For the Symbol MS-4407

Page 93

Chapter 8 SSI Interface Overview This chapter describes the system requirements of the Simple Serial Interface (SSI), which provides a communications link between Motorola decoders (e.g., scan engines, slot scanners, hand-held scanners, two-dimensional scanners, hands-free scanners, and RF base stations) and a serial host. It provides the means for the host to control the decoder or scanner.

Communications All communications between the decoder and host are done over the hardware interface lines using the SSI protocol. See the Simple Serial Interface Programmer’s Guide, p/n 72-40451-xx, for more information on SSI. The host and the decoder exchange messages in packets. (A packet is a collection of bytes framed by the proper SSI protocol formatting bytes.) The maximum number of bytes per packet allowed by the SSI protocol for any transaction is 257 (255 bytes + 2 byte checksum). Decode data can be sent as ASCII data (unpacketed), or as part of a larger message (packeted), depending on the decoder configuration. The SSI performs the following major functions for the host device:

• Maintains a bi-directional interface with the decoder • Allows the host to send commands that control the decoder • Passes data from the decoder to a host device in the formatted SSI packet format or straight decode message. The working environment of the SSI consists of a decoder, a serial cable which attaches to the host device, and in some instances, a power supply. SSI transmits all decode data including special formatting (e.g., AIM ID). The format of the transmitted data can be controlled via parameter settings. The decoder may also send parameter information, product identification information or event codes to the host. All commands sent between the decoder and host must use the format described in the SSI Message Formats section. The SSI Transactions section describes the required sequence of messages in specific cases.

Page 94

8-2

Symbol MS4404/MS4407 Integration Guide

Table 8-1 lists all the SSI Opcodes supported by the decoder. It identifies the SSI partner allowed to send a message of each type. The host transmits Opcodes designated type H. The decoder transmits type D Opcodes, and either partner can transmit Host/Decoder (H/D) types. Table 8-1 SSI Commands

Name

Type

Opcode

Description

AIM_OFF

H

0xC4

Deactivate aim pattern.

AIM_ON

H

0xC5

Activate aim pattern.

BEEP

H

0xE6

Sound the beeper.

CAPABILITIES_REPLY

D

0xD4

Reply to CAPABILITIES_REQUEST; contains a list of the capabilities and commands supported by the decoder.

CAPABILITIES_REQUEST

H

0xD3

Request capabilities report from the decoder.

CMD_ACK

H/D

0xD0

Positive acknowledgment of received packet.

CMD_NAK

H/D

0xD1

Negative acknowledgment of received packet.

DECODE_DATA

D

0xF3

Decode data in SSI packet format.

EVENT

D

0xF6

Event indicated by associated event code.

LED_OFF

H

0xE8

De-activate LED output.

LED_ON

H

0xE7

Activate LED output.

PARAM_DEFAULTS

H

0xC8

Set parameter default values.

PARAM_REQUEST

H

0xC7

Request values of certain parameters.

PARAM_SEND

H/D

0xC6

Send parameter values.

REPLY_ID

D

0xA6

Reply to REQUEST_ID; contains decoder's serial number.

REPLY_REVISION

D

0xA4

Reply to REQ_REV contains decoder's software/hardware configuration.

REQUEST_REVISION

H

0xA3

Request the decoder's configuration.

SCAN_DISABLE

H

0xEA

Prevent the operator from scanning bar codes.

SCAN_ENABLE

H

0xE9

Permit bar code scanning.

SLEEP

H

0xEB

Request to place the decoder into low power.

START_DECODE

H

0xE4

Tell decoder to attempt to decode a bar code.

STOP_DECODE

H

0xE5

Tell decoder to abort a decode attempt.

WAKEUP

H

N/A

Wakeup decoder after it has entered low power mode.

The details of the SSI protocol can be found in the Simple Serial Interface Programmer's Guide (72-40451-xx).

Page 95

SSI Interface

8-3

SSI Transactions General Data Transactions ACK/NAK Handshaking ACK/NAK handshaking may be enabled or disabled. If this parameter is set, all packeted messages must have a CMD_ACK or CMD_NAK response, unless the command description states otherwise. This parameter is enabled by default. This handshaking should remain enabled to provide feedback to the host. Raw decode data and WAKEUP do not use ACK/NAK handshaking since they are not packeted data. Example of a problem which may occur when ACK/NAK handshaking is disabled:

• The host sends a PARAM_SEND message to the decoder to change baud rate from 9600 to 19200. • The decoder cannot interpret the message. • The decoder does not implement the changes requested by the host. • The host assumes that the parameter changes have occurred and acts accordingly. • Communications are lost because the change did not occur on both sides. If the ACK/NAK handshaking is enabled, the following occurs:

• The host sends a PARAM_SEND message. • The decoder cannot interpret the message. • The decoder CMD_NAKs the message. • The host resends the message. • The decoder receives the message successfully, responds with CMD_ACK, and implements parameter changes.

Transfer of Decode Data The Decode Data Packet Format parameter controls how decode data is sent to the host. When this parameter is set, the data is sent in a DECODE_DATA packet. When the parameter is cleared, the data is transmitted as raw ASCII data. NOTE

When decode data is transmitted as raw ASCII data, then ACK/NAK handshaking does not apply regardless of the state of the ACK/NAK handshaking parameter.

Page 96

8-4

Symbol MS4404/MS4407 Integration Guide

ACK/NAK Enabled and Packeted Data The decoder sends a DECODE_DATA message after a successful decode. The decoder waits for a programmable time-out for a CMD_ACK response. If it does not receive the response, the decoder tries to send two more times before issuing a host transmission error. If the decoder receives a CMD_NAK from the host, it may attempt a retry depending on the cause field of the CMD_NAK message. (1)

Decoder Data is captured by decoder

DECODE_DATA

Host (2) CMD_ACK message sent

ACK/NAK Enabled and Unpacketed ASCII Data Even though the ACK/NAK handshaking is enabled, no handshaking occurs because the handshaking applies only to packeted data. In this example the packeted_decode parameter is disabled.

Decoder Data is captured by decoder

ASCII data

Host

ACK/NAK Disabled and Packeted DECODE_DATA In this example ACK/NAK does not occur even though packeted_decode is enabled because the ACK/NAK handshaking parameter is disabled.

Decoder Data is captured by decoder

(1) DECODE_DATA message sent

Host

Page 97

SSI Interface

8-5

ACK/NAK Disabled and Unpacketed ASCII Data Data captured by the decoder is sent to the host.

Decoder

(1)

Data is captured by decoder

Host

ASCII data sent

Communication Summary RTS/CTS Lines All communication must use RTS/CTS handshaking as described in the Simple Serial Interface Programmer’s Guide, p/n 72-40451-xx. If hardware handshaking is disabled or bypassed, the WAKEUP command must be sent prior to all other communications, or else the first byte of a communication message may be lost during the decoder wakeup sequence.

ACK/NAK Option ACK/NAK handshaking can be enabled or disabled. This handshaking is enabled by default and we suggest the option not be disabled. Disabling this handshaking may lead to communication problems, as handshaking is the only acknowledgment that a message was received, and if it was received correctly. ACK/NAK is not used with unpacketed decode data regardless of whether or not this option is enabled.

Number of Data Bits All communication with the decoder must use 8-bit data.

Serial Response Time-out The Serial Response Time-out parameter is used to determine how long to wait for a handshaking response before trying again, or aborting any further attempts. Both the host and decoder should use the same value for this parameter. NOTE

A temporary change may be made to the Serial Response Time-out when the host takes longer to process an ACK, or longer data string. Frequent permanent changes are not recommended due to limited write cycles of non-volatile memory.

Page 98

8-6

Symbol MS4404/MS4407 Integration Guide

Retries When sending data, the host should resend twice after the initial send if the decoder does not respond with an ACK or NAK (if ACK/NAK handshaking is enabled), or response data (e.g., PARAM_SEND, REPLY_REVISION). If the decoder replies with a NAK RESEND, the host resends the data. All resent messages must have the resend bit set in the Status byte. The decoder resends data two times after the initial send if the host fails to reply with an ACK or NAK (if ACK/NAK handshaking is enabled).

Baud Rate, Stop Bits, Parity, Response Time-out, ACK/NAK Handshake If the serial parameters above are changed using PARAM_SEND, the ACK response to the PARAM_SEND is sent using the previous values for these parameters. The new values then take effect for the next transaction.

Errors A communication error is generated by the decoder when:

• The CTS line is asserted when the decoder tries to transmit, and is still asserted on each of 2 successive retries

• Failure to receive an ACK or NAK after initial transmit and two resends.

Things to Remember When Using SSI Communication If hardware handshaking is not used, space messages sufficiently apart. The host must not communicate with the decoder if the decoder is sending. If hardware handshaking is used, frame each message properly with the handshaking signals. Do not try to send two commands within the same handshaking frame. There is a permanent/temporary bit in the PARAM_SEND message. Temporary changes are lost when power is removed from the decoder. Permanent changes are written to non-volatile memory. Frequent changes shorten the life of the non-volatile memory.

Page 99

SSI Interface

8-7

Selecting Time Delay to Low Power Mode using SSI Time Delay to Low Power Mode on page 6-8 provides bar codes to select a general time delay. To program a more specific delay value, use an SSI command according to Table 8-2. Table 8-2 Values for Selecting Time Delay to Low Power

Value

Timeout

Value

Timeout

Value

Timeout

Value

Timeout

0x00

15 Mins

0x10

1 Sec

0x20

1 Min

0x30

1 Hour

0x01

30 Mins

0x11

1 Sec

0x21

1 Min

0x31

1 Hour

0x02

60 Mins

0x12

2 Secs

0x22

2 Mins

0x32

2 Hours

0x03

90 Mins

0x13

3 Secs

0x23

3 Mins

0x33

3 Hours

N/A

N/A

0x14

4 Secs

0x24

4 Mins

0x34

4 Hours

N/A

N/A

0x15

5 Secs

0x25

5 Mins

0x35

5 Hours

N/A

N/A

0x16

6 Secs

0x26

6 Mins

0x36

6 Hours

N/A

N/A

0x17

7 Secs

0x27

7 Mins

0x37

7 Hours

N/A

N/A

0x18

8 Secs

0x28

8 Mins

0x38

8 Hours

N/A

N/A

0x19

9 Secs

0x29

9 Mins

0x39

9 Hours

N/A

N/A

0x1A

10 Secs

0x2A

10 Mins

0x3A

10 Hours

N/A

N/A

0x1B

15 Secs

0x2B

15 Mins

0x3B

15 Hours

N/A

N/A

0x1C

20 Secs

0x2C

20 Mins

0x3C

20 Hours

N/A

N/A

0x1D

30 Secs

0x2D

30 Mins

0x3D

30 Hours

N/A

N/A

0x1E

45 Secs

0x2E

45 Mins

0x3E

45 Hours

N/A

N/A

0x1F

60 Secs

0x2F

60 Mins

0x3F

60 Hours

Page 100

8-8

Symbol MS4404/MS4407 Integration Guide

Simple Serial Interface Default Parameters This section describes how to set up the decoder with a SSI host. When SSI is used, the decoder is programmed via bar code menu or SSI hosts commands. Throughout the programming bar code menus, default values are indicated with asterisks (*).

* Indicates Default

*Baud Rate 9600

Feature/Option

(06h) Option Hex Value for programming via SSI command

NOTE

Most computer monitors allow scanning the bar codes directly on the screen. When scanning from the screen, be sure to set the document magnification to a level where the bar code can be seen clearly, and bars and/or spaces are not merging.

Table 8-3 lists the defaults for the SSI host. There are two ways to change the default values:

• Scan the appropriate bar codes in this guide. These new values replace the standard default values in memory. To recall the default parameter values, scan the *Set All Defaults bar code on page 6-3.

• Download data through the device’s serial port using SSI. Hexadecimal parameter numbers are shown in this chapter below the parameter title, and options are shown in parenthesis beneath the accompanying bar codes. See the Simple Serial Interface (SSI) Programmer’s Guide for detailed instructions for changing parameters using this method. NOTE

See Appendix A, Standard Defaults for all user preferences, hosts, symbologies, and miscellaneous default parameters.

Table 8-3 SSI Default Table

Parameter

Parameter Number

Default

Page Number

Select SSI Host

N/A

N/A

8-10

Baud Rate

9Ch

9600

8-11

Parity

9Eh

None

8-13

Check Parity

97h

Disable

8-14

Software Handshaking

9Fh

ACK/NAK

8-14

Host RTS Line State

9Ah

Low

8-15

Decode Data Packet Format

EEh

Send Raw Decode Data

8-15

Stop Bits

9Dh

1

8-16