The text of a manual · part 10 of 12

Plum A+ Infusion Pump System Operating Manual: Alarms & Troubleshooting

Pages 65 to 68 of 85, from Plum A+ Infusion Pump System Operating Manual

The original document is the PDF (3.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the Abbott 11972-04-01

Page 65

SECTION 6

Alarms and Troubleshooting

6.1 Warning Messages

I turn off Lockout Enabled Battery Service Warnf ng

Warning: Charger Service Warning: Low Battery while a delivery is in progress. Pressinganykeywhfledelfverfng (exceptSTOP)whenLockoutswftch is enabled. The battery capacity is greater than 200 mAjw. The battery voltage is less than the depleted threshold and charge level is higher than the low charge threshold. The charging circuitry is not behaving as expected (charge voltage is not changing) and can reduce battery life. Pump not plugged in to AC power.

PlaceLockoutswitchfndfsabledposftfon[see sectfon4.4RearCaseControls),thenreturnto enabled position. Replace battery.

Silence key is pressed.

Plug into ACpower, only about 15 minutes of battery life remains.

Note: Other error and warning messages will occur when appropriate for the current selection context. For instance, when a change to a required parameter would result in clearing some delivery settings, the warning message will appear when that field is selected. When an input error has just been made, the three quick beeps of an Invalid Key Warning will sound and an appropriate warning messa e will be shown to explain the %y another message upon the next problem. All such messages will be cleared or replaced valid input key press.

6- 1

430-95226-001 (Rev. 5/01)

Page 66

Alarms and Troubleshooting

6.2 Response to Alarms 1.

Disable Lockout by confinning back switch is in the down position.

Press [SILENCE] key. 3. Identify/Observe Alarm condition. 2.

4.

Correct Alarm condition (see chart below).

5.

Press [START] to resume infusion.

Note: Alarm messages begin with an alphanumeric code for tracking purposes only. If troubleshooting does not correct the problem, record the code number and contact technical support. GENERAL ALARMS

El 00 El01 El 02 El 03

Unrecognized cassette No Action Alarm Infuser Idle 2 Mf nutes Memory Failure

E230

Prox Air, Backprime

El 80, E181, El86

Distal Occlusion

Attempting to use unrecognized cassette Pump programming incomplete. Pump is ON without operator input. The pump did not accept the previous programmed data. When the total proximal air detected exceeds the limit. Distal line kinked. Nsfte clotted or positional, pressure limit set too low.

Turn off or insert recognfzed cassette. Press [STAwrl or continue programming. Beginprogramming or turn pump OFF. Reenter all programmed data. Check for clamps, or empty containers on lineA or B. Correct as necessary. Backprime into line B tubing or syringe Check distal tubing, check IV site, reset pressure limit. Note: See Section 8.8 for detailed instructions for avoiding a bolus following a distal occlusion.

El87 El 83, El 85

Neg Distal Occlusion Prox Occl At startup

Pump too high above patient or defective set. Proximal line obstructed, lineA (or B) container disconnected or obstructed during nondelivery (startup).

Lower pump or replace set. Examine line A (or B) for kinks or closed clamps. Open and close door, then immediately open clamp. Continue programming. Replace set if problem persists. Note: If clamps are not opened prior to completion of cassette test, peak proximal occlusion may recur.

E233, E234

Distal Air

E250

Door open while pumping

430-95226-001 (Rev. 5/01)

When air is detected at the distal sensor and exceeds the limit. Pump door opened while operating.

6- 2

Remove and reprime cassette using standard technique. Close door with cassette inserted.

Plum A+ Infusion Pump

Page 67

Resnonse to Alarms set, close door. Backprime into line B tubing or syringe if appropffate. If alarm recurs on retest, replace set. If alarm repeats, replace pump. Plug into AC power.

Failure distal occlusion or air in cassette detected at start up.

E252

Depleted Battery

E253

Lockout Violation

E254

Lockout Enabled

Too much time on battery power. Unlock lockout switch. The use of the [STOP) key dur f ng . &lfvery,oropenfngthedoor,whfle lockout is enabled. PressfngofanykeyexceptlSTOP] Unlock lockout switch. durfngdelfverywhfleanyalarmfs active and lockout is enabled.

LINE A ALARMS

Callback Line A

El61 El 84, El 90, El91 E232

Alarm programmed by user. Change in delivery for Loading Dose to Maintenance Step orfor any step in ‘Multistep” therapy provided the step is not the last SteD.

a Prox Af r A, Backprime

System Operating Manual

Program new VTBI.

Proximal line obstructed, lineA

Examine line A for kinks or closed clamps.

When proximal af r detected exceeds the limit for the line.

Check for clamps, or empty container. Correct as necessary. Backprime into line B tubing or syringe

430-95226-001 (Rev. 5/01)

Page 68

Alarms and Troubleshooting LINE B ALARMS

El60 El 82,

El 88, El89

Line B VTBI complete Prox &cl B

Alarm programmed by user. Change in delivery for piggyback delivery, Loading Dose to Maintenance Step, or for any step in ‘Multistep” therapy provided the step is not the last step. VTBI complete on line B.

Press [SILENCE).

Proximal line obstructed, 1fneB container disconnected or obstructed.

Examine line B for kinks or closed clamps.

Program new VTBI.

I Note: Occlusion caused by empty container, backprime into line B tubing or syringe.

E23 1

13a h-ox Air B, Backprime

L

When proximal air detected exceeds the limit for the line.

Check for clamps, or empty container. Correct as necessary. Backprime into line B tubing or 1 syringe.

MALFUNCTIONS malfunction repeats, replace

430-95226-001 (Rev. 5/01)

Plum A+ Infusion Pump