Plum A+ Infusion Pump System Operating Manual: Basic Operation, pages 37 to 44

Pages 37 to 44 of 85, from Plum A+ Infusion Pump System Operating Manual

The original document is the PDF (3.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the Abbott 11972-04-01

Page 37

Concurrent Deliverv

PROGRAM

1B

4. To enter delivery rate, use numeric keypad.

Wrnd Aati

Mode Concurrent Rate VTBI m ml/hr 0.0 mL

Duration 0O:OO hcmin

When Rate for Line A plus Rate for Line B is greater than valid rate ranges, a “Concurrency violation” message will explain why second line did not start. See Section 9, for ‘. valid deliwry rate ranges. Note:

Note: ‘Concurrency violation” may not be displayed if unit is displaying “VTBI Complete”. The device will not start and an invalid keypress sound will occur when START is pressed.

5. Use [SELECT] key to move down to the VTBI field. 6. To enter VTBI amount, use keypad. Mode Concurrent

Duration En:00 hnmin

Note: While theVTB1 is being entered, the duration is automatically computed. While the computed value is greater than 99:59, dashes (- -) appear in place of hours. To understand the Automatic Calculation feature, see Section 4.7 Examples of Automatic calculation,

7. Use [SELECT] key to move down to the DURATION field. 8. (Optional) Enter values for desired hours and minutes. Note:

When Duration is entered, Rate is automatically computed.

CAUTION: VERIFY RATE. VTBI. AND DURATION PRESSING [START].

BEFORE

9. Press [START] to begin infusion and return to the Main Delivery screen. Example of a concurrent delivery.

System Operating Manual

4- 19

430-95226-001 (Rev.

5101)

Page 38

Basic Operation

4.14 Select Option- Volumes Infused, Pressure/Post Infusion Rate, and Lighting/Contrast 1. To select Options Screen, press

A,[OptionsNol In@

2. Use [SELECT] to highlight an option. 3. Press A [Choose] to select the Volumes Infused screen. Pressure/Past Infusion Rate Lighting/Con~rast

IAl

OPTIONS

IBl

This will bring up the Volumes Infused screen which defaults to Total Volume.

Volumes Infused 393 mL LineA 0.0 mL tine B Total Vol m mL

4. Use [SELECT] to scroll to Line A or B.

5. To reset Line A (or B) Volume Infused, use [SELECT] to highlight and press [CLEAR]. 6. To restore the,cleared quantity, press A [Cancel Back-

430-95226-001 (Rev. 5/01)

4- 20

Plum A+ Infusion Pump

Page 39

Select Option- Volumes Infused, Pressure/Post Infusion Rate, and Lighting/Contrast

7. To reset all valumes, use [SELECT] to highlight Total Vol, then press [CLEAR]. Volumes Infused

8. To restore all quantities, press A‘ [Cancel/Back] softkey. 9. When clearing of quantities is completed, press to return to the Main Delivery screen.

A [Enter]

10. Select A [Options/l/o/ lng from t;he Main Delivery screen to return to the Options screen. 11. Use [SELECT] to highlight the Pressure/Post Infusion Rate option.

Lighting/Contrast

I

12. Press A [Choose]to continue. Note:

Prior to changing Pressure/Post Infusion Rate, pump must be stopped. Al

OPTIONS

IBl

Pressure/Post Infusion Rate Canlinw Rate m 6.0 psi D&l Presrwe Limit Dii

Pressus

0.0 psi

13. Use [SELECT] to highlight Continue Rate or Distal Pressure Limit field. 14. Continue Rate allows a choice when VTBI is completed. Press A [KVO/Rate] KVO- The pump will revert to a fixed KVO of 1 mL/hr or the last programmed Rate, whichever is less. Rate- The pump @ll continue at the Rate programmed.

Note: Option will be for both line A and line B selections. 15. To change Distal Pressure Limit when highlighted, enter value between 1.0 and 15.0 psi. 16. Press A [Enter] to keep changes and return to the Main Delivery screen, or A [Cancel/Back] to restore original values Options screen. Continue Rate default of KVO or Rate is a biomedical setting and must be configured as described in the Technical Service Manual.

Note:

System Operating Manual

4- 21

430-95226-001 (Rev. 5101)

L A

Page 40

Basic Operation

\I

OPTIONS

IB

Volumes Infused Pressure/Post Infusion Rate

17. Select A [Options/Vo/ lnfl from the Main Delivery screen to return to the Options screen. 18. Use [SELECT] key to highlight Cighting/Contrast Option.

19. Press A [Choose] to continue.

[Al

OPTIONS Lighting/Contrast

Display Contrast m

20. Press [SELECT] key to highlight’Backlight Intensity or Display Contrast. 21. While viewing display for desired effect, press either A [Increase Setting] or A [Decrease Setting] to change level. 22. Press A [Enter]to keep change(s) and return to the Main Delivery screen, or A [Cancel/Back]to restore previous settings and return to the Options screen.

430-95226-001 (Rev. 5/01)

4- 22

Plum A+ Infusion Pump

Page 41

Stop and Start with only 1 Line Pumping

4.15 Stop and Start with only 1 Line Pumping A) PUMPING1 STOPPED 1 B

1. Press [STOP] at the Main Delivery Screen.

100

,“,a:;, 0.0

Note: If A [A] (or A [BJ is pressed prior to lSTOP], the line is stopped, the display will show the progr amming screen. Press A [CanceYBack]

8.5 v;:“’ 0.0 to return to the Main Delivery screen.

A

<A A A

B

Options Vol Ini

A, Example of STOPPED delivery.

2. To discontinue infusion, press [ON/OFF]. 3. Clamp secondary tubing if connected. 4. Open door, remove set and close door. Or, press [START] to resume infusion.

4.16 Stop and Start with Both Lines Pumping 1. Press [STOP] at the Main Delivery Screen.

System Operating Manual

430-95226-001 (Rev. 5/01)

Page 42

Basic Operation

2. Press A [Stop A], A [Stop B], or A [Stop All] Note: If A [A] (or A [BJJ is pressed prior to [STOP], the display will show the programming screen. Press A [Cancel/Back] to return to the Main Delivery screen.

q rAISX)PPED 25.0 my;;, 50.0 3.0 “y

Example of stopped infusion.

’

Note: This is an example using A [Stop AU].

20.4

CAUTION: VERIFY RATE, VTBI, PRESSING [START].

AND DURATION BEFORE

3. To resume infusion, press [START]. 4. Select A [Start A], A [Start B], or A [Stati All]. Or, to discontinue infusion, press [ON/OFF]. 5. Clamp secondary tubing if connected. 6. Open door, remove set and close door.

430-95226-001 (Rev. 5/01)

4- 24

Plum A+ Infusion Pump

Page 43

Backpriming

4.17 Backpriming A) STOPPED 1 STOPPED 1 B 5.5

:;tr 0.0

BACKPRIMING Removal of air or fluid from the proximal line and cassette airtrap.

1. To initiate a backprime, press A [Back Prime].

8.9 “Z:” 0.0

Note: Backpriming is available only when all delivery is stopped. Secondary container is necessary to receive the air. Bsck A prime

AA

R%z

A

A, 2. Press and hold A [Back Prime] until fluid pumped from line A to line B has expelled the air from cassette air trap into Secondary container. Note:

Fluid flow is at maximum rate possible.

3. Upon release of A [Back Prime], the cassette test is performed to confirm readiness for pumping. Pressing [START] will resume programmed delivery.

4.18 Clearing Program Settings A I PUMPING i STOPPED I 61

1. Press [STOP] at the Main Delivery Screen. Clear ALL settings when changing patients or multiple programming parameters. Use A [c/ear Program]when changing ONLY Line A or 8.

System Operating Manual

4- 25

430-95226-001 (Rev. 5/01)

Page 44

Basic Operation

Al

PROGRAM Rate m mUhr

VTBI

22.0 mL

Duration 03:20 hr:min

2. Select Line A (or B if active) from the stopped Main Delivery Screen. 3. Select A [C/ear Program]. Note: If Line A or B is programmed with a Therapy (i.e., Dose Calculation), Rate, VTBI and Duration will not display values. The values to be cleared may be viewed in applicable programming screen under A [Therapy].

4. Select A [Yes]. Clear Line A Settings?

Clear ALL settings when changing patients or multiple programming parameters. Use A [C/ear Programlwhen changing ONLY Line A or B.

PROGRAM wind &++a

Rate VTBI m ml/hr 0.0 mL

Example screen of “Cleared Program”.

5. Press A [CanceVBack]to return to the Main Delivery screen or program new settings.

Duration 0O:OO hnmin

4.19 Possible Non-Delivery Programmed Note: Whenever a valid [SIXRT] key is pressed, the device checks the delivery program(s) for non-delivery conditions (such as Line B Piggyback without Line A or delayed start) that could permit a period of non-delivery. If so, the display presents this warning screen. continw anyway?

1. Press A [Yes] to continue with delivery. Or, press A [N O] to cancel.

430-95226-001 (Rev. 5/01)

4- 26

Plum A+ Infusion Pump