The text of a manual · part 5 of 12
Plum A+ Infusion Pump System Operating Manual: Basic Operation, pages 19 to 27
Pages 19 to 27 of 85, from Plum A+ Infusion Pump System Operating Manual
The original document is the PDF (3.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Abbott 11972-04-01
Page 19
Basic Operation
SECTION 4
Line A Flow Indicator tine B Flow Indicator
Status Region ____)
Status Region
Working Region _I_)
Message Region e-b Softkeys Label Region j
+ LCD
Working Region
Messa! ge Region xkIrlr I
AA
Start and Stop Keys - moom
Softkeys - Numeric Keys UP/LEFT e Selection Key
Charging/Line Indicator
DOWN/RIGHT k + Silence Key
ON/OFF Key ___)
- Decimal Point Key
Numeric Field Clear Key Y
Basic Layout of Front Panel Display and Keypad
4.1 GETTING STARTED This section details the Plum A+ instrument setup procedures.
SYSTEM SELF-TESTS CAUTION: Do NOT P L ACE TRE PLUM A+ IN SER VI CE IF IT FAILS THE S YSTEM S ELF -TES TS . Connect the mains cord to mains power, then confirm that the Charging/Line indicator illuminates (above the [ON/OFF] key on the front panel).
4- 1
430-95226-001 (Rev. 5101)
Page 20
Basic Operation A systematic self-testing of the processing, delivery, and safeiy systems is performed whenever the Plum A+ is turned on, to verify readiness for operation. Note:
If the quality of earth grounding source is in doubt, use battery power.
Press the [ON/OFF] key to turn the power on. Check screen display and listen for a beep to indicate the audio is working. Wait for the self-tests to complete. If successful, put a fully primed macro cassette into the cassette door and close the door. See L.oaciing Cassette in Section 3 for details of correct procedure. When the Charging/Line indicator is off and the battery symbol on the display is flashing, this indicates that the Plum A+ is operating on low battery power and should be recharged.
q A
CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITH THE -BATTERY REMOVED. USE OF A PROPERLY MAINTAINBD AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION. To confirm battery is fully charged, connect the Plum A+ to AC power for a minimum of six hours while in the OFF mode. If an alarm occurs during the power on self-test, identify the alarm message, then take corrective action (see Section 6, Alarms and Troubleshooting). Start pump. If the self-test alarm recurs, remove the Plum A+ from service and contact the local Abbott Laboratories representative.
430-95226-001 (Rev. 5/01)
4- 2
Plum A+ Infusion Pump
Page 21
Operating Keys DATA RETENTION Delivery program settings and programming option selections are retained in non-volatile memory. If the Plum A+ has been turned off for longer than four hours, all delivery settings are cleared and programming option selections are restored to their default selections for next use.
4.2 Operating Keys The [ON/OFF] key is used to control the power of the Plum A+. Note: All deliveries must be stopped in order to turn the power off. Note: To confirm~proper shutdown of pump, do not unplug until “double beep” is heard (approximately 5 seconds). lf “double beep” is not heard, the pump will turn itself back on thinking the AC power was lost and revert to battery back-up power.
The [START] key is used to begin infusion and as final confirmation of programming.
The [STOP] key is used to stop the current delivery on one line; used in conjunction with soft keys to stop delivery on both lines.
The [SILENCE] key is used to silence an audible alarm during actions to correct its cause.
System Operating Manual
4- 3
430-95226-001 (Rev. 5/01)
Page 22
Basic Operation
The [SELECT] key is used to move the highlighting cursor between the programming fields.
The numeric keys are used to enter values for any highlighted field requiring numeric data. The [CLEAR] key will cause a highlighted numeric field to be cleared in preparation for data entry. The [DECIMAL POINT] key is used to manually enter numbers other than whole numbers (i.e., 1.2 mL). Softkeys are located at the bottom of the main display and serve a variety of functions. What each key does is indicated by the text in the display above the actual softkey.
430-95226-001 (Rev. 5/01)
4- 4
Plum A+ Infusion Pump
Page 23
lqdicators
4.3 Indicators
CHARGE
The Charge/Line Indicator is illuminated to indicate the battery is currently charging. Note: If the device is plugged into AC power, with a battery installed, and the Charge Indicator is not illuminated, contact technical support.
The fluid drop symbol, when flashing, indicates an active infusion on Line A and/or Line B.
Battery Capacity Symbol (located on right side of message region on display) indicates relative charge with number of white boxes from left to right.
This example indicates the battery is about two-thirds full.
4.4 Rear Case Controls The audio level rotary knob, located on the back, adjusts the alarm audio level.
The toggle switch, located below the audio level control, activates the lockout function when the lever is placed in the “up” position, disabling all front panel keys except [STOP]. Pressing of a key while lockout is active results in the display of the lockout enabled message. This action is recommended to car&m each lockout activation.
Note:
Note: If the lockout switch is put in the “Locked” position while the pump is off, the [ON/OFF] key allows the unit to turn on.
System Operating Manual
4- 5
430-95226-001 (Rev. 5101)
Page 24
00
The connector located below the lockout switch is the Nurse Call Relay Connector, for providing remote notification of a nurse call event. <
The RS-232 DataPort is used to interface with a host computer. Note:
Do not operate while’infusing.
Barcode Wand (optional) is used to scan desired drug names from provided list. It plugs into the Barcode Wand Port, located in the back of the pump.
430-95226-001 (Rev. 5/01)
4- 6
Plum A+ Infusion Pump
Page 25
Power On
4.5 Power On 1. Press [ON/OFF] to turn on the Plum A+ and you will see the System Self-Tests screen. c CAUTION: THE PUMP MUST BE INITIALLY STARTED ON AC POWER, OTHERWISE THE DATAPORT MAY NOT OPERATE PROPERLY.
After self-testing successfully completes, the display will proceed to the next screen. Note: If battery is missing or defective, a “Battery not installed” message with a ‘Continue without battery” prompt will appear (AC power only). CAUTION: Do NOT OPERATE THB PLIRI A+ ON PATIENTS WITH THE BATTERY REMOVED. USE OF A PROPERLY MAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION.
2. To set up the Plum A+, insert selected Plum Set and close cassette holder door lever. Note:
SETUP Clear Settings?
If cassette is already inserted, this screen is skipped.
Example of Clear Settings Screen
3. The on-screen message “Clear Settings?” may appear. Press A [Yes]to clear all settings. Clear ALL settings when changing patients or multiple programming parameters. Use A [c&f Program]when changing ONLY Line A or B. Note: This clears all delivery parameters, such as Rate and VTBI on both lines, as well as setting items, such as the KVO and Nurse Callback options, to their default values. Note: If no settings currently exist or all settings are zero, this screen will be bypassed.
System Operating Manual
4- 7
430-95226-001 (Rev. 5/01)
Page 26
Basic Operation
4.4 Simple Delivery lnlsroppEolsroppEDls1
Example of Main Delivery Screen (with cleared settings). Note: For clarity, whole number font sizes will appear larger than partial number font sizes (numbers on the right-side of the decimal).
1. To Program a delivery rate, press A [A].
2. To enter delivery rate, use numeric keypad.
3. Use [SELECT] key to move down to the VTBI field. Duration 0O:OO hnmin
Note: Upon entry to this screen, and to all lower screens from here on, the Barcode Reader (BCR) is enabled as indicated by “Wand Active” when attached to the Plum A+. To use the BCR, swipe tip of BCR wand across entire barcode, from dot to dot, on the Abbott supplied drug list card. The drug name also can be selected using A [Program Options].
4. To enter VTBI amount, use keypad. Rate VTBI
5.5 mLhr =mL
Duration 0O:OO hrzmin
Note: While the VTBI is being entered, the duration is automatically computed. If the computed value is greater than 99:59, dashes (- -) appear in place of hours.
5. Use [SELECT] key to move down to the DURATION field.
6. Use keypad to change the hours part of duration if desired.
Duration DO5 hEmin
Note: When all parameters are displayed, changing any one parameter will result in other parameters automatically being computed. To understand the Automatic Calculation feature, see Section 4.7 Examples of Automatic Calculation,
7. Use [SELECT] to change from hours to minutes. Note: The maximum hours allowed is 99 and the mmum minutes is 59.
8. Use keypad to change minutes.
430-95226-001 (Rev. 5/01)
4- 8
Plum A+ Iufusien Pump
Page 27
Examnles of Automatic Calculation CAUTION: VERIFY PRESSING [START].
A 1 PUMPING 1 STOPPED I B I lbte 0.0
5.5
RATE, VTBI,
AND DURATION BEFORE
9. Press [START] to begin infusion.
mLlhr
Example of Main Delivery Screen jusf,after being programmed and started. A
A A A A,
B
%z
The Line A Fluid Flow Indicator also will flash to indicate
Note:
pumping.
L A
4.7 Examples of Automatic Calculation
W HILE R UNNING (T ITRATION )
A T S TARTUP
AT KVO \rrBl=O.O DUFUUION=OO:OO
1st Action
2nd Action
[AUTO&UC]
1 st Action
2nd Action
[AUTOCALC]
1 st Action
2nd Action
[AUTOCALC]
Enter RATE
1 Enter VTBI
1 [DURATION]
Change RATE 1 Keep VTBI
) [DURATION]
Enter VTBI
1 Kee0 RATE
I DURAT/ONl
Enter VTBI
Enter DURATION
PATE]
Change RATE Change VTBI [DURATION]
Enter DURATION
Keep RATE
Enter RATE
Enter DURATION
W-W
Change VTBI
Keep RATE
[DURATION]
Change RATE
Enter VTBI
[DURATION]
Change DURATION
Keep VTBI
[RATE]
Enter DURATION
Enter VTBI
[DURATION]
Change VTBl ~~~!fON
PATE]
Enter VTBI
Enter DURATION
[RATE]
I
I
System Operating Manual
4- 9
430-95226-001 (Rev. 5101)