Page 11

hstructions for Use

SECTION 3

‘.

This section explains the Plum A+ setup and administration set use.

3.1 Setup WARNING ARRANGE TUBING, CORDS, AND CABLES TO MINIMIZE THE RISK OF PATIENT STRANGULATION OR ENTANGLEMENT.

UNPACKING CAUTION: PR ODUCT D A MAGE MAY OCCUR UNLESS PROPER CARE IS EXBRCISED DURING UNPACXING AND INSTALLATION. Do NOT USE TRE PLJJM A+ IF IT APPEARS DAMAGED IN ANY WAY. TRE BATTERY MAY NOT BE CRARGED UPON RECEIPT. Inspect the Plum A+ packaging for possible shipping damage. If damage is found, contact the delivery company immediately. Use care when unpacking the Plum A+. Retain the packing slip and save all packing material in case the Plum A+ is damaged or fails the pump self-test and has to be returned to Abbott Laboratories. Inspect the Plum A+ thoroughly for damage. CAUTION:

IF T= PLUM A+ APPEARS TO BE DAMAGED, CONTACT

ABBOTT -ORATORIES.

To set up the Plum A+, connect the mains cord to a properly earthed mains receptacle, unless temporary battery operation is desired.

Note:

Use mains power whenever possible. Store the Plum A+ connected to mains to confirm a fully charged battery for emergencies.

Note:

If the quality of earth grounding source is in doubt, use battery power.

CAUTION: THE PLUM A+ SYSTEM IS DESIGNED TO OPERATE NORMALLY JN TRB PRESENCE OF MOST ENCOUNTERED ELECTROMAGNETIC INTERFERENCE (EMI) CONDITIONS. m TRE EVENT OF EXTREME LEVELS OF INTERFERENCE, SUCH AS ENCOUNTERED NEXT To AN ELECTROSURGICAL GENERATOR. CELLULAR TELEPHONES, OR 2-WAY RADIOS, IT IS POSSIBLE THAT THE NORMAL OPERATION OF A SENSOR OR MICROCOMPUTER MIGHT BE D I SR UPT ED . O PE RA TION OF THE PUMP UND ER THESE CON D ITIONS SROULD BE AVOIDED.

3- 1

430-95226-001 (Rev. 5/01)

Page 12

Instructions for Use TANDEM CARRIER INSTRUCTIONS The Plum A+ may be safely and conveniently mounted on an IV stand or on a tandem carrier mounted on an IV stand. Always mount the tandem carrier (List# 12276-O 1) to the pole before attaching the pump(s) to the tandem carrier. Make sure tandem carrier clamp is attached firmly to pole. Make sure pumps are firmly attached to tandem carrier.

3.2 Administration Sets The Plum A+ is compatible with the wide range of LifeCare Plum series administration sets. Become familiar with the set components before preparing the administration set.

=Ei

Pnrnaty Line-

87

IK!B

Finger Grip

Flow Regulator

4 Pumping Chamber

-~

430-95226-001 (Rev. 5/01)

3- 2

Plum A+ Infusion Pump

Page 13

Administration Sets PREPARING THE ADMINISTRATION SET Note:

For detailed instructions, see administration set packaging.

Use the aseptic technique to prepare the administration set for priming, th$n proceed as follows:

E

Make sure flow regulator is closed.

Insert piercing pin into container outlet with a twisting motion.

Fill drip chamber to about l/2 full or to score mark. Do not completely fill.

Note: To avoid getting Distal Air Alarms, confirm that the fluid bag is positioned higher than the pump. This should be set up prior to priming the cassette.

System Operating Manual

3- 3

430-95226-001 (Rev. 5101)

Page 14

Instructions for Use PRIMING THE ADMINISTRATION SET To prime the administration set, proceed as follows:

1 Invert the cassette.

Turn the flow regulator counter-clockwise until a drop of fluid is seen in the pumping chamber.

Turn the cassette upright, then prime the remainder of the administration set.

Push in the flow regulator to close. Confirm there is no flow.

430-95226-001 (Rev. 5/01)

3- 4

Plum A+ Infusion Pump

Page 15

Administration Sets LOADING THE CASSETTE To load the primed cassette into the Plum A+, proceed as follows:

k.

Open the cassette door by lifting up the handle.

.T,.

Note:

Insert cassette into door guides. Close the cassette door. Confirm there is no flow and no kinks appear in tubing.

If flow is observed, close the tubing clamps and replace the administration set.

Note: The administration set should be changed per CDC guidelines or healthcare provider policy. Discard after use.

System Operating Manual

3- 5

430-95226-001 (Rev. 5101)

Page 16

Instructions for Use PREPARING THE SECONDARY LINE CAUTION: C ONS UL T THE DRUG LABELING TO CO N F IR M DRUG COEXPATIBILITP, CONCE NTRA TION , DELIVERY RATES, AND VOLUMES ARE ALL SUITABLE FOR CONCURRENT DELIVERY OR PIGGYBACK DELIVERY (SECONDARY FOLLOWED AUTOMATICALLY BY PRIMARY) MODES.

,

The Plum A+ features concurrent or piggyback delivery modes when therapy requires administering more than one fluid. In addition to standard containers, the Plum A+ can use syringes or vials on the secondary port. The secondary line can be prepared without removing or repriming the cassette. Before preparing the secondary line, observe the following guidelines: l l l l l l

’

Review the backpriming function Attach the secondary container using an 18- or 19- gauge, 31 mm (or shorter) needle, if the cassette secondary inlet port has a piercing reseal Attach syringe adapter (List 1198648) to syringes smaller than 10 cc. (Syringes 10 cc or larger can be attached directly to the secondary port of the cassette.) Syringes smaller than 3 cc should not be used. Attach a vial adapter (List 1188448) to the line vial cap when using vials on the secondary

When using a syringe adapter, retract the plunger to draw approximately 1 into the syringe to clear air from the adapter filter.

mL of fluid

To prepare the secondary line, use aseptic technique and proceed as follows:

With cassette door closed: loosen and remove white cap, then discard. Add secondary tubing or syringe.

Open all proximal clamps before pressing [START].

Set up pump per operating instructions. (See Se&ions 4 and 5 of this manual.)

430-95226-001 (Rev. 5/01)

3- 6

Plum A+ Infusion Pump

Page 17

Administration Sets Vial or Syringe: Secure the container to the cassette door using the optional container support arm (List 12095-04-01).

E Vial: Attach the vial adapter to the secondary port. Backprime the air from the vial adapter into the vial, if necessary.

System Operating Manual

3- 7

430-95226-001 (Rev. 5/01)

Page 18

Instructions for Use

3.3 DISCONTINUING ELECTRONIC FLOW CONTROL & SETTING GRAVITY FLOW CAUTION:

B EFORE OPE N I N G DOOR , CLOSE CLAMP ON PRIMARY OR S~ONDARY SET OR R EMOVE SECONDARY CONTAINER FROM SECONDARY PORT TO PREVENT MIXING Op_PRIBlARY AND SECONDARY FLUIDS.

If electronic flow needs to be stopped, the precision flow regulator on the cassette can be used to manually set and maintain fluid flow in a gravity mode (see the following graphics).

@ SET GRAVITY RATE BY @ OPEN DOOR & REMOVE CASSETTE NOTE: FLOW REGULATOR CLOSES AUTOMATICALLY WHEN DOOR IS OPENED TO PREVENT ACCIDENTAL FREE FLOW

TURNING FLOW REGUIATOR COUNTER-CLOCKWISE NOTE: CASSElTE SHOULD BE

IN UPRIGHT POSITION

NOTE: IF DOOR OPEN ALARM OCCURS, TURN PUMP OFF

3.4 DISCONTINUING FLUID ADMINISTRATION @ PRESS STOP

@ CLOSE DOOR

0 OPEN DOOR, REMOVE CASSETTE

430-95226-001 (Rev. 5/01)

3- 8

DISCARD SET AND FLUID CONTAINER PER HOSPITAL PROCEDURE

Plum A+ Infusion Pump