Appendix H Troubleshooting Tips If you have problems with your TouchProbe, try the following troubleshooting tips. If you continue to have problems, contact Videx Customer Service by phone at 541-758-0521, by fax at 541-752-5285, or by e-mail at support@videx.com. I can't download data from my TouchProbe. This is one of the most common problems reported by customers, and can be due to a variety of reasons. To narrow down the possible causes, observe whether there is activity on the Receive and Transmit LEDs on the downloader when you attempt to communicate with the TouchProbe. I can't download data from my TouchProbe, and I see the Receive and Transmit LEDs flash when I try to communicate. 1. Verify that the serial cable, PC and base station are properly connected and that the connections are secure (refer to Figure 1-2 on page 10). 2. Be sure that you are either using a serial port cable purchased from Videx (part number TWC-008 or TWC-001), or else a serial port cable using the correct pinouts. Other serial port cables may not allow proper communication. 3. Do not use a serial port adaptor. Some adaptors will interfere with proper communication with the TouchProbe. 4. Be sure that you are using the correct ID within the software. Try using an ID of ten zeroes ("0000000000"); this ID will unlock and enable communication with any TouchProbe (provided that no other factors are interfering with communication). 5. Be sure that you are using the correct baud rate. When TouchProbes are sold or serviced by Videx, they are configured to communicate at 19200 baud. The baud rate can be changed from the Settings-Communications screen within Touch.exe (see page 27), or via the -b switch when using Transfer.exe (see page 93). Refer to

154

Appendix H Toubleshooting Tips

Page 163

page 153 for information on re-configuring the TouchProbe itself to communicate at a different baud rate. 6. If the TouchProbe read head is extremely dirty, that can interfere with communication. Try wiping it with a cloth, and use a q-tip or similar item to clean within the outer rim. 7. Be sure that the center pin (with a 3-sided brass-colored head) has not been removed from the socket within the downloader well. I can't download data from my TouchProbe, and I don't see any activity on the Receive and Transmit LEDs on the Downloader. 1. Verify that you are attempting to use the correct com port number when communicating with the TouchProbe. The com port setting can be changed through the Settings-Communications screen within Touch.exe (see page 27), or via the -p switch when using Transfer.exe (see page 95). 2. Verify that the serial cable, PC and base station are properly connected and that the connections are secure (refer to Figure 1-2 on page 10). 3. Windows computers often load drivers and other software which interfere with serial communication. If you think that there might be a software conflict with the serial port, try the following: a) If another DOS window is open on the Windows task bar, close all windows, and then re-open the TouchProbe communication software. b) If you have installed a Palm or other PDA on the com port, you will need to disable it. (For a Palm, right click on the HotSync icon in the system tray, and remove the check in front of "Local".) c) If you have installed other devic es or software which use the serial port, uninstall them via the Windows control panel. d) Try re-booting the computer; sometimes simply re-booting will solve a serial communication problem. e) Boot the computer from a DOS floppy and attempt to communicate with the TouchProbe. If you can communicate with the TouchProbe after booting from a DOS floppy, but not after booting to Windows, then Windows is installing software, which is interfering with serial communications. Appendix H Toubleshooting Tips

155

Page 164

Note: If you continue to have difficulty downloading the TouchProbe, and the TouchProbe has critical data in it, call Videx Customer Service at 541-758-0521.

How can I configure the TouchProbe to communicate at a different baud rate? When TouchProbes are sold or serviced by Videx, they are configured to communicate at a baud rate of 19200. It is possible to configure the TouchProbe to communicate at other baud rates: 300, 600, 1200, 2400, 4800 and 9600. One way to change this configuration is via the TouchProbe Configuration Utility, which is provided with the TouchProbe Downloader Plus (used for modem downloading of TouchProbes, see Appendix G). It is also possible to change the TouchProbe's baud rate by running transfer.exe from the DOS command line. To do this, run transfer.exe as described in Appendix E, but modify the ID file (for example, probes.txt in the command line example on page 96) to contain: 0000000000,Rn where the value of n sets the new TouchProbe baud rate as follows: n 0 1 2 3 4 5 6

Baud rate 300 600 1200 2400 4800 9600 19200

The TouchProbe's data will be downloaded, and the TouchProbe's baud rate will be set as specified by the n parameter. However, transfer.exe will then "lock up"; press the escape key to stop it from running. 156

Appendix H Toubleshooting Tips

Page 165

Note that to communicate with the TouchProbe at a rate other than 19200, you must specify the baud rate using the -b switch (refer to page 93).

Appendix H Toubleshooting Tips

157

Page 166

Appendix I Error Codes Error codes returned by transfer.exe and written to a status file (-s switch):

158

00

Transfer successful, no errors.

01

Time error: No prompt received after sending time set command.

02

Config error: No prompt received after sending configuration or load text command.

03

Retry error: Packet retried 8 times without successful transfer of consistent data and CRC.

04

Binary file open error: Binary file (usually temporary) image of TouchProbe memory could not be opened for translation. Possibly caused by hardware issues, but more likely because a file was requested that does not exist. Wrong name for instance.

05

Binary file not valid: Data in binary file is not consistent with required format (not a binary file).

06

Invalid record type: The record-type byte in a data record is not a number that has been defined.

07

No end record: The end of the file occurred without finding a valid “End” record.

09

Invalid ID record: An ID record encountered among the data records. Only one ID record is allowed, and it must be at the beginning of the data set.

Appendix I Error Codes

Page 167

10

Invalid record size: A button record header smaller than the required 9 bytes was encountered.

11

Binary file read error: An error reading the binary file occurred during translation. It is possible that the error occurred when the computer wrote the file (corruption or not complete).

12

Output write error: An error writing to the ASCII data file occurred during translation.

13

Error clearing data from probe: Failure to wake TouchProbe to clear data or to get a response to the clear command.

Error codes returned by transfer.exe and written to transfer.err: 0

Operation was successful

2

User cancelled the operation

3

One of the parameters is incorrect

4

Communications have timed out

5

Probe’s ID is not in the list of IDs

6

Error opening serial port

7

Did not get expected response from modem

8

Error reading from or writing to a file

10

Ran out of memory

Appendix I Error Codes

159

Page 168

Index A About menu, 24, 42 About command, 42 about TouchProbe communications, 2 add probe ID, 17–19, 31 Append/Overwrite Header, 70, 71, 80 origin code, 71 Append/Overwrite Tailer, 70, 74, 80

B

D battery, 5, 6 9-volt, 4 backup, 5 replace, 6 baud rate, 27 buffers, 15 button, 2, 4, 8, 33 data, 8, 45, 72 origin code, 72, 73, 81 read/write, 2, 8, 9, 37 DS1992, 8, 9 DS1993, 8, 9 DS1994, 8, 9 read-only, 2, 8, 9, 36 DS1990, 8, 9, 36 DS1991, 8, 9, 36 serial number, 8, 73, 75 specifications, 53

C cable pin outs, 54–60 serial port, 10, 11 cable, downloader (25 pin), 55 cable, downloader (9 pin), 57 160 cable, remote modem, 59 Communications Settings window, 27, 28 config.sys file, 15, 86 Configure Probe screen, 20, 33–38 options, 35 Read options, 36 Write options, 37 configure TouchProbe, 19–21, 33, 38 connect hardware, 11

Index data file, 16, 25–26, 29, 43–46, 62, 86 complex, 80 data, button, 8, 72 data.tmd file, 2, 22, 41 default settings, 34 delete probe ID, 32 disk space, 14 downloader, 9, 10 cable, 55, 57 connection, 10 install, 10 DS1990 button, 8, 9 DS1991 button, 8, 9 DS1992 button, 8, 9 DS1993 button, 8, 9 DS1994 button, 8, 9

E end cap, 4 ERRBIN transfer status message, 42 errors, 98, 155-156 ERRTRAN transfer status message, 42 Exit command, 26

Page 169

F file ASCII text file, 2 ASCII text file, 86 config.sys file, 15, 86 data file, 16, 25–26, 29, 43–46, 86 data.tmd file, 2 history.tmd file, 2 File menu, 16, 24, 25–26 Append/Overwrite Data File command, 26 Exit command, 26 Open Data File command, 25 Set Data File Path command, 25

G GOOD transfer status message, 42

H hard disk, 14 install, 14 hardware specifications , 51–53 Header, 40, 43, 44, 65, 66, 67, 68, 70, 71, 80 Append/Overwrite, 70, 71, 80 origin code, 71 Load Text into Probe, 40 Main, 43, 44, 63, 65, 66, 80 origin codes, 81 Read/Write button, 67 Read/Write Button, 68, 69, 80 history file, 2, 96 history.tmd file, 2, 96

I iButton. See Touch Memory button

ID, probe. See probe ID installing the software, 14

K key ring, 4

L Load Text into Probe Header, 40 Load Text into Probe Tailer, 40

M Main Header, 43, 44, 63, 65, 66, 80 origin code, 66 Main Tailer, 43, 44, 46, 63, 65, 76, 80 menu, 24–42 About, 24, 42 File, 16, 24, 25–26 Settings, 24, 27–29 TouchProbe, 17, 24, 30–42 modem communications, 12, 126–49

N NO RESPONSE transfer status message, 42 NOT CLEARED transfer status message, 42

O origin code, 66, 68, 72, 73 output file, 96

P Power light, 10 probe ID, 18 add probe ID, 17–19

Index

161

Page 170

probe ID, 31, 34, 71, 87 active probe, 34 add probe ID, 31 delete probe ID, 32

Q quick-start exercise, 16 quit, 26

R read a button, 21 Read options, 36 read/write button, 2, 8, 9, 64 Read/Write Button Header, 67, 68, 69, 80 notation, 68, 69, 81 Read/Write Button Tailer, 67, 75, 76, 80 notation, 76, 81 read/write buttons, 36 read-only button, 2, 9, 36 remote modem cable, 59 required hardware, 3 required software, 3 Review Probe Configuration screen, 39 RJ-11, 10, 11

S serial port, 27 serial port cable, 9, 10, 11 Settings menu, 24, 27–29 Communications command, 27 Data Format command, 29 Display command, 29 sound, 115, 116 specifications , 51–53 status file, 97 162

Index

T Tailer, 40, 43, 44, 46, 65, 67, 70, 80 Append/Overwrite, 70, 74, 80 Load Text into Probe, 40 Main, 43, 44, 46, 63, 65, 76, 80 Read/Write Button, 67, 75, 76, 80 notation, 76 touch button, 4 Touch Memory button, 2, 4, 8, 21, 33, 53 communicate with, 4 data, 8 origin code, 81 read a button, 21 read/write button, 2, 8, 9 DS1992, 8, 9 DS1993, 8, 9 DS1994, 8, 9 read-only button, 2, 8, 9 DS1990, 8, 9 DS1991, 8, 9 serial number, 8, 75 Touch Memory Button read/write button, 37 read-only button, 36 DS1990, 36 DS1991, 36 Touch Memory button specifications, 53 TouchProbe, 4 configure probe, 19–21, 38 downloader install, 10 ID, 18 add probe ID, 17–19 ID, 31, 34, 71, 88 add probe ID, 31 delete probe ID, 32