Page 45

SECTION 5

TherapieS (For Advanced Users*) * Advanced users are those who have received additional training on programming and use of functions.

Dose CalculationLoading Dose-

Multistep-

Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep. Allows programming of an initial infusion rate for a specific volume, followed automatically by a maintenance rate from the same container (e.g., a fluid challenge). If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time (mcg/min), from the same container. Allows a sequential program to deliver up to 10 steps; fluid volumes and delivery rates may be programmed for each step. The program may be entered based on Rate and Volume or Volume and Time. If Dose Calc is used, the delivery steps are in the same unit of measure, over the same period of time, from the same container.

5.1 Dose Calc (mcg/kg/min on A) DOSE CALCULATION - Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep. Clear ALL settings when changing patients or multiple programming parameters. Use A [C/ear Programlwhen changing ONLY Line A or B. 1. Press A [A] at the Main Delivery Screen. The Line A Programming screen appears. Rate m mL/hr

2. Press A [Therapy].

Duration 0O:OO hr:min

5- 1

430-95226-001 (Rev. 5/01)

Page 46

Therapies (For Advanced Users*) The Select Therapy screen for Line A appears. Note: Dose Calculation is the only available option if Line A is infusing. Loading Dose

3. Press A [Choosejto continue. ’ ‘:

Multistep

IAI

Proaram Dose Calc

I

The Program Dose Calc screen ap’pears. Note: Screen defaults to the mcg/kg/min option or the previously selected option.

4. Use [SELECT] key to highlight the units to use in the delivery.

5. Press A [Choose] to continue. Note: Previously selected Drug Names and Delayed Start entries will be cleared if entered prior to Therapy selection. To reselect, press A [Program Options1 after selecting desired Therapy and Units. 1Al Program Dose Calc Drug Cone in Container

The Container Concentration Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng) otherwise this screen is skipped. Note: Default choice for possible gram-based Container

Concentration Unit is mg or the previously selected option.

6. Use [SELECT] key to highlight the Container

Concentration units to use in the delivery. 7. Press A [Choose]to continue. The next page of the Program Dose Calc screen appears to provide for entry of delivery program parameters. Weight -kg o.cra mr)*pmrkl

3. Use keypad to enter desired values. Once each value is entered, use [SELECT] to highlight the next programming ‘ield. Note: The Concentration elements and weight value must be entered before the (SELECT] key allows access to remaining fields.

430-95226-001 (Rev. 5/01)

5- 2

Plum A+ Infusion Pump

Page 47

Dose Calc (mca/ka/min on A)

A ( Program Dose Calc wvKIAdi4# 400 mg 70.0 kg

250 mL

-

As the Dose value is entered, the Rate is automatically calculated. To understand the Autom’atic Calculation feature, see

Note:

m 0.0 lr#g!nlin ml_ mm kc* 7.9 ml.&, rA I

9. Continue to enter values using the keypad and advancing to the next field using [SELECT].

Program Dose Calc -MN0

Section 4.7 Examples of Automatic Calculation,

10. Use [SELECT] to scroll down to the VTBI field.

E

I

A I Program Dose Calc WMdccnc mmg

Weight

70.0 kg

250 rnL

KY 0watkel bs

11. Enter a value for VTBI using the keypad. Note: As the VTBI value is entered or changes, Duration is automatically updated. Note: The A [Program Options] path to setting Delayed Start or assignin a drug name from the internal list functions exactly as describe 2 in Basic Operation for the Simple Delivery in Section 4.

CAUTION: VERIFY DOSE, RATE, VTBI, PRESSING [START].

AND DURATION BEFORE

12. After entering the desired VTBI value and any desired A [Program Options],,press [START] to begin infusion.

System Operating Manual

430-95226-001 (Rev. 5/01)

Page 48

Therapies (For Advanced Users*)

4 1 PUMPING 1 STOPPED ~ZZL4&C4 7.9 z;;, 0.0

0.1

Vi:“’ 0.0

1M mql*phnin

The Main Delivery Screen appears, to display the information just entered. Note: The Therapy Type is displayed as well as the Dose Rate and Dose Units.

%, I

Dose

430-95226-001 (Rev. 5/01)

5- 4

Plum A+ Infusion Pump

Page 49

Dose Calc (mg/hr on B)

5.2 Dose Calc (mg/hr on B) A 1 PUMPING 1 STOPPED 1 B

5.5 m?ehr

9-o

O”

Vol Inf mL

0.0

D OSE C ALCULATION - Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep. :, Clear ALL settings when changing patients or multiple programming parameters. Use A [C/ear Programlwhen changing ONLY Line A ‘1 or B.

1. Press A [B] at the Main Delivery screen. The Line B Programming screen appears.

2. Press A [Therapy]. Duration 0O:OO hr:min

The Select Therapy screen for Line B appears. The screen defaults to Dose Calculation or the previously selected option. Note: Dose Calculation is the only available therapy while Line B is running.

3. Press A [Choosejto continue. The Program Dose Calc screen appears. Note: The screen defaults to the mcg/kg/min option or the previously selected option. Note: Previously selected Drug Names and Delayed Start entries will be cleared if programmed prior to Therapy selection. To reselect, press A [Program Options] after selecting desired Therapy and Units.

System Operating Manual

5- 5

430-95226-001 (Rev. 5101)

Page 50

Therapies (For Advanced Users*)

4. Continue to scroll using [SELECT] key until desired field is highlighted. 5. When desired field (mg/hr in this sample) is highlighted, ‘_ press A [Choose].

The Container Concentration Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng) otherwise this screen is skipped.

Drug Cow in Container

Note: Default choice for possible gram-based Container Concentration Unit is mg or the previously selected option.

6. Use [SELECT] key to highlight the Container Beck

,A xA

A,

Concentration units to use in the delivery 7. Press A [Choose]to continue. The next page of the Program Dose Calc screen appears. The screen defaults to Concentration (Cone) and a value must be entered to continue. 8. Use keypad to enter desired values. Once entered, use [SELECT] to highlight the next field.

9. To override the Piggyback Delivery Mode default, scroll down to Mode using [SELECT], then press A [Change

Mode]. Note: Delivery Mode Option is available on Line B, its default is Piggyback, and the Weight field is gone.

430-95226-001 (Rev. 5/01)

5- 6

Plum A+ Infusion Pump

Page 51

Dose Calc (m&r on B)

Program Dose Calc IB v&nd pdive fo0 mL cone 200 w

10. Continue entering desired values using the keypad. Use [SELECT] to advance to the next field.

Mode Concurrent

When entering the RATE, with-a VTBI value entered, the DOSE and DURATION fields are automatically updated. To understand the Automatic Calculation feature, see Section 4.7 Note:

Examples of Automatic Calculation,

The Duration field temporarily than es to dashes when the computed value is larger than what can be /?isplayed. Note:

11. Continue entering the RATE value, if desired. Note: As Hate is entered, the Duration and Dose are automatically adjusted as individual numbers are entered.

Note: If the Duration value is changed, the Rate and Dose are automatically adjusted. Program Dose Calc wAc6u

IB z &&J mL loa

CAUTION: VERIFY DOSE, RATE, VTBI. AND DURATION BEFORE PRESSING [START].

12. To start the delivery, press [START]. Note: To select Delayed Start, press A [Program OptionsJ

PUMPING 1 PUMPING 1.0 5.0 n&e mLI hr

Vol Inf

Upon [START], the Main Delivery Screen shows delivery status for both A and B lines. Note:

See Pelivery Rate Range in Section 9.

rates has recommendations specified in

The ratio of concurrent

Section 8.3.

mL

13. Select A [BJ

Program Dose Calc IB Wmd_ -

14. To update and identify the infusing drug, enter the drug name by using either the BCR or via A [Program Options] then A [Drug List] with an entry for a selection. The drug name will appear in the upper-right corner. Note:

If a drug name was previously entered and confiied by pressing [START], it cannot be changed while line is infusing. The line must be stopped and the Dose Calc screen entered before it can be changed.

System Operating Manual

5- 7

430-95226-001 (Rev. 5/01)

Page 52

Therapies (For Advanced Users*)

4 1 PUMPING 1 PUMPING 1 E D%o c%c%z

15. Press [START] to confirm entries and return to the Main I. Delivery screen.

1.0 Rate 0.5 mL1 hr 5.0 Vd

Inl 0.2 mL

Dose

CAUTION: VERIFY DOSE, BATE, VTBI, AND DURATION BEFORE PRESSING [START].

s,

430-95226-001 (Rev. 5/01)

5- 8

Plum A+ Infusion Pump

Page 53

Loading Dose

5.3 Loading Dose L OADING D OSE - AIIOWS programming of an initial infusion rate for

/A( 0.0

:;tr 0.0

0.0 y: 0.0 a specific volume, followed automatically ,by a maintenance rate from the same container (e.g.. a fluid challenge). If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time (mcg/min). from the same container.

Clear ALL settings when changing patients or multiple programming parameters. Use A [C/ear Program]when changing ONLY Line A or B.

1. Select A [A] (or A [BJ at the Main Delivery screen. The selected Line (A or B) Programming screen appears.

2. Press A [Therapy]. Duration 0O:OO hnmin

Al

PROGRAM

The Select Therapy screen for Line A (or B) appears. VJand AJxka Note:

This example shows line A being programmed.

Loading Dose Multistep

Al

PROGRAM Wsnd_ Doss Calculation

3. Use [SELECT] to scroll down to Loading Dose.

4. Press A [Choose] to continue.

Multistep

System Operating Manual

430-95226-001 (Rev. 5101)

Page 54

Therapies (For Advanced Users*)

I A 1 Proaram Loadina Dose 1

The Program Loading Dose screen appears. Note: The screen defaults to the mL/hr option or the previously selected option.

5. Press A [Choose]to continue, dr use [SELECT] to highlight another option. Chmse Back

<AA AA

6. When desired field is highlighted, press A [Choose]. The next page of the Program Loading Dose screen appears.

2.k mm mL 0.0

“pgp

1 2

0.0

0.0

WE00

0o.m

7. Enter a rate using the keypad from 0.1 to 999 mL/hr depending on delivery mode.

8. After the desired value has been entered, use [SELECT] to highlight the next field.

430-95226-001 (Rev. S/01)

5- 10

Plum A+ Infusion Pump