Page 69

Maintenance and Storage

SECTION 7

7.1 Cleaning, Maintenance, and Storage

‘.

The cleaning, maintenance, and storage of the Plum A+ are described in this section.

CLEANING AND SANITIZING For proper maintenance of the Plum A+, observe the following cleaning and sanitizing guidelines.

Establish a routine weekly schedule for cleaning the Plum A+. To clean the Plum A+, proceed as follows: Turn the Plum A+ off with the [ON/OFF] switch, then disconnect the Plum A+ from mains power. The exposed surfaces of the Plum A+ may be cleaned with a lint-free cloth dampened by one of the recommended cleaning solutions in the following list or mild, nonabrasive soapy water.

Super Edfsonite@

S. M. Edison Co.

Per manufacturer’s recommendation

Vesphene II @ se

Calgon Vestal Laboratories

Per manufacturer’s recommendation

Manu-Klenz@

Calgon Vestal Laboratories

Per manufacturer’s recommendation

Formula C TM

Dfversey Corporation

Per manufacturer’s recommendation

CoverageTMHBV

Sterfs Corporation, A division of Calgon Vestal Laboratories

Per manufacturer’s recommendation

Sporicidfn@

Sporicidin International

Per manufacturer’s recommendation

Household bleach

Various

Per hospital procedures: do not exceed one part bleach in ten parts water

On a routine basis, clean all of the elements behind the cassette door using cotton-tipped swabs saturated with cleaning solution. The cassette door may be unlatched from the door handle to facilitate cleaning. To unlatch the cassette door from its handle, tilt the device back, open the cassette door, then push the door release tab to open the door fully.

7- 1

430-95226-001 (Rev. 5/01)

Page 70

Maintenance and Storage CAUTION: To AVOID MECRANICXL OR ELECTRONIC DAMAGE, DO NOT IMMERSE FLUIDS OR CLEANING SOLUTIONS.

THE PLUM A+ IN ANY

CAUTION: CERTAIN CLEANING AND MTIZING SOLUTIONS M AY SLOWLY DEGRADE COM PONENTS MADE FROM SOME PLASTIC MATERIALS. USING ABRASIVE CLEANERS OR CLEANING SOLUTIONS NOT RECOMMENDED BY ABBOTT LABORATORIES MAY RESULT IN PRODUCT DAMAGE. Do NOT USE COMPOUNDS CONTAINING COMBINATIONS OF ISOPROPYL ALCOHOL AND CAUTION: N EVER USE SHARP ANY PART OF THE PLUM A+.

CAUTION:

OBJECTS SUCH

DIMETWL BEti? AMMONIUM CHLORIDE.

AS FINGER N A I L S, PAPER

CL&S, OR NE EDL ES

TO

CLEAN

Do NOT STERILIZE BY HEAT, STEAM, ETHYLENE OXIDE (M‘O), OR RADIATION.

CAUTION: To AVOID PUMP DAMAGE, CLBANING SOLUTIONS SHOULD ONLY BE USED As DIRECTED. TI3B DISINFECTING PROPERTIES OF CLEANING SOLUTIONS VARY: CONSULT THE .$4ANUFAcTuRBR FOR SPECIFIC INFORMATION.

BATTERY MAINTENANCE CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITa THE BATTERY REMOVED. USE OF A PROPERLY MAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION. CAUTION: IF THE LOW-BATTERY ALARM SOUNDS. CONNECT TRE PLUM A+ TO b%AINS POWER IMMEDIATELY. The Plum A+ is battery powered for emergency backup and temporary portable operation. A fully charged, new battery provides six hours of operation at 125 ml/hr or 500 ml total volume delivered, whichever occurs first. Note: For optimum battery life, the Plum A+ should be operated on battery power for six continuous hours at least once every six months and then charged for a minimum of six hours. The battery charges whenever the Plum A+ is connected to mains power. If the Plum A+ is switched off, recharge takes approximately six hours. Recharge takes longer if the Plum A+ is turned on. As a general rule, the more often the battery is partially discharged and recharged, the sooner it will need to be replaced. Consult a qualified biomedical technician for battery replacement if necessary. To maintain maxImum battery charge and to prolong battery life, connect the Plum A+ to mains power whenever possible.

STORAGE To prolong the life of the Plum A+, observe the following storage precautions: l

Store the Plum A+ away from excessive heat, cold, and humidity l

Store the Plum A+ connected to mains power l

Switch the Plum A+ off using the [ON/OFF] key

SERVICE All servicing or adjustments to the Plum A+ should be referred to qualified technical personnel. A technical service manual may be ordered from the local Abbott Laboratories sales office.

430-95226-001 (Rev. 5/01)

7- 2

Plum A+ Infusion Pump

Page 71

Precautions

SECTION 8

The Plum A+ has been designed and manufactured to be safe, reliable, and easy to use. This section details precautions and possible hazards.

For safe operation of the Plum A+, observe the following precautions and hazards.

8.1 Artifacts Nonhazardous, low-level electrical potentials are commonly observed when fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment such as ECG, EMG. and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infusion device instead of some other source in the environment, set the Infusion device so that it is temporarily not delivering fluid. Disappearance of the abnormality indicates that it was probably caused by the electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring equipment system documentation for setup and maintenance instructions. The Plum A+ system is designed to operate normally in the presence of most encountered electromagnetic interference (EMI) conditions. In the event of extreme levels of Interference, such as encountered next to an electrosurgical generator, it is possible that the normal operation of a sensor or microcomputer might be disrupted. Even in this event, the outcome would likely be a false alarm or detected system malfunction and would not result in a hazard to patient or operator. Use of radio frequency emitting devices such as cellular telephones and a-way radios in close proximity of this device may affect its operation.

8.2 Healthcare Professionals and Patient Related In vitro studies have suggested that packed red blood cells with unusually high hematocrit be diluted with blood-compatible fluids, such as 0.9% sodium chloride injection, to decrease hemolysis and increase flow rate. Setting the primary rate greater than the secondary rate will result in a more rapid infusion of any residual secondary drug remaining in the line and the cassette.

Consult drug labeling to co&ii drug compatibility, concentration, delivery rates, and volumes are suitable for secondary, concurrent and piggyback delivery modes.

8- 1 all

430-95226-001 (Rev. 5101)

E:

Page 72

Precautions Connecting more than one pump Into one patient line may significantly affect the infusion rate of at least one of the pumps. Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement. Before opening the door, close clamp on the primary line or remove the sebondary container from the secondary port to prevent mixing of primary and secondary fluids. Although unlikely, failure of certain robust mechanical components such as the anti-free flow mechanism or valve control springs could cause fluid delivery limited to the contents of the fluid container. Single fault failure of certain electronic/motor control components would result in no more than 5 mL of unexpected fluid delivery. A small amount of fluid is expelled from the set (less than 0.05 ml) each time the door is opened or closed with a set installed. If potent drugs are being used, take appropriate action to guard against overmedication of the patient. Before disconnecting a syringe from the cassette, pull up the plunger slightly to avoid spilling the fluid. For rigid containers (for example, vials), close the upper slide clamp, open the cassette door, then remove and invert the cassette (ports down). Air bubbles may form distal to the cassette as result of normal out-gassing of dissolved air in the fluid. This may occur if chilled solution is in use, if the pump is mounted significantly above the patient, or when using certain fluids known to routinely outgas. In these cases, an air eliminating filter may be used. Repeated opening and closing of the door may defeat the proximal air-in-line alarm and’may cause a distal air-in-line alarm, requiring repriming. The screen displays the VTBI (volume to be infused) in integers when value is above 99.9. Any fraction of a milliliter delivered is not displayed, but is retained in memory.

8.3 Concurrent Flow DELIVERY RATE

When delivering short half-life critical drugs (See Critical Drugs, this section) using the Plum A+ in the Concurrent mode, the following delivery rate guidelines should be observed: l l l

If the critical drug (with half-life less than 6 minutes) is to be infused at less than 2.0 mL/hr, the other infusion should be no faster than 5 times the critical drug’s rate. Dopamine, for example, delivered at 1.5 mL/br should not be accompanied by an infusion programmed any faster than 7.5 mL/hr. If the critical drug (with half-life less than 6 minutes) is to be infused at 2.0 to 5.0 mL/hr the other infusion should be no faster than ten times the critical drug’s rate. Dopamine, for example, delivered at 3.5 ml/hr should not be accompanied by an infusion programmed any faster than 35 mL/hr. If the critical drug (with half-life less than 6 minutes) is to be infused at 5.1 mL/hr or greater, the other infusion can be programmed at any desired rate.

Note:The total of the primary rate plus the secondary rate cannot exceed 500 mL/hr.

430-95226-001 (Rev. 5/01)

8- 2

Plum A+ Infusion Pump

Page 73

Epidural Administration NoteThese guidelines apply only when Infusing short half-1iJe critical drugs in Concurrent mode. patient responses may vary requiring adjustment of delivery rates.

Individual

Delivery Rate Guidelines Short Half-life (less than 6 minutes) Critical Drug Infusion Rate

Maximum Rate of Accompanying Infusion ‘, - I ‘.

0.5 - 1.9 mL/hr

5 Times the Critical Drug Rate

2.0 - 5.0 mL/hr

10 Times the Critical Drug Rate

5.1 or Greater

Any Desired Ratio

CRITICAL D RUGS Examples of drugs with a short half-life (approximately 6 minutes or less when given IV) Include:

l Esmolol l Nitroglycerin l Nitroprusside l Norepinephrine l oxytocin l l Procainamide l Dobutamine l Dopamine l Epinephrine Isoproterenol l l Lidocaine

Trimethaphan

For these drugs, the concurrent flow guidelines should be followed when the infusion rate of the drug will be 5 mL/hr or less. Note: This list of critical drugs is not intended to be all-inclusive of critical drugs or drugs with a short half-life. The clinician should become familiar with the pharmacodynamics of any critical drug before administration. This information is presented to inform clinicians of a rare situation that could be misinterpreted if they are unfamiliar with this phenomenon.

8.4 Epidural Administration Recommended use of the epidural route is to provide anesthesia or analgesia for periods up to 96 hours. This device can be used to administer only those anesthetics/analgesics approved for epidural administration (as indicated or allowed by the drugs’ FDA approved labeling). Epidural administration of drugs other than those Indicated for epidural use could result in serious Injury to the patient. For epidural administration, the use of Abbott Laboratories catheter, List No. 1193 or List No. 6947, pump sets without Y-sites, and “epidural” stickers indicating ongoing epidural administration are recommended. Administration of drugs via the epidural route should be limited to personnel familiar with associated techniques and patient management problems. Proper epidural placement of the catheter is essential since

System Operating Manual

8- 3

430-95226-001 (Rev. 5/01)

E:

Page 74

Precautions catheter migration could result in intravascular or intrathecal administration. Facilities practicing epidural administration must be equipped with resuscitative equipment, oxygen, naloxone, and other resuscitative drugs. Adequate monitoring equipment (e.g., Orrimetry) is recommended for continuous monitoring of the patient during epidural administration. Patients must be observed frequently for side effects in a fullyequipped and staffed environment for at least 24 hours following completion of drug administration by the epidural route. DELAYED RESPIRATORY DEPRESSION FOLLOWING CONTINUOUS EPIDURAL ADMINISTRATION OF PRESERVATIVE-FREE MORPHINE SULFATE HAS BEEN REPORTED. The epidural space has 58 openings through which fluid can exit. Pressure buildup during administration is transient. However, if a large volume of fluid is administered over a short time period, the pressure will take longer to return to normal. If overdelivery occurs during administration, observe the patient closely for signs of spinal cord compression (disorientation, headache, transient neuralgias) and drug overdose.

8.5 Battery Operation CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITEI THE BATTERY REMOVED. USE OF A PROPERLY MAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION. The battery may not be fully charged upon receipt. Connect the Plum A+ to AC power for at least six hours. Use AC power whenever possible. Connect the Plum A+ to AC power during storage to ensure a fully charged battery for emergencies. If quality earth grounding source is in doubt, use battery power. If the low-battery alarm sounds, connect the Plum A+ to AC power immediately.

8.6 Sets and Accessories Only compatible LifeCare Plum Sets can be used with the Plum A+. See individual set instructions for additional information. Administration sets should be changed per CDC guidelines or healthcare provider policy. Discard after use. LifeCare IV Infusion sets with integral nonblood filters are not for use in the administration of blood, blood products, emulsions, suspensions, or any medications not totally soluble in the solution being administered. These medications may be administered through the lower Y-injection site, below the filter. When infusing at low delivery rates (5.0 mL/hr or less) the use of thick-walled microbore Plum sets is recommended. This will reduce the amount of the fluid bolus that may be delivered when a distal line occlusion is released. Use List 11986-48, syringe adapter when using syringes 10 cc or smaller on the secondary line. (Syringes must be larger than 3 cc.) Use a 19-gauge or larger needle or catheter at the venipuncture site for viscous fluids if operating at rates greater than 500 ml/hr. See Section 10 for information on sets and accessories.

430-95226-001 (Rev. 5/01)

8- 4

Plum A+ Infusion Pump

Page 75

Backpriming

8.7 Backpriming Backpriming is not recommended for reconstituting secondary containers containing dry powders. To avoid pressurization when backpriming Into a syringe or a vial, the user must confirm that these containers have sufficient empty space to accept the backprimed fluid.

8.8 General Possible explosion hazard exists if used in the presence of flammable anesthetics. Do not place Plum A+ in service if it fails the self-test. Do not operate the Plum A+ with the case opened. Keep the cassette door securely closed while the Plum A+ is not in use, to avoid cassette door damage. Use the following procedure to avoid the administration of a bolus following a distal occlusion:

If a secondary container is in use. clamp proximal tubing before opening cassette door Open cassette door and remove the cassette Open the flow regulator briefly to dissipate the pressure and then close it Eliminate the source of occlusion Reinsert the cassette and close the cassette door Open all clamps and resume infusion

System Operating Manual

8- 5

430-95226-001 (Rev, 5/01)