Plum A+ Infusion Pump System Operating Manual, pages 76 to 85

Pages 76 to 85 of 85: Specifications; Supplies & Accessories

The original document is the PDF (3.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the Abbott 11972-04-01

Page 76

Specifications

SECTION 9

PHYSICAL Dimensions:

Approximately 8” X 8” X 6”, excluding pole clamp protrusion and power cord storage.

Weight:

Approximately 9.5 Ibs. with battery.

Casing:

High-impact plastic

ELECTRICAL Power Requirements: Fuses: Power Cord: Battery:

120 V-, 50-60 Hz, 35 VA. Fl, F2,25OV-, 0.5 A. (internal) Hospital-grade AC cord. 10 ft long, with transparent plug and retainer plate. One sealed, lead-acid, rechargeable 6 V battery, internal to device.

Battery Life:

With a new fully charged battery, the infuser shall operate for a minimum of six hours at 125 mIJhr or less, or deliver 500 mL at 126 mL/hr or greater on one line.

Recharge:

The battery charges whenever the Plum A+ is connected to AC power. The recharge time is approximately six hours with the device operating at 125 mIJhr on one line.

Electrical Leakage:

Risk current limits meet AAMI/ANSI SCL (ungrounded) 12/78 standard.

NURSE-CALL System:

NURSE-CALL alarm is factory set for Normally- Open (NO) Note: Contact the Technical Services Center to make an internal adjustment to change the device from NorrnallyOpen (NO) to Normally Closed (NC) system.

Circuitry Ratings:

Voltage-30 VDC Max Current- 0.25 Amps Max Contact Rating- 3 Watts Max

VTBI RANGE VTBI Range:

0.1 to 99.9 mL (in 0.1 mL. increments) 100 to 9999 mL (in 1 mL increments)

9- 1

430-95226-001 (Rev. 5101)

Page 77

Specifications ENVIRONMENT Operating Temperature:

5” to 40” C

Storage Temperature:

-20” to 60” C

Atmospheric Pressure:

0 - 10,000 feet (0 - 3,000m) or equivalent pressure

Relative Humidity:

10 - 90% (40” C Max)

DELIVERY RATE RANGE Lines A and B:

0.1 - 99.9 mIJhr (in 0.1 mL increments) 100 - 999 mL/hr (in 1 mL increraents)

Concurrent Delivery: PlumSet: KVO:

0.5 Muir minimum for each line 500 mLfhr cumulative (A+B) maximum 1.0 Muir or the last primary delivery rate, whichever is less

AIR-IN-LINE ALARM PlumSet (Distal):

Bolus 0.1 mL or larger Cumulative 0.26 rnL out of 2.5 mL

PlumSet (Proximal):

430-95226-001 (Rev. 5/01)

Bolus at 0.5 mL, Total 1.0 mL

9- 2

Plum A+ Infusion pump

Page 78

Specifications OCCLUSION ALARM AND LIMITS Distal Occlusion:

The DISTAL OCCLUSION alamr sounds after the distal set tubing or set outlet fitting becomes occluded.

Proximal Occlusion:

The PROXIMAL OCCLUSION alarm sounds if the tubing proximal to the cassette becomes occluded.

Distal Pressure Limit (without alarm):

1 to 15 psig. The maximum pressure k user-selectable. Factory default setting is 6 psig.

Maximum Infusion Pressure:

20 psig

TIME TO DETECT DOWNSTREAM OCCLUSIONS Rate

Occlusion Alarm Pressure Setting

Typical Time to Alarm (Macro)

25 mIfhr

6 psig

38 seconds

47 seconds

15 psig

98 seconds

106 seconds

6 psig

17 minutes

22 minutes

15 psig

45 minutes

49 minutes

ImLJhr

Maximum Time to Alarm (Macro)

* Baseline backpressure is 0 psig *

BOLUS VOLUME RELEASED AFTER DOWNSTREAM OCCLUSIONS ARE CORRECTED Rate

Occlusion Alarm Pressure Setting

Typical Bolus Volume (Macro)

Maximum Bolus Volume (Macro)

6 ~51~

0.23 mL

0.28 mL

15 PQE

0.55 mL

0.63 mL

6 pslg I5 pslg

0.24 mL

0.30 mL

0.59 mL

0.70 mL

25 mJ.Jhr lm.Uhr

* Baseline backpressure is 0 psig *

E

System Operating Manual

9- 3

430-95226-001 (Rev. 5101)

Page 79

DELIVERY ACCURACY The Plum A+ Infusion pump was designed and manufactured to maintain a volumetric delivery rate error of the total fluid delivered of less than or equal to 2 5% over the course of 48 hours at a programmed rate of 1 to 999 mL/hr during normal operating conditions. For use of the device at rates below 1 rnL/‘hr, the delivery rate error is less than or equal to f 10%. Backpressure Effect- Positive backpressure on the distal line may affect delivery accuracy by no more than 0.5% per psig (2% per psig for rates between 0.1 and 0.9 mL/hr) for backpressures up to 15 psig. A typical deviation under these conditions is 0.3%. Negative backpressure may affect delivery accuracy no more than 2% for pressures up to -2 psig (56 inches of water). A typical deviation under these conditions is 0.3%. Filling Head Effect- Variations in filling head (such as container height) majr affect delivery accuracy by no more than 2% for variations in the range -15 to +28 inches. A typical deviation under these conditions is 0.5%. Concurrent Delivery Effect- When both lines (A & B) are delivering, the ratio of delivery for the fluid with the lowest rate may be affected by as much as 5% for ratios up to 9 to 1, For higher ratios, the absolute percentage of delivery for the lowest rate may be affected by no more than 0.5%. When air is present in the bubble trap, the absolute percentage of delivery for the lowest rate may be affected by up to 2.0%. When variations in container height are present, the absolute percentage of delivery for the lowest rate may be affected by up to 2.5% for up to 24 inches of container height differences. TRUMPET CURVES The Trumpet Curve Graphs following the Example show representative maximum and minimum percent flow rate deviation from the programmed rate over time. This information was developed in accordance with IEC 60601-2-24: 1998, Sub-Clause 50.102. Refer to this standard for detailed information. wi Hog p& W T : The graphs following the Example plot flow rates at 30 second intervals for the first 2 hours and for the 96th hour of delivery. The graph plots mean delivery rate error (Average of 3 pumps) for the 2nd hour and the 96th hour as a straight line. The graph also presents maximum and minimum average delivery rate error for this interval plotted by averaging delivery errors over intervals of 2, 5, 11, 19 and 3 1 minutes (“Trumpet Curve”).

430-95226-001 (Rev. 5/01)

9- 4

Plum A+ Infusion Pump

Page 80 read from a scan by OCR; it may hold reading errors

Specifications

EXAMPLE

From the Trumpet Curve Graph sample that follows, find the 5 minute interval (A) at the horizontal axis and read the corresponding points (B) and (C) on the vertical axis. The values are approximately +2.8% and -0.5%.

This means that at the rate of 25 mL/hr the average maximum flow rate fluctuation for any 5 minute time interval during the 2nd hour of operation was within the limits of +2.8% and -0.5% from the nominal rate. The average delivery rate error over the entire 2nd hour was +1.6% (D). :

For other time intervals look at other points at the horizontal axis and determine corresponding limits as above.

25 MUHR, 0 mm Hg DELIVERY HEAD, START UP

Qa

15.0% = et oo 4 Max Positive Error over 5 cr 10.0% © Minute Interval = +2.8% mc 5.0% 7 ' J ' bE 0.0% 7 i egy, 4 Max Negative Error over 5 2° . Minute Interval = -0.5% W 410.0% + ‘Average Delivery (Rate Error = +1.6% -15.0% H 0 5 10 15 20 25 30

Observation Interval (minutes)

| Max, + - Min -@ Mean

A trained professional can use the resulting graphs to select a pump with the appropriate startup and flow characteristics to suit the clinical application. :

Note: As an example of how the trumpet curves can be used, consider the maximum and minimum deviations at the 5 minute average interval. The upper curve provides the maximum expected delivery rate error over a 5 minute interval, the lower curve provides the minimum expected delivery rate error over a 5 minute interval. An example would be Dopemine administered at 5 pgm/kg/min. At 5 minutes, the average drug delivery error would be within the range of +2.8% and -0.5% of the expected nominal rate.

System Operating Manual 9-5 430-95226-001 (Rey. 5/01)

Page 81 read from a scan by OCR; it may hold reading errors

Specifications

50

25 MU/HR, 0 mm Hg DELIVERY HEAD, START UP (Typical)

45 40 4 35 4 30 |

20 4 15 4 10 5

Rate (mi/hr)

-10

10 20 30 40 50 60 70 80 90 Time (minutes)

|, - Actual Rate _ - - - Programmed Rate

25 ML/HR, 0 mm Hg DELIVERY HEAD, START UP

(Typical) 15.0% se 10.0% 4 & 5.0% 4 E Ne er Ps 0.0% = + + 8 5.0% ® 2.-10.0% + 15.0% 0 5 10 15 20 25 30 Observation interval (minutes) | Max, Min, - Mean | 25 ML/HR, 0 mm Hg DELIVERY HEAD, START UP (Composite of 3 Pumps) 15.0% ~ 10.0% + xe 5 5.0% + ee W 0.0% =< - F Q -6.0% 4 & -10.0% + -15.0% 0 5 10 15 20 25 30

Observation Interval (minutes)

= Max, +- Min -+ - Mean |

430-95226-001 (Rev. 5/01)

Plum A+ Infusion Pump

Page 82 read from a scan by OCR; it may hold reading errors

Specifications

25 ML/HR, 0 mm Hg DELIVERY HEAD , 96 HOURS (Typical) 50 45 4 40 4 = 354 = 304, E 25 + sa Dee PA LOAD foe SLBA Aer Ae fprormaly Jae ae NO A IN AA ee A © ra 20 + Mm 154 104, 5 4 0 : : : . 7 : . . . 7 0 10 20 30 40 50 Time (minutes) {, - Actual Rate Programmed Rate | 25 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS (Typical) 15.0% J 10.0% 7 5 5.0% + : uw 0.0% ts T + - r, e 8 5.0% 4 o OL -10.0% 4 -15.0% 0 5 10 15 20 25 30 Observation Interval (minutes) [ - - Max -* - Min, - Mean |

25 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS (Composite of 3 Pumps)

15.0%, ~ 10.0% 4 x 5 5.0%) 9 ~~~. Ui 0.0%, < 8 5.0% ® O- 50.0%

15.0%

0 5 10 45 20 25 30 Observation Interval (minutes) [ - - Max, - Min, - Mean |

System Operating Manual 9-7 430-95226-001 (Rev. 5/01)

Page 83 read from a scan by OCR; it may hold reading errors

Specifications

1 ML/HR, 0 mm Hg DELIVERY HEAD, START UP (Typical) 2.0 1.8 5 1.6 4 am 144 E12. | | | E 1.0 47> ed ~ aA ene RTT ME Poe top or = 08 ia @ 064 GS 0.44 © 024 0.0 T T T 7 T -0.2 4 -0.4 0 10 20 30 40 50 ~= - «-60 70 80 go 100 110 120 Time (minutes) Actual Rate Programmed Rate | 1 MU/HR, 0 mm Hg DELIVERY HEAD, START UP (Typical) 15.0% $e10.0% 4 5 5.0% 5 NN bad UW 0.0% ee t 5, - 3) -5.0% + o O.-10.0% + -15.0% i) 5 10 15 20 25 30 Observation Interval (minutes) 4 ML/HR, 0 mm Hg DELIVERY HEAD, STARTUP (Composite of 3 Pumps) 15.0% Se 10.0% | 5 5.0% 4 NN Wi 0.0% - c ., 8 5.0% 4 @ -10.0% 4 -15.0% 0 § 10 45 20 25 30 Observation Interval (minutes) [ - - Max, = Min, - Mean |

430-95226-001 (Rev. 5/01) 9-8 Plum A+ Infusion Pump

Page 84 read from a scan by OCR; it may hold reading errors

Specifications

1 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS (Typical) : 2.0 1.8 4 (164 £ 1.47 = 124 & o64 0.4 4 0.2 0 10 20 30 50 Time (minutes) [, - Astual Rate _ - - - Programmed Rate _| 1 MUHR, 0 mm Hg DELIVERY HEAD, 96 HOURS (Typical) 15.0% SS 10.0% 4 S 5.0% | Wi 0.0% = 2, - + , = ; 4 @ -10.0% + -15.0% 0 5 10 15 20 25 30 Observation Interval (minutes) | - - Max, > Min, + - Mean | 4 MU/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS (Composite of 3 Pumps) 15.0% > 10.0% 5 = 5 5.0% + ee uw 0.0%, + ' r= 8 5.0% + a o OW -10.0% 4 -15.0% Q 5 410 15 20 25 30 Observation Interval (minutes) | - - Max, - Min - - Mean | System Operating Manual 9.9 430-95226-001 (Rev. 5/01)

Page 85

SECTION IO

Supplies and Accessories

IO. 1 Administration Fluids ADMINISTRATION FLUIDS l All parenteral fluids l Whole Blood l Blood fractions ENTERAL AND HIGH VISCOSITY FLUIDS l Dextrose 20% to 50% concentration l Isomil Powder l Similac Powder l Ensure plus HN l TwoCal HN l Packed red cells CONTAINERS Dual Chamber Parenteral flexible container (Nutrimix) Large Volume Parenteral flexible plastic containers, including premixed Large Volume Parenteral glass containers, including premixed and nutritional Part Fill Parenteral flexible plastic containers, including ADD-Vantage Part Fill Parenteral glass containers Small Volume Parenterals Syringes, vials. includes part fills (may require special adapters) Top Filled Enteral bags Ready to hang Enteral solution containers

10.2 Accessories Syringe Adaptor- List# 11986-48 Vial Adaptor- List# 1188448 Mini Pole- List# 12096-04-91 Secondary Container Support- List# 12095-04-61 Bar Code Reader (Available as a Field-Installed Infuser Option)-

List# 11972-04-01

Tandem Carrier- List# 12270-O 1 lo- 1

430-95226-001 (Rev. 5101)