
4 5
12
Page 13
MATRIX 400™ QUICK GUIDE
5.
Press the "Fine Focusing" button to activate the Fine Focusing Tool. The reader continuously acquires images and gives visual feedback on the focusing quality in the Focusing Tool window.
Rotate the Focusing ring on the lens. The Current Focus Quality Bar (green) together with the vertical optimal focus line (green) increase together until the optimal focus is reached; the vertical optimal focus line stops. Continue rotating the Focusing ring on the lens a little farther; the Current Focus Quality Bar decreases (red) see below.
13
Page 14
MATRIX 400™ QUICK GUIDE
Rotate the Focusing ring in the opposite direction. The Current Focus Quality Bar (green) increases towards the vertical optimal focus line (green) until the optimal focus is reached; the Current Focus Quality Bar touches the vertical optimal focus line (indicating the best focus).
Tighten the Locking Knobs on the lens and press the "Close" button to return to the Setup Wizard.
6.
Select a Code Setting Mode choice and press the "Code Setting" button. Using the Grade A Barcode Test Chart, this step performs image density calibration in order for Matrix 400™ to function correctly and to the fullest extent of its capabilities. The Setup Result section of the Setup Wizard window shows the code type results and the image density calibration settings.
6
14
Page 15
MATRIX 400™ QUICK GUIDE
7.
Place the application specific code in front of the reader at the same reading distance and repeat steps 3, 4, and 6.
3 4
6
Setup Result
8.
Select a Saving Options choice and press the "Save" button.
9.
Close the Setup Wizard.
If your application has been configured using the VisiSet™ Setup Wizard, your reader is ready. If necessary you can use VisiSet™ for advanced reader configuration. NOTE
15
Page 16
MATRIX 400™ QUICK GUIDE
STEP 9 – TEST MODE Use a code suitable to your application to test the reading performance of the system. 1.
Enter the Test function by pressing and holding the X-PRESS™ push button until the Test LED is on.
2.
Release the button to enter the Test function. Once entered, the Bar Graph on the five LEDs is activated and if the reader starts reading codes the BarGraph shows the Good Read Rate. In case of no read condition, only the STATUS LED is on and blinks. READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 17 – X-PRESS™ Interface: Test Function
3.
To exit the Test, press the X-PRESS™ push button once.
By default, the Test exits automatically after three minutes. NOTE
The Bar Graph has the following meaning: READY
≥ 95%
LEARN
GOOD
≥ 75%
SETUP
TRIGGER
≥ 60%
FOCUS
COM
≥ 40%
TEST
STATUS
≥ 20%
16