Page 28

Basic Operation

4.8 Titration r

I

R A

1. Select Line A [A] (or A [BJ at the Main Delivery Screen.

I AI

,AlAlAlA]

2. Enter the desired Rate using the keypad. Note: When Rate is changed, the duration is automatically recomputed using the current VTBI. To understand the Automatic Calculation feature, see Section 4.7 Examples of Automatic

Calculation,

Duration 03:20 h

CAUTION: VERIFY PRESSING [START].

RATE. VTBI.

AND DURATION

BEFORE

3. After any change, press [START] to initiate new Rate and return to the Main Delivery Screen.

A

\A A A A,

B

%z

Note: Volume Infused is an incrementing value which is cleared in A [Options/V01 Inn.

THE FOLLOWING 3 SCREENS ARE EXAMPLES OF CHANGED VALUES. EXAMPLE OF CHANGED RATE. Duration 03:20 hrmin

4. From the Main Delivery Screen, press A [A]. 5. Use keypad to change the RATE value. The display time is only accurate to f one minute. As shown in this example, the delivery shows 3 hours and 20 minutes (not 3 minutes and 20 seconds).

Note:

Note: [START] accepts the change. Delivery is changed to the new RATE and DURATION and display returns to Main Delivery Screen.

430-95226-001 (Rev. 5/01)

4- 10

Plum A+ Infusion Pump

Page 29

Titration

EXAMPLE OF CHANGED VTBI VALUE. 6. Use [SELECT] key to highlight the VTBI field. Duration 06:40 hrzmin

7. Use the keypad to change the VTBI:value. When the VTBI is changed, the Duration is recomputed with respect to the current Rate. To understand the Automatic Calculation feature, see Section 4.7 Examples of Automatic Calculation,

Note:

Note: [STAKE?] accepts the change. Delivery is changed to the new VTBI and DURATION and display returns to Main Delivery Screen.

EXAMPLE OF CHANGED DURATION FIELD. 8. Use [SELECT] to highlight the Duration field, then use the keypad to change it’s value. Duration DO:mhnmin

When DURATION is changed, RATE is automatically recomputed using the current VTBI value. To understand the Automatic Calculation feature, see Section 4.7JZxampIes OfAutomatic Note:

Calculation,

Note:

[SI’ART] accepts the change. Delivery is changed to the new DURATION and RATE and display returns to Main Delivery Screen.

CAUTION: VERIFY PRESSING [START].

System Operating Manual

4- 11

RATE. VTBI,

AND DURATION

BEFORE

430-95226-001 (Rev. 5101)

Page 30

Basic Operation

4.9 Simple Delivery using Drug List 1 A I SfOPPED STOPPED I B I

0.0

Jp,& 0.0

0.0

““,:“’ 0.0

1. Select Line A [A] (or A [f3j) at the Main Delivery Screen, Note: To select a Drug Name using a Therapy (i.e., Dose Calc) you must first select which line you want (A or B), next select A [Therapy], and then A [Program Options]. If you select a Drug Name first, it will be cleared upon entry in Therapy mode.

The Line A Programming Screen appears. Rate m mVhr

Duration 0O:OO hrzmin

2. Enter a RATE using the keypad. 3. Use [SELECT] to highlight the VTBl field and enter a value. The Duration will be automatically calculated.

4. Before or after entering the simple delivery program, press A [Program Options]. VTBI

WmL

Duration 50:00 hrmin

A1

Program Options wud&aivl

Delay Start for lO0 hcmin

Callback

The Default screen for Program Options appears.

5. Press A [Drug List].

No

430-95226-001 (Rev. S/01)

4- 12

Plum A+ Infusion Pump

Page 31

Simple Delivery using Drug List The Drug List Screen appears, in mainly alphabetical order. Note: Default entry is “No Drug Selected”. This selection may be used later to remove an incorrect or unwanted drug name from the display.

Alfentanil Alpnstadil Alteplase (d-PA) Amikacin

Program Drug List Wnd_ Cydogporin cytarabine Dexmetatomidine Diltiiem Diuretic Dotutamine

6. Use [SELECT] to highlight desired’drug name. 7. When list does not contain desired generic drug name or group name press A [Page Down] (or A [Page Up]) until it shows, then use [SELECT] to highlight choice. 8. Press A [Enter]to indicate selection and return to the Program Options screen. Or, press A [Cancel/Back] if no choice is acceptable.

A1 aopamm

Program Options wmld*ntw

Delay Start form:00 hcmir: No

Callback

Drug name is now visible in the upper-left corner.

9. Press A [Enter] or A [Cancel/Back] to return to the Line Programming screen. CAUTION: VERIFY PRESSING [START].

RATE, VTBI,

AND DURATION

BEFORE

10. Press [START] to confirm the programming, begin the delivery, and return to the Main delivery screen. A 1 PUMPING 1 STOPPED 1 B Dc+mlwd

5.0 r&k,

0.1

VZ:“’ 0.0

Notice the drug name remains in the upper-left corner.

6.0

System Operating Manual

4- 13

430-95226-001 (Rev. 5/01)

Page 32

Basic Operation

4.10 Simple Delivery Using Delayed Start 1. Before or after entering the simple delivery program, 1 Rate

5.0 mL/hr I

VTBI

BmL

Duration 50:00 hr:min

A1

Program Options Wud_ press A [Program Options].

j

,

Note: To program a Delayed Start using a Therapy (i.e., Dose Calc) you must first select which line you want (A or B), next select A [Therapy], and then A F'rogrcun Options]. If you program the Delayed Start first, it will be cleared upon entry into Therapy mode.

The Default screen for Program Options appears.

Delay Start for BOO hr:min

2. Enter values for desired hours and/or minutes. Delay Start for Ooa hrzmin

3. Press A [Enter-to confirm delayed start. CAUTION: VERIFY RATE, VTBI, PRESSING [START].

AND DURATION BEFORE

4. Press [START] to confirm programming.

t A I DELAYED I STOPPED I B 1

5.0 mt%r O-O

O-O

Vd’nf mL

Line A is delayed for the time entered. Note: Delayed line drip indicator is solid, not flashing. It will begin flashing when actually pumping.

0.0

430-95226-001 (Rev. 5/01)

4- 14

Plum A+ Infusion Pump

Page 33

Piggyback Delivery

4.1 I Piggyback Delivery I A I PUMPING1 STOPPED 1 B I

5.5 z;Er 0.0

20.0 vy 0.0

PIGGYBACK DELIVERY - Infusion will stop Line A and infuse Line B until VTBI completes. Line A will automatically restart. 1. With pumping started on Line A, t&rogram a Piggyback delivery, press A [Bl.

L

‘1

A

The Delivery Mode field is highlighted. Note:

If an entry was previously programmed, RATE is highlighted. If no previous entries, MODE is highlighted.

Duration 0090 hr:min

2. If field reads “Concurrent”, press A switch to “Piggyback”.

[Change Mode]to

3. Press [SELECT] to highlight the RATE field. Note: Upon entry to this screen, and to all lower screens from here, the Barcode Reader (BCR) is enabled as indicated by “Wand Active” when attached to the Plum A+. To use the BCR, swipe tip of BCRwand across entire barcode, from dot to dot, on the Abbott supplied drug list card. The drug name also can be selected using A IPrwrcun

Options].

4. To enter delivery rate, use numeric keypad.

5. Use [SELECT] key to move down to the VTBI field. Duration 0O:OO hr:min

System Operating Manual

4- 15

430-95226-001 (Rev. 5101)

Page 34

Basic Operation

6. To enter VTBI amount, use keypad. Mode Piggyback

Duration 00:30 hcmin

Note: While the VTBI is being entered, the duration is automatically computed. If the computed value is greater than 99:59, dashes (- -) appear in place of hours. To understand the Automatic Calculation feature, see Section 4.7 Examples of Automatic Calculation,

7. Use [SELECT] key to move down to the DURATION field.

8. (Optional) Use keypad to change duration. Mode Piggyback 133 mL/hr Rate ml 100 mL Duration OOa hnmin

When all parameters are displayed, changing any one parameter will result in other parameters automatically being computed. CAUTION: V E RI FY RATE. VTBI. AND DURATION PRESSING [START].

BEFO

RE

9. Press [START] to begin infusion. Note: If Line A has a Delay Start remaining, a message will appear warning of a possible non-delivery (see section 4-18).

10. Press A /Yes] to continue. A i DELAYED 1 PUMPING 1 B 1

5.5 m%“,,

20.0

““,:” 0.1

Example of Main Delivery Screen just after being programmed and started.

133

430-95226-001 (Rev. 5/01)

4- 16

Plum A+ Infusion Pump

Page 35

Piggyback with Nurse Callback

4.12 Piggyback with Nurse Callback PROGRAM v&d &%++a

(B

Mode Piggyback 200 mUhr Rate m mL VTBI

1. Press A [Program Options]. , To set a nurse callback, the pump must be stopped. This function can be selected either before or after entering the Piggyback delivery program. Note:

Duration DC:30 hcmin

1

Program Options B wuld MIU

Delay Start for 0O:oO htzmin CallbaCk I

PROGRAM wmdA0.r.m

2. Use [SELECT] to highlight “Callback” field and press A [Yes/No], if necessary to obtain “Yes”, then press A [Enter]. Note: Callback default ofYes or No is a biomedical setting and must be configured as described in the Technical Service Manual.

1B

Mode Piggyback 200 mL/hr Rate VT31

The default screen for Program Options appears.

m mL

CAUTION: VERIFY PRESSING [START].

RATE. VTBI,

AND DURATION

BEFORE

3. Press [START] to confirm program setting and return to Main Delivery screen.

Duration 00:30 hcmin

At end of Piggyback delivery, with Callback enabled for the line, Callback alarm is issued. 4. Press [SILENCE] to clear alarm.

System Operating Manual

4- 17

430-95226-001 (Rev. 5/01)

Page 36

Basic Operation

4.13 Concurrent Delivery l-

I

-PUMPING/ 5.5

E,

0.1

““nl:“’ 0.0

O-0

I m

\rIAlAlAj

CONCURRENT DELIVERY - Simultaneous delivery of two fluids at e. ~ independent flow rates. CAUTION:

II

C ONSULT DRUG LABEL&G TO CONFIRM DRUG COMPATIBILITY, CONCENTRATION, DELIVERY RATES, AND VOLUME ARE ALL SUITABLE FOR CONCURRENT AND PIGGYRACK DELIVERY MODES. S EE S ECTION 8, FOR INF O R M A TION ON C ONC URRENT FLOW (SEE

SECTION 9 FOR DELJVER Y RATE RANGES).

Note:

In concurrent mode with ceitain alarm conditions, one line may stop while the other line may continue to deliver.

1. With pumping started on line A, to program a Concurrent delivery rate, press A [f3J PROGRAM (B w*

Mode bxq Rate 0.0 mUhr 0.0 mL VrBl Duration 0O:OO hcmin

The Delivery Mode field is highlighted. Note: If an entry was previously programmed, RATE is highlighted. If no previous entries, MODE is highlighted.

2. If field reads “Piggyback”, press A [Change Mode]to switch to “Concurrent”. Note:

Upon entry to this screen, and to alI lower screens from here, the Barcode Reader (BCR) is enabled as indicated by “Wand Active” when attached to the Plum A+. To use the BCR. swipe tip of BCR wand across entire barcode, from dot to dot, on the Abbott supplied drug list card. The drug name also can be selected using A fProgram OptionsJ

Verify change to “Concurrent” mode. Mode B 0.0 mUhr Rate

3. Use [SELECT] to highlight Rate field.

0.0 mL Vi-B1 Duration 00: 00 hrrmin

430-95226-001 (Rev. 5/01)