Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 10 of 15

AdaptaScan Bar Code Readers User Manual (Cat. No. 2755-SN3, -SN5, -SN8): 6 - Troubleshooting and Maintenance

Pages 66 to 72 of 96, from AdaptaScan Bar Code Readers User Manual (Cat. No. 2755-SN3, -SN5, -SN8)

The original document is the PDF (2.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Accu-Sort 2755-SN3, Accu-Sort 2755-SN5, Allen-Bradley 2755-SN5, Accu-Sort 2755-SN8 and Allen-Bradley 2755-SN8

Page 66

Chapter

6

Troubleshooting and Maintenance Chapter Objectives

This chapter describes how to isolate and correct the most common operating problems and perform routine maintenance tasks. Section

Page

Using the Troubleshooting Chart

6–1

Technical Support

6–1

Equipment Required

6–1

Troubleshooting Chart

6–2

LED Indicators

6–3

Cleaning the Scan Window

6–5

Replacing the Scan Window

6–6

Technical Support

If you should require assistance or need additional information on operating the AdaptaScan Reader, Configuration Software, or accessories, contact your local Allen-Bradley support office or Allen-Bradley Technical Support at (216) 646-6800.

Equipment Required

Other than verifying that the correct power source is connected to the wiring base (use a voltmeter), no electronic diagnostic equipment is required for troubleshooting.

Using the Troubleshooting Chart

Table 6.A is the AdaptaScan troubleshooting chart. This chart lists the most common operating problems, the probable causes, and steps to correct the problem.

Page 67

6–2

Troubleshooting and Maintenance

Troubleshooting Chart Table 6.A Troubleshooting Chart Problem

Probable Cause(s)

Corrective Action(s)

Reader does not power up. No LED indicators 1. Improper connection to power source. illuminate. 2. Power source not providing 12 to 28 VDC.

1. Check power source.

No communications with personal computer.

1. Communications fault.

1. Check cable connection.

2. No power supply connection at DeviceNet connector

2. Power must be supplied. See Chapter 3.

3. Computer fault.

2. Check for green LED in wiring base. 3. Check the wiring base connector for bent pins.

3. Refer to computer user manual.

4. Broken or bent pins on Configuration Port of Reader

4. Check Configuration Port of Reader for damaged pins. If pins are damaged, return Reader.

No laser beam emits from scan window.

1. No laser light trigger signal.

1. Verify Reader Laser Light trigger source for proper configuration and operation.

Unable to read a label.

1. A label has not been defined.

1. Define a label and symbol.

2. Decoding functions not configured correctly.

2. Check decoder configuration to verify that parameters are set appropriately for your application. After making changes, SEND the configuration and try again.

3. Label(s) out of specification.

3. Use good quality labels that are within specifications.

4. Scan line not on the label.

4. Reposition the scan line.

5. Reader is at incorrect angle or distance from bar code symbols.

5. Use autofocus to adjust the read range. Use the ON SYMBOL LED to verify that the Reader is scanning a possible bar code symbol.

No communications with host.

No communications with network.

1. Baud rates not set properly.

1. Verify that Reader and host are set at the same Baud rate.

2. Loose or incorrect wiring.

2. Verify cable connections using connection diagrams in Chapter 3.

3. Broken or bent pins on 24-pin Reader connector.

3. Check Reader connector for broken or bent pins.

4. Reader DeviceNet address or DH485 node and maximum node are not set properly.

4. Verify DeviceNet Address or DH485 node number settings.

5. Termination switch not set properly.

5. Check position of termination switch in wiring base. Set to correct position.

1. Communications fault.

1. Check status of NETWORK LED, refer to section “LED Indicators”.

2. Loose or incorrect wiring.

2. Verify cable connections using connection diagrams in Chapter 3.

3. Broken or bent pins on 24-pin Reader connector.

3. Check Reader connector for broken or bent pins.

4. Reader DeviceNet address or DH485 node and maximum node are not set properly.

4. Verify DeviceNet Address or DH485 node number settings.

5. Termination switch not set properly.

5. Check position of termination switch in wiring base. Set to correct position.

Page 68

Troubleshooting and Maintenance

LED Indicators

Use the 7 LED indicators to isolate operating problems. Figure 6.1 LED Indicators

On Symbol Condition

Indication

Yellow

Bar code symbol is being scanned and decoded.

Flashing

Bar code symbol is being read at less than 100% rate.

Off

Not reading bar code symbol.

Trigger/Read Condition

Indication

Yellow

Decoder trigger is active.

Green

Valid bar code symbol read.

Off

No trigger or valid bar code symbol read.

I/O 1 and I/O 2 Condition

Indication

Yellow

Corresponding input or output is in an On (closed contact) state.

Off

Corresponding input or output is in Off (open contact) state.

6–3

Page 69

6–4

Troubleshooting and Maintenance

Network Condition

Indication

Green

Normal DeviceNet operating state.

Flashing Green

Communication link established but no data transfer.

Red

DeviceNet communication fault detected.

Flashing Red

One or more DeviceNet devices are not responding. Reader may not be able to perform all configured functions. No DeviceNet communications established.

Off

Laser On Condition

Indication

Yellow

The scan beam is On (active).

Off

The scan beam is Off (inactive).

Module Condition Green

Indication

Flashing Green

Normal operating condition. Power is applied to the Reader and no faults have been detected. Power-up initialization or incorrect configuration.

Red

Processor fault (unrecoverable).

Flashing Red

Minor fault (recoverable). Occurs when downloading firmware.

Off

No power applied to Reader.

Page 70

Troubleshooting and Maintenance

6–5

Cleaning Scan Window ATTENTION: Other than cleaning the scan window, no user maintenance of the Reader is required. Do not open the enclosure!

!

For optimum performance the scan window should be clean. ATTENTION: Only use reagent grade alcohol to clean the window. Do not use organic solvents. Do not use abrasive materials, such as disposable paper wipes to clean the scan window. Disposable wipes usually contain glass fibers which will scratch and cloud the window.

!

To clean the scan window: 1. Turn the Reader off.

!

ATTENTION: Do not attempt to clean the window while the Reader is turned on. Although momentary exposure to the laser light is not harmful, precautions should be taken to avoid looking into the beam.

2. Verify that the Module LED on the Reader is off. 3. Dust off the scan window and adjacent areas with clean air. 4. Clean the window using a reagent grade alcohol or a similar cleaner for precision optics and lint free lens cleaning paper. To avoid smearing and fingerprints, wipe the window in one direction and then discard the tissue. Repeat as required. Note: Clean the window in sections if the entire surface contains contaminants. 5. Apply power to the Reader. The Module LED on the Reader should be on.

Page 71

6–6

Troubleshooting and Maintenance

Replacing Scan Window

The scan window fits into an opening behind the front of the Reader. A bezel and gasket create a NEMA Type 4 seal. The table below lists numbers for ordering a glass or plastic window kit. Do not substitute other material for a damaged window. The windows have an optical coating necessary to Reader performance. Window Kits

Description

2755-NW45

Replacement Glass Window Kit Includes bezel/window assembly with anti-reflective, optical glass window.

2755-NW44

Replacement Plastic Window Kit Includes bezel/window assembly with hard coated, anti-reflective, optical quality, plastic window.

Figure 6.2 Scan Window Removal/Replacement

#6-32 Bezel Screws (6)

Bezel/Window Assembly

ATTENTION: When servicing the Reader, do not touch components inside window. Do not touch the internal mirrors! Fingerprints diffuse the laser beam. The internal mirrors cannot be serviced in the field.

!

To remove the scan window: 1. Shut off power to the Reader.

!

ATTENTION: Disconnect power to the Reader before attempting to replace the scan window to avoid potential eye damage from the laser beam.

2. Clean area around window bezel.

Page 72

Troubleshooting and Maintenance

6–7

3. Remove the six #6-32 screws from the bezel/window assembly. 4. Carefully pull bezel/window assembly away from the front of the Reader. 5. Discard the old bezel/window assembly. To install the replacement scan window: 1. Do not smudge the window when removing the bezel/window assembly from package. If necessary, refer to page 6–5 for cleaning instructions. 2. Place the bezel/window assembly against the front of the Reader (with wider side at top of Reader), aligning the 6 screw holes with the 6 holes on the Reader body. Top of Scan Window

3. Tighten the six #6-32 screws to a torque of 10 inch–pounds (1.1 NSm) in a side-to-side sequence providing even compression of the sealing gasket. 4. Refer earlier chapters in this manual to install, align and verify operation of the Reader before resuming full service of Reader. Observe all warnings and cautions.