Chapter 5 References This section contains explanations and examples of selected barcode features. See the Configuration section for the actual barcode labels used to configure the reader.
RS-232 Parameters RS-232 Only Baud Rate Baud rate is the number of bits of data transmitted per second. Set the reader's baud rate to match the baud rate setting of the host device. With an improper baud rate setting, data may not reach the host correctly.
Stop Bits The stop bit(s) at the end of each transmitted character marks the end of transmission of one character and prepares the receiving device for the next character in the serial data stream. The number of stop bits selected (one or two) depends on the number the receiving terminal is programmed to accommodate. Set the number of stop bits to match host device requirements.
Parity This feature specifies parity required for sending and receiving data. A parity check bit is the most significant bit of each ASCII coded character. Select the parity type according to host device requirements. • Select None when no parity bit is required. • Select Odd parity and the parity bit value is set to 0 or 1, based on data, to ensure that an odd number of 1 bits are contained in the coded character. • Select Even parity and the parity bit value is set to 0 or 1, based on data, to ensure that an even number of 1 bits are contained in the coded character.
Handshaking Control The data interface consists of an RS-232 port designed to operate either with or without the hardware handshaking lines, Request to Send (RTS), and Clear to Send (CTS). Handshaking Control includes the following options: • RTS, RTS is asserted during transmissions. CTS is ignored. • RTS/CTS, RTS is asserted during transmissions. CTS gates transmissions.
Product Reference Guide
231
Page 242
References • •
RTS/XON/XOFF, RTS is asserted during transmissions. CTS is ignored. XON and XOFF gate transmissions. RTS On/CTS, RTS is always asserted. CTS gates transmissions.
RTS/CTS Scan Control, RTS is asserted during transmissions. CTS gates transmissions and controls enable and disable state of scanner.
RS-232/USB COM Parameters Intercharacter Delay This parameter specifies the intercharacter delay between the end of one character and the beginning of the next. The delay can be set within a range of zero (0) to 990 milliseconds in 10ms increments. A setting of zero specifies no delay. To set the delay: 1. Determine the desired setting in milliseconds. 2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc. 3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Go to page 25 and scan the barcode: SELECT INTERCHARACTER DELAY SETTING. 5. Scan the appropriate two digits from the keypad in Appendix D, Keypad, that represent the duration which was determined in the steps above. You will hear a two-beep indication after the last character. If you make a mistake before the last character, scan the CANCEL barcode to abort and not save the entry string. You can then start again at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit. This completes the procedure. See Table 28 for some examples of how to set this feature. Table 28. Intercharacter Delay Setting Examples STEP
ACTION
EXAMPLES
1
Desired Setting
50ms
150ms
600ms
850ms
2
Divide by 10 (pad with leading zeroes to yield two-digits)
05
15
60
85
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT INTERCHARACTER DELAY SETTING
5 6
232
Scan Two Characters From Appendix D, Keypad
‘0’ and ‘5’
‘5’ and ‘0’
‘6’ and ‘0’
‘8’ and ‘5’
Scan ENTER/EXIT PROGRAMMING MODE
WLS9600
Page 243
RS-232 Parameters
ACK NAK Options This enables/disables the ability of the reader to support the RS-232 ACK/NAK protocol. When configured, the reader and/or host sends an “ACK” when it receives data properly, and sends “NAK” when the data is in error. Options are: • Disable • Enable for label transmission, The reader expects an ACK/NAK response from the host when a label is sent. • Enable for host-command acknowledge, The reader will respond with ACK/NAK when the host sends a command. • Enable for label transmission and host-command acknowledge
ACK Character This setting specifies an ASCII character or hex value to be used as the ACK character. ASCII characters or any hex value from 0 to 0xFF can be selected. Setting to previously defined characters such as XON, XOFF, or host commands conflicts with normal operation of these characters. 8-bit data is not recognized when the option Data Bits has been set as 7 Data Bits.
1. Determine the desired character or value. 2. Use the ASCII Chart on the inside back cover of this manual to find the hex equivalent for the desired character/value. 3. Go to page 28 and scan ENTER/EXIT PROGRAMMING MODE to enter Programming Mode. 4. Scan the barcode: SELECT ACK CHARACTER SETTING. 5. Scan the appropriate two alpha-numeric characters from the keypad in Appendix D, Keypad, that represent the desired character/value in step 1 above. The second character will cause a two-beep indication. 6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit. See Table 29 for some examples of how to set this feature. Table 29. ACK Character Setting Examples STEP
ACTION
EXAMPLES
1
Desired Character/Value
ACK
$
@
>
2
Hex equivalent from ASCII Chart
0x06
0x24
0x40
0x3E
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT ACK CHARACTER SETTING
5
Scan Two Characters from Appendix D, Keypad
6
Product Reference Guide
‘0’ and ‘6’
‘2’ and ‘4’
‘4’ and ‘0’
‘3’ AND ‘E’
Scan ENTER/EXIT PROGRAMMING MODE
233
Page 244
References
NAK Character This setting specifies an ASCII character or hex value to be used as the NAK character. ASCII characters or any hex value from 0 to 0xFF can be selected. Setting to previously defined characters such as XON, XOFF, or host commands conflicts with normal operation of these characters. 8bit data is not recognized when the option Data Bits has been set as 7 Data Bits.
To set this feature: 1. Determine the desired character or value. 2. Use the ASCII Chart on the inside back cover of this manual to find the hex equivalent for the desired character/value. 3. Go to page 28 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT NAK CHARACTER SETTING. 5. Scan the appropriate two alpha-numeric characters from the keypad in Appendix D, Keypad, that represent the desired character/value in step 1 above. The second character will cause a two-beep indication. 6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 30 for some examples of how to set this feature. Table 30. NAK Character Setting Examples STEP
ACTION
EXAMPLES
1
Desired Character/Value
NAK
$
@
>
2
Hex equivalent from ASCII Chart
0x15
0x24
0x40
0x3E
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT ACK CHARACTER SETTING
5 6
234
Scan Two Characters From Appendix D, Keypad
‘1’ and ‘5’
‘2’ and ‘4’
‘4’ and ‘0’
‘3’ AND ‘E’
Scan ENTER/EXIT PROGRAMMING MODE
WLS9600
Page 245
RS-232 Parameters
ACK NAK Timeout Value This option specifies the amount of time the reader waits for an ACK character from the host following label transmission. The selectable timeout range is 200 milliseconds to 15,000ms (15 seconds) in 200ms increments. A selection of 0 disables the timeout. To set this value: 1. Determine the desired setting in milliseconds. 2. Divide the desired setting by 200 (setting is in 200ms increments). Pad the result with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc. 3. Go to page 29 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT ACK NAK TIMEOUT VALUE SETTING. 5. Scan the appropriate two digits from the keypad in Appendix D, Keypad, that represent the duration which was determined in the steps above. You will hear a two-beep indication after the last character. If you make a mistake before the last character, scan the CANCEL barcode to abort and not save the entry string. You can then start again at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 31 for some examples of how to set this feature. Table 31. ACK NAK Timeout Value Setting Examples STEP
ACTION
EXAMPLES
1
Desired Setting
200ms
1,000ms (1 sec.)
5200ms (5.2 sec.)
15,000ms (15 sec.)
2
Divide by 200
01
05
26
75
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT ACK NAK TIMEOUT VALUE SETTING
5
Scan Two Characters From Appendix D, Keypad
6
Product Reference Guide
‘0’ and ‘1’
‘0’ and ‘5’
‘2’ and ‘6’
‘7’ and ‘5’
Scan ENTER/EXIT PROGRAMMING MODE
235
Page 246
References
ACK NAK Retry Count This feature specifies the number of times the reader retries a label transmission due to a retry condition. The selectable range is from 1 to 254 retries. A selection of 0 disables the count, and a selection of 255 specifies unlimited retries. To set this feature: 1. Determine the desired setting. 2. Pad the number with leading zeroes to yield three digits. For example: 0 = 000, 5 = 005, 20 = 020, etc. 3. Go to page 29 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT ACK NAK RETRY COUNT SETTING. 5. Scan the appropriate three digits from the keypad in Appendix D, Keypad, that represent the number which was determined in the steps above. You will hear a two-beep indication after the last character. If you make a mistake before the last character, scan the CANCEL barcode to abort and not save the entry string. You can then start again at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 32 for some examples of how to set this feature. Table 32. ACK NAK Retry Count Setting Examples STEP
ACTION
EXAMPLES
1
Desired Setting
Disable Retry Count
3 Retries
54 Retries
Unlimited Retries
2
Pad with leading zero(es)
000
003
054
255
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT ACK NAK RETRY COUNT SETTING
5 6
236
Scan Three Characters From Appendix D, Keypad
‘0’, ‘0’ and ‘0’
‘0’, ‘0’ and ‘3’
‘0’, ‘5’ and ‘4’
‘2’, ‘5’ and ‘5’
Scan ENTER/EXIT PROGRAMMING MODE
WLS9600
Page 247
RS-232 Parameters
Disable Character Specifies the value of the RS-232 host command used to disable the reader. ASCII characters or any hex value from 0 to 0xFF can be selected. Setting to previously defined characters such as XON, XOFF, or host commands conflicts with normal operation of these characters. 8bit data is not recognized when the option Data Bits has been set as 7 Data Bits.
To set the value: 1. Determine the desired character or value. A setting of 0xFF indicates the Disable Character is not used (not available). 2. Use the ASCII Chart on the inside back cover of this manual to find the hex equivalent for the desired character/value. 3. Go to page 31 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT DISABLE CHARACTER SETTING. 5. Scan the appropriate two alpha-numeric characters from the keypad in Appendix D, Keypad, that represent the desired character/value in step 1 above. The second character will cause a two-beep indication. 6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 33 for some examples of how to set this feature. Table 33. Disable Character Setting Examples STEP
ACTION
EXAMPLES
1
Desired character/value
‘d’
‘}’
‘D’
Disable Command Not Used
2
Hex equivalent from ASCII Chart
0x64
0x7D
0x44
0xFF
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT DISABLE CHARACTER VALUE SETTING
5
Scan Two Characters From Appendix D, Keypad
6
Product Reference Guide
‘6’ and ‘4’
‘7’ and ‘D’
‘4’ and ‘4’
‘F’ AND ‘F’
Scan ENTER/EXIT PROGRAMMING MODE
237
Page 248
References
Enable Character Specifies the value of the RS-232 host command used to enable the reader. ASCII characters or any hex value from 0 to 0xFF can be selected. Setting to previously defined characters such as XON, XOFF, or host commands conflicts with normal operation of these characters. 8bit data is not recognized when the option Data Bits has been set as 7 Data Bits.
To set this feature: Determine the desired character or value. A setting of 0xFF indicates the Enable Character is not used (not available). 1. Determine the desired character or value. 2. Use the ASCII Chart on the inside back cover of this manual to find the hex equivalent for the desired character/value. 3. Go to page 31 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT ENABLE CHARACTER SETTING. 5. Scan the appropriate two alpha-numeric characters from the keypad in Appendix D, Keypad, that represent the desired character/value in step 2 above. The second character will cause a two-beep indication. 6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 34 for some examples of how to set this feature. Table 34. Enable Character Setting Examples STEP
ACTION
EXAMPLES
1
Desired character/value
‘e’
‘}’
‘E’
Enable Command Not Used
2
Hex equivalent from ASCII Chart
0x65
0x7D
0x45
0xFF
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT DISABLE CHARACTER VALUE SETTING
5 6
238
Scan Two Characters From Appendix D, Keypad
‘6’ and ‘5’
‘7’ and ‘D’
‘4’ and ‘5’
‘F’ AND ‘F’
Scan ENTER/EXIT PROGRAMMING MODE
WLS9600
Page 249
Keyboard Interface
Keyboard Interface Wedge Quiet Interval Specifies the amount of time the reader looks for keyboard activity before it breaks the keyboard connection in order to transmit data to host. The range is from 0 to 990ms in 10ms increments. This feature applies ONLY to the Keyboard Wedge interface.
1. Determine the desired setting in milliseconds. 2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc. 3. Go to page 40 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Prog. Mode. 4. Scan the barcode: SELECT WEDGE QUIET INTERVAL SETTING. 5. Scan the appropriate two digits from the keypad in Appendix D, Keypad, that represent the duration which was determined in the steps above. You will hear a two-beep indication after the last character. If you make a mistake before the last character, scan the CANCEL barcode to abort and not save the entry string. You can then start again at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit. This completes the procedure to set the Wedge Quiet Interval. See Table 35 for some examples of how to set this feature. Table 35. Wedge Quiet Interval Setting Examples STEP
ACTION
EXAMPLES
1
Desired Setting
10ms
150ms
600ms
850ms
2
Divide by 10 (and pad with leading zeroes)
01
15
60
85
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT WEDGE QUIET INTERVAL SETTING
5
Scan Two Characters From Appendix D, Keypad
6
Product Reference Guide
‘0’ and ‘1’
‘1’ and ‘5’
‘6’ and ‘0’
‘8’ and ‘5’
Scan ENTER/EXIT PROGRAMMING MODE
239
Page 250
References
Intercharacter Delay This parameter specifies the intercharacter delay between the end of one character and the beginning of the next. The delay can be set within a range of zero (0) to 990 milliseconds in 10ms increments. A setting of zero specifies no delay. This feature applies ONLY to the Keyboard Wedge interface.
To set the delay: 1. Determine the desired setting in milliseconds. 2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc. 3. Go to page 41 and scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming Mode. 4. Scan the barcode: SELECT INTERCHARACTER DELAY SETTING. 5. Scan the appropriate two digits from the keypad in Appendix D, Keypad, that represent the duration which was determined in the steps above. You will hear a two-beep indication after the last character. If you make a mistake before the last character, scan the CANCEL barcode to abort and not save the entry string. You can then start again at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming Mode. This completes the procedure. See Table 36 for some examples of how to set this feature. Table 36. Intercharacter Delay Setting Examples STEP
ACTION
EXAMPLES
1
Desired Setting
50ms
150ms
600ms
850ms
2
Divide by 10 (and pad with leading zeroes to yield twodigits)
05
15
60
85
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT INTERCHARACTER DELAY SETTING
5 6
240
Scan Two Characters From Appendix D, Keypad
‘0’ and ‘5’
‘1’ and ‘5’
‘6’ and ‘0’
‘8’ and ‘5’
Scan ENTER/EXIT PROGRAMMING MODE
WLS9600