Page 55

Loading Dose

A ( Program Loading Dose Wsnd A&w

1 2

Rate vrsl

“;FmT

MO

& w:c6

0.0

0.0 lYJ:W mUhr

9. Enter Volume To Be Infused (VTBI) from 0.1 to 9999 mL using the keypad. After a Rate and VTBI value are entered, the Duration is automatically computed. Manually changing the Duration field will automatically recompute the Rate. To understand the Automatic Calculation feature, see Section 4.7 Exaniples of Automatic

Note:

Calculation,

10. After the values for the loading dose Step-l are entered, use [SELECT] to highlight the maintenance dose Step-2. Note: The loading dose therapy has two steps. The first is the loading dose followed by the maintenance dose. These words will appear on the Main Screen as each step is delivered.

6

I

A I Program Loading Dose WUdAC4iV8

Rate mm 1

2

VTBI ml wo 55.0 m

0.0

Duratim tuxlin M):ffi w:oo

Al Program Loadzdse 1

Rate miAw 1 2 v-ml mL

II. Enter the Rate using the keypad.

12. Use [SELECT] to highlight the next programming field.

“h% C&X w:w

System Operating Manual

5- 11

430-95226-001 (Rev. 5101)

Page 56

Therapies (For Advanced Users*)

A I Program Loading Dose WandMive

13. Enter the desired VTBI using the keypad. Note: After the desired Rate and VTBI values are selected, the Duration is automatically updated.

14. Press A [Program Options] to enable setting a Delayed Start or to enable the Nurse Callback.option.

/Program A1

Optioiow

Nurse Callback enables an alarm when a change in the delivery on a line occurs such as completing Step- 1 and starting Step-2. Or on line B, a piggyback delivery is completed and line A is starting its delivery.

Note:

Delay Start for lO0 hnmin

15. Enter a delay time (less than or equal to 24 hours) using the keypad. Delay Start for OOa hcmin

Note:

Not available if a delivery is taking place on the line.

16. After selecting a desired value, use [SELECT] to highlight the callback feature.

430-95226-001 (Rev. 5/01)

5- 12

Plum A+ Infusion Pump

Page 57

Loadina Dose

A)

Program Options wllnd-

17. Select using A [Yes/No]. Callback changes from NO to YES.

Delay Start for OD:O5 hr:min

Callback m

Note: Nurse Callback cannot be changed while a delivery is taking place on the line.

18. Press A [Enter] to accept the changes. A [Cancel/Back] returns the fields to their original value and displays the previous screen. Note: A 1

Program Loading Dose WM(I*

CAUTION: VERIFY DOSE, RATE, VThI, AND DURATION BEFORE PRESSING [START].

19. Press [START] to begin delivery.

A 1 DELAYED 1 STOPPED 1 B

The Main Delivery Screen appears.

Mhprrr

Note: The words “DELAYED” and “Loading Dose” appear in the upper-left comer. These indicate the options which have been selected.

500 m!% “’

0.0 “$” 0.0

I A I PUMPING

STOPPED I B I

-Do-

500 mL/hr Iwe 0.0 0.5

After the programmed delay period has expired, the Plum A+ starts “Pumping”. This is indicated where “Delayed” previously was displayed.

“$” 0.0

System Operating Manual

430-95226-001 (Rev. 5/01)

Page 58

Therapies (For Advanced Users*) After the first step has completed, the Main Delivery Screen changes to the second step. A flashing “Callback Line A” message and the audio alarm sound indicates activation of that option.

Note:

20. The user should acknowledge the Callback message and stop the audio alarm by pressing [SILENCE].

Upon completion of delivery, the screen shows a flashing “Line A VTBI complete” message and the audible alarm sounds. Note: The KVO rate will change to the rate of 1.0 mL/hr or less depending on delivery rate and “Pumping” mode changes to “KVO”. If Continue Rate option was selected, pumping continues at original rate.

21. Press [SILENCE] to stop alarm sound. Note: Warning message continues to flash and audio alarm will return after a two minute period of silence, unless delivery is stopped or VTBI of last line is changed.

430-95226-001 (Rev. 5101)

5- 14

Plum A+ Infusion Pump

Page 59

Multistep Programming

5.4 Multistep Programming MULTISTEP- Allows a sequential program to deliver up to 10 steps: fluid volumes and delivery rates may be Rrogrammed for each step. The program may be entered based on Rate +ndVolume or Volume and Time. If Dose Calc is used, the delivery steps’ are in the same unit of measure, over the same period of time, from the same container. Clear ALL settings when changing patients or multiple programming parameters. Use A [C/ear Program]when cianging ONLY Line A or B.

1. Select A [A] at the Main Delivery screen. Al

PROGRAM W&_ Rate m mL/hr

VTBI

0.0 mL

The Line A Programming screen appears.

2. Press A [Therapy].

Duration 0O:OO hr:min

Al

PROGRAM whnd&WO

The Select Therapy screen appears.

Loading Dose Multistep

Al

I

PROGRAM WMdAC%Va Dose Calculation

Use [SELECT] to scroll down to Multistep. Press A [Chooselto continue.

Loading Dose

System Operating Manual

430-95226-001 (Rev. 5/01)

Page 60

Therapies (For Advanced Users*) The Program Multistep screen appears.

Program Multistep

Note: Screen defaults to the mL/hr option or the previously selected option.

AI

5. Use [SELECT] to highlight the desired setting. This example is using mcg/hr.

Program Multistep

6. Press A [Choose]to go to the second page of the Program Multistep screen.

AI

The Container Concentration Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng) otherwise this screen is skipped.

Program Multistep Drug Cone in Container

Note: Default choice for possible gram-based Container Concentration Unit is mg or the previously selected option.

Back

7. Press A [Choose]to continue, or use [SELECT] to highlight another option.

Ax.A

AI

1 s

8. When desired field is highlighted, press A [Choose]. 9. The default field is the concentration field (Cone). Enter value using keypad.

Program Multistep WMd_ z

?%?

O.oW IE oo:W zt

OOZW

2 z!ir

0.0

The user must enter the concentration drug amount and diluent volume before any other entries. Note:

::: 0.0 0.0 0.0 0.0 0.0

430-95226-001 (Rev. 5/01)

Plum A+ Infusion Pump

Page 61

Multistep Programming

A)

Program Multistep WMd*aMI cow

: 3 i

50.0 mg m rnL mgmr hcmin 00: 00: 00 00 mL 0.0 0.0 0.0 0.0 mLRn

_.

I?303 oo:w

0.0 0.0

0.0 0.0

%z olxa 0.030

10. Use [SELECT] to highlight the next programming field (Diluent-Amount). 11. Enter value using keypad.

12. Use [SELECT] to highlight the next field. 13. Enter Dose using keypad. Note: The Rate value is automatically computed. To understand the Automatic Calculation feature, see Section 4.7 Exa~npZes ofAutomatic

Calculatfon,

AI

Program Multistep v&a MM ceix 50.0

AI caw mg so0 rnL

Program Multistep WMd*ruvs 50.0 mg

500 mL

System Operating Manual

14. Use [SELECT] to highlight the Duration or VTBI field. The VTBI value is automatically computed when the Duration value is entered. If VTBI is entered, the Duration value is automatically computed.

Note:

15. Use [SELECT] to highlight the Rate field. If Rate is changed, Dose and Duration (or VTBI) are automatically re-computed, depending on the prog ramming order.

Note:

5- 17

430-95226-001 (Rev. 5/01)

6.

Page 62

Therapies (For Advanced Users*)

16. Use [SELECT] to scroll past the Rate. The cursor will drop down to Step-2 Dose field. 17. Enter the set of desired values for Step-2 of the program. Continue process for each additionaf.Step.

Note: When Step-4 is programmed;:using [SELECT] to advance will highlight another item on this screen.

r

18. When Step-4 is programmed, press A [More Steps] to program up to 10 Steps. Note:

Using [SELECT] will not advance you to the additional steps.

19. To set up Steps 5 through 10, enter desired values as for Steps 1 through 4. 20. To see the previous steps, press A [Prev Steps].

AI

Program Options wand*cvn

Delay Start for DO0 hr:min

21. To program a Delay Start Time or set up a Nurse Callback for this program, press A [Program Options]. 22. Enter a value using the keypad. You can choose Drug Name, Callback, or Delay Start as previously shown.

23. Press A [Enter]to accept the changes.

430-95226-001 (Rev. 5/01)

5- 18

Plum A+ Infusion Pump

Page 63

Multistep Programming

A 1 PUMPING ( STOPPED 1 B

CAUTION:

VERIFY DOSE, RATE, VTBI, PRESSING [START].

AND DURATION BEFORE

%Pl 2.0

,“,a:;, 0.0

24. Press [START] to begin the delivery. The Main Delivery

0.5

“2:” 0.0 m, mr

Screen displays Step-l in the upper-left corner. ‘_

Dose

Note: In units other than mL/hr, the Dose rate and units also will be displayed.

A 1 PUMPING 1 STOPPED 1 B 8-3 2.5 mzehr 0.0

37.5 v”,:“’ 0.0

Example of Multistep Screen Pumeing on Step-3.

25. To change a program parameter while pumping, press A i-A%

250 Dose -1

AI Program

ME2 1

Note: The **” in a Step number field indicates the step is delivering or “d” indicates each completed step.

26. Enter a value usi.ng the keypad. Note: The Concentration drug amount or Diluent volume cannot be changed while a delivery is in progress.

AI cone

Program Multistep WMd*mve 50.0 mg

5X mL

System Operating Manual

If Dose value is changed, the Duration and Rate will automatically change. Note: Only the current delivering step and future steps can be changed.

5- 19

430-95226-001 (Rev. 5/01)

Page 64

Therapies (For Advanced Users*) CAUTION: VERIFY DOSE, RATE, VTBI, AND DURATION BEFORE PRESSING [START].

27. Press [START] to begin infusion with the changed program and return to the Main Deljvery Screen. ‘. Note:

The Values are now updated on the Main Delivery Screen.

The pump reverts to KVO when last Step completes. The last step can be restarted by entering a new VTBI value.

430-95226-001 (Rev. 5/01)

5- 20

Plum A+ Infusion Pump