Page 13

INSTALLATION

2

2

INSTALLATION

2.1

PACKAGE CONTENTS

Verify that the C-BOX 150 and all the parts supplied with the equipment are present and intact when opening the packaging; the list of parts includes: 1)

C-BOX 150

2)

Installation manual

3)

C-BOX 150 configuration CD-ROM

4)

2 mounting screws

5)

Grounding cable

2 1

3

4 5

Figure 1 - C-BOX 150 Package Contents

3

Page 14

C-BOX 150

2

2.2

OPENING THE DEVICE

To install the C-BOX 150 or during normal maintenance, it is necessary to open it by unscrewing the four cover screws:

The C-BOX 150 must be disconnected from the power supply during this operation. CAUTION

Figure 2 - Opening the C-BOX 150

It is possible to perform the following operations: •

Proceed with the cable connections (see paragraph 2.4.2).

•

Mount the C-BOX 150 to a wall or panel.

4

Page 15

INSTALLATION

2.3

2

MECHANICAL INSTALLATION

12 [0.47]

The diagram below gives the overall dimensions of the C-BOX 150 and may be used for its installation.

5 [0.20]

89.8 [3.54]

40 [1.57]

150 [5.91]

114.5 [4.51]

161 [6.34]

90 [3.54]

5 [0.20]

12 [0.47]

66 [2.60]

Ø3.5 [Ø0.14] mm inch

Figure 3 - Overall Dimensions

5

Page 16

C-BOX 150

2

C-BOX 150 can be installed to operate in different positions. The two screw holes inside the housing of the C-BOX 150 are for mechanical fixture (Figure 4). To mount the C-BOX 150: 1)

Open the C-BOX 150 by unscrewing the 4 cover screws. If necessary, using the two mounting holes inside the device as a pattern, mark the panel with an appropriate object and then drill two holes in the panel.

2)

Align the C-BOX 150 and insert two screws and screw them into the panel until tight (see Figure 4).

Figure 4 - Mounting C-BOX 150

6

Page 17

INSTALLATION

2.4

2

ELECTRICAL CONNECTIONS AND HARDWARE SETUP

The following figure shows the typical layout.

C-BOX 150

SCANNER

SCANNER AUXILIARY INTERFACE

SCANNER CONFIGURATION PC SYSTEM CABLES

Figure 5 – System Layout

A PC can be connected to the C-BOX 150 (and consequently to the scanner auxiliary interface) through the internal 9-pin connector. This allows both scanner configuration and monitoring of the data transmitted by the scanner (see the scanner Reference Manual for more details). The scanner auxiliary interface signals are also available on the internal spring clamp connectors. The scanner configuration and data monitoring are possible once the warning LED is turned OFF. The internal 9-pin connector can also be used to configure the C-BOX 150 with a PC, through the WinHost utility. After making system cabling and switch settings, connect the scanner to the 25-pin connector on the left side of the C-BOX 150 housing. Switch ON the C-BOX 150 power switch (see Figure 6). By default, after power on, an automatic connection procedure takes place between the C-BOX 150 and the scanner. During this phase, requiring a few seconds, the warning LED is turned ON. Once the procedure had been completed successfully, the warning LED is turned OFF. To disable this automatic connection procedure, refer to WinHost Help Online.

7

Page 18

C-BOX 150

2

After system functioning has been verified, close the C-BOX 150 using the 4 cover screws making sure the rubber seal is fitted correctly between the parts of the housing.

NOTE

2.4.1

It is recommended to use the dedicated grounding cable (included in the package) to improve the C-BOX 150 noise immunity. Simply connect its open end to the scanner cable’s fixing screw and the other end to the ground, as shown in the figure below.

Power Supply

Power is supplied to the C-BOX 150 through the pins provided on the spring clamp connector. The power switch (see Figure 6) switches the power supply ON or OFF for both the C-BOX 150 and the connected scanner.

ON

OFF Figure 6 - Power Switch ON/OFF Positions

8

Page 19

INSTALLATION

2

USER INTERFACE

C-BOX 1

VS

2

GND

V+ (10 - 30 Vdc) GND

Figure 7 - Power Supply Connections

The power supply must be between 10 and 30 Vdc only.

Pin 1 is also electrically connected to pins 3 and 5, just as pin 2 is electrically connected to pins 4 and 6. This is useful for external trigger/inputs connections.

NOTE

The C-BOX 150 is protected against polarity reversal. A dedicated red LED signals this condition until correct polarity is restored. NOTE

2.4.2

System Wiring

The connection and wiring procedure for C-BOX 150 is described as follows: 1)

Open the C-BOX 150 as described in paragraph 2.2.

2)

Verify that the C-BOX 150 power switch is off (see Figure 6).

3)

Unscrew the compression connectors and pass all the system cables through them into the C-BOX 150 housing.

4)

To connect the power and input/output signals: •

Prepare the individual wires of the system cables by stripping the insulation back approximately 1 cm.

9

Page 20

C-BOX 150

2 •

Using a device such as a screwdriver, push down on the lever directly next to the clamp (see Figure 8).

•

Insert the wire into the clamp and release the lever.

The wire will now be held in the spring clamp.

Figure 8 - System Cable Connections

The wiring used can be solid or stranded but must meet the following specifications. All positions:

10

24 - 16 AWG

0.2 - 1.5 mm²