Page 8
SECTION 2
2.1 Features The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospitalwide, as well as alternate site and home healthcare. standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield@ needleless protection systems makes the Plum A+ a convenient and cost-effective infusion system.
THERAPIES l
Dose Calculation l
Loading Dose l
Multistep Programming
LINE PROGRAMMIN G OPTIONS . Therapy Selection l Nurse Call Back l Delayed Start Setting l Drug Label Library . Concurrent delivery l Titration l Micro 0.1-99.9 mL/hr (in 0.1 mL increments) flow rate range for both lines l Macro loo-999 mL/hr (in 1 mL increments) flow rate range for both lines l Automated Secondary drug delivery (Piggyback) PLUMSET CAPABILITIES l Anti Free-Flow Protection l Direct Connection for syringe delivery AIR MANAGEMENT l Air Trap l Air Removal/Backpriming l Air Detection-Proximal l Air Detection-Distal
2- 1
430-95226-001 (Rev. 5/01)
Page 9
Device Description BATTERY Battery Gauge l Six Hour Battery Recharge Time l Long battery life (6 hours) for emergency backup and l tempprary portable operation
BIOMEDICAL l Serial Communication l Upgradability (Field) l Variable Rate Cap . Alarm History l Plug-in Bar Code Reader for drug identification (Optional) l Nurse Call Relay Connector
1
OPTIONS l Volumes Infused (A, B, Total Volume) l KVO at dose end (1.0 ml/hr or less depending on delivery rate) or Continue Rate (CR) to continue at the current rate l Variable Distal Pressure Settings OTHER FEATURES Nonpulsatile volumetric accuracy Microprocessor control Large liquid crystal display (LCD) screen Panel back illumination on mains power Lockout switch Standard fullfill, partfill, syringe and vial use Parenteral and nonparenteral (enteral) fluid delivery Blood and blood products delivery Wide range of Standard and Specialty administration sets
430-95226-001 (Rev. 5/01)
2- 2
Plum A+ Infusion Pump
Page 10
User Qualification
2.2 User Qualification The Plum A+ is intended for use at the direction or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the pump and the administration of parenteral, enteral, and epidural fluids and drugs and whole blood or red blood cell components. The training should emphasize preventing related IV complications, including appropriate precautions to prevent accidental infusion of air. The epidural route can be used to provide anesthesia or analgesia.
WARNING ADMINISTER ONLY ANESTHETICS/ANALGESICS APPROVED FOR EPIDURAL ADMINISTRATION (AS INDICATED OR ALLOWED BY THE DRUGS’ FDA APPROVEiD LABELING). EPIDURAL ADMINISTRATION OF DRUGS OTHER THAN THOSE INDICATED FOR EPIDURAL USE COULD RESULT IN SERIOUS INJURY TO THE PATIENT.
2.3 System Self-Tests The pump performs a suite of System Self-Tests when power is applied to con.fii readiness for use. Failure during the Self-Tests is reported as a Malfunction Condition.
System Operating Manual
2- 3