Page 1 read from a scan by OCR; it may hold reading errors

Vi WEBSCAN

BARCODE VERIFIERS

TruCheck DPM Tower hd WEBSCAN

A WEBSCAN TruChe sa

Easy, Accurate, Trusted.

User Manual

Address: 1341 Sherman Drive Unit D, | Longmont CO 80501 USA

Phone: 877.WEBSCAN or 303.485.6811 Fax: 303.485.6353

Email: support@webscaninc.com Web: www.webscaninc.com

Page 2

Congratulations on the purchase of your Webscan barcode verifier! With this system, you can reach a new level of quality assurance. This manual will guide you in using your Webscan barcode verifier and in fully understanding the available features. In addition, it will assist you in understanding the verification results shown on the display and printed reports. All Webscan verifiers are operated by the same TruCheck™ Software, but some operating instructions will differ based on the particular unit that you have. This manual will point out how to operate each unit. For information about a specific symbology, please see the “Annex” entry for that symbology.

CONTENTS

Page 3 read from a scan by OCR; it may hold reading errors

9.2 How to Calibrate your Camera Verifier

10. Settings

10.1 1D Settings

10.2 2D Settings

10.3 Report and Print Options 1D and 2D

10.4 User Passwords

10.5 User Information

10.6 Revision Log

10.7 About Window

11. Explanation of Grading Standards and their Parameters

12. Using Multi-Mode

13. Annex and Compliance Statement

13.2 CDRH Compliance Statement

13.3 FCC Class “A” Statement

13.4 Routine Care and Maintenance

13.5 Engineering Specs

14. Compliance to Global and Industry Standards

14.2 ASCII Code Representation

14.3 Symbology Specific Verifications

15. Repairs and Other Services Available from Webscan

15.1 Webscan Service

15.2 Repair/Upgrade

15.3 Troubleshooting

15.4 Standard Warranty

15.5 What is Not Covered:

15.6 FCC Class “A” Statement:

15.7 Purchasing a calibration card

15.8 Seminar Service

15.9 Validation Protocol

Revision 1.12

A

28

29

29

33

36

38

38

38

39

40

50

54

54

55

55

55

57

58

59

64

64

64

64

65

65

65

66

67

68

Page 4

Congratulations on the purchase of your TruCheck™ barcode verifier! With this system, you can reach a new level of barcode quality assurance. The Webscan TruCheck™ family of laser and CCD imager based barcode verifiers evaluate 1D and 2D barcodes and grade them according to international and industry standards established by ANSI, ISO/IEC, CEN, GS1 and other organizations. Detailed diagnostic information is available on screen and reports of verification results are produced for printout or electronic archival. The TruCheck™ system is as easy to use as it is powerful. The main features of the system are:

Accurate barcode verification according to global standards

CCD Imager with integrated illumination or Laser-based barcode profiler (scanner)

Repeatable and accurate results

Calibrated and traceable to NIST standards

Intuitive User Interface

Detailed reporting

Storage of verification results in Adobe PDF™, HTML, MS Excel, and CSV

Consistent results independent of operator training or skill

This manual will guide you in using the TruCheck™ verifier and in fully understanding the available features. In addition, it will assist you in understanding the verification results shown on the display and printed reports. *TruCheck is a trademark of Webscan, Inc.

Check the verifier package for the following items.

TruCheck™ Imager or Laser-based verifier unit

TruCheck™ Software Installation CD (can also be downloaded from Webscan's website)

Power Supply (if applicable)

Quick Set Up Sheet (if applicable)

One of several available Calibration Cards

If any of the above items are missing, please contact Webscan. (Contact information can be found in the “if you need service” section near the end of the manual.) NOTE: The calibration card is usually an optional orderable item and might not be included in your package. However, a calibration card is required for proper operation of the TruCheck™ barcode verifier, so be sure that you have one before attempting to use the barcode verifier for the first time.

Page 5

TruCheck™ USB installation program takes you through the installation with instructions on every screen. Before installing TruCheck™ USB: Close all other applications. Make sure the TruCheck™ USB hardware is not plugged into the computer. NOTE: If the TruCheck™ USB hardware is plugged in before or during the software installation, the computer will install the TruCheck™ USB as an “unknown” device and it will not function. Log into your computer with administrator privileges if you are installing on Windows 7, 8, or Vista. Make sure that the .NET framework version 3.5 is installed on the PC. If the framework is not installed on the PC, you can do so by browsing to the “dotNET Installer” folder on the CD and running “dotNet Installer v3.5” executable. Make sure to use the Windows Update function to check for update information pertaining to the .NET framework. To install TruCheck USB Software: Insert the TruCheck™ USB CD-ROM into the CD-ROM drive. The installation program should start automatically. If it does not start, locate your CD-ROM drive in Windows Explorer and double-click the Setup.exe program at the top-level of the CD-ROM. Follow the on-screen prompts to complete the installation.

The following information shows the proper setup procedures for your new TruCheck™ USB. Before proceeding, locate the verifier and any accessories you purchased. NOTE: The TruCheck™ USB software should already be installed in accordance with the preceding section of this manual before plugging in the TruCheck™ USB hardware to the computer. Please connect the power supply cable to the cable coming out of the verifier with the barrel connector. Next, plug the power supply into the AC outlet. Connect the USB connector directly to a USB port on the computer. The USB port on the computer must support USB 2.0. Follow the on-screen prompts to complete the hardware installation. Once the USB cable is attached to the computer, you should see the “Found New Hardware” dialogue. With laser verifiers, you will be prompted twice. Wait for the “Found new hardware, your hardware is installed and ready to use” window to pop up. If prompted that the driver has not passed Window’s logo testing, select “continue anyway”. If prompted for the location of the drivers or any other files required during installation of the hardware drivers, please use the following folders: the “Drivers” folder on the CD or the path “C:\Program Files\Webscan Inc\TruCheckUSB\Drivers”

Page 6

A Windows-based PC running an Intel Core 2 or iSeries processor (i.e. Core i3, i5, i7) or equivalent. 2GB RAM minimum, 4 GB RAM recommended Hard drive with at least 200 MB free space Screen resolution 1024x768 recommended - 800x600 minimum Embedded USB 2.0 (system will not work with USB 1.0, USB 1.1, or some USB 2.0 add on cards) Direct connection to a PC USB 2.0 ‘A’ connector is recommended, not through an external USB hub.

Windows 10, Window 8, Windows Vista, Windows 7, or Windows XP installed. It is recommended that the latest service pack and all critical updates for Windows XP are loaded and configured before installing the Webscan application. Dot Net 3.5 SP1 must be installed and activated on the machine.

Minimize or eliminate other USB data intensive applications before running the TruCheck software. I.e., applications streaming USB audio or video.

Software Updates are available online at www.webscaninc.com under the resources tab. Select “Software Update Request” and fill out the request form. You can also contact Webscan Customer Service to request the upgrade by e-mail at support@webscaninc.com. We recommend joining our mailing list or following us on Facebook or Twitter to receive update notifications. Email marketing@webscaninc.com to subscribe to our mailing list.

http://www.facebook.com/webscaninc http://www.twitter.com/webscaninc

Page 7

Intelligent Mail barcodes are used in the USPS mailstream to increase the efficiency of deliveries. Intelligent Mail barcodes combine routing ZIP code information and tracking information into a single 4-state code. It encodes data from POSTNET and PLANET barcodes into a single barcode while providing a greater range of tracking data.

POSTNET barcodes encode a ZIP code or ZIP+4 code into half and full-height bars. Most often the delivery point is added, usually being the last two digits of the address or PO box number. In 2013, POSTNET barcodes were replaced with Intelligent Mail barcodes (IMB) by the USPS. IMB symbols encode data in 4-state bars which can be one of: tracker, descender, ascender, or full height bars. Correctly printing IMB symbols is critical for commercial mailers, as the USPS imposes penalties (in the form of higher postal rates, or non-approval of mail systems) for non-conforming symbols. However, it is almost impossible to know if an IMB system will meet USPS requirements. Webscan barcode verifiers fill this void by applying tests to IMB symbols according to requirements specified in USPS-B-3200. Support for Postnet and Intelligent Mail Barcode (IMB) is available as an optional feature in Webscan barcode verifiers. To enable this feature, a license file must be purchased and loaded into the verifier unit (under the Help Menu, Activate Feature). The license file is tied to the "Board Serial Number" found in the Help About window in the software. Postnet and IMB symbols are verified in accordance with USPS-B-3200 and Merlin* test criteria and reported with Pass (meets USPS requirements), Good (does not pass all USPS-B3200 requirements but does pass Merlin test criteria), Warning (does not pass all USPS-B-3200 requirements and generates a “Warning” message under "Merlin" test criteria), or Failure (does not pass all USPS-B-3200 requirements and generates a “Failure” message under “Merlin” test criteria). A visual representation of each passing and failing feature of the symbol is shown on the results screen and itemized in the report for each bar of the symbol. Bar spacing, rotation, minimum and maximum height, minimum and maximum width, bar continuity, pitch, clear zone reflectance, and symbol skew, are all graded and reported.** Also, the orientation of the symbol (forward or backward/upside down), including mirror image, is automatically detected and reported. * Merlin test criteria are not published by USPS and have been derived according to the best available information. ** Decoding method is not specified by USPS to interpretation of bar orientation and other parameters are subject to Webscan’s implementation. Please contact support@webscaninc.com for more information and assistance in setting up your software.

Page 8

Support for Maxicode (UPS Barcode) is available as an optional feature. To enable this feature, a license file must be purchased and loaded into the verifier unit (under the Help Menu, Activate Feature). The license file is tied to the "Board Serial Number" found in the Help About window in the software. Maxicode is verified in accordance with ISO/IEC 16023 and ISO/IEC 15415, with some necessary modification documented by ISO/IEC JTC1/WG1 committee. Please contact support@webscaninc.com for more information and assistance in setting up your software.

Page 9

Before we begin, you may need to calibrate your verifier. For more information on calibration, please refer to the calibration chapter.

Notice the large “Go Live” button. When clicked, the camera will capture live images, which will be shown in the main window. This image is useful for positioning the imager over the bar code that you wish to verify. Place the imager over the symbol to verify. NOTE: The "Start" button on the imager will do the same thing as clicking the “Go Live” button. If more than one code is in the field of view, you can draw a box around the code you want to verify. For more information, please see

.

NOTE: Holding the imager with your right index finger on the “Start” button and the cable away from you will make it easier to aim the verifier and position it over the symbol to verify. You will notice that the “Go Live” button now says “Verify”. When the bar code is positioned correctly, click on that button again to begin the verification. After successful verification, you should see a similar screen. The main screen features a static image of the symbol verified. The center portion contains the overall grade and barcode type. The quality parameters are listed to the right. Below the main screen is a box showing the encoded data. Depending on the symbology and grading standard, more or different detail will be displayed. The tool bar buttons and tabs are useful for in-depth analysis of barcodes and will be explained later in the Results section.

Page 10

Specifying regions in the live image may be required for the following reasons:

Multiple codes in the field of view

Defining a section of a linear code

In multi-mode, identifying the regions in a particular label format to be verified

If the code is off center

Speed up the verification process on verifiers with large fields of view; defining a small sampling area will increase the speed.

In order to define a region, position your cursor where you want the region to start, hold the left mouse button down and begin drawing a box by moving the mouse. Release the button when you are done. NOTE: When defining a region on 1D linear code draw your region within the symbol. If the region extends outside of the code, the verifier will attempt to scan that region, resulting in un-decoded scans. In the 1D settings menu, you may optionally select linear auto detect and the selected region can extend beyond the code without affecting the grade. For 2D matrix symbols and linear stacked symbols, the region should be drawn surrounding the entire code. The verifier will need the entire finder pattern, including quiet zones, within the selection to properly grade the code.